Skip to main content
Journal of Clinical Laboratory Analysis logoLink to Journal of Clinical Laboratory Analysis
. 2022 May 19;36(7):e24490. doi: 10.1002/jcla.24490

Analysis of current status of quantitative detection of biomarkers for liver fibrosis in Clinical labs in China

Chao Zhang 1,, Chuanbao Zhang 1,
PMCID: PMC9279982  PMID: 35587485

Abstract

Aim

To explore the quality control and implementation of the quantitative detection of liver fibrosis biomarkers, laminin (LN), collagen IV (Col Ⅳ), procollagen III amino‐terminal propeptide (PⅢNP), hyaluronic acid (HA), and cholyglycine (CG), in China.

Methods

Two quality control products were measured in different laboratories using different measurement methods and reagents, and the acquired results were subjected to analysis. The quantitative detection technique was based on the conventional assessment criteria, with a target value ±30% being employed.

Results

Hundred labs were involved in the External Quality Assessment with 88 laboratories completing the assessment, and the pass rates were 84%, 80.2%, 67.5%, 77.3%, and 58.3% for HA, LN, PⅢNP, Col Ⅳ, and CG, respectively. Chemiluminescence immunoassay was used most for HA (90.1%), LN (90.1%), PⅢNP (87.9%), and Col Ⅳ (82.9%) determination, whereas the chemiluminescence immunoassay (31.6%), latex‐enhanced immunoturbidimetry (36.7%), and homogeneous enzyme immunoassay (26.7%) were used for CG determination. The coefficients of variation for HA, LN, PⅢNP, Col Ⅳ, and CG in different laboratories were 3.3%–19.49%, 1.74%–38.81%, 1.97%–41.29%, 2.85%–41.69%, and 2.71%–41.8%, respectively.

Conclusion

The clinical quantitative detection of liver fibrosis biomarkers is highly performed in China. The existing problems are that there are many manufacturers producing reagents and instruments, the quality of reagents is uneven, the specificity and sensitivity of reagents are greatly different, the comparability of results of various systems is poor, and the accuracy and consistency between different systems are lacking. All above underscores the critical importance of EQA in improving and monitoring the identification of biomarkers for liver fibrosis.

Keywords: external quality assessment, internal quality control, liver fibrosis biomarkers, quality control products, quantitative detection and analysis


First, a questionnaire survey was conducted to investigate the current status of quantitative detection of liver fibrosis markers in clinical laboratories. Then, external quality assessment is carried out by detecting quality control products.

graphic file with name JCLA-36-e24490-g001.jpg

1. INTRODUCTION

The characteristics of chronic liver disease include increasing fibrosis, which may result in portal hypertension, cirrhosis, liver synthesis damage, and hepatocellular carcinoma. 1 , 2 Cholestatic and hepatotoxic chronic liver injuries are the two most prevalent forms of chronic liver injury. 3 Hepatotoxic injury is induced by persistent liver cell damage, such as HBV and/or HCV infection, and non‐alcoholic or alcoholic steatohepatitis. 4 , 5  More than 240 million individuals worldwide are infected with chronic HBV, with China being one of the countries with a high incidence of HBV. 6 , 7  Therefore, a large number of people with chronic HBV infection develop liver fibrosis in China.

Liver fibrosis is defined as the production of fibrous scarring caused by the accumulation of extracellular matrix (ECM) proteins (mainly type I and III crosslinked collagen) that substitute injured normal tissue. 1 , 3  To be frank, the stage of liver fibrosis could only be determined via the biopsy of the liver at present. Nevertheless, a biopsy of the liver can cause serious problems to patients in up to 3%, as well as mortality occurring in 0.03% of the cases. 5 For monitoring chronic liver disease progression, assessing fibrosis staging is critical, which determines prognosis and optimal treatment time, monitors treatment response, and assesses disease progression to reduce morbidity and mortality associated with cirrhosis sequelae. Thus, accurate non‐invasive tests for liver fibrosis diagnosis and staging are required. Non‐invasive fibrosis testing of fibrosis is gaining traction in clinical practice to diagnose liver fibrosis. For fibrosis assessment criteria, blood‐based indicators provide considerable benefits than standard biopsy of the liver, such as accessibility, cost, and safety. Serum markers estimate fibrosis using markers that measure ECM degradation/fibrogenesis. An extensive spectrum of chronic liver disorders, such as non‐alcoholic and alcoholic fatty liver disease, as well as chronic viral hepatitis, has been studied and verified with the aid of this technique.

Currently, five major liver fibrosis indicators are utilized in clinical laboratories, namely laminin (LN), hyaluronic acid (HA), procollagen III's N‐terminal peptide (PIIINP), cholyglycine (CG), and collagen IV (Col IV). 8 , 9 , 10 , 11 , 12 , 13 However, none of these biomarkers was included in the formal External Quality Assessment (EQA) plan. This means it is impossible to assess whether the measurement is accurate. The reliability and utility of disease diagnosis and treatment are determined by the accuracy of the clinical outcomes. The comparability and accuracy of findings across various methods employed in different laboratories, as a consequence, are critically essential considerations. EQA is a method in which the same samples are examined by various labs, and the findings are obtained, analyzed, and subjected to a comparison made by an external, independent organization. The detection and calibration capabilities of clinical laboratories are determined by the National Center for Clinical Laboratories (NCCL) via interlaboratory comparisons, and the laboratory progress is monitored. As a consequence, EQA contributes to the guarantee of comparability and accuracy of outcomes. Now that clinical laboratories are providing detection services for liver fibrosis biomarkers, it is necessary for the NCCL to conduct EQA services for them to ensure the accuracy and comparability of test results.

2. MATERIAL AND METHODS

2.1. Respondents

In the EQA survey for the quantitative identification of liver fibrosis biomarkers, the responders comprised clinical labs that had consented to be involved in it.

