Abstract
Using the van Deemter model, the efficiency of three stationary phase systems in the analysis of a mixture of synthetic peptides was evaluated: (i) monolithic, (ii) packed, and (iii) core–shell columns, and it was shown that the efficiency of the monolithic column is superior to the others, specifically using it, the lowest values of Hmin (0.03 and 0.1 mm) were obtained, and additionally its efficiency was not significantly affected by increasing the flow. Using the concept of the gradient retention factor (k*), a method for chromatographic separation of a peptide complex mixture was designed, implemented, and optimized and then transferred from a packed column to a monolithic one. The results showed that it was possible to separate all components of the mixture using both evaluated columns; moreover, the analysis time was reduced from 70 to 10 min, conserving the critical pair resolution (1.4), by the transfer method using the k* concept. The method developed was tested against a mixture of doping peptides, showing that this method is efficient for separating peptides of various natures. This investigation is very useful for the development of methods for the analysis of complex peptide mixtures since it provides a systematic approach that can be extrapolated to different types of columns and instrumentation.
1. Introduction
Peptides are a family of molecules involved in many biological processes, and their diversity is mainly due to their primary structure and structural conformation.1 Peptides have great advantages as drugs because they have unique therapeutic properties, are usually harmless, and act on specific targets of physiological processes.2 Peptides that fulfill functions of regulation, modulation, and/or activation of the immune system, transport, biomarkers, and antimicrobial and anticancer activities, among others, have been described.3,4 Recent advances in peptide therapeutics include progress from synthetic methods and in the use of naturally occurring peptides, which have intrinsic weaknesses, to a sophisticated system of drug development based on drug discovery and design, peptide synthesis, and structural modification for activity evaluation.4−7
Reversed-phase high performance liquid chromatography (RP-HPLC) is a powerful and versatile technique for the monitoring, characterization, and/or purification of complex peptide mixtures from synthetic processes or protein hydrolysates in biotic and abiotic materials.1,8−11 General methods reported for peptide analysis include the use of silica-based C18 columns and ACN/H2O gradient elution using TFA as an ion pairing agent.3,4 However, the analysis of complex peptide mixtures by means of RP-HPLC is challenging since it is a procedure that requires a lot of time and involves a high consumption of solvents, which increases both costs and the amount of waste, with high environmental impact.8,9,11 RP-HPLC methods used for peptides usually have run times longer than 30 min.8,12,13 In RP-HPLC, the separation of analytes is mediated by partitioning and/or adsorption phenomena, in which the hydrophobic part of the peptide interacts with the hydrocarbon chain of the stationary phase.14 The wide variety of stationary phases available on the market has provided tools for developing methods that lead to better separation and reduce analysis time. The van Deemter equation makes it possible to determine the efficiency of a chromatographic system, which depends mainly on the chromatographic column, the flow of the mobile phase, and the physicochemical properties of the analyte. In this model, the chromatographic efficiency is given by eq 1, where height of the theoretical plate (HETP) is the height equivalent theoretical plate; A, B, and C are constants related to the eddy diffusion coefficient, longitudinal diffusion, and mass transfer resistance, respectively; and u is the linear velocity. The diffusion coefficient of Eddy (A) is constant, unaffected by the flow of the mobile phase and depends on the particle size and uniformity of the column packing. The term B/u represents the longitudinal diffusion of the analyte, and its contribution to HETP decreases as the flow of the mobile phase increases, while that the contribution of mass transfer resistance (Cu) to HETP increases as the mobile phase flow increases.15
![]() |
1 |
This equation allows the calculation of the optimal flow of the mobile phase required to achieve the highest efficiency (lowest HETP value) of the column for a given analysis.
Analytical methods for the separation of small molecules have been reported in the literature, and interactions of analytes with stationary phases have been studied.16 However, in the case of peptides, sometimes a method must be developed for a specific peptide, so that there are few reports of methods for the separation of complex mixtures of peptides that can be applied routinely. Chromatographic methods for peptides are often designed and optimized based on previous experience.17 However, this strategy consumes a lot of time and resources, which is undesirable in the design and development of chromatographic methods applied to the resolution of complex peptide mixtures. The design, implementation, and transfer of a chromatographic method is a complex process that requires consideration of factors such as the physicochemical properties of peptides (primary structure, solubility, molecular weight, length, hydrophobicity, amphipathicity, basicity, acidity, etc.), column properties (particle length, diameter and porosity, type of silica, etc.), stationary phase nature (C4, C8, C18, phenyl, cyanide, etc.), elution system (isocratic, linear gradient, segmented gradient, etc.), mobile phase composition (water/ACN, water/MeOH, buffer, etc.), counterion (TFA, FA, etc.), temperature, and injection volume, among others. Therefore, the design, implementation, optimization, and transfer of a chromatographic method to analyze peptide mixtures is a valuable tool for identifying and/or quantifying peptides in complex matrices. Other options, such as in silico prediction using computational models, could minimize the number of experiments and speed up method development. However, so far this tool has been used primarily to predict the tR of the analytes in a given chromatographic system.17 The analysis of peptides and proteins using conventional HPLC equipment is not a simple task, mainly due to the heterogeneity of the mixture because of the physicochemical properties and concentration of the molecules that make it up.18 In addition, optimization of HPLC methods for peptide analysis aims to reduce the analysis time and improve or maintain resolution. When the mixture is heterogeneous, gradient elution is generally the best option. RP-HPLC gradient elution is based on changing the composition of the mobile phase as the run passes, usually increasing the concentration of the organic phase (solvent B: ACN, MeOH, THF, etc.). Gradient elution is defined by several parameters, such as gradient time (tG), initial and final % B, linear or stepped gradient, etc. With this in mind, the gradient retention factor (k*) becomes a great tool for optimizing the chromatographic run and transferring methods among different chromatographic systems, maintaining the selectivity profile.
