Vector spherical harmonic analysis and experimental validation of spherical shells illuminated with broadband, millimeter wave Gaussian beams: applications to corneal sensing

Faezeh Zarrinkhat 1,2,*, Joel Lamberg 2, Aleksi Tamminen 2, Mariangela Baggio 2, Irina Nefedova 2, Juha Ala-Laurinaho 2, Elsayed E M Khaled 3,4, Juan Rius 1, Jordi Romeu 1, Zachary Taylor 2
PMCID: PMC9352287  PMID: 35991916

Abstract

Coupling to longitudinal modes of thin spherical shells, under Gaussian-beam illumination, was explored with a theoretical method based on Fourier-optics analysis and vector spherical harmonics and was scrutinized with an experimental setup. For the theory part, the illumination frequency band was fixed between 100–600 GHz and the outer spherical shell radius of curvature and thickness are 7.5 mm and 0.5 mm, respectively. The shell material was either the lossless cornea or an aqueous effective media representing the cornea. Six different beam-target strategies were introduced being potential candidates for maximum coupling. Two dispersion-tuned beam ensembles with strongly frequency-dependent phase center location have been created with a fixed incident beam 1/e radius and radius of curvature called forward strategies. These computations of different alignments were continued with four beam ensembles of frequency-invariant phase center, constructed from fits to experimental data, oriented at four different axial locations with respect to the spherical shell center of curvature, they are called reverse strategies. Coupling efficiency for all strategies was calculated for different targets including perfect electrical conductor (PEC) sphere, PEC core covered by a cornea loss-free layer and cornea. All scattering strategies contrasted to scattering from equivalent planar targets as a reference with maximum coupling. The results show that, under an ideal calibration, forward strategies are a closer approximation to the plane-wave condition for the cornea. An experimental setup was assembled to explore the simulation approach in a frequency range between 220 GHz to 330 GHz. Two different quartz samples with permittivity of 4.1 were mounted on a water core, acting for a cornea. The first and second quartz radius and thickness were 7.5 mm and 0.5 mm and 8 mm and 1 mm, respectively. An adequate agreement between theory and experiment was confirmed. A particle optimisation swarm algorithm was applied to extract the thickness and permittivity of quartz from the measured back-scattered field for reverse strategies.

1. Introduction

The sub-millimeter wave and THz frequency sensing of the cornea leverages the layered tissue structure for corneal water content and thickness quantification. Changes in corneal tissue water content (CTWC) and corneal central thickness (CCT) are correlated with human eye diseases and disorders. Existing clinical measurement approaches are not sufficiently accurate for the early detection of these conditions. A promising method for non-invasive, in situ tissue characterization, is THz spectroscopy. The cornea is  0.6 mm thick and is bounded by air and an optically thick body of water on its anterior (outer surface ) and posterior (bottom inner) surface, respectively. At THz frequencies, this structure presents as a lossy thin-film lying atop a lossy half-space. Resolution of the cornea’s lossy longitudinal modes via frequency-domain reflectometry in a band sufficiently low (e.g. 220 GHz - 330 GHz) for significant penetration allows simultaneous estimates of CTWC and CCT. However, since the cornea is spherical, an efficient coupling to longitudinal modes nominally requires normal incidence across the interrogated area and thus a converging spherical phase front whose curvature matches the corneal surface curvature [13]. However, as the average corneal radius of curvature (RoC) is 7.5 mm and the midband free-space wavelength at 100 GHz - 600 GHz is 1 mm, approximate phase-front matching occurs near the beam confocal point where the distinctly non-spherical phase-front curvature is rapidly changing and leads to low longitudinal modes coupling.

A typical, non-contact measurement system block diagram is shown in Fig. 1. A transceiver optical co-locates a source and detector to enable normal incidence illumination and reception. The output beam is collimated and focused onto the cornea and back-scattered reflection is referenced to some calibration standard reflection to compute broadband, complex sample reflectivity. Fits between the measured data and a model are used to estimate water content or water content gradients. Additionally, spectra can be used to estimate thickness and axially varying hydration simultaneously.

Fig. 1.

Fig. 1.

Scheme of the problem. Modelling cornea as a planar layered structure under plane wave illumination as well as a spherical layered structure under Gaussian illumination is shown.

The nominal output beam profile of a THz system is TEM00 Gaussian and different strategies regarding the placement of focused beam waist with respect to corneal RoC have been evaluated [4]. Restated from above, while the Gaussian beam phase front radius of curvature can match the cornea RoC on-axis, there is no configuration that achieves perfect matching across the field of view. Nevertheless, spectra are almost exclusively analyzed with Fresnel’s equations at normal incidence informed by effective media permittivities to model depth-dependent water content and stratified media theory to model the depth-dependent and thus provide an aggregate reflection coefficient of the stack. Implicit in this analysis is a planar thin film on semi-infinite planar half-space, with plane wave illumination. The radius of curvature mismatch at the cornea’s air-anterior interface and posterior-aqueous humour interface and resulting walk-off losses within the cornea itself cannot be accommodated.

In summary, THz spectroscopy of cornea is performed via focused Gaussian beam illumination on lossy hemispherical shell enclosing a lossy homogenous aqueous core while analysis assumes planar illumination on a planar, infinite transverse extent lossy thin film on top of a semi-infinite aqueous half space. The above-listed constraints raise two interrelated questions about obtaining maximum coupling: ( 1 ) where should the corneal center of curvature (CoC) be positioned relative to beam geometry? ( 2 ) Is there a frequency dependence on Gaussian beam parameters that further helps to maximize coupling? More succinctly, given a set of constraints on the beam, how does one maximize coupling to the cornea’s sub-millimeter wave longitudinal modes? It’s worthy to say maximum coupling is considered as convergence to plane-wave on planar surface coupling due to the perfect match of plane wave’s phase-front with the planar surface.

To best address, the above questions, a method based on Fourier optics (FO) and vector spherical harmonics (VSH) is used to computationally explore longitudinal coupling in structures resembling cornea. The VSH is the vector solution of the wave equation in spherical coordinates. In 1908, Gustave Mie was one of the pioneers who used VHS to address the incident and scattered field from a sphere illuminated by a plane wave [5]. Later, Davis modified the theory for Gaussian beam illumination [6]. The early proposed method was quite intricate and cumbersome. In 1993, Khaled et al. [7] presented a method using Fourier analysis [8] to model Gaussian beam as the angular spectrum of plane waves, and for computing scattering coefficients, they employed the T-matrix method [9].

In fact, the T-matrix relates the scattering coefficients of a target illuminated by a plane wave with the scattering coefficients of the target illuminated by a Gaussian beam. To compute the scattering coefficients of layered sphere under the plane wave illumination many approaches were presented [10,11,12,13], here Yang [13] algorithm was employed. This classical method will serve as a new application in our paper allowing us to scrutinize the cornea as a multilayered spherical structure. FO allows for a closed-form expression of the steady-state scattering solution (multiple reflections within the corneal layer) while VSH provides a convenient representation of the fields with respect to the target geometry. The goals of the paper are to address different beam target strategies which are potential candidates for maximum coupling and investigate these strategies theoretically and experimentally for spherical targets. Additionally explored was the efficiency of using a metal sphere as a calibrator [14] in an experimental set-up.