2.2. Questionnaire survey

In response to the query on "Investigations on the Quantitative Detection of Biomarkers of Liver Fibrosis," the clinical laboratories provided responses using the EQA system established by NCCL. In addition to the essential data, the questionnaire comprised of eight queries, including the quantitative identification of liver fibrosis biomarkers, the type of the sample, method of detection, the brand of the reagent, the brand of the instrument, the unit of the concentration used by the laboratory, and their readiness to be involved in the EQA survey (serum matrix) for the quantitative identification of liver fibrosis biomarkers (Table S1). After the data were obtained, a summary of the current state of the quantitative detection of liver fibrosis indicators in clinical labs was prepared.

2.3. EQA survey

The NCCL handled the notification of the EQA investigation and the submission of the application. The quality control items were provided to the labs that had submitted their applications. 202011 and 202012 were the two samples (comprising HA, LN, PⅢNP, Col Ⅳ, and CG biomarkers) with different concentrations (cut‐off values and high values).

2.4. Criteria of EQA evaluation

For the purpose of conducting the present research, the statistically robust mean ± 30% was used as the evaluation criterion. The detection result was considered as “qualified” if the deviation of the two samples was within the set range.

2.5. Statistical analysis

Based on the response and submitted findings, the accumulated number and the proportion of the questionnaire responses were calculated. The quantitative identification outcomes of the liver fibrosis biomarkers were classified in accordance with the reagents and evaluation made. Moreover, every reagent manufacturing company was named as reagents 1, 2, 3, and so on (Table S2). The coefficient of variation (CV), mean value, and bias between the target and the mean values of the reagent groups were derived when no less than two laboratories were in a single group.

3. RESULTS

3.1. The proportion of detection projects of liver fibrosis biomarkers

In the questionnaire, the proportion of the various detection items was evaluated. As shown in Table 1, 49.6% (112/226) of the laboratories quantitatively detected four liver fibrosis biomarkers (HA, LN, PⅢNP, and Col Ⅳ); 39.8% (90/226) of the laboratories detected five liver fibrosis biomarkers (HA, LN, PⅢNP, Col Ⅳ, and CG); 10.6% (24/226) of the laboratories detected only the CG biomarker. We assessed the proportion of distinct detection techniques with the aid of a questionnaire. The results revealed that the proportion of chemiluminescence immunoassay (CLIA) for HA, LN, PⅢNP, and Col Ⅳ detection was 86.6%. Other detection methods included latex‐enhanced immunoturbidimetry, up‐conversion luminescence immunoassay, (ULIA), electrochemiluminescence, (ECLIA), and radioimmunoassay (RIA), accounting for approximately 13.4% of the total detection methods. The common methods for CG detection were CLIA, latex‐enhanced immunoturbidimetry, and homogeneous enzyme immunoassay (HEI), accounting for 28.1%, 37.7%, and 28.9%, respectively. The other methods including ECLIA and RIA represented around 5.3% of the overall estimate.

TABLE 1.

Summary of questionnaire information

Project Amount Proportion
HA, LN, PⅢNP, Col Ⅳ 112 49.6%
HA, LN, PⅢNP, Col Ⅳ, and CG 90 39.8%
CG 24 10.6%
Method Amount Proportion (%)
HA, LN, PⅢNP, Col Ⅳ Chemiluminescence immunoassay, CLIA 175 86.6
Latex‐enhanced immunoturbidimetry 17 8.4
Up‐conversion luminescence immunoassay, ULIA 5 2.5
Electrochemiluminescence, ECLIA 3 1.5
Radioimmunoassay, RIA 2 1.0
CG Chemiluminescence immunoassay, CLIA 32 28.1
Electrochemiluminescence, ECLIA 3 2.6
Radioimmunoassay, RIA 1 0.9
Latex‐enhanced immunoturbidimetry 43 37.7
Homogeneous enzyme immunoassay, HEI 33 28.9
Others 2 1.8

The proportion of detection projects of liver fibrosis biomarkers and composition ratios of detection methods.

3.2. Group statistics on the basis of the methods of detection

For the quality control materials that were distributed, two concentrations (cut‐off value and high value) were established. For the purpose of collecting more data, the quality control materials were distributed to the labs that quantitatively detected four or five biomarkers. Moreover, the clinical labs that were involved in the present research employed a variety of methods, procedures, and reagents. “Other category” was used to classify laboratories that did not describe their detection techniques in their application. As illustrated in Table 2, robust CV, standard uncertainty and the robust standard deviation are all in accordance with the ISO13528 standards. The results revealed significant differences in the robust mean values among the various identification methods of the laboratories, indicating the necessity for grouped statistics. For HA detection using RIA, the quality control substance concentration increased with an increase in robust CV (24.1% vs 70.8%), which can be attributed to identification procedures defects and small trial size. With increased concentration, robust CV of chemiluminescence immunoassay‐HRP for HA, LN, PⅢNP, and Col Ⅳ showed a downward trend, while the chemiluminescence immunoassay‐Acridinium Ester showed an opposite trend. We tried to explain that maybe CLIA‐HRP is prone to be stable in HA, LN, Col Ⅳ, and PⅢNP detection when the sample concentration is increased, while CLIA‐AE is more suitable for detection at lower concentrations. Conversely, for Col IV detection, the robust CV of CLIA‐AE and microplate decreased with an increase in concentration. However, the detection of PⅢNP, HA, LN, and Col Ⅳ via ULIA was relatively stable. For CG detection, the robust CV of CLIA decreased with increased sample concentration. The robust CV of immunoturbidimetry increased significantly. We suspect that CLIA may have high reliability in detecting high concentrations.

TABLE 2.

