Abstract
Oxytocin has become a popular analyte in behavioral endocrinology in recent years, due in part to its roles in social behavior, stress physiology, and cognition. Urine samples have the advantage of being non-invasive and minimally disruptive to collect, allowing for oxytocin measurements even in some wild populations. However, methods for urinary oxytocin immunoassay have not been sufficiently optimized and rigorously assessed for their potential limitations. Using samples from oxytocin knockout (KO) and wildtype (WT) mice, we find evidence of considerable interference in unextracted urine samples, with similar distributions of measured oxytocin in both genotypes. Importantly, although this interference can be reduced by a reversed-phase solid-phase extraction (SPE), this common approach is not sufficient for eliminating false-positive signal on three immunoassay kits. To better understand the source of the observed interference, we conducted epitope mapping of the Arbor Assays antibody and assessed its cross-reactivity with known, biologically active fragments of oxytocin. We found considerable cross-reactivity (0.5–52% by-molarity) for three fragments of oxytocin that share the core epitope, with more cross-reactivity for longer fragments. Given the presence of some cross-reactivity for even the tripeptide MIF-1, it is likely that many small protein metabolites might be sufficiently similar to the epitope that at high concentrations they interfere with immunoassays. We present a new mixed-mode cation-exchange SPE method that minimizes interference—with knockout samples measuring below the assay’s limit of detection—while effectively retaining oxytocin from the urine of wildtype mice. This method demonstrates good parallelism and spike recovery across multiple species (mice, dogs, sifakas, humans). Our results suggest that immunoassays of urine samples may be particularly susceptible to interference, even when using common extraction protocols, but that this interference can be successfully managed using a novel mixed-mode cation exchange extraction. These findings imply that previous conclusions based on urinary oxytocin measurements—especially those involving unextracted samples—may need to be reassessed.
Keywords: Oxytocin, Solid-phase extraction, ELISA, Urine, Matrix interference, Knockout mice
1. Introduction
Oxytocin has been the focus of considerable research effort over the last several decades, especially in behavioral endocrinology and pharmacology. Early work demonstrated its importance for maternal behavior (Pederson and Prange, 1979; Van Leengoed et al., 1987) and pair-bonding (reviewed in Carter et al., 1995). Recently, evidence has been mounting for oxytocin’s role in a myriad of social behaviors and cognitive processes, as well as stress physiology (reviewed in Feldman, 2012; Carter, 2014; Froemke and Young, 2021).
Making sense of the oxytocin literature can be challenging, partially due to the variety of experimental methods used. For example, administration of exogenous oxytocin can produce dosage-dependent results (Bales et al., 2007; Cardoso et al., 2013; Peters et al., 2014), and the effects of administration are not necessarily the same as the effects of endogenous oxytocin (Crockford et al., 2014). While administration studies may be fruitful for pharmacology, studies of endogenous oxytocin are crucial for understanding the biology of the oxytocin system, its function, and its evolution. However, studies of endogenous oxytocin concentrations have often struggled to reliably measure oxytocin, and there are disagreements regarding the best approaches for measuring oxytocin, especially by immunoassay (e.g. McCullough et al., 2013; Lefevre et al., 2017; MacLean et al., 2019). Furthermore, endogenous oxytocin measurements and patterns of change are often uncorrelated across biological matrices, especially urine (Amico et al., 1987; Feldman et al., 2011; Weber et al., 2017). While there are good reasons that measurements might differ across matrices— specifically different rates of clearance and windows of integration (e.g. urine accumulates oxytocin over the entire time since the last void) (Amico et al., 1987; Mitsui et al., 2011)—it is unclear to what extent measurements across matrices are functionally similar. Complicating these comparisons, sample preparation techniques—such as extraction—that have been developed and validated in one matrix are often assumed to work equally well in other matrices, despite their different compositions.
Using blood plasma samples from oxytocin knockout (KO) and wildtype (WT) mice, we previously demonstrated that while some commercially available immunoassay kits (e.g. Enzo Life Sciences) are susceptible to interference, others (e.g. Arbor Assays) are not, even with unextracted (diluted 1:8) mouse plasma samples (Gnanadesikan et al., 2021). Here, we turn this approach towards improving and validating urinary oxytocin (uOT) measurements. While plasma samples are often available in laboratory research settings, blood collection is often impossible or disruptive. In studies of wild populations, blood collection usually involves trapping animals, and even in captive settings or human studies, it may affect behaviors of interest. Thus, the advantages of non-invasive hormone measures for behavioral endocrinology have long been appreciated both in general (Strier and Ziegler, 2005) and for oxytocin research specifically (Crockford et al., 2014). However, developing reliable methods has been challenging.
Urine samples provide a promising avenue for non-invasive behavioral endocrinology for several reasons: 1) urine can be caught in a container, on an absorbent material, or pipetted up off surfaces without handling the animal; 2) urination may occur throughout the day, enabling collection at multiple time points; 3) many species produce volumes of urine that are sufficient to assay multiple analytes in the same sample.
1.1. Urinary oxytocin and behavior across species
Many studies have already utilized uOT measurements and found interesting associations with social behavior. In humans, uOT has been associated with perceptions of a romantic partner’s bonding behavior, regardless of gender (Algoe et al., 2017), and with performance on a facial visual search task in men (Saito et al., 2014). In children, a mother’s touch and speech after a stressful event increased uOT while decreasing salivary cortisol (Seltzer et al., 2010). Interestingly, in mothers, uOT was higher after interacting with other people’s children compared to their own children, perhaps due to oxytocin’s role in the onset—rather than maintenance—of maternal behavior, or perhaps simply due to a novelty effect (Bick and Dozier, 2010).
Increasingly, uOT has been used in studies of human-animal interactions. Nagasawa et al., (2009, 2015) demonstrated that uOT in dogs and their owners was associated with mutual gazing. Mitsui et al. (2011) suggested that oxytocin could be used as a biomarker of positive emotion, finding that dogs’ uOT increased after feeding, exercising, and petting, but not drinking water. In contrast, Marshall-Pescini et al. (2019) found that affiliative physical contact between dogs and people did not produce consistent changes in uOT levels measured in either the dogs or their owners. Wirobski, Range et al. (2021) found that in pet dogs, uOT was associated with physical contact with their owner; however, in hand-raised but pack-living dogs and wolves, there was no association between measured uOT and physical contact with a bonded human, suggesting the importance of the animal’s life experience and the strength of the human-animal bond.
Primatologists have also made use of uOT measurements in a variety of lab and field studies. Using SPE and high-performance liquid chromatography (HPLC), Seltzer and Ziegler (2007) found that a captive male common marmoset’s uOT was lower during a 48-hour social isolation period than during the subsequent visual reintroduction to his mate. In captive tamarins, Snowdon et al. (2010) found high levels of individual uOT variation, with pair-bonded mates exhibiting similar concentrations that were partially explained by their affiliative behaviors. Similarly, in captive common marmosets, strongly bonded pairs exhibited synchronous fluctuations in uOT concentrations (Finkenwirth et al., 2015), and uOT was associated with infant care behaviors such as infant-licking and proactive food sharing in all group members (Finkenwirth et al., 2016). In field studies of baboons, higher uOT concentrations have been associated with a female’s estrous stage and their proximity maintenance with consort partners (Moscovice and Ziegler, 2012). In field studies of chimpanzees, uOT has been associated with social bonding (Crockford et al., 2013), food sharing (Wittig et al., 2014), and intergroup conflict (Samuni et al., 2017). Lastly, in field studies of bonobos, female-female sexual behavior was associated with increased oxytocin levels, compared with both a baseline condition and inter-sexual copulation (Moscovice et al., 2019).
Thus, uOT studies have already documented a variety of intriguing associations with social behavior. However, there are differences across studies and species regarding the specific behavioral associations found, and generalizations remain elusive. Furthermore, some studies have failed to find the expected effects (e.g. Marshall-Pescini et al., 2019), and there is likely publication bias in this regard.