2. Background and Gaussian-beam analysis

At this point, the constraints on the illuminated beam parameters are explored, addressing the questions in the introduction, whether there is a frequency dependence on Gaussian-beam parameters assisting coupling enhancement. Also, to determine which beam-cornea alignment are the most likely optimal strategies to explore.

2.1. Cornea as a spherical scatterer and spherical cavity

Interestingly, a cornea can be considered as a scatterer in a wide frequency range. A plot of the size parameter range for cornea in the THz band is shown in Fig. 2(a) with wavelength along the horizontal axis and particle radius along the vertical axis. The family of oblique lines is parameterized by a fixed size parameter, ka=2πn0/λ , where a , k , λ/n are respectively, the radius, the wavenumber, and the wavelength in the medium in which the particle resides ( n0=1 for air refractive index). The fixed size parameter contours of 0.0002 , 0.2 , and 2000 serve as approximate thresholds between the varying scattering regimes. The Mie scattering is a proper candidate to describe the scattering behavior of particles with size parameters lying in the range 0.2<ka<2000 .

Fig. 2.

Fig. 2.

(a) Cornea scattering region, (b) stability analysis of the cornea as a spherical cavity, coupling efficiency computed by ABCD matrix for beam illumination (c) on the apex, and (d) on the center of the cornea.

The vacuum wavelength range 3 mm - 0.3 mm corresponding to the 100 GHz - 1 THz band is indicated by the gray shaded vertical box. The nominal range for human cornea RoC, 7 mm - 8 mm is denoted by the thin purple horizontal shaded rectangle. Its intersection reports the approximate corneal size parameter range subtended by the anterior segment. The size parameter of the posterior segment was estimated by subtracting the nominal CCT range from the anterior segment RoC and computing the wavelength in the complex aqueous corneal medium. Bruggeman’s effective media theory [15] is used to compute the aggregate permittivity of a mixture of 60% collagen and 40% free water.

These intracorneal size parameters are labeled "anterior segment" and "posterior segment" in Fig. 2(a). The dispersion arising from the liquid water component is evident in the curvature of the shaded region. At 100 GHz, both the anterior and posterior segments lie comfortably in the Mie scattering regime. At 1 THz, the posterior segment is quite close to the approximate geometric scattering limit and, depending on the illumination profile, indicating maybe GO is sufficient to approximate the expected back-scatter.

Moreover, the cornea can be treated as a spherical cavity. It implies certain constraints on the beam. Only particular ranges of the cavity outer radius R1 , inner radius R2 , and the distance between them L , produce stable resonators. An unstable cavity will increase the beam size without limitation, consequently, it will get larger than the cavity size and will be lost completely.

A stability analysis of the cornea for nominal RoCs and thicknesses are displayed in Fig. 2(b) where g1=1L/R1=0.9333 and g2=1L/R2=0.9286 . The unstable region g1g2>1 is indicated by the shaded region and contours of constant g1g2 by the dotted lines. For clarity, both RoC sign pairs are plotted; ( +Rc,Rc+t ) and ( Rc,+Rct ). In a region about the corneal apex, the anterior and posterior surfaces are concentric and thus g1g2=1 for all combinations of Rc and tc . The corneal thickness increases towards the periphery which can be described as an anterior segment RoC increase with respect to the posterior segment RoC and thus a divergence from concentricity. This combination is unstable and is represented by the overlap between the g1g2 curves and shaded unstable region.

Thus, the corneal cavity is susceptible to beam walk-off which reduces the interference between the primary reflection at the air/cornea interface and multiple reflections from paths through the cornea. Two examples of beam walk-off are shown in Fig. 2(c) using ABCD matrix computations [16]. The cornea was modeled via the effective media theory described above but only the real part of the permittivity was applied. Ray transfer matrices were used to compute the parameters of the beam scattered from the anterior segment and then the beams scattered after one, two, and three round trips through the cornea. The plots show the coupling coefficients (Eq. (11), [4]) as a function of frequency between the primary reflection beam and the round trip beams. The superior coupling of the RoC-matched beams compared to the incident waist radius beams suggests that the phase front matching will enhance interference and thus aid in the coupling and extraction of longitudinal modes.

2.2. Gaussian-beam analysis

The problem geometry definitions are displayed in Fig. 3. A converging beam is an incident, from the left, on the cornea. The incident plane is labeled plane 1 and the incident beam transverse radius and RoC are denoted ω1 and R1 , respectively. The beam waist is labeled ω0 and located at plane 0 where the parameters of plane 0 and plane 1 are related via free-space propagation in the absence of the cornea. The physical distance between plane 0 and plane 1 is labeled d01 and the distance between the corneal CoC and plane 0 is z0 .

Fig. 3.

Fig. 3.

Geometry of the problem, cornea modelled as a layered sphere under Gaussian beam illumination by two different beam ensemble definitions including forward and reverse strategies.

Two beam ensemble definitions were considered where the incident beam RoC was fixed such that R1=Rc frequencies. The first was termed "forward" (shown in Fig. 3) where ω1 was set at some constant value and then ray transfer matrices and complex beam parameters were used to compute d01 and ω0 as a function of frequency:

d01(λ)=Rc(πω12)2(πω12)2+(λRc)2, (1)
ω0(λ,d01)=((Rc+d01)πω12)2+(λd01Rc)2(πω1Rc)2. (2)

The second approach was termed "reverse" or "varied confocal distance" (shown in Fig. 3) where ω0 as a function of wavelength is defined and the standard Gaussian beam equation for axially dependent RoC was used to find d01 and then ω1 . Note that d01 is the solution to a second order polynomial thus two solutions are possible. These are labeled d01,NF and d01,FF in Eq. (3) and (4), respectively where the d01,FF>d01,NF for zc<Rc/2 . The iF subscript stands for i=N,F indicating FF far-field and NF near-field, respectively:

d01,iF(λ)=Rc±Rc24zc22, (3)
ω1,iF(λ,d01,iF)=ω01+d01,iF2/zc2. (4)

The d01 and ω0 in the forward direction over a range of ω1 are reported in Fig. 4(a,b) for 100 GHz - 1 THz. In both plots, the ( f , ω1 ) pairs that produce waist radii that violate the paraxial limit ω0<λ/2 are eliminated. Further (f, ω1 ) pairs that produce d01<Zc are denoted. Input beam radii ω1=2.1 mm and 3.1 mm are denoted by the white dotted lines and are considered in the next section.

Fig. 4.

Fig. 4.

(a) Forward d01 and (b) forward ω0 over a range of ω1 is plotted. (c) Reverse d01,NF and (d) reverse ω1,NF over a range of ω0 is plotted.

Parameters spaces for d01,NF and ω1,NF computed in the reverse direction are displayed in Fig. 4(c,d) with the ( f , ω0 ) pairs in violation of the indicated paraxial approximation. The large area in the upper right-hand corner of the plot correspond to ( f , ω0 ) pairs where Zc>Rc/2 and thus complex valued d01 has been set aside. Inspection of Eq. (4) shows that the contour lines in the d01,NF plot of Fig. 4 correspond to contours of constant zc . In other words, if one defines ω0 such that Zc is invariant to frequency then, as evidenced by Eq. (3), d01,NF and d01,FF are invariant to frequency and thus the beam RoC evolves equally along the axis for all spectral components.