The result of grouped statistics according to detection methods

Batch number Group Total number Robust mean Robust SD Standard uncertainty Robust CV (%)
HA 202011 All 81 119.54 23.3 3.236 19.49
Chemiluminescence immunoassay‐AMPPD 8 97.2 26.66 12.594 27.42
Chemiluminescence immunoassay‐HRP 22 128.39 9.97 2.657 7.77
Chemiluminescence immunoassay‐Acridinium Ester 13 118.36 21.7 7.524 18.34
Chemiluminescence immunoassay‐ABEI 25 121.61 14.58 3.645 11.99
Chemiluminescence immunoassay‐Microplate 5 54.55 4.1 2.958 7.51
Up‐conversion luminescence immunoassay, ULIA 3 102.74 5.24 3.782 5.1
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 2 137.1 33.04 29.201 24.1
Latex‐enhanced immunoturbidimetry 1
Others 1
202012 All 81 303.35 66.8 9.278 22.02
Chemiluminescence immunoassay‐AMPPD 8 339.91 44.37 20.965 13.06
Chemiluminescence immunoassay‐HRP 22 343.84 20.89 5.568 6.08
Chemiluminescence immunoassay‐Acridinium Ester 13 360.37 113.51 39.353 31.5
Chemiluminescence immunoassay‐ABEI 25 288.06 23.93 5.982 8.31
Chemiluminescence immunoassay‐Microplate 5 87.69 13.33 9.618 15.2
Up‐conversion luminescence immunoassay, ULIA 3 126.98 1.44 1.04 1.14
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 2 341.9 239.6 211.774 70.08
Latex‐enhanced immunoturbidimetry 1
Others 1
LN 202011 All 81 146.94 57.03 7.921 38.81
Chemiluminescence immunoassay‐AMPPD 8 120.43 40.49 19.127 33.62
Chemiluminescence immunoassay‐HRP 22 124.14 8.98 2.394 7.24
Chemiluminescence immunoassay‐Acridinium Ester 13 115.95 57.95 20.091 49.98
Chemiluminescence immunoassay‐ABEI 25 189.48 7.04 1.761 3.72
Chemiluminescence immunoassay‐Microplate 5 131.99 17.45 12.59 13.22
Up‐conversion luminescence immunoassay, ULIA 3 81.91 1.98 1.432 2.42
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 2 326 49.72 43.942 15.25
Latex‐enhanced immunoturbidimetry 1
Others 1
202012 All 81 457.63 178.68 24.816 39.04
Chemiluminescence immunoassay‐AMPPD 8 404.71 54.62 25.804 13.5
Chemiluminescence immunoassay‐HRP 22 395.11 23.49 6.261 5.95
Chemiluminescence immunoassay‐Acridinium Ester 13 363.45 266.71 92.464 73.38
Chemiluminescence immunoassay‐ABEI 25 607.1 34.2 8.55 5.63
Chemiluminescence immunoassay‐Microplate 5 334.05 19.9 14.359 5.96
Up‐conversion luminescence immunoassay, ULIA 3 73.31 2.81 2.026 3.83
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 2 1213.85 39.53 34.942 3.26
Latex‐enhanced immunoturbidimetry 1
Others 1
PⅢNP 202011 All 83 12.36 4.98 0.683 40.28
Chemiluminescence immunoassay‐AMPPD 8 10.32 3.92 1.852 37.99
Chemiluminescence immunoassay‐HRP 22 10.86 1.49 0.397 13.73
Chemiluminescence immunoassay‐Acridinium Ester 13 4.3 1.29 0.448 30.06
Chemiluminescence immunoassay‐ABEI 25 14.57 1.08 0.269 7.38
Chemiluminescence immunoassay‐Microplate 5 156.64 31.6 22.802 20.17
Up‐conversion luminescence immunoassay, ULIA 3 61.96 4.87 3.512 7.86
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 1
Latex‐enhanced immunoturbidimetry 3 4.19 3.01 2.175 71.99
Others 1
202012 All 83 30.85 12.74 1.747 41.29
Chemiluminescence immunoassay‐AMPPD 8 23.96 9.27 4.378 38.67
Chemiluminescence immunoassay‐HRP 22 35.22 3.87 1.032 11
Chemiluminescence immunoassay‐Acridinium Ester 13 10.23 4.86 1.686 47.55
Chemiluminescence immunoassay‐ABEI 25 29.37 2.68 0.671 9.13
Chemiluminescence immunoassay‐Microplate 5 475.43 355.83 256.801 74.84
Up‐conversion luminescence immunoassay, ULIA 3 59.72 4.49 3.242 7.52
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 1
Latex‐enhanced immunoturbidimetry 3 10.23 0.69 0.5 6.77
Others 1
Col Ⅳ 202011 All 88 78.34 32.66 4.352 41.69
Chemiluminescence immunoassay‐AMPPD 8 144.65 10.95 5.173 7.57
Chemiluminescence immunoassay‐HRP 22 96.01 5.4 1.438 5.62
Chemiluminescence immunoassay‐Acridinium Ester 13 74.29 26 9.014 25
Chemiluminescence immunoassay‐ABEI 25 48.77 3.51 0.895 7.19
Chemiluminescence immunoassay‐Microplate 5 98.79 23.82 17.188 24.11
Up‐conversion luminescence immunoassay, ULIA 3 129.65 6.18 4.458 4.76
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 2 57.15 7.78 6.875 13.61
Latex‐enhanced immunoturbidimetry 8 61.82 25.69 10.704 41.55
Others 1
202012 All 88 306.87 110.22 14.686 35.92
Chemiluminescence immunoassay‐AMPPD 8 454.64 189.92 89.731 41.78
Chemiluminescence immunoassay‐HRP 22 387.81 17.78 4.739 4.59
Chemiluminescence immunoassay‐Acridinium Ester 13 282.14 71.14 24.665 25.22
Chemiluminescence immunoassay‐ABEI 25 258.77 28.29 7.219 10.93
Chemiluminescence immunoassay‐Microplate 5 426.51 40.5 29.231 9.5
Up‐conversion luminescence immunoassay, ULIA 3 327.88 9.73 7.024 2.97
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 2 100.1 13.95 12.332 13.94
Latex‐enhanced immunoturbidimetry 8 135.41 73.86 30.776 54.55
Others 1
CG 202011 All 60 2.47 0.65 0.106 26.46
Chemiluminescence immunoassay‐AMPPD 2 2.02 1.14 1.007 56.39
Chemiluminescence immunoassay‐HRP 2 2.15 0.15 0.134 7.05
Chemiluminescence immunoassay‐ABEI 15 2.7 0.2 0.067 7.46
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 1
Immunoturbidimetry 7 2.28 0.5 0.238 22.1
Latex‐enhanced immunoturbidimetry 15 2.03 0.23 0.074 11.3
Homogeneous enzyme immunoassay, HEI 16 2.46 0.53 0.167 21.72
Others 1
202012 All 60 14.12 5.9 0.952 41.8
Chemiluminescence immunoassay‐AMPPD 2 10.2 3.37 2.977 33.02
Chemiluminescence immunoassay‐HRP 2 17.16 1.24 1.099 7.25
Chemiluminescence immunoassay‐ABEI 15 12.89 0.31 0.103 2.38
Electrochemiluminescence, ECLIA 1
Radioimmunoassay, RIA 1
Immunoturbidimetry 7 14.84 8.22 3.885 55.42
Latex‐enhanced immunoturbidimetry 15 9.69 2.3 0.743 23.76
Homogeneous enzyme immunoassay, HEI 16 17.76 7.32 2.287 41.22
Others 1