1.2. Challenges for immunoassay of urinary neuropeptides
Compounding these challenges in interpreting the urinary oxytocin literature, the specificity and reliability of uOT techniques has not always been tested rigorously. Even when extensive validation work has been conducted (e.g. Seltzer et al., 2010; Wirobski, Schaebs et al., 2021), assessment of potential false-positive signal has not been possible, lacking a “ground truth” (i.e. known concentrations). Additionally, although many studies have employed some form of extraction, it remains unknown to what extent these methods are sufficiently selective to retain oxytocin while eliminating potentially interfering molecules. Urine contains a large quantity and variety of metabolites (Bouatra et al., 2013) that we expect might interfere with immunoassays, and thus extraction and immunoassay specificity may be particularly important for this matrix. Are inconsistencies across studies due to meaningful differences in species, behaviors, and sampling windows, or are they methodological artifacts due to differences in extraction and immunoassay? To shed light on these methodological questions, in this study we tested multiple methods on both oxytocin KO and WT mice, searching for techniques that minimized interference in the KO samples while maximizing signal in the WT samples.
The specificity of immunoassays can be affected by many aspects of their chemistry. The antibody is particularly important, as different antibodies may bind different portions of a molecule. This region of an antigen that is bound by the antibody is known as the epitope. Aiming to shed light on potential sources of interference, we worked with Pepscan Presto BV (Lelystad, The Netherlands) to perform epitope mapping of the Arbor Assays antibody. Motivated by these results, we also tested the cross-reactivity of the Arbor Assays antibody with three known, biologically active fragments of oxytocin (Uvnäs Moberg et al., 2019), as well as the effects of two extraction methods on retention of these fragments.
Lastly, using a mixed-cation exchange method that minimized interference in knockout mouse samples, we tested its analytical validity on the Arbor Assays kit across a variety of species (mice, sifakas, dogs, and humans). We focused on the Arbor Assays kit because it is the predominate platform that we use in our work (Gnanadesikan et al., 2021), although we ran a small number of samples on the Cayman Chemical and Enzo Life Sciences kits for comparison. We used samples from species that the investigators work with regularly in their research and that represent different orders of mammals with potentially interesting biological differences (i.e. size, diet, urine concentration).
2. Methods
2.1. Mouse subjects and sample collection
As in our previous work (Gnanadesikan et al., 2021), Oxftm1Zuk mice (Nishimori et al., 1996) were maintained on a C57BL/6J background and bred at Florida State University; all breeding and handling procedures were approved by the Institutional Animal Care and Use Committee (IACUC) following the Guide for the Care and Use of Laboratory Animals (Protocols 2020–21 and 202 000 042). Oxt+/− breeder pairs were continuously housed. The first morning of the appearance of a litter was noted as postnatal day 0 (P0). Mice were weaned, tagged, and tailed for genotyping on P21 and group housed by sex. After genotyping, same-sex mice were re-housed by genotype, so that only mice of the same genotype were housed together. Mice were housed on a 12:12 L:D cycle in open wire-top caging with wood chip bedding and provided ad libitum food (LabDiet Rodent 5001) and water. In total, samples from 159 mice of this strain (WT: 73, KO: 86) were used in this study. Additionally, adult C57BL/6J (B6) wildtype mice were purchased from The Jackson Laboratory (total n = 20).
Urine samples were passively collected from adult male and female mice during routine handling on the day of planned euthanasia for other purposes. During live handling to confirm the mouse identification (ID number) by ear tag prior to euthanasia, the mouse was held over a microcentrifuge tube to collect urine. After confirming the mouse ID number, the tube was closed, labeled with the ID number, and frozen. Samples were shipped frozen to the University of Arizona on dry ice and stored at − 80 °C until time of assay. Each urine sample, identifiable by ID number, was assigned a genotype code by E.A.D.H. to enable experimental design, assay layout planning, and sample pooling (within genotype when necessary), while retaining sample blinding for G.E.G. for assays directly comparing knockout and wildtype individuals.
2.1.1. Genotyping
DNA from tail samples was genotyped for Oxt alleles using a common forward primer (5’-TCAGAGATTGAACAAGACGCC) and specific reverse primers for WT (5’-TCAGAGCCAGTAAGCCAAGC) and KO (5’-ACTTGTGTAGCGCCAAGTGC). Using a hot start and 40 cycles of 30 s each of 94 °C, 57 °C, and 72 °C, the primers generated a wildtype allele of approximately 500 bp and a knockout allele of approximately 180 bp, resolved by gel electrophoresis.
2.1.2. Samples in each analysis
For a summary of mouse samples used in this study, see table A3. The first experiment was conducted on unextracted urine samples: 10 wildtype Oxt+/+ samples (7 female, 3 male) and 14 knockout Oxt−/− samples (8 female, 6 male). The second experiment, as well as the methods development work reported in the appendix used a series of sample pools. In total, these pools used samples from 76 wildtype individuals (mix of C57BL/6J and Oxt+/+) and 62 knockout individuals (Oxt−/−). For experiment three, some individual samples were pooled across 2–3 individuals within the same—blinded—genotype code, due to limited sample volume. All samples were from male mice. Six individual Oxt+/+ wildtype samples were used; 10 individual Oxt−/− knockout samples were pooled as necessary to create 6 pooled samples of 1–3 individuals each.
2.2. Sifaka samples
Verreaux’s sifaka (Propithecus verreauxi) samples were collected from a habituated wild population at Ankoatsifaka Research Station in Kirindy Mitea National Park, Madagascar, under IACUC Protocol #13–170 from the University of Arizona and IACUC Protocol #2017–152 from the University of Texas at Austin. Fresh urine samples were collected from known, identified individuals by catching the stream directly with foil, catching the stream with foil off of tree trunks, or pipetting urine off of foliage with disposable pipettes. Samples were pipetted into cryovials, which were immediately placed into a thermos with either instant ice packs or tubes of frozen water until transferred into a liquid nitrogen tank within 4 h. Samples were kept frozen in a liquid nitrogen dewar during transport from the field site to Antananarivo and while awaiting shipment to the US. They were shipped to the University of Arizona in a charged liquid nitrogen shipper and frozen at − 80 °C until analysis at the Laboratory of the Evolutionary Endocrinology of Primates (LEEP).
We created one pool (of two samples) for assessment of parallelism and a second pool (adding seven additional samples) to assess spike recovery. Sample information is provided in the appendix (table A5).
2.3. Dog and human samples
Dog (Canis lupus familiaris) and human (Homo sapiens) urine pools were created through volunteer collections among laboratory staff/students and their pet dogs, as well as using extra volume collected for ongoing studies, under University of Arizona IACUC Protocol #16–175 and IRB Protocol #1 808 883 345.
2.4. Solid-phase extractions
2.4.1. Hydrophilic-Lipophilic Balanced (HLB)
Hydrophilic-Lipophilic Balanced reversed-phase extractions were conducted using the Oasis PRiME HLB 1 cc cartridges (Waters Corporation, Milford, MA, USA, Part Number: 186 008 055) and a positive pressure manifold (Biotage PRESSURE+48). All samples were diluted into an equal volume of 0.1% trifluoroacetic acid (TFA) in water, vortexed for at least 30 s, and then centrifuged. Exact speeds and times varied, in part due to sample volumes; all samples were centrifuged at either 10 000 rpm (RCF = 9632 x g) for 5 min or 5000 rpm (RCF = 4696 x g) for 15 min.
SPE cartridges were conditioned first with 1 ml acetonitrile (ACN) with 0.1% TFA, and then with 1 ml 0.1% TFA in water. The entire sample supernatant was loaded onto the cartridge. Each cartridge was then washed with 1 ml 10% ACN, 0.1% TFA. Finally, samples were eluted with 2 ml 30% ACN, 0.1% TFA. This procedure is identical to that reported in Gnanadesikan et al. (2021), although 2 ml of eluant was used in all cases, to maximize recovery. It is worth noting that the ACN content of the elution (30%) is lower than is commonly used in other extraction protocols (e.g. 60% or 85%). In reversed-phased SPE, higher organic content is expected to elute more molecules—especially more hydrophobic molecules—which could include additional sources of interference. Thus, elution with lower ACN content should theoretically result in a more selective extraction.