2. Materials and Methods
2.1. Reagents and Materials
Rink amide resin, Fmoc-amino acids, Triton-X, piperidine, N,N-dimethylformamide (DMF), dichloromethane (DCM), dicyclohexylcarbodiimide (DCC), 1-hydroxy-6-chloro-benzotriazole (6-Cl-HOBt), ninhydrin, potassium cyanide (KCN), ethanol, pyridine, phenol, trifluoroacetic acid (TFA), ethyl ether, triisopropylsilane (TIS), ethanedithiol (EDT), acetonitrile (ACN), methanol (MeOH), and Supelco solid-phase extraction columns were purchased from Sigma-Aldrich (St. Louis, MO). Fmoc-Dip-OH was obtained from AAPPTec (Louisville, KY). A Zorbax Eclipse XDB-C18 column was purchased from Agilent Technologies (California), a SunShell C18 column was purchased from ChromaNik Technologies Inc (Osaka, Japan), and a Chromolith High Resolution RP-18e column was purchased from Merk (New Jersey).
2.2. Peptide Synthesis (SPPS-Fmoc/tBu)
The peptides were obtained by means of manual solid-phase peptide synthesis, using the Fmoc/tBu strategy (SPPS-Fmoc/tBu) following previous reports.19 Briefly, Rink amide resin or Fmoc-amino acids anchored to solid support were treated with 5% piperidine in DMF for 10 min at RT (2×) to remove the Fmoc group. The Fmoc-amino acids were preactivated by mixing the amino acid with DCC/6-Cl-HOBt (1/1 equiv and 5 excesses with respect to resin substitution), and then, the preactivated Fmoc-amino acid was mixed with the resin or resin peptide for 2 h at RT. Once all of the amino acids were incorporated, the peptide was separated from the solid support by treating the resin peptide with TFA/H2O/TIS/EDT (92.5/2.5/2.5/2.5% w/w) for 8 h at RT. Finally, the reaction mixture was filtered and the solution was treated with ethyl ether to precipitate the peptide, and then, the solid was washed five times with ethyl ether.
2.3. RP-SPE Purification
The peptides were purified by means of solid-phase extraction in reverse phase mode (RP-SPE) Supelclean C18 columns (5 g, 45 μm, 60 Å)1. Briefly, a RP-HPLC analysis of the crude peptide was performed, and based on this result, using a mathematical model, the elution program was transferred to a RP-SPE. The peptide was loaded into the column and eluted by increasing the concentration of solvent B (TFA 0.05% in ACN), and the fractions containing the pure peptide were collected and lyophilized.
2.4. Chromatographic Conditions
The peptides (1 mg/mL) were dissolved in solvent 95% A (TFA 0.05% in water) and 5% B (TFA 0,05 in ACN), and a 10 μL sample was injected into an Ultimate 3000 HPLC (Thermo Scientific, Massachusetts) with a binary pump of up to 620 bar (9000 psi) capacity, a column oven compartment, and a diode array detector. For gradient analyses, the composition of B and the workflow varied according to the assay. For the isocratic analysis, a mixture of H2O–ACN (82:18) containing TFA 0.05% was used. Two channels (channel 1:210 nm, channel 2:280 nm) were used as working wavelengths. The column temperature was kept at 30 °C, and the temperature of the samples was maintained at 15 °C for all of the tests.
3. Results and Discussion
3.1. Van Deemter Model
Chromatographic methods for separating peptide mixtures were designed and implemented, and their performance was evaluated in three chromatographic columns (packed, monolithic, and core–shell), which are described in Table 1. The columns were functionalized with C18 chains, so the hydrophobicity (H) value was quite similar; however, the stationary phase technology of each of them differs in pore size, steric resistance (S*), and other parameters, such as H-B basicity (B) and ion-exchange capacity (C). Considering that these analytes are charged and their interaction with them is structurally complex, the stationary phase is not ruled only by H.
Table 1. Parameters and Characteristics of Columns and Their Stationary Phasesa.
column | Sunshell C18 | Eclipse XDB-C18 | Chromolith RP-18e |
---|---|---|---|
length (mm) | 500 | 500 | 500 |
pore size (Å) | 90.0 | 80.0 | 150 |
particle size (μm) | 2,60 | 5.00 | |
H | 1.09 | 1.07 | 0.99 |
S* | 0.028 | 0.020 | 0.019 |
A | –0.12 | –0.060 | 0.017 |
B | –0.056 | –0.030 | –0.001 |
C (2,8) | –0.43 | 0.050 | 0.21 |
C (7,0) | –0.84 | 0.080 | 0.28 |
description | core–shell | packed | monolithic |
Key selectivity parameters reported in the web platform https://www.hplccolumns.org: H (column hydrophobicity), S* (column steric resistance), A (column hydrogen-bond acidity), B (column hydrogen-bond basicity), C (2,8) (column cation exchange capacity at pH 2,8), C (7,0) (column cation exchange capacity at pH 7.0).