If there is only one laboratory, robust standard deviation, standard uncertainty, and robust CV cannot be calculated.

3.3. Group statistics premised on detection reagents

Besides the detection techniques, the responses were classified according to the detection reagents used in view of the fact that the same detection technique can use different reagents; therefore, the classification by reagents increases accuracy. As shown in Table 3, reagents 1 and 13 represented the majority of the market share, indicating that they are frequently employed in most Chinese clinical labs for the quantitative identification of liver fibrosis biomarkers. Furthermore, the pass rates of the reagents 1 and 13 groups (100%) were higher than those of the other groups, suggesting that the larger sample size could increase the statistical accuracy. Although the reagent 1, 7, 10, 13, and 14 groups used CLIA for the detections of Col Ⅳ, LN, HA, and PⅢNP, the results varied, thereby suggesting the need for grouping based on the use of reagents and improving the commutability of quality control materials. For CG detection, reagents 1, 13, and 26 were commonly used in most Chinese clinical laboratories. Reagents 1 and 13 were used in the CLIA method, and reagent 26 was used in the HEI method. With increased sample concentration, CLIA’s robust CV decreased, whereas that of HEI was elevated, which was consistent with the findings of grouping by detection methods (Table 2).

TABLE 3.

The result of grouped statistics according to detection reagents

Batch number Group Total number Robust mean Robust SD Standard uncertainty Robust CV (%)
HA 202011 All 81 119.54 23.3 3.236 19.49
Reagent 1 22 131.94 7.81 2.131 5.92
Reagent 2 2 137.1 33.04 29.201 24.1
Reagent 5 3 102.74 5.24 3.782 5.1
Reagent 7 5 42.17 1.39 0.778 3.3
Reagent 8 3 63.91 0.8 0.709 1.25
Reagent 9 3 106.87 9.92 8.767 9.28
Reagent 10 5 101.61 5.36 2.995 5.27
Reagent 11 3 99.39 4.9 3.533 4.93
Reagent 13 25 122.45 13.49 3.372 11.01
Reagent 14 5 265.2 11.58 6.473 4.37
202012 All 81 303.35 66.8 9.278 22.02
Reagent 1 22 349.89 16.54 4.511 4.73
Reagent 2 2 341.9 239.6 211.774 70.08
Reagent 5 3 126.98 1.44 1.04 1.14
Reagent 7 5 63.93 9.96 5.567 15.58
Reagent 8 3 140.03 1.6 1.418 1.15
Reagent 9 3 230.33 33.53 29.64 14.56
Reagent 10 5 311.77 16.08 8.988 5.16
Reagent 11 3 375.08 6.61 4.773 1.76
Reagent 13 25 289.44 22.98 5.744 7.94
Reagent 14 5 705.18 42.11 23.541 5.97
LN 202011 All 81 146.94 57.03 7.921 38.81
Reagent 1 22 123.44 7.72 2.107 6.26
Reagent 2 2 326 49.72 43.942 15.25
Reagent 5 3 81.91 1.98 1.432 2.42
Reagent 7 5 127.6 2.23 1.244 1.74
Reagent 8 3 251.22 130.25 115.122 51.85
Reagent 9 3 91.58 3.39 2.998 3.7
Reagent 10 5 63.27 3.26 1.822 5.15
Reagent 11 3 87.38 1.76 1.269 2.01
Reagent 13 25 189.18 7.18 1.794 3.79
Reagent 14 5 235.67 15.85 8.86 6.73
202012 All 81 457.63 178.68 24.816 39.04
Reagent 1 22 394.83 18.76 5.118 4.75
Reagent 2 2 1213.85 39.53 34.942 3.26
Reagent 5 3 73.31 2.81 2.026 3.83
Reagent 7 5 360.99 35.81 20.016 9.92
Reagent 8 3 707.88 372.38 329.136 52.61
Reagent 9 3 266.35 3.79 3.345 1.42
Reagent 10 5 117.53 6.56 3.668 5.58
Reagent 11 3 392.16 17.59 12.697 4.49
Reagent 13 25 608.39 32.44 8.111 5.33
Reagent 14 5 624.26 25.49 14.252 4.08
PⅢNP 202011 All 83 12.36 4.98 0.683 40.28
Reagent 1 22 10.54 1.18 0.321 11.17
Reagent 2 2 233.95 26.38 23.318 11.28
Reagent 5 3 61.96 4.87 3.512 7.86
Reagent 7 5 182.71 31.28 17.488 17.12
Reagent 8 3 16.42 9.85 8.704 59.97
Reagent 9 3 10.07 0.1 0.085 0.95
Reagent 10 5 3.21 0.16 0.089 4.96
Reagent 11 3 7.36 0.4 0.287 5.39
Reagent 13 25 14.78 0.99 0.247 6.69
Reagent 14 5 4.2 0.24 0.132 5.65
Reagent 16 2 2.69 0.92 0.66 34.02
202012 All 83 30.85 12.74 1.747 41.29
Reagent 1 22 34.76 3.13 0.854 9
Reagent 2 2 1858.1 1460.67 1291.059 78.61
Reagent 5 3 59.72 4.49 3.242 7.52