2.4.2. Mixed-mode Cation Exchange (MCX)
Mixed-mode Cation Exchange extractions were performed using Oasis PRiME MCX 1 cc cartridges (Waters Corporation, Milford, MA, USA, Part Number: 186 008 917) and a positive pressure manifold (Biotage PRESSURE+48). All samples were diluted into an equal volume of loading buffer (200 mM ammonium formate with phosphoric acid to achieve pH 5). Samples were vortexed for at least 30 s and then centrifuged, as above, at either 10 000 rpm (9632 × g) for 5 min or 5000 rpm (4696 × g) for 15 min
The MCX cartridges were first conditioned with 1 ml methanol (MeOH) and 1 ml of deionized water (Thermo Scientific #751 628). The entire supernatant of the sample was then loaded onto each cartridge. Cartridges were washed with successive 1 ml washes: 1 ml deionized water, 4 ml wash buffer (60% loading buffer, 40% MeOH), 1 ml deionized water. Finally, each sample was eluted in 2 ml of elution buffer (50% MeOH, 50% ammonium hydroxide solution). The pH of the elution buffer was 12 in the three-way kit comparison; the subsequently optimized protocol—used for the individual-level analyses—uses pH 11. We experimented with various combinations of loading, wash, and elution buffers and volumes to optimize this protocol, as reported in the appendix.
2.4.3. All extracted samples
The eluted samples were centrifuged at 3000 rpm (1690 × g) for 5 min to ensure that liquid on the side of the tubes settled at the bottom. All samples were stored at − 80 °C overnight and lyophilized the next day using a CentriVap (Labconco model #7 810 016) with the following settings: CentriVap unheated, cold trap – 80–85 °C, vacuum 0.3–0.4 mbar. When lyophilization was complete, and immediately before assay, samples were reconstituted in assay buffer. Individual samples were reconstituted in 250 μl assay buffer; certain pools for multi-kit comparisons or validations used larger volumes, with concentration factors specified below. Reported concentrations are corrected for the concentration or dilution factors inherent in the extraction.
2.4.4. Extraction efficiency
Extraction efficiencies were determined by extracting spiked buffer. We used the stock oxytocin solution from the Arbor Assays kit and diluted it to 1000 pg/ml in Arbor Assay’s oxytocin assay buffer. Extractions were performed on 250 μl aliquots, while an additional aliquot was frozen for assay unextracted. Reported extraction efficiencies were calculated as (extracted measurement / unextracted measurement) × 100. Extraction efficiency was also assessed in dog urine by adding spiked buffer (10%, ~1000 pg/ml) to a dog urine sample (90%), and measuring the sample with and without the spike, as well as the unextracted spiked buffer. Reported extraction efficiencies were calculated as (spiked urine measurement)/ (0.1 × unextracted spike + 0.9 × urine measurement) × 100. Extraction efficiency was not assessed in other species at this time; however we encourage future studies to assess extraction efficiency in the relevant species and matrix.
2.5. Oxytocin assays
Except where otherwise noted, oxytocin was measured using the Arbor Assays Oxytocin Enyzme Immunoassay (EIA) kit (Catalog #K048-H5), following the recommended assay protocol (although not using the extraction solution provided). The reported sensitivity for this kit is 17.0 pg/ml and the lower limit of detection is 22.9 pg/ml; the limit of detection of a given experiment is dependent on (multiplied or divided by) the dilution or concentration factor inherent in the extraction, specified below. Arbor Assays reports that cross-reactivity is 94.3% for isotocin, 88.4% for mesotocin, and less than 0.15% for vasotocin and arginine vasopressin.
In the three-way kit comparison, we also used the Enzo Life Sciences (Catalog #ADI-900–153A0001) and Cayman Chemical (Catalog #500 440) kits. The manufacturer-reported sensitivities are 15 pg/ml for Enzo and 20 pg/ml for Cayman. The reported cross-reactivities for the Enzo kit are: mesotocin 7.0%, Arg8-vasotocin 7.5%, and < 0.02% for all other reported compounds including Arg8-vasopressin. Reported cross-reactivities for the Cayman kit are 100% for mesotocin and isotocin and < 0.01% for all other reported compounds including Arg8-vasopressin.
2.5.1. Coefficients of variation
For unextracted samples of both genotypes (experiment 1), the mean intra-assay coefficient of variation (CV) between duplicates was 7.1%. For MCX-extracted samples (experiment 3), the mean intra-assay CV for samples that measured above the assay’s limit of detection (n = 6) was 7.4%. Results were not compared across plates (except for the multi-kit comparison).
2.6. Epitope mapping
Epitope mapping of the Arbor Assays antibody was conducted by Pepscan Presto BV (Lelystad, The Netherlands). The basic approach involves synthesizing peptide fragments and testing an antibody’s affinity for each sequence (Geysen et al., 1984). First, Pepscan constructed arrays of peptide fragments to assess the linear and conformational epitope binding; these arrays consisted of all possible 10-amino-acid fragments of oxytocin/neurophysin I prepropeptide, the larger precursor protein from which oxytocin is derived. Next, they performed fine-mapping of the epitope region by using single amino acid substitutions for each of the 9 amino acids in oxytocin to determine how each substitution affected antibody binding. The binding of the antibody to each of the synthesized peptides was tested in a Pepscan-based immunoassay. This approach enables a comparison of the antibody’s affinity for different peptide sequences, which can shed light on the antibody’s specificity; sequences that do not demonstrate high levels of antibody binding are unlikely to interfere with immunoassay measurement, while sequences that show binding levels similar to the identified epitope may cause interference.
2.7. Oxytocin fragments
Given that fragments of oxytocin are known to be formed through multiple metabolic processes (Uvnäs Moberg et al., 2019), we hypothesized that these fragments might cross-react with immunoassay antibodies. We focused on three biologically active fragments of oxytocin that have been studied previously: OT 2–9, OT 4–9, and OT 7–9, where the numbers indicate the residues of oxytocin (a nonapeptide) contained in the fragment. OT 7–9 is more commonly known as melanocyte-inhibiting factor 1 (MIF-1). In order to assess the cross-reactivities of these fragments on the Arbor Assays kit, we purchased MIF-1 from Cayman Chemical (Catalog #24 476) and had OT 2–9 and OT 4–9 synthesized by Biomatik (Ontario, Canada). Since oxytocin is cyclic, with residues 1 and 6 connected by a disulfide bridge, OT 2–9 and OT 4–9 were synthesized with a cystine at position 6 (cysteine-cysteine, connected by a disulfide bridge), which is consistent with known metabolic pathways (Uvnäs Moberg et al., 2019). For more information on these fragments, see Fig. 4.
We reconstituted a single 10 mg aliquot of MIF-1 in 10 ml of deionized water, freezing reconstituted aliquots for future assays. The synthesized fragments came in 1 mg aliquots, so we used a new aliquot for each assay; we reconstituted each aliquot in 1 ml deionized water to create a 1 mg/ml solution, thus variation in these aliquots could contribute to variation in measured cross-reactivities across assays. On the day of each assay, the 1 mg/ml solution was diluted in assay buffer to the starting concentration of the assay. In a first experiment, we created 1 μg/ml starting solutions and then conducted five 1:10 part-to-whole (p:w) serial dilutions for each fragment to identify what orders of magnitude would result in measurements within the standard curve. Informed by these results, we conducted a second assay with serial dilutions targeting the middle of the standard curve (see table A6 for input concentrations). Reported mean cross-reactivities were calculated based on those dilutions that measured within 30–70% of the assay’s maximum binding (B0) as the percentage of the fragment’s known concentration that was measured by the assay. Although immunoassay results are usually reported by weight (e.g. pg/ml), since the fragments are very different sizes, we report cross-reactivities by both molarity and weight. The cross-reactivities-by-molarity represent a direct comparison based on the number of peptides present, while cross-reactivities-by-weight may be helpful in considering what fraction of a measurement might be due to each of these fragments. In a third assay, we explored whether the HLB and MCX extractions removed these fragments. Each fragment was prepared in Arbor Assay’s oxytocin assay buffer to a concentration that we expected would result in approximately 30% of the maximum binding, based on the previous assays. An aliquot was frozen overnight to measure unextracted, and two aliquots were extracted using the HLB and MCX solid phase extraction methods. The extractions were performed on 250 μl of sample, the eluate was lyophilized, and the sample was reconstituted in 250 μl of assay buffer at the time of assay. As a baseline for comparison, the same procedure was applied to a spiked buffer sample composed of Arbor Assay’s oxytocin assay buffer spiked with ~1000 pg/ml of the kit standard.