2.3. Gaussian-beam strategies

According to previous discussions, the final candidate strategies are displayed in Fig. 5 where the left column shows the evolution of the beam radius and beam waist location with respect to the corneal geometry and the middle column shows the evolution of the beam RoC superimposed on the corneal anterior and poster surface locations. The right column reports the spectral dependence of beam waist ω0 , incident beam radius ω1 , and beam waist location Z0 .

Fig. 5.

Fig. 5.

Definition and orientation of Gaussian beams and their relationship to corneal geometry for (a-c) S1 , (d-f) S2 , (g-i) S3 , (j-l) S4 , (m-o) S5 and (p-r) S6 strategies.

Strategy 1 ( S1 ) fixes the input beam radius to ω1=2.1 mm yielding a diameter similar to that of modern ultrasound pachymeter probes [17]. Additionally, S1 allows analysis of illumination where the low-frequency incidence occurs in the near-field region of the beam and the high-frequency incidence in the far-field. This is evident by the black trace denoting the confocal point location (Fig. 5) and requires significant dispersion visible by the beam waist locations reported in the left and right columns.

The beams in strategy 2 ( S2 ) were also defined in the forward direction with ω1=3.1 mm fixed for all frequency. This ensemble locates the incidence location in the far-field region for all beams but is still sufficiently small to avoid ω0<λ/2 . The S2 beam waists are slightly smaller than S1 but the dispersion (variation in Z0 ) is significantly reduced.

The reverse analysis was utilized for strategies 3 - 6 ( S3 - S6 ) with the confocal distance fixed at zc=2.62 mm. This confocal distance is below the Rc/2 threshold and yields physically realizable, paraxial approximation compatible, providing a set of beams for the frequency range 100 GHz - 1 THz.

The S4 places the phase front match in the super confocal (far-field) region of the beam as evidence by the beam overlapping beam RoC plots in Fig. 5. The beam waists are collocated and the waist radii are all larger than λ/2 although they approach the limit at 100 GHz. The S3 is the same beam ensemble as S4 but places the phase front match in the subconfocal (near field) region of the beam. The beam radius on the cornea is significantly smaller ( ω1ω0 ) which should reduce phase front mismatch error in the beam periphery but the beam RoC magnitude is rapidly increasing for increasing z instead of decreasing suggesting a substantial mismatch at the posterior surface.

Strategies 5 and 6 ( S5 , S6 ) were evaluated for comparison to two strategies commonly reported in the literature. The S5 places the beam waists at the corneal CoC. This is the typical arrangement for imaging via a Gaussian beam telescope optical train and produces phase fronts that are slightly larger in RoC than the corneal RoC. The S6 places the beam waist at the corneal apex thus mimicking the common approach reported by many groups using THz time-domain spectroscopy. The beam radius on target is minimized at the cost of a significant RoC mismatch. These strategies are labeled "reference" throughout this work and are iterations of the reverse strategies.

3. Angular spectrum based analysis

An approach based on Angular Spectrum, for calculating the coupling efficiency of sub-millimeter wave illumination on the cornea, was presented in [18] which utilized the methodology described in [7]. This method is applicable for any Gaussian beam incident on a spherical surface and is valid for spheres with RoC in order, or larger than the illumination wavelength. The advantage of this method compared to PO and full-wave approaches are the ability to solve directly the steady state scattering of both homogeneous and coated spheres. This ability significantly reduces computation time as many issues such as "ray splitting" at dielectric interfaces and finely discretized surface current densities can be avoided. Thus, it is possible to explore several illumination strategies on the scattering profile of the cornea when it is modeled as a spherical shell encasing a lossy dielectric sphere.

The applied approach first decomposes the Gaussian beam to its plane-wave spectrum representation and then expresses these components as VSH [7]. The incident Gaussian beam is represented as:

Ei=mnD[aeMe1+aoMo1+beNe1+boNo1], (5)

where ae , ao , be , and bo are the incident field coefficients (addressed in appendix A) and the VSH of the first kind M1 and N1 are the vector solution of the wave equation [10]. The parameter D=ϵm(2n+1)(nm)!4n(n+1)(n+m)! is a normalization factor and ϵm=1 for m=0 and ϵm=2 for m>0 . The radial mode number and azimuthal modes are represented by n and m , respectively, and vary over the range m=0:Nstop , also n=m:Nstop+1 ( Nstop is addressed in Appendix B). For any arbitrary beam-target alignment, the scattered field from the spherical targets can be written [7] as:

Es=mnD[feMe3+foMo3+geNe3+goNo3], (6)

where fe , fo , ge , and go are scattered field expansion coefficients and M3 and N3 [11] are the VSH of the third kind. The scattered fields expansion coefficients are functions of the incident beam coefficients and are calculated by the T-matrix method as [19]:

[fefogego]=[T110000T220000T330000T44][aeaobebo], (7)

where, T is a diagonal matrix and for the case of a coated sphere, its elements are obtained from the scattering coefficients of a coated sphere under plane-wave illumination ( anc and bnc ) [19]:

T11=T22=anc,T33=T44=bnc. (8)

For the calculation of coefficients anc and bnc , Khaled in [7] and [19] utilized the algorithm introduced by Toon and Ackerman [10]. In this work, the algorithm employed by Yang [12,13] was applied to the coefficients calculations. The equations for the algorithm is addressed in Appendix B. Electromagnetic analysis for a layered sphere with more than one layer is the advantage of the Yang algorithm.

Figure 6 shows the field distribution of incident Gaussian beam and back-scattered field from cornea computed with 5 and 6, in a plane perpendicular to the optical axis. The plane dimension is 20 mm by 20 mm. The source beam was polarized in the x - z direction. The cornea was modeled as a single-layered spherical shell encapsulating a homogeneous, pure-water sphere. The water permittivity was obtained by the double-Debye model [20], and the shell (cornea) modeled with the effective medium theory via the Bruggeman mixing model [15]. The corneal shell consisted of 60 % water and 40 % collagen with dispersion-free and real permittivity 2.9 . Core radius was 7.5 mm and shell thickness was 0.5 mm.

Fig. 6.

Fig. 6.

Field distribution of source and back-scattered field of the cornea is simulated with Eqs. (5) and (6) at 220 GHz. Source amplitude and phase (rad) in x , y , and z direction is shown in (a,d), (b,e), and (c,f), respectively. Amplitude and phase (rad) of the back-scattered field from the cornea in x , y , and z direction is shown in (g,j), (h,k), and (i,l), respectively. Field distribution is calculated on a plane perpendicular to the optical axis with dimension 20 mm × 20 mm. The source beam is assumed to be polarized in the x and z direction.

Source fields distribution plots based on Eqs. (5) and (6) are almost consistent with the analytical field plot of the Gaussian beam. Since the source beam is polarized in x and y directions, the Ey field amplitude (Fig. 6(b)) is in micro order, much smaller than the Ex and Ez amplitude (Fig. 6(a,c)). The back-scattered field amplitude diminishes about  40 % in x and Z directions. Likewise, if the cornea is considered as a planar structure illuminated by a plane wave, the reflection amplitude would be also around 40 % at 220 GHz. The back-scattered field phase is shifted 180° as expected.