To evaluate the efficiency of the columns, the van Deemter model was used. This model makes it possible to determine the equivalent height of the theoretical plate (HETP) for each column and to determine for which of them the analytical method could perform best. In addition, it allows one to determine which diffuse phenomena are contributing the most to the loss of efficiency of the system and how it affects the optimal workflow.15 Two synthetic peptides were selected for the columns’ evaluation; they are derived from bovine lactoferricin (LfcinB): (peptide 1), a peptide of six residues, RRWQWR (MW: 986,132 g/mol), and (peptide 2), a branched peptide of 24 residues, specifically the dimer (RRWQWRFKKLG)2-K-Ahx (MW: 3441, 964 g/mol). Peptide 2 is heavier and more voluminous than peptide 1. The two peptides have quite different physicochemical properties and were used to make the mixture more complex.
To obtain the van Deemter curve experimentally, a solution containing NaNO3 (3.00 mg/mL), peptide 1 (0.67 mg/mL), and peptide 2 (0.72 mg/mL) was prepared using the mobile phase H2O/ACN as a solvent (82:18 v/v) containing 0.05% v/v TFA. A chromatographic separation of the mixture was performed using isocratic elution, injecting 10 μL of the solution, and the working wavelength was 210 nm. Successive chromatographic runs were performed while varying the flow from 0.02 to 4 mL/min. The maximum pressure allowed for each column was 250 bar (2900 psi), so that the runs could be replicated on any conventional HPLC. It is important to note that to apply the van Deemter model to high molecular weight analytes such as peptides, it is necessary to use rather low flows (0.02 mL/min) to estimate the contribution of the term B to the height of the theoretical plate. Another important point to bear in mind is that the American (USP) and European (EP) pharmacopoeias recommend determining the number of theoretical plates by measuring the peak width at 4σ and 2.35σ. However, in isocratic peptide runs, due to their structural complexity, the peaks are usually not symmetrical, so the results do not ideally fit the van Deemter model. For this project, the number of theoretical plates was determined using the statistical moment, which is a nonlinear fit that works very well for nonsymmetrical Gaussian peaks.20
The obtained results for the van Deemter model for each peptide in the three evaluated columns can be seen in Figure 1. It is evident that the three columns present high efficiency for both analytes, with optimal workflows close to or below 12 mm/s (0.1 mL/min) and theoretical plate values below 0.4 mm. As can be seen, the monolithic column performed best for both analytes since the height of the HETP did not show a significant variation in the workflows evaluated, and HETP was the lowest, which implies a greater number of theoretical plates, thus representing a greater efficiency. This column has the highest porosity, so it is possible to increase the workflow up to 4.00 mL/min (297 mm/s) without exceeding a pressure of 250 bar (2500 psi); this allows transfer of the method, without loss of efficiency, to conventional HPLC equipment, which usually has maximum pressure limits between 300 and 400 bar.
Figure 1.
Van Deemter plots for (A) peptide 1 (RRWQWR) and (B) peptide 2 ((RRWQWRFKKLG)2-K-Ahx). Blue (Acclaim packed column), green (SunShell C18 core–shell column), red (Chromolith RP monolithic column).
The core–shell column showed greater efficiency than the packed column for peptide 1, while both columns showed similar efficiency with peptide 2, suggesting that the efficiency of the core–shell column depends on the molecular weight and conformational structure of the peptide (Table 2).
Table 2. Experimental Van Deemter Parameters Obtained to Determine Column Efficiency for Peptides 1 and 2.
peptide N°: sequence | Chromolith | SunShell | Acclaim | |
---|---|---|---|---|
1: RRWQWR | A (mm) | 0.043 | 0.14 | 0.24 |
B (mm/s) | 0.049 | 0.0011 | 0.20 | |
C (mm × s) | 4.11 × 10–5 | 2.0 × 10–4 | 1.2 × 10–4 | |
uopt (mm/s) | 34.74 | 2.74 | 40.59 | |
Hmin (mm) | 0.046 | 0.14 | 0.25 | |
R2a | 0.96 | 0.97 | 0.95 | |
2: (RRWQWRFKKLG)2-K-Ahx | A (mm) | 0.053 | 0.24 | 0.32 |
B (mm/s) | 1.71 | 0.90 | 0.087 | |
C (mm × s) | 0.0036 | 0.0019 | 0.002 | |
uopt (mm/s) | 68.88 | 0.24 | 6.24 | |
Hmin (mm) | 0.10 | 0.32 | 0.34 | |
R2a | 0.95 | 0.99 | 0.97 |
Pearson’s correlation coefficient between the experimental values and the theoretical model.
The van Deemter parameters showed higher values for peptide 2 in the three evaluated columns; for peptide 1, a decrease in the Hmin value was observed, suggesting that the efficiency is possibly affected by the physicochemical properties of the peptide, such as molecular weight, load, and hydrophobicity, among others. The results show that the value of term A had the greatest variation in the three columns for both analytes. This behavior is in accordance with the fact that term A corresponds to the Eddy diffusion coefficient and that the main difference in the columns is the nature of the solid support and its functionalization. The monolithic column showed lower A values (4 to 6 times, compared to the other columns). This can be attributed mainly to the stationary phase, which consists of a monolithic structure containing macropores and micropores that allows the analytes to interact with the stationary phase more efficiently in the evaluated flow range.
