Skip to main content
NIST Author Manuscripts logoLink to NIST Author Manuscripts
. Author manuscript; available in PMC: 2023 Mar 3.
Published in final edited form as: Nat Phys. 2021;17:10.1038/s41567-021-01226-y. doi: 10.1038/s41567-021-01226-y

Absolute 13C/12C Isotope Amount Ratio for Vienna Pee Dee Belemnite from Infrared Absorption Spectroscopy

Adam J Fleisher 1,4,, Hongming Yi 1,4,5, Abneesh Srivastava 1, Oleg L Polyansky 2,3, Nikolai F Zobov 3, Joseph T Hodges 1
PMCID: PMC9982939  NIHMSID: NIHMS1873171  PMID: 36873572

Abstract

Measurements of isotope ratios are predominantly made with reference to standard specimens that have been characterized in the past. In the 1950s, the carbon isotope ratio was referenced to a belemnite sample collected by Heinz Lowenstam and Harold Urey1 in South Carolina’s Pee Dee region. Due to the exhaustion of the sample since then, reference materials that are traceable to the original artefact are used to define the Vienna Pee Dee Belemnite (VPDB) scale for stable carbon isotope analysis2. However, these reference materials have also become exhausted or proven to exhibit unstable composition over time3, mirroring issues with the international prototype of the kilogram that led to a revised International System of Units4. A campaign to elucidate the stable carbon isotope ratio of VPDB is underway5, but independent measurement techniques are required to support it. Here we report an accurate value for the stable carbon isotope ratio inferred from infrared absorption spectroscopy, fulfilling the promise of this fundamentally accurate approach6. Our results agree with a value recently derived from mass spectrometry5, and therefore advance the prospects of SI-traceable isotope analysis. Further, our calibration-free method could improve mass balance calculations and enhance isotopic tracer studies in CO2 source apportionment.


Small variations in the isotopic abundances of common nuclei such as carbon, oxygen and hydrogen occur naturally. The isotope-delta notation is a useful expression for precision stable isotope analysis and is generally defined as δ iE = R(iE/jE)sample/R(iE/jE)ref − 1, where i is the heavier isotope mass number of element E (e.g., 13C), j is the lighter isotope mass number of E (e.g., 12C), and R(iE/jE)sample and R(iE/jE)ref are the absolute isotope ratios of an unknown sample and a reference material, respectively (e.g., R(13C/12C)sample and R(13C/12C)ref)7. The δ notation—expressed per mille using the symbol ‰, where 1 ‰ ≡ 1 × 10−3—allows for the convenient expression of exceedingly small differences in isotope composition, and thus an accessible discussion of subjects as diverse as variability in atmospheric composition, climatology, geochemistry, ecology, and dietary evolution and networking.

To measure small differences on a δ scale, ultraprecise analytical techniques such as isotope ratio mass spectrometry (IRMS)8 and Fourier transform infrared spectroscopy9, based on the identical treatment of sample and reference, are commonly used. Although δ values can be determined with a precision of ≤0.01 ‰, absolute isotope ratio measurements of R(iE/jE)sample with commensurate accuracy have not been possible because of either systematic uncertainties in the values of R(iE/jE)ref or the achievable uncertainty in primary methods such as calibration using synthetic isotope mixtures. For example, the most recent evaluation of the VPDB reference ratio that underpins the δ 13C scale, R(13C/12C)VPDB, reported uncertainties that were more than 100-fold greater than the achievable δ-value precision5. The lack of independent experimental techniques for stable carbon isotope ratio measurements has generally impeded the assessment of uncertainty in R(13C/12C)ref values and confounded efforts to correct for their respective drifts over time3.

The δ notation has additional limits. For example, δ values are approximately a linear function of mole fraction only for very small differences between R(iE/jE)sample and R(iE/jE)ref, and multiple reference materials are required to establish linearity2. For samples with large differences between R(iE/jE)sample and R(iE/jE)ref (e.g., samples enriched or depleted in 13C, or extraterrestrial samples), δ values are a highly nonlinear function of mole fraction7, and precision is generally degraded when measuring sample values far from the scale definition (δ iE = 0). For several applications, direct measurements of R(iE/jE)sample or atom (mole) fraction would be advantageous. These include tracer studies for enriched compound- or position-specific isotope analysis10 and the calculation of mixing ratios in complex systems using a mass balance equation (e.g., systems with two or more sources and/or sinks of variable isotopic composition7). From a standards and metrology perspective, direct measurements of R(iE/jE)sample would enable detailed studies of drifts in primary reference materials3 and SI traceable isotope ratio determinations5. Accurate measurements of R(iE/jE)sample that are traceable to intrinsic physical invariants would also maintain compatibility with historical δ notation scales by underpinning the isotope ratios of the primary and secondary reference materials. This is highly desirable, given that the ratio-of-ratios approach conveniently expressed by the δ notation is inherently susceptible to temporal instability and physical inhomogeneity in the required traceability chain and two-point calibration schemes. Finally, reliable scales that eliminate artefacts altogether are desirable when measuring isotope ratios in extreme environments like those encountered during solar system exploration. There, carrying bulky and consumable specimens can be cost-prohibitive and pre-launch calibration routines are necessary to validate measurements like the isotopic composition of the modern Martian atmosphere11.