4. Physical-optics analysis

The six different strategies were simulated with an in-house developed physical-optics (PO) script to verify the results of the presented Fourier-optics method. The two-way propagation from an emitting screen to the homogenized PEC sphere and back was simulated with the Gaussian-beam parameters introduced in section 2. In PO, the field outside the radiating aperture was calculated as in [21], [22]:

E(r)=SG(rr)×Jms(r)dA, (9)

where r and r are the locations at the radiating aperture and at the observation point respectively, G is the scalar Green’s function, Jms is the magnetic surface current density, and dA is the differential area element. The integral was applied first from the virtual Gaussian-beam waist ( rw0rscreen ) to the screen to define the initial distribution. The two pass propagation from screen to sphere and back to the screen was computed ( rsphererscreen , rscreenrsphere ).

The PO simulated electric field was oriented to ensure no shadowing between the surfaces occurred with an edge taper sufficient to limit the spill-over loss. The coupling coefficient from the PO simulation is overlaid with that from the Fourier-optics method, validate it for homogeneous targets such as PEC sphere. For all strategies, the coupling coefficient with FO method and PO differ by less than 1% across the 220 GHz to 330 GHz. The coupling coefficient from the physical-optics simulation is consistently less than that from the Fourier method and is likely due to the remaining spill-over loss necessary to avoid shadowing in this geometry.

5. Results and discussion

The coupling efficiency of the six different strategies addressed in section 3 was compared for different targets and compared with the plane wave reflection from a multilayered planar surface as the maximum coupling efficiency case. It was computed with Fresnel equations and superimposed on the Gaussian-beam equations to evaluate divergence from the plane-wave condition. The cornea was modeled as described in section 3 and Fig. 6.

To compare different strategies for various objects, coupling efficiency was defined as Eq. (11). It calculated the coupling between the back-reflected field and the incident field in a reference plane located at z=40 mm. This distance is chosen according to experimental setup. The integration range for x and y starts from 4ω1 to 4ω1 and * denotes the complex conjugate. For accurate calculations of coupling efficiency, the field at the edge of the reference plane should be vanishingly small, thus the reference plane was increased until this criteria was met.

CE=Ei.EsdxdyEi.Eidxdy, (10)

where Ei and Es are computed by Eq. (5) and Eq. (6). The coupling efficiency is defined as the absolute value of coupling coefficient |CE| and phase of CE is defined as arctan(image(CE)real(CE))180°π .

Normally, the number of modes ( Nstop ) used in CE calculations were determined by the introduced equation in appendix B. To achieve a more robust result, each simulation accuracy was checked in terms of convergence of coefficients to a specific value. For each target, Nstop were reported in a table. Simulation time increases exponentially by increasing the number of modes. The frequency band 100 - 600 GHz was chosen regarding the simulation time and stability of the algorithms as well as constraints introduced in section 2.

5.1. Homogeneous PEC sphere coupling efficiency magnitude

In this subsection, the coupling efficiency of the previously described designs in section 2. for the PEC sphere was calculated using the proposed methodology in section 3. The Gaussian-beam illuminating PEC was assumed to scatter more energy back at the screen than any other material of equal RoC and thus served as a reference/calibration target for the further exploration of PEC coated with the lossless cornea, and cornea simulations.To compute the scattering coefficients of a PEC sphere illuminated by a Gaussian beam Eq. (8) was used. For applying the T-matrix method it was first necessary to compute the scattering coefficients of the PEC sphere when illuminated by a plane wave. Here, the approach in [23] and presented in appendix C was applied. The PEC sphere radius was set to Rc=7.5 mm and the beam waist radii locations were distributed to match strategies S1 to S6 as described in section 2. The number of modes to reach enough accuracy for the PEC sphere is reported inTable 1. The simulation time for each strategy was about 17 hours on UPC computational cluster.

Table 1. Number of modes for PEC sphere and coated PEC sphere simulations.

Frequency (GHz) 100-445 445-600
Nstop 75 95

According to Fig. 7, the forward strategies S1 and S2 , with dispersion set to achieve frequency wavefront matching, behave similarly and increase from 97.84% to 99.92% and from 97.83% to 99.89% . Both reverse strategies S3 and S4 coupling efficiencies behave almost in the same way and range from 96.46% to 100% and 96.77% to 99.94% across the band. The reference strategies S5 and S6 behave likewise and unlike the other strategies decrease over the frequency band from 95.71% to 94.59% and 96.08% to 94.59% , respectively.

Fig. 7.

Fig. 7.

PEC sphere coupling efficiency magnitude comparison is presented for (a) Forward strategies, S1 , and S2 , (b) reverse strategies, S3 , and S4 , and (c) reference strategies, S5 , and S6 . Plane wave reflection from equivalent PEC planar is indicated by black-dotted line and radius of sphere equals Rc=7.5 mm.

Overall, for homogeneous PEC sphere being a target, S1 comes closest to the plane-wave condition although it behaves closely to S2 . Forward strategies reveal higher coupling compared to reverse strategies ( 1.1% ) and reverse strategies display more coupling compared to reference strategies ( 0.45.5% across frequency band).

5.2. Lossless cornea sitting on a PEC core coupling efficiency

Lossless dielectric spherical shells backed by PEC were simulated to explore the effect of spectral phase-front variation and mismatch. A target was constructed of 0.5 -mm thick lossless cornea spherical shell encapsulating a PEC sphere of 7 mm RoC. The layer refractive index was set as the real part permittivity of the cornea computed by effective media theory [15]. The coupling efficiency was computed with Eq. (11). Then, the calibrated coupling efficiency was computed by normalizing the PEC-backed lossless cornea shell coupling efficiency with the coupling efficiency from a 7.5 mm PEC sphere (Fig. 8). The scattering coefficients of a coated PEC were addressed in appendix C. The number of modes considered to reach stability for coated PEC was similar to PEC sphere modes inTable 1 and simulation time was the same.

Fig. 8.

Fig. 8.

Calibrated lossless cornea sitting on a PEC core coupling efficiency magnitude and phase comparison are presented for (a,d) Forward strategies, S1 , and S2 , (b,e) reverse strategies, S3 , and S4 , and (c,f) reference strategies, S5 , and S6 . Plane wave reflection from equivalent layered planar is indicated by black-dotted line. The shell permittivity is set according to real part of cornea permittivity and its thickness is 0.5 mm sitting on 7 mm RoC PEC sphere.

This operation mimics experimental calibration routines [4] and enables analysis of the absolute phase following deconvolution of the free space path influence on complex coupling angle. The PEC sphere calibrated lossless cornea shell coupling efficiency magnitude and phase are displayed in Fig. 8 where magnitude and phase of forward strategies S1 and S2 , reverse strategies S4 and S5 , and reference strategies S5 and S6 are plotted in panels (a), (b) and (c), respectively. All trends are referenced to the equivalent plane-wave condition.

Interestingly, as illustrated in Fig. 8, with calibrating all strategies, phase behaviors were almost consistent with plane wave condition, implying a proper phase matching. In fact, they matched perfectly if plane wave phase figure shifts about 10 GHz toward low frequencies. These results suggest that good results can be obtained with any most of the strategies provided the target reflection is calibrated with reflections from an appropriately sized, metallic sphere. Moreover, from the magnitude aspect, forward strategies provided better coupling between the incident and back-scattered fields than reverse ones, however, S4 featured closer to S1 and S2 (less than 1% difference). It could be said S3 and S5 were also close enough to plane wave condition (less than 2.2% difference) and certainly, S6 was far from proper coupling ( 1.58.8% deviation).