2.8. Analytical validations
Analytical validations were conducted with wildtype mouse, sifaka, dog, and human sample pools. Each species differed in the amount of sample extracted (and thus starting concentration/dilution factor) but were otherwise treated identically. Parallelism was assessed via serial dilution (2:3, p:w) of urine pools and calculation of the coefficient of variation on corrected concentrations at each dilution (Andreasson et al., 2015). Starting concentration factors for each parallelism were 1.33x for mouse and sifaka, 5.3x for dog, and 16x for human.
Spike recovery was assessed using samples at a single dilution (ratio of volume applied to the cartridge for SPE to the volume reconstituted): 2:5 for mouse, 1:2 for sifaka, 4x for dogs, and 8x for humans. All spike recovery extractions were reconstituted in 250 μl assay buffer; sample volumes were 100 μl for mouse, 125 μl for sifaka, 1 ml for dog, and 2 ml for humans. Spiked samples consisted of 90% sample matrix at the specified dilution and 10% synthetic oxytocin in assay buffer (kit standards 1–5). Percent recovery was calculated as (observed / expected) x 100, where the expected values were measured independently by adding each spike to assay buffer.
2.9. Statistical software
Statistical analyses were conducted in R version 4.1.1 (R Core Team, 2021). For the unextracted urine comparison in experiment 1, distributions were not significantly different from a normal distribution (Shapiro-Wilk’s WWT = 0.948, pWT = 0.644; WKO = 0.955, pKO = 0.677), so Welch’s t-test was used to compare the means. For the HLB extractions in experiment 2 and the three-way kit comparison in experiment 3, single measurements were made on pooled samples and no statistical tests were used. For the individual-level results in experiment 3, sample sizes were small (n = 6) for each genotype, so we used a Mann-Whitney test to avoid normality assumptions. See the appendix for a discussion of power analyses.
3. Results
3.1. Experiment 1: unextracted urine
Given our past success assaying unextracted mouse plasma samples (Gnanadesikan et al., 2021), we began by assessing the genotype contrast in unextracted diluted mouse urine samples on the Arbor Assays EIA kit. Due to limited sample volume for some individuals, some samples were assayed at a 1:10 (p:w) dilution into Arbor Assay buffer, while others were measured at a 1:25 dilution. Based on a serial dilution of a single wildtype C57BL/6J urine sample, we expected both of these dilutions to be within range of the assay (see appendix, table A1).
Knockout samples were indistinguishable from wildtype samples (Mean ± SD: WT = 930.8 ± 512.7 pg/ml, KO = 885.7 ± 295.6 pg/ml; Welch’s t(13.5) = 0.248, p = 0.808; Fig. 1). We interpreted this as indicating a high level of interference in the immunoassay and proceeded to explore various extraction options to eliminate or minimize this interference.
3.2. Experiment 2: basic HLB extraction
Since solid-phase extraction is commonly used to remove interfering molecules from biological matrices, we evaluated the performance of a relatively simple hydrophilic-lipophilic balanced (HLB) extraction. Similar reversed-phase extraction techniques using different cartridges and buffers have frequently been used in oxytocin measurements across matrices and species (e.g. Szeto et al., 2012; Crockford et al., 2013; Bienboire-Frosini et al., 2017), and we had previously validated this specific HLB method in mouse plasma (Gnanadesikan et al., 2021). With urine samples, this extraction resulted in a clear genotype contrast, but there was still considerable interference in the knockout samples (WTpool = 611.9 pg/ml; KOpool = 188.03 pg/ml). Subsequent replications found similar results (not reported). We assessed whether an increased wash volume would eliminate the observed interference, but instead found that we lost signal in both genotype pools, and the genotype contrast was not improved (WTpool = 116.49 pg/ml; KOpool = 72.81 pg/ml).
3.3. Experiment 3: improved MCX extraction method
In order to improve our extraction selectivity, we developed and optimized a new method using a mixed-mode cation exchange (MCX) cartridge. This enabled us to manipulate the retention using both pH and organic content of the loading, wash, and elution buffers. For details of the methods development, see the appendix (Section 3, figures A1–A5).
Using this new method, we then compared the MCX and HLB extraction methods across three different oxytocin immunoassay kits, hypothesizing that differences in the kit chemistries—especially antibodies—might lead to different susceptibilities to interference. We found that the HLB extraction was susceptible to varying levels of interference across kits, but the MCX extraction minimized this interference on all three kits (KO measurements 13—39 pg/ml, corrected concentrations 21—66 pg/ml; WT measurements 124–171 pg/ml, corrected concentrations 206–286 pg/ml; Fig. 2; table A2).
While these results were much improved, we wanted to further minimize the KO interference, with the aim of increasing the WT:KO ratio. We achieved this on the MCX cartridge by decreasing the pH of the elution buffer from 12.0 to 11.0 for a more selective elution. We then tested individual samples (some pooled) to examine the genotype contrast again on the Arbor Assays kit. All knockout samples measured below the assay’s limit of detection. As such, these values should not be interpreted (and in some cases had high CVs due to low measurement); nevertheless, the difference between groups is statistically significant (Mann-Whitney W = 0, p = 0.002; Fig. 3). We therefore used a pH 11.0 elution buffer for all subsequent assays.
Using spiked buffer samples, we also assessed the extraction efficiency of the MCX method; in one assay across three samples, the extraction efficiency was 83%, while in another assay across two samples, the extraction efficiency was 92%. Measured in spiked dog urine, extraction efficiency was 121%.
3.4. Epitope mapping
Given the considerable interference that we observed in unextracted and HLB-extracted urinary samples, we decided to further probe the properties of the anti-oxytocin polyclonal rabbit antibody used in the Arbor Assays EIA. We therefore worked with Arbor Assays and Pepscan to perform both linear and conformational epitope mapping of this antibody. Epitope mapping tests an antibody’s affinity to a library of synthesized peptide fragments, thereby determining its epitope(s) and assessing its specificity.
The epitope mapping results identified a single epitope within oxytocin/neurophysin I prepropeptide, corresponding to residues 2–9 of oxytocin (YIQNCPLG). Further testing of this region— by assessing single amino acid substitutions—revealed that the core epitope is PLG (Fig. 4, A6). Almost any change to these residues resulted in highly disrupted binding, although the magnitude of this change was dependent on the exact substitution; substitutions outside of this core region had less impact on antibody binding (see supplementary data file for raw epitope mapping results).
The linear and loop arrays produced similar results, as did versions allowing the free cysteines to bridge, suggesting that the conformational structure of oxytocin—including the disulfide bridge between the cysteine residues—is not important for epitope recognition, consistent with earlier studies reporting immunoreactivity with the linear form of oxytocin after a reduction/alkylation procedure (Brandtzaeg et al., 2016).
3.5. Oxytocin fragment cross-reactivities and extractions
Given what we learned about the epitope of the Arbor Assays antibody, and especially that the core epitope consists of residues 7–9, we investigated the cross-reactivity of this antibody with known, biologically active fragments of oxytocin (Uvnäs Moberg et al., 2019). Specifically, we purchased OT 2–9, OT 4–9, and OT 7–9, the latter more commonly known as melanocyte-inhibiting factor 1 (MIF-1). Across multiple dilutions, the average cross-reactivities-by-molarity measured within the range of 30–70% of the assay’s maximum binding (B0) were 52% for OT 2–9 (also 52% by-weight), 22% for OT 4–9 (32% by-weight), and 0.5% for MIF-1 (2% by-weight; Figs. 4 and 5; table A6).