3.2. Chromatographic Method Development Using the k* Concept
From previous results, the monolithic column was selected to develop the optimized chromatographic method for the analysis of peptide mixtures. Considering the heterogeneity of the analytes present in a mixture of peptides, the first step is to define whether the elution should be isocratic or gradient. Normally, for the development of methods in isocratic analysis, a retention factor (k) between 2 and 10 is desired, and on this premise, the method is optimized. It is advisable to use gradient elution to solve complex peptide mixtures since chromatographic methods are fast and only one chromatographic run is required. Another advantage is that the efficiency of the system is not affected because the composition of the mobile phase changes during the run, so k also changes, preventing the broadening of the peaks of all analytes. For this reason, to develop the method, it is necessary to address the concept of gradient retention factor k*, given by eq 2.(21)
![]() |
2 |
Here, tG is the gradient run time (expressed in minutes), F is the flow rate (expressed in mL /min), Vm is the column void volume (expressed in mL), Δφ is the gradient range, and S is a constant value that depends on the analyte’s molecular weight. By manipulating the variables of eq 2, keeping constant the value of k*, it is possible to transfer methods to chromatographic systems with different columns to optimize the analysis time without compromising the resolution of the critical peak pair.
Theoretically, if the variables of eq 2 are changed while keeping the value of k* unchanged, the resolution for most analytes should not change significantly; on the other hand, if the value of k* is increased, the resolution will also increase. Using eq 2, we previously optimized a routine chromatographic method for peptide analysis by RP-HPLC.24 In this method, H2O containing TFA (0.05% v/v) was used as solvent A and ACN containing TFA (0.05% v/v) as solvent B. The elution program was a linear gradient from 5 to 70% B in 45 min. This method has worked very well for the analysis of a wide variety of synthetic peptides; however, its biggest disadvantages are the analysis time (70 min) and that it does not adequately solve complex peptide mixtures. The aim of this investigation was to adapt this method to the analysis of peptide mixtures, reducing the analysis time and having the resolution of the critical pair close to or greater than 1.5.
Our results using isocratic elution indicated that differences in the physicochemical properties of peptides 1 and 2 significantly influenced the efficiency of the system. Therefore, a complex mixture of peptides was designed as a model to develop the chromatographic method by gradient elution: peptides 1 and 2 and four synthetic peptides were chosen, which differed in terms of length, structure, volume, and hydrophobicity, among other factors. In Table 3, the peptide sequences and its properties are presented. The peptide sequences were modeled on the ADMETlab 2.0 platform (https://admetmesh.scbdd.com/) to establish some synthetic and physicochemical properties of each peptide. The parameters selected for the characterization in silico of each peptide in this study were molecular weight (MW), synthetic accessibility score (SAscore), aqueous solubility value (S), and n-octanol/water distribution coefficient (P). The parameters MW and SAscore give information on the size and shape of the molecule, which are related to the conformation and structural complexity, while the parameters S and P give information on amphipathic, lipophilic, and hydrophilic properties.
Table 3. Physicochemical Properties of Synthesized Peptidesa.
sequence | code | MW (g/mol) | SAscore | log S(g/L) | log P(g/L) |
---|---|---|---|---|---|
RRWQWR | 1 | 984.55 | 5.15 | –0.35 | 0.25 |
(RRWQWRFKKLG)2K-Ahx | 2 | 3341.97 | 9.42 | 1.69 | 5.01 |
KQNLAEG | 3 | 757.41 | 4.36 | 1.25 | –1.09 |
RRWQWRFKKLG | 4 | 1558.91 | 6.59 | 0.22 | 3.88 |
RRWQWRWQWR | 5 | 1641.87 | 6.76 | –0.81 | 2.96 |
RWQW-Dip-WQWR | 6 | 1552.76 | 6.52 | –0.56 | 5.60 |
CYFQNCPRG | A | 1086.50 | 5.23 | –2.62 | –1.45 |
Mrp-YFQNCPdRG | B | 1071.44 | 6.15 | –1.81 | –0.41 |
HdWAWdFK | C | 873.41 | 3.63 | –3.48 | 2.01 |
dWAWdF | D | 607.29 | 4.68 | –2.84 | 2.63 |
MW: molecular weight; S: aqueous solubility value; P: n-octanol/water distribution coefficient; SAscore: synthetic accessibility score.
The mixture consisted of five monomeric peptides and one dimeric peptide; the molecular weight range of the peptides was wide and ranged from 757. 41 to 3341. 97 g/mol. In addition, the SAscore range was between 4.36 and 9.42, which indicates the wide variability in the complexity of the molecules and their synthetic ease/difficulty; e.g. the SAscore value for dimeric peptide (2) is 9.42, while for its monomeric analogue (4), the value is 6.59, indicating the greatest synthetic difficulty for the dimeric peptide. In addition, peptides (2) and (6) contain the unnatural amino acids Ahx (6-aminohexanoic acid) and Dip (diphenylalanine), respectively, which confer hydrophobicity to these sequences. The log S and log P values of the six peptides indicate that these molecules have significant differences in their hydrophobic and hydrophilic properties; e.g., peptide (1) has the lowest hydrophobicity (P = 0.08 g/L), while peptide (6) has the highest hydrophobicity (P = 395 461 g/L), the difference in the P value of these peptides being 6 orders of magnitude.