Reagent 7 5 618.64 108.46 60.633 17.53
Reagent 8 3 29.43 14.28 12.623 48.53
Reagent 9 3 17.24 0.64 0.567 3.72
Reagent 10 5 6.21 0.25 0.139 3.99
Reagent 11 3 16.8 0.94 0.675 5.57
Reagent 13 25 29.38 2.14 0.536 7.29
Reagent 14 5 9.38 0.19 0.103 1.97
Reagent 16 2 9.56 0.69 0.5 7.25
Col Ⅳ 202011 All 88 78.34 32.66 4.352 41.69
Reagent 1 22 94.82 4.01 1.094 4.23
Reagent 2 2 57.15 7.78 6.875 13.61
Reagent 5 3 129.65 6.18 4.458 4.76
Reagent 7 5 100 5.45 3.048 5.45
Reagent 8 3 161.97 9.82 8.682 6.07
Reagent 9 3 70.49 23.37 20.66 33.16
Reagent 10 5 107.02 12.58 7.035 11.76
Reagent 11 3 151.01 2.14 1.546 1.42
Reagent 13 25 47.65 4.12 1.051 8.65
Reagent 14 5 56.4 4.83 2.702 8.57
Reagent 17 7 64 17.15 8.75 26.79
202012 All 88 306.87 110.22 14.686 35.92
Reagent 1 22 385.22 17.66 4.818 4.59
Reagent 2 2 100.1 13.95 12.332 13.94
Reagent 5 3 327.88 9.73 7.024 2.97
Reagent 7 5 401.93 52.77 29.498 13.13
Reagent 8 3 404.9 38.82 34.311 9.59
Reagent 9 3 200.4 44.48 39.314 22.2
Reagent 10 5 408.7 30.17 16.864 7.38
Reagent 11 3 637.21 6.36 4.593 1
Reagent 13 25 254.99 23.07 5.886 9.05
Reagent 14 5 235.62 6.72 3.758 2.85
Reagent 17 6 146.89 38.78 19.791 26.4
CG 202011 All 60 2.47 0.65 0.106 26.46
Reagent 1 5 3.44 0.71 0.395 20.58
Reagent 7 4 1.3 0.02 0.021 1.85
Reagent 13 15 2.74 0.18 0.056 6.4
Reagent 17 3 2.2 1.06 0.935 48.09
Reagent 19 2 1.88 0.29 0.255 15.37
Reagent 24 4 1.94 0.07 0.046 3.77
Reagent 25 4 4.06 0.1 0.085 2.37
Reagent 26 10 2.46 0.19 0.074 7.61
Reagent 29 3 2.1 0.13 0.095 6.24
202012 All 60 14.12 5.9 0.952 41.8
Reagent 1 5 24.02 4.17 2.333 17.37
Reagent 7 4 7.82 0.45 0.397 5.74
Reagent 13 15 12.93 0.35 0.113 2.71
Reagent 17 3 29.4 25.68 22.695 87.33
Reagent 19 2 11.96 7.1 6.272 59.36
Reagent 24 4 8.21 1.14 0.714 13.91
Reagent 25 4 27.78 2.92 2.58 10.51
Reagent 26 10 17.28 3.04 1.2 17.58
Reagent 29 3 8.71 0.54 0.392 6.24

There was only one case of reagent 3, 4, 6, 12, 15 for detection of HA, LN, PⅢNP, and Col Ⅳ, so statistical calculation could not be carried out.

There was only one case of reagent 2, 4, 12, 18, 20, 21, 22, 23, 27, 28 for detection of CG, so statistical calculation could not be carried out.

3.4. Group statistics premised on the principle of grouping

According to the statistical data shown in Tables 2 and 3, it is reasonable to hypothesize that the detection reagents and techniques were employed in their entirety. The reagent produced by a particular manufacturer is typically matched to the detection technique used by that manufacturer. Regardless of the data for grouping based on equipment, procedures, or reagents, grouping was conducted according to the number of labs involved: greater than or equal to 18 or 12 were in one group under ISO 13528. The “other” category was formed if they did not fit into any of the other groupings. According to ISO 13528, the findings were examined using the robust mean, robust standard deviation, and robust CV of the involved labs. However, owing to the limited number of participating labs, the grouping criterion was unsuitable. Therefore, the participating laboratories were classified into separate groups based on the number of laboratories using the same reagent ≥ 5 (Table 4). For the detection of PⅢNP, the lowest values were obtained using reagent 7 and the highest using reagent 14. However, for the detection of LN, the results obtained on using reagent 7 and reagent 14 were opposite. Furthermore, reagents 7 and 14 were both used in the CLIA method but labeled with different substrates. This can be attributed to the different sensitivity of the various reagents for the detection items. When the EQA of the quantitative detection of liver fibrosis biomarkers is properly performed, the variability of the results may be minimized by expanding the number of labs that were included in the present research.