Finally, we tested both the HLB and refined MCX extractions on each of the three oxytocin fragments, as well as oxytocin itself. While both extractions retained > 90% of oxytocin itself, extraction considerably reduced the measured oxytocin concentrations for all three fragments. Both extractions removed most of the fragments, reducing the assay’s measurement by 1–2 orders of magnitude. The extractions were more effective at eliminating OT 4–9 and MIF-1; for OT 2–9—the molecule most similar to oxytocin—the MCX extraction removed considerably more of the fragment (measured 117.15 pg/ml) than did the HLB extraction (measured 283.05 pg/ml; see Fig. 6).
3.6. Analytical validations: mouse, sifaka, dog, and human
Using the optimized MCX method, we conducted analytical validations on pooled samples from four species/strains: wildtype mice (Musmusculus), Verreaux’s sifaka (Propithecus verreauxi), domestic dogs (Canis lupus familiaris), and humans (Homo sapiens).
The wildtype mouse pool diluted parallel to the standard curve (CV of corrected concentrations = 10.1%; figure A8) and exhibited good spike recovery for samples measured at a 2:5 dilution factor (mean ± SD = 114 ± 7%). Similarly, pooled sifaka urine samples demonstrated good parallelism (CV of corrected concentrations = 17.5% over full range; 5.11% for lower 5 dilutions; figure A9) and spike recovery measured at a 1:2 dilution (mean ± SD = 114 ± 6%). A dog urine pool diluted parallel to the standard curve (CV of corrected concentrations = 6.0%; figure A10) and exhibited acceptable spike recovery when assessed at a 4x concentration (mean ± SD = 119 ± 9%).
Lastly, a human urine pool diluted relatively parallel to the standard curve, especially for higher concentrations (figure A11). The CV of corrected concentrations over the entire range (16x–2.1x) was 18%; excluding the fourth dilution, which had a duplicate CV of 25%, this rose to 19%. However, the last two dilutions measured somewhat higher than expected, thus for the first three dilutions alone, the CV of corrected concentrations was 6.6%. Human urine also exhibited acceptable spike recovery when measured at an 8x concentration (mean ± SD = 123 ± 8%). It should be noted, however, that this pool had a specific gravity of 1.023 and a corrected concentration of approximately 7 pg/ml. Human population means for specific gravity are generally 1.01–1.02 (Goldberger et al., 1995; Miller et al., 2004). Thus, even with an 8x concentration, some samples will be near or below the limit of detection of this assay, and higher concentration factors may be necessary for more dilute samples. More experimentation is needed to determine whether an ideal protocol exists for samples of all concentrations, or what the best methods are for compensating for the wide variation in specific gravity (and consequently oxytocin concentrations) in human urine. Relatedly, in some subsequent experiments, spike recovery for certain human urine pools was above acceptable thresholds, so further optimization may be necessary for human samples.
4. Discussion
Our results suggest that many existing urinary oxytocin methods may be suboptimal. We found that unextracted urine samples—measured at a 1:10 or 1:25 dilution in assay buffer—were indistinguishable on the Arbor Assays kit across oxytocin knockout and wildtype mice. A basic reversed-phase SPE on the Waters Oasis PRIME HLB cartridge improved the genotype contrast, but still exhibited considerable interference, which was observed at varying levels across three different commercially available immunoassay kits. This interference was eventually minimized with the development of a new SPE method using the Waters Oasis PRiME MCX cartridge, which utilizes a mixed-mode cation exchange chemistry. Although we have not tested all of the cartridge-extraction-kit combinations that have previously been used in the literature, our findings across three different commercially available kits suggest that there is strong potential for interference in immunoassays of urinary oxytocin, even following sample extraction. It is notable that the concentrations we report using the MCX method are lower than those reported with some other methods. For example, Wirobski, Range et al. (2021) reported pet dog concentrations in the range of 77 – 1500 pg/ml (Mean ± SD = 440 ± 290) before correcting for specific gravity (SG), and 130 – 950 pg/ml (Mean ± SD = 430 ± 190) after doing so. The dog pool that we used for both parallelism and spike recovery analyses had concentrations of 40 – 50 pg/ml (no SG correction), and in another study in progress, pet dog samples have generally measured < 100 pg/ml before SG-correction, and typically < 250 pg/ml after SG-correction.
In methods development work, higher concentrations are often assumed to be better (e.g. higher extraction efficiency or improved storage). This assumption is tempting, in part because many studies suffer from an inability to detect analytes at low concentrations. However, this assumption may lead us astray without a careful consideration of potential sources of interference, which may be particularly abundant in urine. On the other hand, it has also been assumed—without sufficient empirical evidence—that high oxytocin concentrations are biologically implausible. This and other recent work (Gnanadesikan et al., 2021) suggest that some methods, at least, are not susceptible to interference in mouse samples.
Feldman et al. (2011) found that in human samples, while plasma and salivary measures of oxytocin were correlated, extracted urinary measurements were not. Weber et al. (2017) also found no correlation between unextracted plasma and unextracted urinary oxytocin measurements in premature human infants. While there are many possible reasons for these findings, including that different biological matrices are thought to integrate over different periods of time, another plausible interpretation in light of our results is that the urine measurements in these studies were susceptible to interference. This interpretation is supported by studies that have found correlations between plasma and extracted urinary oxytocin in certain conditions and time windows (Amico et al., 1987; Francis et al., 2016).
It should be noted that many of the previously demonstrated associations between urinary oxytocin measurements and social behavior may be robust, despite these potential limitations, especially if employing a change-from-baseline approach. Nevertheless, many studies may have inflated concentrations due to the type of interference we demonstrated here. This interference may also contribute to negative findings if the noise introduced by interference obscures an underlying effect, which may be especially important for studies of individual differences.
While the success of our method across a wide variety of mammalian species—including members of the orders Primates, Rodentia, and Carnivora—is promising, the extraction method as reported here may not be optimal for all species. It should be noted, however, that this method worked well even in wild sifaka samples, which are highly concentrated due to their extremely limited water intake; therefore we do not expect highly concentrated urine to be problematic. Nevertheless, we strongly encourage anyone who aims to extend this work to other species and populations to perform a careful analytical validation with their own samples, particularly parallelism and spike recovery, although these tests alone may not be sufficient to determine the absence of interference.
Our epitope mapping results illuminate one of the core components of any immunoassay: the antibody. The importance of the core epitope—spanning residues 7–9—is consistent with the manufacturer’s reports of low cross-reactivity with mesotocin and vasopressin, since both of these related nonapeptides differ from oxytocin at the eighth residue. It is worth noting that a novel form of oxytocin has been documented in some species of New World monkeys; this form of oxytocin has a substitution at residue 8, with proline replacing the ancestral leucine (Lee et al., 2011). Given our epitope mapping results, we would expect binding for this form of oxytocin to be significantly reduced. Interestingly, although the identity of each residue in the rest of oxytocin (1–6) is less important for antibody binding, cross-reactivity drops rapidly for smaller fragments.