The simulation results suggest that the lossless cornea shell is acting as a lens and improving the wavefront match to the 7 mm RoC inner PEC surface relative to the reference 7.5 mm RoC PEC sphere. Non-sequential ray-tracing simulations of a Gaussian beam in Zemax OpticStudio also suggest that the presence of a lossless dielectric layer on top of the PEC (hemisphere) sphere might slightly improve the coupling. In Fig. 9, the power distribution on the detector is reported for S1 at two different frequencies, 200 GHz and 600 GHz. At 200 GHz the root mean square (RMS) spot radius is 3.40 mm for the coated sphere and 3.33 mm for the PEC sphere. At 600 GHz the RMS spot radius variation is much more constrained: 3.13 mm for the coated sphere and 3.12 mm for the PEC sphere. This trend agrees with Fig. 8(a) where the coupling shows a peak at 200 GHz.

Fig. 9.

Fig. 9.

Incoherent irradiance for S1 strategy in 200 GHz for (a) PEC sphere and (c) coated PEC sphere and in 600 GHz for (b) PEC sphere and (d) coated PEC sphere, indicating higher RMS of coated sample compare to calibration sphere, resulting above 1 values of coupling efficiency amplitude in Fig. 8. Panel (e) supports the idea by a histogram of normalized frequency of rays in each angle, emphasizing on loss-free shell assists the incoming beam to come closer to the optical axis (lower angles).

In a quasi-optical, mono-static ( S11 ) measurement, the maximum signal is obtained when the beam is retro-directive, i.e. the scattered beam mirrors the illumination beam. In the case of a spherical target, this can be visualized by discretizing the wavefront as a spatial collection of converging rays. The maximum signal is achieved when all converging rays are normal to the spherical surface. However, in this case, the converging beam is Gaussian with a non-spherical phase front thus normal incidence is achieved only on-axis. The loss-free shell assists the incoming beam to come closer to the optical axis (as shown in Fig. 9(e)), that’s why the above one values for coupling efficiency appeared in Fig. 8.

5.3. Calibrated cornea coupling efficiency

The approach described in [18], and outlined in Appendix B, yielded the magnitude and phase coupling between the incidence and back-scattered electric fields for cornea structure. The number of modes used for simulation is reported inTable 2 and simulation time was 25 hours on UPC computational cluster. As mentioned earlier, the best strategy for performing an experiment that matches with the theory developed in previous works [4] is the one that Gaussian-beam illumination on a spherical surface acts such as a plane-wave illumination on a plane resembling a maximum coupling. Magnitude and phase of relative calibrated coupling efficiency for cornea is plotted in the Fig. 10. In this context, relative calibrated means subtract calibrated cornea coupling efficiency from equivalent plane wave condition coupling efficiency.

Table 2. Number of modes for cornea simulations.

Frequency (GHz) 100-200 200-380 380-510 510-600
Nstop 65 75 100 125

Fig. 10.

Fig. 10.

Relative calibrated cornea coupling efficiency magnitude and phase comparison are shown for (a,d) Forward strategies, S1 , and S2 , (b,e) reverse strategies, S3 , and S4 , and (c,f) reference strategies, S5 , and S6 .

Calibration with PEC sphere mirrors current experimental methodology where the reflectivity of a phantom or the ex-vivo cornea is normalized by a spherical reference reflector of approximately equal RoC. The reflectivity is obtained on an absolute scale when the optical path is deconvolved from the reflection by the normalization. These results support the previous hypothesis in the literature that, under the right alignment conditions, Gaussian-beam illumination on cornea approximates plane-wave illumination on planar stratified media.

Surprisingly, these strategies behave almost the same as a plane-wave illumination on the planar surface and show a maximum of 2.3% magnitude deviation in low frequency for strategy S6 . Similarly, for calibrated phase coupling efficiency (indicated in Fig. 10(d-f)) of preceding strategies attributes almost the same as plane-wave illumination on the plane, with limited deviation at low frequency for strategy S3 (a maximum of 1.2° deviation).

These results suggest that, under ideal alignment and calibration conditions, the efficiency of coupling to corneal longitudinal modes is nearly indistinguishable amongst the six different strategies. This behavior is almost similar to the loss-free shell results, although the lossless cornea shell deviates from plane wave condition more than cornea case, indicating that lossy cornea steady state reflectivity minimizes walk-off losses and any illumination condition, regardless of wavefront match/mismatch is sufficient for proper coupling as far as equal surface geometry is feasible for calibration.

6. Experimental exploration

In order to explore the theory was introduced in section 3. and was used for investigation in previous sections, a setup (as shown in Fig. 11(a)) is implemented. Gaussian beam telescope introduced in [4] and shown in Fig. 11(b) is applied to illuminate quartz dome sitting on a water core (resembling a cornea). It is shown previously that a back-scattered measured field from a 0.5 mm quartz sitting on an air core acts as a plane wave illumination on plane [24]. Here, two sizes of quartz are considered in this experiment, quartz size 1 , Q1 with diameter 15 mm and thickness 0.5 mm, and quartz size 2 , Q2 with diameter 16 mm and thickness 1 mm. These dimensions are chosen according to real cornea size and their permittivity is around the real part permittivity of the human cornea (around 4.1 ). The quartz dome is set on a holder which is mounted on mechanical arms which move in x, y, and z-direction, illustrated in 11(a). The back-scattered field is measured in a frequency range from 220 GHz to 330 GHz. As shown in Fig. 11(c), a metal ring is used to hold the quartz dome on the water core and an absorber ring (Fig. 11(d)) covers it to prevent unwanted scattered fields from metal. Figure 11(e) presents the confocal distance of the output Gaussian beam in the quasioptical system. The output beam waist radius of the quasioptical system is measured and fitted with the fundamental Gaussian-beam model. The fit waist radii in E- and H-planes are used to calculate the confocal distance: zc,E=πω0,H2/λ , zc,H=πω0,H2/λ . Due to the slightly different E- and H-plane beam width of the feed horn and polarization effects in the quasioptics, the output beam is astigmatic and the confocal distance is determined separately for the two planes.

Fig. 11.

Fig. 11.

(a) Measurement setup, (b) Gaussian beam telescope illuminate sample, (c) holding quartz dome on a water half-space by a metal ring, (d) use ring absorber to eliminate unwanted back-scattering from metal holder, (e) confocal distance of the output Gaussian beam, (f) OCT image of Q1 , and (g) Q2 .

Alignment was initiated with a course 2 D transverse (x-y) sweep of the sample about the approximate focal axis of the Gaussian beam telescope. The location corresponding to peak collected back-scattered intensity was identified and then a finer 2 D scan was performed about the peak to further refine the peak location. Next, the sample was fixed on the peak location, and the target was swept 20 mm in the vertical (z-axis) direction over a distance of 20 mm. Complex S11 were acquired in steps of 0.1 mm for a total of 201 axial points. For data analysis, the reflection from the sample at each point was Fourier transformed to the time/space domain, time-gated and transformed back to the frequency domain. This procedure performed repeated for PEC spheres of radius 8 mm and radius 7.5 mm and for the water backed quartz dome phantoms.