TABLE 4.

The result and passing rates of grouped statistics according to the grouping principle

Batch number Group Total number Robust mean Robust SD Standard uncertainty Robust CV Total number Number of passing labs Pass rate (%)
HA 202011 All 81 119.54 23.3 3.236 19.49 19 14 70
Reagent 1 22 131.94 7.812 2.131 5.92 22 21 100
Reagent 7 5 42.17 1.391 0.778 3.3 5 5 100
Reagent 10 5 101.61 5.358 2.995 5.27 5 5 100
Reagent 13 25 122.45 13.486 3.372 11.01 25 24 96
Reagent 14 5 265.2 11.58 6.473 4.37 5 5 100
202012 All 81 303.35 66.799 9.278 22.02 19 10 50
Reagent 1 22 349.88 16.536 4.511 4.73 22 21 100
Reagent 7 5 63.93 9.959 5.567 15.58 5 4 80
Reagent 10 5 311.77 16.079 8.988 5.16 5 5 100
Reagent 13 25 289.44 22.975 5.744 7.94 25 25 100
Reagent 14 5 705.18 42.111 23.541 5.97 5 5 100
LN 202011 All 81 146.94 57.032 7.921 38.81 19 6 30
Reagent 1 22 123.44 7.723 2.107 6.26 22 21 100
Reagent 7 5 127.6 2.226 1.244 1.74 5 5 100
Reagent 10 5 63.27 3.26 1.822 5.15 5 5 100
Reagent 13 25 189.18 7.175 1.794 3.79 25 25 100
Reagent 14 5 235.67 15.85 8.86 6.73 5 5 100
202012 All 81 457.63 178.677 24.816 39.04 19 8 40
Reagent 1 22 394.83 18.764 5.118 4.75 22 21 100
Reagent 7 5 360.99 35.805 20.016 9.92 5 5 100
Reagent 10 5 117.53 6.562 3.668 5.58 5 5 100
Reagent 13 25 608.39 32.442 8.111 5.33 25 25 100
Reagent 14 5 624.26 25.494 14.252 4.08 5 5 100
PⅢNP 202011 All 83 12.36 4.978 0.683 40.27 21 7 31.8
Reagent 1 21 10.54 1.178 0.321 11.17 22 20 95.2
Reagent 7 5 182.71 31.284 17.488 17.12 5 5 100
Reagent 10 5 3.21 0.159 0.089 4.96 5 5 100
Reagent 13 25 14.78 0.989 0.247 6.69 25 25 100
Reagent 14 5 4.2 0.237 0.132 5.65 5 4 80
202012 All 83 30.85 12.736 1.747 41.29 21 4 18.2
Reagent 1 21 34.76 3.129 0.854 9 22 21 100
Reagent 7 5 618.64 108.464 60.633 17.53 5 5 100
Reagent 10 5 6.21 0.248 0.139 3.99 5 5 100
Reagent 13 25 29.38 2.142 0.536 7.29 25 20 80
Reagent 14 5 9.38 0.185 0.103 1.97 5 4 80
Col Ⅳ 202011 All 88 78.33 32.66 4.352 41.69 22 8 36.4
Reagent 1 22 94.82 4.011 1.094 4.23 22 21 100
Reagent 7 5 100 5.452 3.048 5.45 6 5 83.3
Reagent 10 5 107.02 12.584 7.035 11.76 5 5 100
Reagent 13 25 47.65 4.121 1.051 8.65 5 5 100
Reagent 14 5 56.4 4.834 2.702 8.57 25 23 95.8
Reagent 17 6 64 17.146 8.75 26.79 5 5 100
202012 All 88 306.87 110.217 14.686 35.92 22 10 45.5
Reagent 1 22 385.22 17.664 4.818 4.59 22 21 100
Reagent 7 5 401.93 52.768 29.498 13.13 6 4 66.7
Reagent 10 5 408.7 30.168 16.864 7.38 5 5 100
Reagent 13 25 254.99 23.069 5.886 9.05 5 5 100
Reagent 14 5 235.62 6.722 3.758 2.85 25 24 100
Reagent 17 6 146.89 38.783 19.791 26.4 5 5 100
CG 202011 All 60 2.47 0.654 0.106 26.46 30 19 63.3
Reagent 1 5 3.44 0.707 0.395 20.58 5 4 80
Reagent 13 15 2.74 0.175 0.056 6.4 15 15 100
Reagent 26 10 2.46 0.187 0.074 7.61 10 10 100
202012 All 60 14.12 5.9 0.952 41.8 30 8 26.7
Reagent 1 5 24.02 4.173 2.333 17.37 5 5 100
Reagent 13 15 12.93 0.35 0.113 2.71 15 15 100
Reagent 26 10 17.28 3.037 1.2 17.58 10 9 90

All included labs were divided into groups and subjected to analyses according to the reagents they employed, with the findings being reviewed in relation to the target value ±30%. The desired outcome is a robust average value, with a dispersion of ±30% being the maximum permitted for the degree of dispersion of the data. A satisfactory result was judged to be achieved if the detection value fell within a certain range (Table 4). In the 202011 sample detection, the pass rates of HA, LN, PⅢNP, Col Ⅳ, and CG in the “other group” were 70%, 30%, 31.8%, 36.45%, and 63.3%, whereas those of reagents 1, 7, 10, 13, and 14 were nearly 100%. Therefore, if laboratories using the same reagent can be individually grouped, the pass rates are relatively high. This situation can also be seen in the detection of high‐concentration samples. The pass rate of the separate group was remarkably higher than that of the mixed group. The coefficient of variation for HA, LN, PⅢNP, Col Ⅳ, and CG in different laboratories were 3.3%–19.49%, 1.74%–38.81%, 1.97%–41.29%, 2.85%–41.69%, and 2.71%–41.8%, respectively. Generally, the lower CV appeared in the separate group, while the higher CV appeared in the mixed group. Therefore, these results signify the importance of grouping. With the increase in concentration, the pass rates of HA, LN, PⅢNP, Col Ⅳ, and CG, apart from the other group, showed no significant difference. The pass rates of HA, PⅢNP, and CG decreased with increasing concentrations, whereas that of LN and Col Ⅳ increased but not by much.