This demonstrated cross-reactivity is important for several reasons. First, measurements of oxytocin by immunoassay likely include some amount of signal from these bioactive fragments; this interpretation is supported by the observation that injection of radiolabelled oxytocin (in marmosets) results in urinary samples that produce multiple radioactive fractions after HPLC, including ones other than the standard (Seltzer and Ziegler, 2007). Second, although the cross-reactivity for MIF-1 seems low, it is a very small fragment, with a sequence that is common in a variety of peptides. This cross-reactivity may partially explain why immunoassay of urinary samples is more difficult than for plasma samples, given that urine is a waste product, full of thousands of known and unknown metabolites (Bouatra et al., 2013). A BLASTP (Altschul et al., 1990) search of the NCBI Protein Reference Sequences database (O’Leary et al., 2016), restricted to Homo sapiens, on oxytocin 4–9 reveals 678 unique sequences with 100% identity (i.e. containing QNCPLG), and many more with partial identity. This result suggests that a large number of metabolites might have sufficient sequence similarity to cross-react with the Arbor Assays anti-oxytocin antibody, although we do not know which specific fragments are present in urine or what metabolic processes produce them. Importantly, oxytocin has been shown to be one of the least abundant—while still detectable—metabolites in human urine, while other metabolites can be more abundant by up to 10 orders of magnitude (Bouatra et al., 2013). Thus, even peptide fragments with low cross-reactivities could interfere when present in high enough concentrations. Although this presents challenges for urinary oxytocin work, the method we report here minimizes measurement of other small peptide fragments, including oxytocin metabolites and other interfering molecules.
It should also be noted that while other immunoassay kits use different antibodies, given that the oxytocin molecule is so small, other antibodies likely have a similar susceptibility to interference, even if their epitope and/or core epitope is slightly different. For example, we previously demonstrated that the Enzo Life Sciences kit is susceptible to interference with plasma samples in situations where the Arbor Assays kit is not, presumably reflecting differences in their epitopes or specificity (Gnanadesikan et al., 2021). Similarly, Wirobski, Schaebs et al. (2021) found that the Arbor Assays kit performed better than the Enzo kit for human and canine urinary oxytocin, with better spike recoveries, lower intra-assay CVs, and increased sensitivity. Nevertheless, given the fundamental constraint posed by the small size of the oxytocin molecule, improved immunoassay of urine samples will most likely rely on improved extraction techniques, such as the solid-phase extraction method that we report here.
Whether immunoreactivity from oxytocin’s metabolites should be considered interference depends on the perspective of the end user. Indeed, analytes measured in excretory products such as urine and feces often involve binding to metabolites of the target analyte rather than the parent molecule, and in some cases this may be desirable (Palme et al., 2013). It has recently been proposed that many of oxytocin’s functions may derive from active fragments of the nonapeptide (Uvnäs Moberg et al., 2019), suggesting that oxytocin metabolites may be relevant biomarkers for some aspects of the oxytocinergic system. Our results indicate that several bioactive fragments of oxytocin are nearly entirely eliminated using solid-phase extraction, making this an important consideration for the protocols described here. Given the abundance of interfering molecules in urine, it will be challenging for immunoassays to measure oxytocin metabolites accurately in urinary samples; however, this may be possible in other matrices (e.g. plasma, CSF) and may contribute to divergent results with extracted and unextracted samples in some studies.
Some oddities of urinary oxytocin measurement have been reported previously. Schaebs et al. (2019) performed a validation of dog and wolf samples on the Enzo kit, but found that despite good parallelism, the spike recoveries and extraction efficiencies were concerningly high. Reyes et al. (2014) reported that 18% of their samples produced out-of-range high measurements on the Enzo kit even after solid-phase extraction. Interestingly, when assayed at a two-fold dilution, not only were the samples within range, but the corrected oxytocin concentrations were also within range, suggesting some form of interference when assayed at higher concentrations. Reyes et al. (2014) hypothesize that this may be due to dehydration and overly concentrated samples, supported by the finding that out-of-range samples had higher creatinine values. Higher creatinine was also associated with higher vasopressin measurements, higher specific gravity, and more acidic samples. Our work suggests that either high concentrations of metabolites or the altered pH of dehydrated samples could explain this result, and the MCX method—with the pH of buffers optimized for both retaining oxytocin and minimizing interference—improves assay performance considerably.
Lastly, we would like to caution that this and our previous work (Gnanadesikan et al., 2021) emphasize the need for detailed reporting of immunoassay methods, not only in validation papers, but also accompanying any empirical result. This reporting should include the details of extraction methods and immunoassay kits, sample and buffer volumes, dilution factors, and other aspects of the sample treatment and processing protocols. Attention to these details will aid in both interpretation and replication of reported results.
5. Conclusions
Our immunoassay results indicate that many existing methods of measuring urinary oxytocin may be susceptible to interference. Our epitope mapping and fragment cross-reactivity results suggest this may be attributable to the abundance of metabolites in urine. Nevertheless, we show that a more selective extraction method, utilizing a mixed-cation exchange (MCX) chemistry, is successful at minimizing this interference. This method has the potential to improve urinary oxytocin studies in a wide variety of species and to open doors for reconsidering the relationships between urinary oxytocin and other physiological and behavioral measures.
Supplementary Material
Acknowledgements
We would like to acknowledge the Pepscan team for their work on the epitope mapping, as well as Arbor Assays for their contributions. In particular, Bobbi O’Hara at Arbor Assays was extremely helpful at multiple stages of this work. We also benefited significantly from conversations with Terence Gorman at Waters, who advised us throughout the process of developing and optimizing the MCX method. Sifaka samples were collected and transported with funding by the National Science Foundation under award number BES-1 719 654, with the permission of CAFF/CORE, the Ministry of Water and Forests, and Madagascar National Parks, and the assistance of the University of Antananarivo, MICET, the Ankoatsifaka Research Station, the Sifaka Research Project staff, and Meredith Lutz, Diary Razafimandimby, Fanomezantsoa Razafimalala, Mc Antonin Andriamahaihavana, Sylvia Rahobilalaina, Hannah Carbonneau, and Allison Hays. We would also like to acknowledge the positive and constructive comments of two anonymous reviewers in improving this paper. This material is based upon work supported by the Eunice Kennedy Shriver National Institute of Child Health & Human Development of the National Institutes of Health under award number R21HD095217, by the National Institute of Mental Health under award number MH114994, and by the National Science Foundation Graduate Research Fellowship Program under grant number DGE-1 746 060. Any opinions, findings, and conclusions or recommendations expressed in this material are those of the authors and do not necessarily reflect the views of the National Institutes of Health or the National Science Foundation.
Funding
This study was funded in part by the National Institutes of Health (R21HD095217 to E.L.M. and S.R.T. and MH114994 to E.A.D.H.). Sifaka sample collection was funded by a University of Texas at Austin Research Grant and the National Science Foundation (BES-1 719 654) to R.J.L. Sifaka sample analysis was funded by a National Science Foundation grant (BES-1 719 655) to S.R.T. G.E.G. was additionally funded by the National Science Foundation Graduate Research Fellowship Program (DGE-1 746 060), the P.E.O. Scholar Award, and FRISCO. The Laboratory for the Evolutionary Endocrinology of Primates (LEEP) was funded by the University of Arizona College of Social and Behavioral Sciences, School of Anthropology, Institute for the Environment, Provost’s Office, and Bio-5 Institute. Arbor Assays partially funded the epitope mapping analyses.
Footnotes
CRediT authorship contribution statement
Gitanjali E. Gnanadesikan: Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Visualization, Writing – original draft. Elizabeth A.D. Hammock: Conceptualization, Resources (mouse samples), Writing – original draft. Stacey R. Tecot: Conceptualization, Funding acquisition, Resources (laboratory, sifaka samples), Project administration, Writing – original draft. Rebecca J. Lewis: Resources (laboratory, sifaka samples), Writing – review & editing. Russ Hart: Conceptualization, Funding acquisition, Methodology, Writing – review & editing. C. Sue Carter: Conceptualization, Writing – review & editing. Evan L. MacLean: Conceptualization, Funding acquisition, Methodology, Project administration, Resources (laboratory equipment), Writing – original draft.
Declaration of Competing Interest
R.H. founded Arbor Assays and is a board member. The authors declare no other competing interests.
Appendix A. Supporting information
Supplementary data associated with this article can be found in the online version at doi:10.1016/j.psyneuen.2022.105827.