Examples of range the Fourier transformed data are shown in the Fig. 12(a) and 12(b) for the 7.5 mm RoC PEC sphere and 7.5 mm, 0.5 mm thick water backed dome respectively. The vertical axes corresponds to the approximate physical distance of the target with respect to a reference plane located in the beam path and shows the axial range traversed by the z-axis during scanning. The horizontal axes show the calculated range from the extender output flange to the target as computed with the Fourier transform of the complex S11 data. In Fig. 12(a) the (r,z) location of the peak corresponds to the subconfocal point as reported in [4] and the superconfocal point and waist location can be found by traversing the (r,z) space by the known separation distances. The same analysis was applied to the phantom data in Fig. 12(b). Once the locations corresponding to strategies 3 , 4 , 5 , and 6 are identified in these data, the quartz domes back-scatter can be calibrated with the PEC sphere data. An example of this is shown in Fig. 12(c).

Fig. 12.

Fig. 12.

Examples of range of the Fourier transformed data for (a) the 7.5 mm RoC PEC sphere, and (b) 7.5 mm, 0.5 mm thick water backed dome and, (c) Normalized S11(t) as function of r . Purple line shows the spherical mirror calibration data from where the strongest return came (arguably close to the subconfocal distance indicated by green line), sample location which results in best fit to the theoretical is shown with blue line and its pick is in yellow.

In Fig. 13, the measured back-scattered field of the sample is depicted (dashed lines) and compared with VSH simulation results (solid lines). All the plots are calibrated by the same size PEC sphere, both for simulation and measurement. Figure 13(a) and (b) displays the amplitude and phase of quartz size 1 (Diameter 15 mm and thickness 0.5 mm) and Fig. 13(c) and (d) demonstrates the amplitude and phase of quartz size 2 (Diameter 16 mm and thickness 1 mm). There is an adequate agreement between simulation and measurement in different reverse illumination strategies, representing realistic strategies S3 , S4 , S5 , and S6 ), in order to illuminate a sample as S1 and S2 a custom lens with engineered spectral dispersion is required. Resonances and peak locations in the measured spectrum are consistent with the theory. Also, as the Gaussian beam focus moves from the center of the sample ( |Z0|=0 mm) to the apex of the sample ( |Z0|=7.5 mm), a slight shift toward the higher frequency is observed both in the measurement and simulation. The shift is more detectable in thick quartz and may be due to the Gouy phase shift and corresponding superluminal velocity which have not been corrected for in this analysis. This result confirms that, with correct calibration, all six illumination strategies produce similar results. Note that the thicker, fabricated quartz dome was 0.96 mm thick and the thinner one was 0.5 mm. Both had an estimated ϵ4.0 . These values were used in the VSH simulation. The extraction of quartz dome permittivity and thickness were performed with particle swarm optimization (PSO). The PSO algorithm searches for the best fit to a plane wave model by minimizing a merit function. Strategies S3 , S4 , S5 , and S6 were investigated representing realistic strategies. The merit function combined two terms: the phase error and amplitude error. The used merit function for each scenario is:

MF=0.331Ni=iNΓMeasΓ21Ni=iNΓMeas2+0.661Ni=1NΓMeasΓ21Ni=1NΓMeas2, (11)

where ΓMeas are the measured reflection coefficients reported in Fig. 13, Γ is the plane wave model reflection coefficient, and N is the number of frequency points. The merit function is a sum of the mean squared error normalized with the average power of the merit function metric. The extracted thicknesses and permittivities, reported in Fig. 14 (a,b,c,d), are ordered by the distance from the beam waist and CoC. For the 0.5 mm thick dome there is a monotonic increasing permittivity and decreasing distance as the waist is brought closer to the apex. The opposite is observed for the 1 mm dome. The complementary behaviour of permittivity and thickness suggest a conservation of optical path length. The superluminal velocity is axial location dependent and monotonically decreases from the beam waist which is coincides with the monotonic behaviour seen here. Further corrections may help address non-unique behaviour of the merit function space where an ensemble of permittivity/thickness pairs produces a range of optical path lengths with equal goodness of fit.

Fig. 13.

Fig. 13.

Measurement data are compared with VSH simulation for reverse strategies for (a,b) quartz size 1 amplitude and phase, and (c,d) quartz size 2 amplitude and phase. As the beam focuses closer to apex a slight shift toward higher frequency is detected. The shift is more evident in thicker quartz. This is in agreement with simulation results.

Fig. 14.

Fig. 14.

(a) Extracted thickness and (b) extracted permittivity for Q1 are obtained by PSO fitting method considering strategies S3 , S4 , S5 , and S6 . (c) Extracted thickness and (d) extracted permittivity for Q2 are obtained by PSO fitting method considering strategies S3 , S4 , S5 , and S6 .

7. Conclusion

A theory based on Fourier-optics analysis and VSH was employed to explore the coupling efficiency between the incident and back-scattered fields from homogeneous and coated spherical targets at different beam-target alignments. In this method, Gaussian beam approximated by an ensemble of plane waves and scattering coefficients were calculated via the T-matrix method.

PEC spheres are targets for calibration in different simulations. PEC coupling efficiency magnitude in six different strategies was explored. The S2 alignment shows the highest matching with plane-wave illumination on a planar surface. Also, the influence of +0.2 -mm RoC change was investigated. The S6 strategy shows the highest sensitivity (close to 3.3% ) to RoC discrepancies.

To explore the effect of spectral phase front variation and mismatch, coated PEC structure was studied. Due to the lossless layer in this assembly, walk-off loss takes part in the coupling, and beam profile influences the calibrated coupling efficiency magnitude, although phase behavior remains insensitive to the illumination profile and only shows 10 GHz shift with respect to low frequencies than the plane wave condition. The S2 design revealed the highest consistency with plane wave on plane illumination.

Surprisingly, the calibrated coupling efficiency magnitude for S1 - S4 strategies (in coated PEC) is above one. The reason is the refraction of the Gaussian beam while passing through the loss-free layer, as an incident beam propagates closer to the optical axis, the beam reflection enhances for certain kinds of Gaussian beams. This is in striking contrast to the stratified medium theory and theories considering plane waves - the specifics of the Gaussian beam and target RoC need to be considered carefully before applying the stratified medium model.

Angular spectra and vector spherical harmonic analyses were used to investigate the corneal coupling efficiency. The cornea was modeled as a layered media on top of the aqueous core and illuminated by a Gaussian beam. It was shown that engineering the sub-millimeter wave beam such that incident beam 1/e radius fixed to ω1=3.1 mm and radius of the Gaussian beam matches with 7.5 mm cornea radius at each frequency (strategy S2 ) leads to the highest coupling among other strategies.

Illumination profiles defines the coupling efficiency behavior unless the corneal coupling efficiency is calibrated the coupling efficiency of PEC sphere of equal size. In all strategies, coupling efficiency magnitude and phase behave according to the plane-wave illumination on the planar half-space. It implies that the cornea is lossy enough that walk-off loss does not influence coupling efficiency.

An experimental setup is assembled to validate the VSH theory. The experimental back-scattered spectrum is compatible with theory, peak and resonance location are the same for different size of domes and a shift toward higher frequency observed both in experiment and simulation. A PSO analysis is applied to extract the thickness and permittivity of quartz from measured back-scattered field for different realistic strategies.