3.5. Total pass rates

After calculating the pass rate of each detection project, the total pass rates of all projects were calculated. The total pass rates of HA, LN, PⅢNP, Col Ⅳ, and CG were 84.0%, 80.2%, 67.5%, 77.3%, and 58.3%, respectively (ie pass for both low and high concentrations). Therefore, the pass rates of PⅢNP, Col Ⅳ, and CG need improvement (Table 5).

TABLE 5.

The total passing rates

Project Total number Number of passing laboratories Pass rate (%)
HA 81 68 84.0
LN 81 65 80.2
PⅢNP 83 56 67.5
Col Ⅳ 88 68 77.3
CG 60 35 58.3

4. DISCUSSION

EQA (referred to as “proficiency testing”) is employed for the purpose of assessing the capacity of a laboratory to perform tests. It is an essential tool for identifying problems in a clinical laboratory and designing the appropriate solutions. EQA is an essential external monitoring technique for quality assurance, especially with the lack of a reference method and reference material. Several novel protein detection markers have recently been introduced into the clinical testing environment. This contributes to China's fast expansion of the medical diagnostic industry; however, it also highlights the significance of conducting the necessary interlaboratory quality assessment method. Therefore, the NCCL carried out an EQA survey for the quantitative identification of liver fibrosis biomarkers, which contributes to the preparation for a formal EQA program for the identification of liver fibrosis indicators.

There are three accepted methods for determining the evaluation criteria of EQA. The first method is based on clinical use‐value, which is based on the experience of clinicians (eg glycosylated hemoglobin). The second method is based on biological variation comprising a few indices that have been studied and can be found on the Westgard (https://www.westgard.com/minimum‐biodatabase1.htm) and EFLM (https://biologicalvariation.eu/) websites (eg cholesterol). The third method is based on existing detection technology. 14 , 15 , 16 , 17 , 18  The detection of liver fibrosis biomarkers uses the third method. For the data collected, the pass rates according to the target values of ±10%, 15%, 20%, 25%, and 30% were calculated. These results help to determine the appropriate criteria for existing detection capabilities.

According to the results of the EQA study conducted in China, the CLIA technique accounted for the greatest proportion of the quantitative detection of HA, LN, PⅢNP, and Col Ⅳ, representing around 90%, whereas, for the identification of CG, the CLIA, HEI, and latex‐enhanced immunoturbidimetry methods accounted for approximately one‐third each (Table 1). This suggested that despite the fact that there are multiple detection techniques, they are considerably concentrated. On analyzing the reported data, the results revealed significant differences based on the detection techniques employed, even in different orders of magnitude (Table 2). Even with the similar detection techniques but varied reagents, the detection values varied 3–100‐fold (Tables 3 and 4; reagents 1, 7, 10, 13, and 14). Although using a similar reagent and procedure, the numerical variance across various labs differed by more than twofold. These differences could be attributed to the following reasons. The reagents used were different, which adopt different methodologies (some manufacturers use the competition method, whereas others use the sandwich method). Additionally, the binding sites and affinity for quality control products along with sensitivity differ as different reagent manufacturers select different antibodies. The nature of the raw materials of quality control products (natural antigen or recombinant antigen) also contributes to the varied results. The quality control products employed in the present research were obtained from human serum, which might include a number of different antibody proteins in addition to the indicators for liver fibrosis. Hence, it is reasonable to hypothesize that the significant disparities in test findings across labs are due to changes in reagents, procedures, and characteristics, such as anti‐interference effects and specificity. In accordance with the existing regulations for medical device registration, the detection system that is formed of the reagents and their calibration products provided by the reagent manufacturers, as well as the “suitable equipment” listed on the kit's specifications, can serve as a supporting system, which may be defined as the “open” supporting system. The reasons for the substantial differences in outcomes between the “closed” and “open” matching methods are complicated. The development of various analytical systems, such as the data reading techniques, configuration of absorbance wavelengths, and built‐in calibrations, could be a possibility. Besides, plenty of invalid data were removed (such as wrong filling, missing file, and unit error) when analyzing the results, suggesting that the data reporting behavior of laboratory staff needs to be standardized. Therefore, the human factor in the data submission process cannot be ignored. According to the present EQA data, it is impossible to isolate a specific factor.

A large number of labs engaged in the quantitative identification of liver fibrosis biomarkers in China. Although the pass rate for HA and LN was relatively high in this survey, it does not suggest that the quality of this marker satisfied the clinical standards (Tables 4 and 5) in view of the fact that the evaluation criterion of target value set at ±30% is not very strict. At the same time, we found that the pass rate of the individual group (grouped according to ISO 13528) was substantially greater than the one of the other groups. This not only shows the necessity for grouping but also suggests that reagents with a small market share are at a disadvantage in grouping (Table 4). In addition to objective factors, many subjective factors also affect the results including data input error, reagent or method selection error, unit error, incorrect sample sequence use, internal quality control, and personnel operation error. All these factors restrict the accuracy of EQA results.