References
- Algoe SB, Kurtz LE, Grewen K, 2017. Oxytocin and social bonds: the role of oxytocin in perceptions of romantic partners’ bonding behavior. Psychol. Sci 28 (12), 1763–1772. 10.1177/0956797617716922. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Altschul SF, Gish W, Miller W, Myers EW, Lipman DJ, 1990. Basic local alignment search tool. J. Mol. Biol 215 (3), 403–410. 10.1016/S0022-2836(05)80360-2. [DOI] [PubMed] [Google Scholar]
- Amico JA, Ulbrecht JS, Robinson AG, 1987. Clearance studies of oxytocin in humans using radioimmunoassay measurements of the hormone in plasma and urine. J. Clin. Endocrinol. Metab 64 (2), 340–345. 10.1210/jcem-64-2-340. [DOI] [PubMed] [Google Scholar]
- Andreasson U, Perret-Liaudet A, van Waalwijk van Doorn LJ, Blennow K, Chiasserini D, Engelborghs S, Teunissen CE, 2015. A practical guide to immunoassay method validation. Front. Neurol 6 (Aug), 1–8. 10.3389/lheur.2015.00179. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bales KL, van Westerhuyzen JA, Lewis-Reese AD, Grotte ND, Lanter JA, Carter CS, 2007. Oxytocin has dose-dependent developmental effects on pair-bonding and alloparental care in female prairie voles. Horm. Behav 52 (2), 274–279. 10.1016/j.yhbeh.2007.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bick J, Dozier M, 2010. Mothers’ concentrations of oxytocin following close, physical interactions with biological and nonbiological children. Dev. Psychobiol 52 (1), 100–107. 10.1002/dev.20411. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bienboire-Frosini C, Chabaud C, Cozzi A, Codecasa E, Pageat P, 2017. Validation of a commercially available enzyme immunoassay for the determination of oxytocin in plasma samples from seven domestic animal species. Front. Neurosci 11 (SEP) 10.3389/fnins.2017.00524. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bouatra S, Aziat F, Mandal R, Guo AC, Wilson MR, Knox C, Wishart DS, 2013. The human urine metabolome (sep). PLoS ONE 8 (9), e73076. 10.1371/journal.pone.0073076. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brandtzaeg OK, Johnsen E, Roberg-Larsen H, Seip KF, MacLean EL, Gesquiere LR, Wilson SR, 2016. Proteomics tools reveal startlingly high amounts of oxytocin in plasma and serum. Sci. Rep 6 (0317), 1–7. 10.1038/srep31693. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cardoso C, Ellenbogen MA, Orlando MA, Bacon SL, Joober R, 2013. Intranasal oxytocin attenuates the cortisol response to physical stress: a dose-response study. Psychoneuroendocrinology 38 (3), 399–407. 10.1016/j.psyneuen.2012.07.013. [DOI] [PubMed] [Google Scholar]
- Carter CS, 2014. Oxytocin pathways and the evolution of human behavior. Annu. Rev. Psychol 65 (1), 17–39. 10.1146/annurev-psych-010213-115110. [DOI] [PubMed] [Google Scholar]
- Carter CS, Courtney Devries A, Getz LL, 1995. Physiological substrates of mammalian monogamy: The prairie vole model. Neurosci. Biobehav. Rev 19 (2), 303–314. 10.1016/0149-7634(94)00070-H. [DOI] [PubMed] [Google Scholar]
- Crockford C, Wittig RM, Langergraber K, Ziegler TE, Zuberbuhler K, Deschner T, 2013. Urinary oxytocin and social bonding in related and unrelated wild chimpanzees. Proc. R. Soc. B: Biol. Sci 280 (1755), 1–8. 10.1098/rspb.2012.2765. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Crockford C, Deschner T, Ziegler TE, Wittig RM, 2014. Endogenous peripheral oxytocin measures can give insight into the dynamics of social relationships: a review. Front. Behav. Neurosci 8 (March), 1–14. 10.3389/fnbeh.2014.00068. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Feldman R, 2012. Oxytocin and social affiliation in humans. Horm. Behav 61 (3), 380–391. 10.1016/j.yhbeh.2012.01.008. [DOI] [PubMed] [Google Scholar]
- Feldman R, Gordon I, Zagoory-Sharon O, 2011. Maternal and paternal plasma, salivary, and urinary oxytocin and parent-infant synchrony: considering stress and affiliation components of human bonding. Dev. Sci 14 (4), 752–761. 10.1111/j.1467-7687.2010.01021.x. [DOI] [PubMed] [Google Scholar]
- Finkenwirth C, van Schaik C, Ziegler TE, Burkart JM, 2015. Strongly bonded family members in common marmosets show synchronized fluctuations in oxytocin. Physiol. Behav 151, 246–251. 10.1016/j.physbeh.2015.07.034. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Finkenwirth C, Martins E, Deschner T, Burkart JM, 2016. Oxytocin is associated with infant-care behavior and motivation in cooperatively breeding marmoset monkeys. Horm. Behav 80, 10–18. 10.1016/j.yhbeh.2016.01.008. [DOI] [PubMed] [Google Scholar]
- Francis SM, Kirkpatrick MG, de Wit H, Jacob S, 2016. Urinary and plasma oxytocin changes in response to MDMA or intranasal oxytocin administration. Psychoneuroendocrinology 74, 92–100. 10.1016/j-psyneuen.2016.08.011. [DOI] [PubMed] [Google Scholar]
- Froemke RC, Young LJ, 2021. Oxytocin, neural plasticity, and social behavior. Annu. Rev. Neurosci 44, 359–381. 10.1146/annurev-neuro-102320-102847. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Geysen HM, Meloen RH, Barteling SJ, 1984. Use of peptide synthesis to probe viral antigens for epitopes to a resolution of a single amino acid. Proc. Natl. Acad. Sci. USA 81 (13 I), 3998–4002. 10.1073/pnas.81.13.3998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gnanadesikan GE, Hammock EA, Tecot SR, Carter CS, MacLean EL, 2021. Specificity of plasma oxytocin immunoassays: a comparison of commercial assays and sample preparation techniques using oxytocin knockout and wildtype mice. Psychoneuroendocrinology 132 (July), 105368. 10.1016/j.psyneuen.2021.105368. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goldberger BA, Loewenthal B, Darwin WD, Cone EJ, 1995. Intrasubject variation of creatinine and specific gravity measurements in consecutive urine specimens of heroin users. Clin. Chem 41 (1), 116–117. 10.1093/dinchem/41.1.116. [DOI] [PubMed] [Google Scholar]
- Lee AG, Cool DR, Grunwald WC, Neal DE, Buckmaster CL, Cheng MY, Parker KJ, 2011. A novel form of oxytocin in new world monkeys. Biol. Lett 7 (4), 584–587. 10.1098/rsbl.2011.0107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lefevre A, Mottolese R, Dirheimer M, Mottolese C, Duhamel JR, Sirigu A, 2017. A comparison of methods to measure central and peripheral oxytocin concentrations in human and non-human primates. Sci. Rep 7 (1), 1–10. 10.1038/s41598-017-17674-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- MacLean EL, Wilson SR, Martin WL, Davis JM, Nazarloo HP, Carter CS, 2019. Challenges for measuring oxytocin: the blind men and the elephant? Psychoneuroendocrinology 107 (May), 225–231. 10.1016/j.psyneuen.2019.05.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Marshall-Pescini S, Schaebs FS, Gaugg A, Meinert A, Deschner T, Range F, 2019. The role of oxytocin in the dog–owner relationship. Animals 9 (10), 1–19. 10.3390/ani9100792. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McCullough ME, Churchland PS, Mendez AJ, 2013. Problems with measuring peripheral oxytocin: Can the data on oxytocin and human behavior be trusted? Neurosci. Biobehav. Rev 37 (8), 1485–1492. 10.1016/j.neubiorev.2013.04.018. [DOI] [PubMed] [Google Scholar]