The parameters of eq 2 used for the design and development of the gradient elution methods can be seen in Table 4 and Figure 1. Initially, the mixture was analyzed under the conditions of method I, previously reported,24 using the Agilent Zorbax XDB-C18 packed column (150 × 4.6 mm; particle size 3 μm). As can be seen, the chromatographic profile shows six peaks, which correspond to the six peptides of the mixture, where it is possible to distinguish the critical peak pair between peptides (2) and (5), with a resolution (Rs) of 1.47 (Figure 2.I). The chromatographic profile shows that after 30 min, no signals were observed, so the next step was to optimize the method by changing the value of ΔΦ but maintaining the gradient slope, thus reducing the analysis time without affecting Rs (method II A). Rs was used as a comparison parameter for each method and to determine whether the resolution was maintained or improved, especially that of the critical peak pair, and selectivity was also evaluated.
Table 4. Variables of Equation 2 of the Chromatographic Methods Evaluated.
column | method | ΔΦ | F (mL/min) | tG (min) | tGF | Rsa | P (Bar) |
---|---|---|---|---|---|---|---|
XDB-C18 | I | 0.65 | 1 | 45 | 45 | 1.47 | 144 |
II A | 1 | 30 | 1.48 | 144 | |||
Chromolith | II B | 0.43 | 1 | 30 | 30 | 1.46 | 48 |
III | 2 | 15 | 1.46 | 74 | |||
IV | 3 | 10 | 1.49 | 110 | |||
V | 5 | 6 | 1.44 | 183 |
Rs calculated only for the critical pair corresponding to peptides 2 and 5.
Figure 2.
Chromatographic profiles of the methods developed for the analysis of the mixture of synthetic peptides, 1: RRWQWR, 2: (RRWQWRFKKLG)2K-Ahx, 3: KQNLAEG, 4: RRWQWRFKKLG, 5: RRWQWRWQWR, and 6: RWQW-Dip-WQWR. I–V correspond to the methods described in Table 4.
The chromatographic profile (Figure II A) shows that methods I and II B resolve the mixture in the same way, the Rs of the critical pair being practically the same. However, to continue optimizing the method, the workflow rate must be increased, which is not recommended because the pressure reached with method II A is 144 bar (2080 psi). So, the next step was to transfer method II A to the monolithic column, which allows increasing the workflow; as can be seen, the chromatographic profile of method II B (Figure II B) is the same as that observed for method II A, and the Rs of the critical pair was 1.46.
According to eq 2, if the pressure of the system does not increase significantly, we can increase the flow rate (F) and decrease the time of the gradient (tG), so that the product tGF remains constant. With this strategy, we reduce the time of analysis without decreasing the Rs since the numerator of the equation remains constant and therefore k* as well. Methods III (2 mL/min;74 Bar), IV (3 mL/min;110 Bar), and V (5 mL/min;183 Bar) were designed to progressively increase the workflow up to 5 mL/min. The bimodal porosity of the stationary phase of the monolithic column allows increasing the flow without significantly increasing the pressure of the system.14 As can be seen in Figure 2 and Table 5, as the flow increased, the analysis time decreased: method II required an analysis time of 12.5 min; method IV, 7.2 min; and method V, 6.1 min. In this way, five chromatographic methods were designed and optimized for the separation of a complex mixture of peptides. The results show that it was possible to reduce the analysis time from 70 min (initial method) to 25 min (methods I and II A) using a packed column. In addition, the method was transferred from a packaged column to a monolithic column (method II B), and from these three, methods were designed whose analysis times were 12.5 (method III), 7.2 (method IV), and 6. 1 min (method V). The chromatographic profile observed for all methods was similar without affecting the resolution of the critical pair. Method V has the shortest analysis time and a system pressure of less than 200 bar, which makes it advantageous compared to the other methods (Figure 2, Table 5). It is important to highlight that it was not necessary to change the temperature during chromatographic method optimization, which is a great advantage facilitating its implementation and transfer.
Table 5. Peptide Retention Times for Each Chromatographic Method.
retention
time tR (min) |
||||||
---|---|---|---|---|---|---|
method | 3 | 1 | 4 | 2 | 5 | 6 |
I | 5.2 | 13.1 | 16.1 | 18.4 | 18.6 | 25.2 |
II | 5.2 | 13.2 | 16.3 | 18.5 | 18.7 | 25.2 |
III | 1.4 | 6.6 | 8.4 | 9.6 | 9.8 | 12.5 |
IV | 0.6 | 4.3 | 5.2 | 5.7 | 5.8 | 7.2 |
V | 0.5 | 3.8 | 4.5 | 4.9 | 5.1 | 6.1 |
Sequences of peptides 1, 2, 4, 5, and 6 are derived from the LfcinB sequence and contain hydrophobic amino acids (Trp) and positively charged amino acids (Arg).4,22 These peptides are amphipathic because the side chains of the Arg are oriented toward the same side forming a polar face, while the side chains of the Trp are oriented toward the other side forming a hydrophobic face. In polar (aqueous) media, the hydrophobic faces of peptides tend to interact with each other in such a way that the polar faces are exposed to the polar environment, so these hydrophobic interactions promote the aggregation of peptides. When the peptide mixture was dissolved in solvent A (0. 05% TFA in water) and stored at 4 °C for two days, precipitate formation was observed, possibly due to peptide aggregation. Subsequently, the peptide mixture was dissolved in a mobile phase containing 95% A (0. 05% TFA in water) and 5% B (0. 05% TFA in ACN) and was not observed in precipitate, suggesting that peptide aggregation can be prevented using mobile phases containing solvent B in a concentration equal to or greater than 5%. Our results showed that the mixture containing peptides 2 (linear) and 4 (dimeric), which have the same sequence, was efficiently separated with high resolution, suggesting that the methodology developed here is unaffected by the aggregation of amphipathic peptides.