5. CONCLUSION

The present research is the first report to reveal the current state of the quantitative identification of liver fibrosis biomarkers in China, as far as we know. Despite the small number of labs involved in the present research, our findings indicate that there are a few drawbacks in the quantitative detection of liver fibrosis biomarkers. Commercially accessible quantitative detection kits need methodological study, and quality control procedures must be improved for the purpose of better understanding the reasons for differences in detection findings. Overall, the present research establishes the basis for the development of a formal EQA for the quantitative identification of liver fibrosis biomarkers. At the same time, it also provides guidance for the selection of instruments, methods, and reagents for laboratory liver fibrosis marker detection.

AUTHOR CONTRIBUTIONS

Chao Zhang was involved in acquisition and analysis of data. Chao Zhang and Chuanbao Zhang participated in management of data. Chao Zhang designed the study and drafted the manuscript.

CONFLICT OF INTEREST

The authors declare no competing interests.

Supporting information

Table S1‐S2

ACKNOWLEDGMENTS

We thank Autobio for providing quality control products.

Zhang C, Zhang C. Analysis of current status of quantitative detection of biomarkers for liver fibrosis in Clinical labs in China. J Clin Lab Anal. 2022;36:e24490. doi: 10.1002/jcla.24490

Funding information

This work was supported by the National Natural Science Foundation of China (no. 82003809) and Beijing Hospital Nova Project (BJ‐2020‐087).

Contributor Information

Chao Zhang, Email: cbzhang@nccl.org.cn.

Chuanbao Zhang, Email: zc_mdy@163.com, Email: cbzhang@nccl.org.cn.

DATA AVAILABILITY STATEMENT

Not applicable.

REFERENCES

  • 1. Adams LA. Biomarkers of liver fibrosis. J Gastroenterol Hepatol. 2011;26:802‐809. [DOI] [PubMed] [Google Scholar]
  • 2. Caballeria L, Toran P, Caballeria J. Markers of hepatic fibrosis. Med Clin (Barc). 2018;150:310‐316. [DOI] [PubMed] [Google Scholar]
  • 3. Kisseleva T, Brenner D. Molecular and cellular mechanisms of liver fibrosis and its regression. Nat Rev Gastroenterol Hepatol. 2021;18:151‐166. [DOI] [PubMed] [Google Scholar]
  • 4. Dhar D, Baglieri J, Kisseleva T, Brenner DA. Mechanisms of liver fibrosis and its role in liver cancer. Exp Biol Med (Maywood). 2020;245:96‐108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Masuzaki R, Kanda T, Sasaki R, et al. Noninvasive assessment of liver fibrosis: current and future clinical and molecular perspectives. Int J Mol Sci. 2020;21(14):4906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Tong S, Revill P. Overview of hepatitis B viral replication and genetic variability. J Hepatol. 2016;64:S4‐S16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Tang LSY, Covert E, Wilson E, Kottilil S. Chronic hepatitis B infection: a review. JAMA. 2018;319:1802‐1813. [DOI] [PubMed] [Google Scholar]
  • 8. El‐Mezayen HA, Habib S, Marzok HF, Saad MH. Diagnostic performance of collagen IV and laminin for the prediction of fibrosis and cirrhosis in chronic hepatitis C patients: a multicenter study. Eur J Gastroenterol Hepatol. 2015;27:378‐385. [DOI] [PubMed] [Google Scholar]
  • 9. Enomoto H, Bando Y, Nakamura H, Nishiguchi S, Koga M. Liver fibrosis markers of nonalcoholic steatohepatitis. World J Gastroenterol. 2015;21:7427‐7435. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Neuman MG, Cohen LB, Nanau RM. Hyaluronic acid as a non‐invasive biomarker of liver fibrosis. Clin Biochem. 2016;49:302‐315. [DOI] [PubMed] [Google Scholar]
  • 11. Mak KM, Mei R. Basement membrane type IV collagen and laminin: an overview of their biology and value as fibrosis biomarkers of liver disease. Anat Rec (Hoboken). 2017;300:1371‐1390. [DOI] [PubMed] [Google Scholar]
  • 12. Adams LA, Wang Z, Liddle C, et al. Bile acids associate with specific gut microbiota, low‐level alcohol consumption and liver fibrosis in patients with non‐alcoholic fatty liver disease. Liver Int. 2020;40:1356‐1365. [DOI] [PubMed] [Google Scholar]
  • 13. Loomba R, Adams LA. Advances in non‐invasive assessment of hepatic fibrosis. Gut. 2020;69:1343‐1352. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Kenny D, Fraser CG, Hyltoft Petersen P, Kallner A. Consensus agreement. Scand J Clin Lab Invest. 1999;59(7):585. [PubMed] [Google Scholar]
  • 15. Kallner A, McQueen M, Heuck C. The Stockholm Consensus Conference on quality specifications in laboratory medicine, 25‐26 April 1999. Scand J Clin Lab Invest. 1999;59(7):475. [DOI] [PubMed] [Google Scholar]
  • 16. Ceriotti F, Fernandez‐Calle P, Klee GG, et al. Criteria for assigning laboratory measurands to models for analytical performance specifications defined in the 1st EFLM Strategic Conference. Clin Chem Lab Med. 2017;55:189‐194. [DOI] [PubMed] [Google Scholar]
  • 17. Kallner A. Quality specifications based on the uncertainty of measurement. Scand J Clin Lab Invest. 1999;59:513‐516. [DOI] [PubMed] [Google Scholar]
  • 18. Sandberg S, Thue G. Quality specifications derived from objective analyses based upon clinical needs. Scand J Clin Lab Invest. 1999;59:531‐534. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1‐S2

Data Availability Statement

Not applicable.


Articles from Journal of Clinical Laboratory Analysis are provided here courtesy of Wiley

RESOURCES