Acknowledgments

We would like to thank Prof. Ari Sihvola for his valuable advice to understand and implement the theory of problem.

Appendix A. Gaussian beam coefficients derived by Fourier analysis

For a polarized Gaussian beam in x and z direction, after a cumbersome calculation, the incident beam coefficients in Eq. (5) are obtained as:

ae=Fθxθy4inaθxye,ao=Fθxθy4inaθxyo,be=Fθxθy4in+1beθxyo,bo=Fθxθy4in+1bθxyo, (12)

where F=(kω0p)2/4π in which p=π/180 . The θx and θy are angles respect to x and y axis. Also,

aθxye=Aθxy(sinϕcosmϕτmcosϕsinmϕΠ/cosθ),aθxyo=Aθxy(mcosϕcosmϕΠ/cosθ+sinϕsinmϕτ),bθxye=Aθxy(msinϕsinmϕΠ+cosϕcosmϕτ/cosθ),bθxyo=Aθxy(cosϕsinmϕτ/cosθmsinϕcosmϕΠ). (13)

In the above equations, the Π=Πmn=Pnm(cosθ)sinθ and τ=τmn=ddθPnm(cosθ) are auxiliary functions which can be obtained by recursion relation [9]:

Πmn=(2n1)cosθΠmn1(n+m1)Πmn2nm,τmn=ncosθΠmn(n+m)Πmn1. (14)

The first two starting values of Πmn are generated by the closed-form expressions when m0 :

Πmn=0n<m,Πmn=(2m)!sinm1(θ)2mm!n=m, (15)

and while m=0 :

Π0,0=1sinθ,Π0,1=cosθsinθ. (16)

On the other hand, the Aθxy=TGsinθxsinθy where T=eik(s.v)=ei(kxx0+kyy0+kzz0) indicates the relocation of waist radius location relative to center of sphere and G=e(ksω02)2=eω024(kx2+ky2) . Besides, PW=eik(s.r)=ei(kxx+kyy+kzz) illustrates a plane wave. The vectors s , v , and r are s=cosθxix+cosθyiy+cosθiz , v=x0ix+y0iy+z0iz , and r=xix+yiy+ziz . The wave vector defines as k=kxix+kyiy+kziz and k=|k|=[kx2+ky2+kz2]1/2=2π/λ is the wave number in the wavelength λ . For any arbitrary k , θx , θy , θ , and ϕ are defined the way that the components of k are:

kx=kcosθx=ksinθcosϕ,ky=kcosθy=ksinθsinϕ,kz=kcosθ. (17)

Appendix B. Coated sphere illuminated by a plane wave

The scattering coefficients of a coated sphere illuminated by a plane wave is presented by [13] and summarized in [12]. The method starts with solving the Helmholtz equation in spherical coordinate and fulfillment of the boundary condition at each layer. The scattering coefficients are derived as below:

anc=[Hna(mlxl)/ml+n/xl]ψn(xl)ψn1(xl)[Hna(mlxl)/ml+n/xl]ζn(xl)ζn1(xl),bnc=[Hnb(mlxl)/ml+n/xl]ψn(xl)ψn1(xl)[Hnb(mlxl)/ml+n/xl]ζn(xl)ζn1(xl), (18)

where ψn and ζn are Riccati-Bessel functions. The ml and xl are refractive index and size parameter of the l th layer. The Hna and Hna are given by following expressions:

Hna(m1x1)=Dn1(m1x1),Hna(mlxl)=G2Dn1(mlxl)QnlG1Dn3(mlxl)G2QnlG1,Hnb(m1x1)=Dn1(m1x1),Hna(mlxl)=Gˆ2Dn1(mlxl)QnlGˆ1Dn3(mlxl)Gˆ2QnlGˆ1,G1=mlHna(ml1xl1)ml1Dn1(mlxl1),G2=mlHna(ml1xl1)ml1Dn3(mlxl1),Gˆ1=ml1Hnb(ml1xl1)mlDn1(mlxl1),Gˆ1=ml1Hnb(ml1xl1)mlDn3(mlxl1). (19)

Next equations determine the logarithmic derivatives of the Riccati–Bessel functions, Dn1(z) and Dn3(z) , the ratio Qnl , ψn and ζn . A thorough explanation can find in [13] and [12].

DNmax1(z)=0+0i,Dn11(z)=nz1Dn1(z)+nz,D03(z)=i,Dn3(z)=Dn1(z)+iψn(z)ζn(z),ψ0(xl)ζ0(xl)=12[1(cos2a+isin2a)exp(2b)],ψn(xl)ζn(xl)=ψn1(xl)ζn1(xl)[nzDn11(z)][nzDn13(z)],ψ0(xl)=sin(xl),ψn(xl)=ψn1(xl)[nxlDn11(xl)],ζ0(xl)=sin(xl)icos(xl),ζn(xl)=ζn1(xl)[nxlDn13(xl)],Q0l=exp(2ia1)exp(2b1)exp(2ia2)exp(2b2)exp(2[b2b1]),Qnl=Qn1l(xl1xl)2[z2Dn1(z2)+n][z1Dn1(z1)+n][nz2Dn13(z2)][nz1Dn13(z1)], (20)

where z=a+ib , z1=mlxl1=a1+ib1 and z2=mlxl=a2+ib2 . For all recurrence relation n=1,..,Nmax except Dn1 which is a downward recurrence. The maximum number of the modes Nmax play an crucial rule to stability of the problem. It is a function of the size parameter and Nmax=max(Nstop,|mlxl|,|mlxl1|)+15 when l=1,2,..,L and

Nstop={xl+4xl1/3+10.02xl<8xl+4.05xl1/3+28xl<4200xl+4xl1/3+24200xl<20,000. (21)

As mentioned in section 3, to compute the scattering coefficients of a coated sphere while illuminated by a plane wave, a method different from [7] and [19] was utilized. Khaled et al. used Toon and Ackerman [10] algorithm while in here the algorithm described by Yang [13] is applied. The advantage of the Yang method is providing the possibility of analysis of a coated sphere with more than one shell.

Appendix C. PEC and coated PEC sphere illuminated by a plane wave

The scattering coefficients of a sphere illuminated by a plane wave are computed by solving the vector-wave equation in spherical coordinates and evoking the boundary condition (Mie theory [11]). In the case of a PEC sphere, the electric field at the PEC boundary was forced to zero and the simplified form of scattering coefficients for a PEC sphere are obtained as:

an=ψn(x)ζn(x)bn=ψn(x)ζn(x), (22)

where the size parameter x=kr and r is the radius of the PEC sphere.

With the mutual approach, solving a vector-wave equation, applying boundary condition and forcing the electric field to be zero inside the PEC core, the simplified scattering coefficients for a coated PEC sphere are obtained as:

an=ψn(x1)[ψn(m2x1)Anχn(m2x1)]m2ψn(x1)[ψn(m2x1)Anχn(m2x1)]ζn(x1)[ψn(m2x1)Anχn(m2x1)]m2ζn(x1)[ψn(m2x1)Anχn(m2x1)],bn=m2ψn(x1)[ψn(m2x1)Bnχn(m2x1)]ψn(x1)[ψn(m2x1)Bnχn(m2x1)]m2ζn(x1)[ψn(m2x1)Bnχn(m2x1)]ζn(x1)[ψn(m2x1)Bnχn(m2x1)], (23)

where

An=ψn(m2x)χn(m2x),Bn=ψn(m2x)χn(m2x). (24)

The PEC core radius r1 is coated with a layer with outer radius of r2 and refractive index of m2 . The Riccati-Bessel functions are χn(z)=zyn(z) , ψn(z)=zjn(z) , and ζn(z)=zhn(1)(z) in which jn(z) , yn(z) , and hn(1)(z) are Bessel function of the first, Bessel function of the second kind and Hankel function of the first kind, respectively. The primes denote the differentiation with respect z .