To establish the applicability of method V, this method was evaluated for the separation of a mixture containing four peptides used in doping; GHRP-4 and GHRP-6 peptides exhibit antioxidant activity, while vasopressin and its analogue desmopressin exhibit antidiuretic activity.25−27 The applicability of this method for the characterization of these peptides, classified by WADA in groups S2 and S5 of the list of prohibited substances, is of particular interest in doping control.
Figure 3 shows the chromatographic profile obtained for the mixture of doping peptides using method V. The signals of the four peptides can be observed, as well as the fact that they are separated with high values of Rs, indicating that this method is suitable for the analysis of mixtures containing peptides with different physicochemical properties, such as molecular weight, hydrophobicity, hydrophilicity, amphipathicity, and structural conformation, among others. Note that for the transfer of this method, it is necessary to consider the correction of the dwell volume (vD) or dwell time (tD).21,23 The equipment used in this study has a vD of 2 mL.Table 6 shows a comparative chart between the method optimized in this research (method V) and other recently reported methods for peptide analysis by conventional HPLC. As can be seen, many of the commonly used methods report analysis times of between 40 and 70 min, while method V achieves an efficient separation of peptides of different chemical natures in just 7 min, which shows that the k* concept is a very useful tool for method transfer and the Chromolith monolithic column is an excellent option for the analysis of these molecules.
Figure 3.
RP-HPLC analysis of a complex mixture of four peptides used in doping, using optimized method V.
Table 6. Comparative Chart between the Method Optimized in This Research (Method V) and Other Recently Reported Methods for Peptide Analysis by Conventional HPLC.
column | Δ%B (gradient) | chromatographic conditions | analysis time (min) | refs |
---|---|---|---|---|
XBridge BEH130 C18 (10 × 150 mm) | 0–90% B, ACN 0.1%, TFA | HPLC (Waters Company), 5 mL/min Detection at 215 nm | 50 | Wen et al.28 |
Shim-pack Inertsil ODS–SP (4.6 × 250 mm, 5 μm) | 0–100% B, A (0.1% TFA in H2O), B (0.1% TFA in ACN) | HPLC Waters 2695, 1 mL/min, Sephadex G-15 gel chromatography | 40 | Ma et al.29 |
C4 and C18 columns | linear gradients, of 0.1% TFA in ACN | HPLC using C4 and C18 columns | 40 | Laxman et al.30 |
XBridge OST C18 (5 μm, 10 mm × 250 mm) | 0–50% B, A (0.1% TFA in H2O), B (0.1% TFA in 80% ACN) | HPLC waters 2695, 5 mL/min, three-step gradient | 70 | Kim et al.31 |
Agilent Eclipse XDB-C18 (4.6 × 150 mm, 3.5 μm) | 5–70% B, A (0.05% TFA in H2O), B (0.05% TFA in ACN) | Agilent 1200 HPLC, 1.0 mL/min | 70 | León-Calvijo et al.24 |
Chromolith high resolution RP-18e (4.6 × 50 mm) | 5–48% B, A (0.05% TFA in H2O), B (0.05% TFA in ACN) | ultimate 3000 HPLC (Thermo Scientific) flow: 5 mL/min | 7 | current method |
4. Conclusions
The efficiency of the analysis of peptides (1) and (2) by means of the RP-HPLC of three columns (packed, core–shell, and monolithic) using the van Deemter model was evaluated. It was found that the monolithic column presented the lowest HETP values and therefore the best efficiency for the analysis of these peptides. Additionally, this column allowed working at flows of up to 5 mL/min without affecting efficiency. Using the k* concept, it was possible to develop and optimize five methods for analyzing a complex mixture containing peptides from a previously reported method. Also, it was possible to transfer the method from a packed column to a monolithic column, and it was possible to go from an analysis time of 70 min to the one of 6.1 min (7 times faster) while maintaining a similar chromatographic profile and the same resolution of the critical peak pair.
Acknowledgments
This research was funded by MINCIENCIAS Grant Number Code 66986, RC 845-2019. Project: “Obtención de un prototipo peptídico promisorio para el desarrollo de un medicamento de amplio espectro para el tratamiento del cáncer de colon, cuello uterino y próstata”.
Author Contributions
The manuscript was written through contributions of all authors. All authors have given approval to the final version of the manuscript.
The authors declare no competing financial interest.