- Miller RC, Brindle E, Holman DJ, Shofer J, Klein NA, Soules MR, O’Connor KA, 2004. Comparison of specific gravity and creatinine for normalizing urinary reproductive hormone concentrations. Clin. Chem 50 (5), 924–932. 10.1373/dinchem.2004.032292. [DOI] [PubMed] [Google Scholar]
- Mitsui S, Yamamoto M, Nagasawa M, Mogi K, Kikusui T, Ohtani N, Ohta M, 2011. Urinary oxytocin as a noninvasive biomarker of positive emotion in dogs. Horm. Behav 60 (3), 239–243. 10.1016/j.yhbeh.2011.05.012. [DOI] [PubMed] [Google Scholar]
- Moscovice LR, Ziegler TE, 2012. Peripheral oxytocin in female baboons relates to estrous state and maintenance of sexual consortships. Horm. Behav 62 (5), 592–597. 10.1016/j.yhbeh.2O12.08.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Moscovice LR, Surbeck M, Fruth B, Hohmann G, Jaeggi AV, Deschner T, 2019. The cooperative sex: Sexual interactions among female bonobos are linked to increases in oxytocin, proximity and coalitions. Horm. Behav 116 (August), 104581 10.1016/j.yhbeh.2019.104581. [DOI] [PubMed] [Google Scholar]
- Nagasawa M, Kikusui T, Onaka T, Ohta M, 2009. Dog’s gaze at its owner increases owner’s urinary oxytocin during social interaction. Horm. Behav 55 (3), 434–441. 10.1016/j.yhbeh.2008.12.002. [DOI] [PubMed] [Google Scholar]
- Nagasawa M, Mitsui S, En S, Ohtani N, Ohta M, Sakuma Y, Kikusui T, 2015. Oxytocin-gaze positive loop and the coevolution of human-dog bonds. Science 348 (6232), 333–336. 10.1126/science.1261022. [DOI] [PubMed] [Google Scholar]
- Nishimori K, Young LJ, Guo Q, Wang Z, Insel TR, Matzuk MM, 1996. Oxytocin is required for nursing but is not essential for parturition or reproductive behavior. Proc. Natl. Acad. Sci. USA 93 (21), 11699–11704. 10.1073/pnas.93.21.11699. [DOI] [PMC free article] [PubMed] [Google Scholar]
- O’Leary NA, Wright MW, Brister JR, Ciufo S, Haddad D, McVeigh R, Pruitt KD, 2016. Reference sequence (RefSeq) database at NCBI: current status, taxonomic expansion, and functional annotation. Nucleic Acids Res. 44 (D1), D733–D745. 10.1093/nar/gkv1189. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Palme R, Touma C, Arias N, Dominchin MF, Lepschy M, 2013. Steroid extraction: Get the best out of faecal samples. Wien. Tierärztl. Mon 100 (9–10), 238–246. [Google Scholar]
- Pederson CA, Prange AJ, 1979. Induction of maternal behavior in virgin rats after intracerebroventricular administration of oxytocin. Proc. Natl. Acad. Sci. USA 76 (12), 6661–6665. 10.1073/pnas.76.12.6661. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Peters S, Slattery DA, Uschold-Schmidt N, Reber SO, Neumann ID, 2014. Dose-dependent effects of chronic central infusion of oxytocin on anxiety, oxytocin receptor binding and stress-related parameters in mice. Psychoneuroendocrinology 42, 225–236. 10.1016/j.psyneuen.2014.01.021. [DOI] [PubMed] [Google Scholar]
- R Core Team. (2021). R: A Language and Environment for Statistical Computing [Computer software manual]. Vienna, Austria. Retrieved from https://www.r-project.org/. [Google Scholar]
- Reyes TL, Galinsky AM, Hoffmann JN, You HM, Ziegler TE, McClintock MK, 2014. Social peptides: Measuring urinary oxytocin and vasopressin in a home field study of older adults at risk for dehydration. J. Gerontol. - Ser. B Psychol. Sci. Soc. Sci 69, S229–S237. 10.1093/geronb/gbul04. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Saito A, Hamada H, Kikusui T, Mogi K, Nagasawa M, Mitsui S, Hiraki K, 2014. Urinary oxytocin positively correlates with performance in facial visual search in unmarried males, without specific reaction to infant face. Front. Neurosci 8 (8 JUL), 1–7. 10.3389/fnins.2014.00217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Samuni L, Preis A, Mundry R, Deschner T, Crockford C, Wittig RM, 2017. Oxytocin reactivity during intergroup conflict in wild chimpanzees. Proc. Natl. Acad. Sci 114 (2), 268–273. 10.1073/pnas.1616812114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schaebs FS, Marshall-Pescini S, Range F, Deschner T, 2019. Analytical validation of an Enzyme Immunoassay for the measurement of urinary oxytocin in dogs and wolves. Gen. Comp. Endocrinol 281 (May), 73–82. 10.1016/j.ygcen.2019.05.015. [DOI] [PubMed] [Google Scholar]
- Seltzer LJ, Ziegler TE, 2007. Non-invasive measurement of small peptides in the common marmoset (Callithrix jacchus): a radiolabeled clearance study and endogenous excretion under varying social conditions. Horm. Behav 51 (3), 436–442. 10.1016/j.yhbeh.2006.12.012. [DOI] [PubMed] [Google Scholar]
- Seltzer LJ, Ziegler TE, Pollak SD, 2010. Social vocalizations can release oxytocin in humans. Proc. R. Soc. B: Biol. Sci 277 (1694), 2661–2666. 10.1098/rspb.2010.0567. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Snowdon CT, Pieper BA, Boe CY, Cronin KA, Kurian AV, Ziegler TE, 2010. Variation in oxytocin is related to variation in affiliative behavior in monogamous, pairbonded tamarins. Horm. Behav 58 (4), 614–618. 10.1016/j.yhbeh.2010.06.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Strier KB, Ziegler TE, 2005. Advances in field-based studies of primate behavioral endocrinology. Am. J. Primatol 67 (1), 1–4. 10.1002/ajp.20165. [DOI] [PubMed] [Google Scholar]
- Szeto A, Ph D, Mccabe PM, Ph D, Nation DA, Ph D, Ph D, 2012. Evaluation of enzyme immunoassay and radioimmunoassay methods for the measurement of plasma oxytocin. Psychosom. Med 73 (5), 393–400. 10.1097/PSY.0b013e31821df0c2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Uvnäs Moberg K, Handlin L, Kendall-Tackett K, Petersson M, 2019. Oxytocin is a principal hormone that exerts part of its effects by active fragments. Med. Hypotheses 133 (August), 109394. 10.1016/j.mehy.2019.109394. [DOI] [PubMed] [Google Scholar]
- Van Leengoed E, Kerker E, Swanson HH, 1987. Inhibition of post-partum maternal behaviour in the rat by injecting an oxytocin antagonist into the cerebral ventricles. J. Endocrinol 112 (2), 275–282. 10.1677/joe.0.1120275. [DOI] [PubMed] [Google Scholar]
- Weber A, Harrison TM, Sinnott L, Shoben A, Steward D, 2017. Plasma and urinary oxytocin trajectories in extremely premature infants during NICU hospitalization. Biol. Res. Nurs 19 (5), 549–558. 10.1177/1099800417718266. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wirobski G, Range F, Schaebs FS, Palme R, Deschner T, Marshall-Pescini S, 2021. Life experience rather than domestication accounts for dogs’ increased oxytocin release during social contact with humans (dec). Sci. Rep 11 (1), 14423. 10.1038/S41598-021-93922-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wirobski G, Schaebs FS, Range F, Marshall-Pescini S, Deschner T, 2021. Analytical and physiological validation of an enzyme immunoassay to measure oxytocin in dog, wolf, and human urine samples. Sci. Rep 11 (1), 1–11. 10.1038/s41598-021-92356-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wittig RM, Crockford C, Deschner T, Langergraber KE, Ziegler TE, Zuberbühler K, 2014. Food sharing is linked to urinary oxytocin levels and bonding in related and unrelated wild chimpanzees. Proc. R. Soc. B Biol. Sci 281 (1778), 1–10. 10.1098/rspb.2013.3096. [DOI] [PMC free article] [PubMed] [Google Scholar]