Funding

Agencia Estatal de Investigación10.13039/501100011033 (PID2019-107885GB-C31/AEI/10.13039, PRE2018-084326); Academy of Finland10.13039/501100002341 (327640).

Disclosures

The authors declare that there are no conflicts of interest related to this article.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

References

  • 1.Taylor Z. D., Garritano J., Sung S., Bajwa N., Bennett D. B., Nowroozi B., Tewari P., Sayre J. W., Hubschman J., Deng S. X., Brown E. R., Grundfest W. S., “Thz and mm-wave sensing of corneal tissue water content: In vivo sensing and imaging results,” IEEE Trans. Terahertz Sci Technol 5(2), 184–196 (2015). 10.1109/TTHZ.2015.2392628 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Taylor Z. D., Garritano J., Sung S., Bajwa N., Bennett D. B., Nowroozi B., Tewari P., Sayre J., Hubschman J. P., Deng S., Brown E. R., Grundfest W. S., “THz and mm-wave sensing of corneal tissue water content: electromagnetic modeling and analysis,” IEEE Trans. Terahertz Sci Technol. 5(2), 170–183 (2015). 10.1109/TTHZ.2015.2392619 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Sung S., Selvin S., Bajwa N., Chantra S., Nowroozi B., Garritano J., Goell J., Li A. D., Deng S. X., Brown E. R., Grundfest W. S., Taylor Z. D., “Thz imaging system for in vivo human cornea,” IEEE Trans. Terahertz Sci Technol. 8(1), 27–37 (2018). 10.1109/TTHZ.2017.2775445 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Tamminen A., Pälli S. V., Ala-Laurinaho J., Salkola M., Räisänen A. V., Taylor Z. D., “Quasioptical system for corneal sensing at 220-330 ghz: Design, evaluation, and ex vivo cornea parameter extraction,” IEEE Trans. Terahertz Sci Technol 11(2), 135–149 (2021). 10.1109/TTHZ.2020.3039454 [DOI] [Google Scholar]
  • 5.Mie G., “Beiträge zur optik trüber medien, speziell kolloidaler metallösungen,” Ann. Phys. 330(3), 377–445 (1908). 10.1002/andp.19083300302 [DOI] [Google Scholar]
  • 6.Davis L. W., “Theory of electromagnetic beams,” Phys. Rev. A 19(3), 1177–1179 (1979). 10.1103/PhysRevA.19.1177 [DOI] [Google Scholar]
  • 7.Khaled E. E. M., Hill S. C., Barber P. W., “Scattered and internal intensity of a sphere illuminated with a Gaussian beam,” IEEE Trans. Antennas Propag. 41(3), 295–303 (1993). 10.1109/8.233134 [DOI] [Google Scholar]
  • 8.Goodman J. W., Introduction to Fourier Optics (Roberts, 2005). [Google Scholar]
  • 9.Barber P. W., Hill S. C., Light Scattering by Particles: Computational Methods (World Scientific, 1990). [Google Scholar]
  • 10.Toon O. B., Ackerman T. P., “Algorithms for the calculation of scattering by stratified spheres,” Appl. Opt. 20(20), 3657–3660 (1981). 10.1364/AO.20.003657 [DOI] [PubMed] [Google Scholar]
  • 11.Bohren C., Huffman D. R., Absorption and Scattering of Light by Small Particles (Wiley Science Paperback Series, 1998). [Google Scholar]
  • 12.Pe na U. P. O., “Scattering of electromagnetic radiation by a multilayered sphere,” IEEE Trans. Antennas Propag. 180, 2348 (2009). 10.1016/j.cpc.2009.07.010 [DOI] [Google Scholar]
  • 13.Yang W., “Improved recursive algorithm for light scattering by a multilayered sphere,” Appl. Opt. 42(9), 1710–1720 (2003). 10.1364/AO.42.001710 [DOI] [PubMed] [Google Scholar]
  • 14.Zarrinkhat F., Baggio M., Lamberg J., Tamminen A., Nefedova I., Ala-Laurinaho J., Khaled E. E. M., Rius J. M., Romeu J., Taylor Z., “Calibration alignment sensitivity in corneal terahertz imaging,” Sensors 22(9), 3237 (2022). 10.3390/s22093237 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Sihvola A., Electromagnetic Mixing Formulas and Applications (Institute of Electrical Engineers, 1999). [Google Scholar]
  • 16.Goldsmith P. F., Quasioptical Systems: Gaussian Beam Quasioptical Propagation and Applications (IEEE Press, 1998). [Google Scholar]
  • 17.Lackner B., Schmidinger G., Pieh S., Funovics M. A., Skorpik C., “Repeatability and reproducibility of central corneal thickness measurement with pentacam, orbscan, and ultrasound,” Optom. Vis. Sci. 82(10), 892–899 (2005). 10.1097/01.opx.0000180817.46312.0a [DOI] [PubMed] [Google Scholar]
  • 18.Zarrinkhat F., Lamberg J., Tamminen A., Baggio M., Ala-Laurinaho J., Khaled E. E. M., Rius J. M., Robert J. R., Taylor Z., “Fourier analysis of submillimeter-wave scattering from the human cornea,” in EuCAP (2021), pp. 1–5. [Google Scholar]
  • 19.Khaled E. E. M., Hill S. C., Barber P. W., “Light scattering by a coated sphere illuminated with a gaussian beam,” Appl. Opt. 33(15), 3308–3314 (1994). 10.1364/AO.31.003308 [DOI] [PubMed] [Google Scholar]
  • 20.Xu J., Plaxco K., Allen S., Bjarnason J., Brown E., “0.15-3.72 thz absorption of aqueous salts and saline solutions,” Appl. Phys. Lett. 90(3), 031908 (2007). 10.1063/1.2430781 [DOI] [Google Scholar]
  • 21.Kong J., Electromagnetic Wave Theory (John Wiley and Sons, 1986). [Google Scholar]
  • 22.Orfanidis S. J., Electromagnetic Waves and Antennas (Rutgers University New Brunswick, 2002). [Google Scholar]
  • 23.Ruppin R., “Scattering of electromagnetic radiation by a perfect electromagnetic conductor sphere,” J. Electromagn. Waves Appl. 20(12), 1569–1576 (2006). 10.1163/156939306779292390 [DOI] [Google Scholar]
  • 24.Zarrinkhat F., Lamberg J., Baggio M., Tamminen A., Ala-Laurinaho J., Khaled E. E. M., Rius J. M., Robert J. R., Taylor Z., “Experimental exploration of longitudinal modes in spherical shells at 220 ghz – 330 ghz: applications to corneal sensing,” in CLEO/Europe-EQEC, (2021), p. 1. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.


Articles from Biomedical Optics Express are provided here courtesy of Optica Publishing Group