References
- Insuasty Cepeda D. S.; Pineda Castañeda H. M.; Rodríguez Mayor A. V.; García Castañeda J. E.; Maldonado Villamil M.; Fierro Medina R.; Rivera Monroy Z. J. Synthetic Peptide Purification via Solid-Phase Extraction with Gradient Elution: A Simple, Economical, Fast, and Efficient Methodology. Molecules 2019, 24, 1215 10.3390/molecules24071215. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lau J. L.; Dunn M. K. Therapeutic Peptides: Historical Perspectives, Current Development Trends, and Future Directions. Bioorg. Med. Chem. 2018, 26, 2700–2707. 10.1016/j.bmc.2017.06.052. [DOI] [PubMed] [Google Scholar]
- Cavaco M.; Fraga P.; Valle J.; Andreu D.; Castanho M. A. R. B.; Neves V. Development of Breast Cancer Spheroids to Evaluate Cytotoxic Response to an Anticancer Peptide. Pharmaceutics 2021, 13, 1863 10.3390/pharmaceutics13111863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Insuasty-Cepeda D. S.; Barragán-Cárdenas A. C.; Ochoa-Zarzosa A.; López-Meza J.; Fierro-Medina R.; García-Castañeda J. E.; Rivera-monroy Z. J. Peptides Derived from (RRWQWRMKKLG) 2- K- Ahx Induce Selective Cellular Death in Breast Cancer Cell Lines through Apoptotic Pathway. Int. J. Mol. Sci. 2020, 21, 4550 10.3390/ijms21124550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fosgerau K.; Hoffmann T. Peptide Therapeutics: Current Status and Future Directions. Drug Discovery Today 2015, 20, 122–128. 10.1016/j.drudis.2014.10.003. [DOI] [PubMed] [Google Scholar]
- Wang L.; Wang N.; Zhang W.; Cheng X.; Yan Z.; Shao G.; Wang X.; Wang R.; Fu C. Therapeutic Peptides: Current Applications and Future Directions. Signal Transduction Targeted Ther. 2022, 7, 48 10.1038/s41392-022-00904-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Insuasty-Cepeda D. S.; Barragán-Cárdenas A. C.; Ochoa-Zarzosa A.; López-Meza J. E.; Fierro-Medina R.; García-Castañeda J. E.; Rivera-Monroy Z. J. Peptides Derived from (Rrwqwrmkklg)2- k-Ahx Induce Selective Cellular Death in Breast Cancer Cell Lines through Apoptotic Pathway. Int. J. Mol. Sci. 2020, 21, 4550 10.3390/ijms21124550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Erak M.; Bellmann-Sickert K.; Els-Heindl S.; Beck-Sickinger A. G. Peptide Chemistry Toolbox – Transforming Natural Peptides into Peptide Therapeutics. Bioorg. Med. Chem. 2018, 26, 2759–2765. 10.1016/j.bmc.2018.01.012. [DOI] [PubMed] [Google Scholar]
- Kang L.; Alim A.; Song H. Identification and Characterization of Flavor Precursor Peptide from Beef Enzymatic Hydrolysate by Maillard Reaction. J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2019, 1104, 176–181. 10.1016/j.jchromb.2018.10.025. [DOI] [PubMed] [Google Scholar]
- Wali A.; Mijiti Y.; Yanhua G.; Yili A.; Aisa H. A.; Kawuli A. Isolation and Identification of a Novel Antioxidant Peptide from Chickpea (Cicer Arietinum L.) Sprout Protein Hydrolysates. Int. J. Pept. Res. Ther. 2021, 27, 219–227. 10.1007/s10989-020-10070-2. [DOI] [Google Scholar]
- Zhong H.; Abdullah; Zhang Y.; Deng L.; Zhao M.; Tang J.; Zhang H.; Feng F.; Wang J. Exploring the Potential of Novel Xanthine Oxidase Inhibitory Peptide (ACECD) Derived from Skipjack Tuna Hydrolysates Using Affinity-Ultrafiltration Coupled with HPLC–MALDI-TOF/TOF-MS. Food Chem. 2021, 347, 129068 10.1016/j.foodchem.2021.129068. [DOI] [PubMed] [Google Scholar]
- Kim J. M.; Liceaga A. M.; Yoon K. Y. Purification and Identification of an Antioxidant Peptide from Perilla Seed (Perilla Frutescens) Meal Protein Hydrolysate. Food Sci. Nutr. 2019, 7, 1645–1655. 10.1002/fsn3.998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pfaunmiller E. L.; Paulemond M. L.; Dupper C. M.; Hage D. S. Affinity monolith chromatography: a review of principles and recent analytical applications. Anal. Bioanal. Chem. 2014, 405, 2133–2145. 10.1007/s00216-012-6568-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sunde H.; Ryder K.; Bekhit A. E. D. A.; Carne A. Analysis of Peptides in a Sheep Beta Lactoglobulin Hydrolysate as a Model to Evaluate the Effect of Peptide Amino Acid Sequence on Bioactivity. Food Chem. 2021, 365, 130346 10.1016/j.foodchem.2021.130346. [DOI] [PubMed] [Google Scholar]
- Ehab Ibrahim A.; Hashem H.; Elhenawee M.; Saleh H. Monolithic and Core-Shell Particles Stationary Phase Morphologies in Protein Analysis; Peptide Mapping of Erythropoietin Hormone and Determination of Carbetocin. Ann. Pharm. Fr. 2020, 78, 206–216. 10.1016/j.pharma.2020.01.008. [DOI] [PubMed] [Google Scholar]
- Field J. K.; Euerby M. R.; Petersson P. Investigation into Reversed Phase Chromatography Peptide Separation Systems Part III: Establishing a Column Characterisation Database. J. Chromatogr. A 2020, 1622, 461093 10.1016/j.chroma.2020.461093. [DOI] [PubMed] [Google Scholar]
- Porto D. L.; da Silva A. R. R.; de Santana Oliveira A.; Nogueira F. H. A.; Pedrosa M. D.; Aragão C. F. S. Development and Validation of a Stability Indicating HPLC-DAD Method for the Determination of the Peptide Stigmurin. Microchem. J. 2020, 157, 104921 10.1016/j.microc.2020.104921. [DOI] [Google Scholar]
- Howard J. W.; Kay R. G.; Pleasance S.; Creaser C. S. UHPLC for the Separation of Proteins and Peptides. Bioanalysis 2012, 4, 2971–2988. 10.4155/bio.12.283. [DOI] [PubMed] [Google Scholar]
- Deborah M.; Layne J.; Chitty M.; Farkas T. Optimising Core-Shell UHPLC Columns for Improving Protein and Peptide Separations. Chromatogr. Today 2011, 33–36. [Google Scholar]
- Vergel Galeano C. F.; Rivera Monroy Z. J.; Rosas Pérez J. E.; García Castañeda J. E. Efficient Synthesis of Peptides with 4-Methylpiperidine as Fmoc Removal Reagent by Solid Phase Synthesis. J. Mex. Chem. Soc. 2014, 58, 386–392. [Google Scholar]
- Gritti F.; Guiochon G. The van Deemter Equation: Assumptions, Limits, and Adjustment to Modern High Performance Liquid Chromatography. J. Chromatogr. A 2013, 1302, 1–13. 10.1016/j.chroma.2013.06.032. [DOI] [PubMed] [Google Scholar]
- Pápai Z.; Pap T. L. Determination of Chromatographic Peak Parameters by Non-Linear Curve Fitting Using Statistical Moments. Analyst 2002, 127, 494–498. 10.1039/b111304f. [DOI] [PubMed] [Google Scholar]
- Meyer V. R.Practical High-Performance Liquid Chromatography, 2010.
- León-Calvijo M. A.; Leal-Castro A. L.; Almanzar-Reina G. A.; Rosas-Pérez J. E.; García-Castañeda J. E.; Rivera-Monroy Z. J. Antibacterial Activity of Synthetic Peptides Derived from Lactoferricin against Escherichia Coli ATCC 25922 and Enterococcus Faecalis ATCC 29212. Biomed Res. Int. 2015, 2015, 1–8. 10.1155/2015/453826. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gil J.; Cabrales A.; Reyes O.; Morera V.; Betancourt L.; Sánchez A.; García G.; Moya G.; Padrón G.; Besada V.; González L. J. Development and Validation of a Bioanalytical LC-MS Method for the Quantification of GHRP-6 in Human Plasma. J. Pharm. Biomed. Anal. 2012, 60, 19–25. 10.1016/j.jpba.2011.11.007. [DOI] [PubMed] [Google Scholar]
- Thomas A.; Solymos E.; Schänzer W.; Baume N.; Saugy M.; Dellanna F.; Thevis M. Determination of Vasopressin and Desmopressin in Urine by Means of Liquid Chromatography Coupled to Quadrupole Time-of-Flight Mass Spectrometry for Doping Control Purposes. Anal. Chim. Acta 2011, 707, 107–113. 10.1016/j.aca.2011.09.027. [DOI] [PubMed] [Google Scholar]
- Judák P.; Esposito S.; Coppieters G.; Van Eenoo P.; Deventer K. Doping Control Analysis of Small Peptides: A Decade of Progress. J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2021, 1173, 122551 10.1016/j.jchromb.2021.122551. [DOI] [PubMed] [Google Scholar]
- Wen C.; Zhang J.; Feng Y.; Duan Y.; Ma H.; Zhang H. Purification and identification of novel antioxidant peptides from watermelon seed protein hydrolysates and their cytoprotective effects on H2O2-induced oxidative stress. Food Chem. 2020, 327, 127059 10.1016/j.foodchem.2020.127059. [DOI] [PubMed] [Google Scholar]
- Ma B.; Guo Y.; Fu X.; Jin Y. Identification and antimicrobial mechanisms of a novel peptide derived from egg white ovotransferrin hydrolysates. LWT 2020, 131, 109720 10.1016/j.lwt.2020.109720. [DOI] [Google Scholar]
- Laxman K.; Reddy B. P. K.; Robinson A.; Srivastava R.; Ravikanth M. Cell Penetrating Peptide Conjugated BF2-Oxasmaragdyrins as NIRF Imaging and Photo-Thermal Agents. ChemMedChem 2020, 15, 1783 10.1002/cmdc.202000401. [DOI] [PubMed] [Google Scholar]
- Kim J. M.; Liceaga A. M.; Yoon K. Y. Purification and identification of an antioxidant peptide from perilla seed (Perilla frutescens) meal protein hydrolysate. Food Sci. Nutr. 2019, 7, 1645–1655. 10.1002/fsn3.998. [DOI] [PMC free article] [PubMed] [Google Scholar]