Optical methods to measure R(iE/jE)sample directly—circumventing the δ notation—have been proposed for decades6. Several notable efforts were focused on improving measurement precision and studied infrared transitions over a narrow wavelength range1215. That approach is problematic even when the absorption cross-sections—intrinsic molecular quantities which scale the isotope amounts—are well known16. Transitions involving different isotopic substitutions of the same molecule that occur at wavelengths near to one another often originate from very different rotational states, and consequently substantial temperature biases are introduced. Those biases can be mitigated through a high degree of temperature control17,18, or by studying pairs of transitions with nearly identical lower-state rotational energy19. While these prior proposals and proof-of-principle demonstrations collectively suggest that accurate optical measurements of R(iE/jE)sample may be possible without calibrations, each prior work was limited by either the aforementioned temperature biases or by uncertainties in the chosen absorption cross-sections. As a result, the absolute isotope ratios of reference materials inferred from accurate infrared absorption spectroscopy measurements have never been compared with SI-traceable mass spectrometry values.

Here we overcome both experimental hurdles using accurately known absorption cross-sections for carbon dioxide (CO2) transitions from two distinct wavelength regions—effectively eliminating temperature biases. The results are direct measurements of R(iE/jE)sample by infrared absorption spectroscopy with an uncertainty commensurate with calibrated mass spectrometry. By also performing state-of-the-art IRMS on our CO2-in-air gas samples, we infer R(13C/12C)VPDB = 0.011125 ± 0.000043. The value of R(13C/12C)VPDB derived from our unified measurement approach is in excellent agreement with other recently reported values5 and differs significantly from the historic value of Craig20. Consequently, our results demonstrate the potential for perpetual and direct SI-traceability in stable carbon isotope analysis, thus circumventing limitations associated with curation and further propagation of the new physical artifact standards that underpin the δ 13C VPDB scale21.

Our accurate isotope ratio infrared spectroscopy (AIR-IS) instrument, illustrated in Fig. 1A, is a robust platform for measuring the mole fraction of both 13CO2 and 12CO2, and therefore the stable carbon isotope ratio of a gas sample. Typical relative measurement precision of R(13C/12C)sample for a 3 min spectral acquisition was 0.2 × 10−3, with a best recorded relative precision of 0.13 × 10−3 [comparable to the World Meteorological Organization / International Atomic Energy Agency (WMO/IAEA) extended compatibility goal for δ 13C of ± 0.1 ‰22]. We simultaneously probed 13CO2 and 12CO2 rotational-vibrational transition pairs with nearly identical lower-state quantum numbers and lower-state energies, and therefore effectively identical Boltzmann population factors (Supplementary Information, Sections S1 and S3). Using diode lasers centered near wavelengths of 2.0 μm (13CO2) and 1.6 μm (12CO2) and high-reflectivity mirrors to form a single sample cell, the AIR-IS instrument created overlapping intra-cavity mode volumes for both wavelengths. This arrangement, combined with synchronous acquisition of the spectra, largely eliminated the influence of spatial-temporal temperature and pressure variations on the measured peak area ratios.

Fig. 1. Accurate isotope ratio infrared spectroscopy (AIR-IS).

Fig. 1.

A. The spectrometer synchronously interrogated two infrared transitions using diode lasers at wavelengths nominally equal to 2.0 μm (13C16O2) and 1.6 μm (12C16O2), respectively. Frequency agile, rapid-scanning was accomplished via two electro-optic modulators (EOM) driven by individual radiofrequency function generators. The length-stabilized optical enhancement cavity and sample cell provided effective path lengths of 12 km and 4.5 km at 2.0 μm and 1.6 μm, respectively. Laser light transmitted by the cavity was split at a dichroic mirror (DM) and then directed onto one of the two photoreceivers (PR). Additional abbreviations: AOM, acousto-optic modulator; M, reflective mirror; L, lens; PZT, piezo-electric transducer; PD, photodetector; PC, personal computer. B-C. Representative high-resolution cavity ring-down absorption spectra of 13C16O2 (B) and 12C16O2 (C) for a sample of NIST Standard Reference Material 1720 Northern Continental Air (χCO2 = 393.23 μmol/mol; δ 13C = −8.6 ‰). Measured absorption coefficients (α) in units of inverse length are plotted in the upper panels as a function of laser detuning (B, red dots, 2.0 μm laser; C, blue dots, 1.6 μm laser). Also shown are the individual fitted spectral models (solid lines) and listed are the known transition wave numbers, vibrational band assignments, and rotational transition assignments (Supplementary Information, Section S1). The lower panels show the observed-minus-calculated (O–C) fit residuals for each measured CO2 transition.

The absorption spectra of 13C16O2 and 12C16O2 in air at mole fractions within the range of natural terrestrial and marine abundances are shown in Fig. 1BC. These spectra were measured using cavity ring-down spectroscopy and correspond to a reference sample of North American continental air with a value of δ 13C = −8.6 ‰ realized by the WMO Global Atmosphere Watch Central Calibration Laboratory at the National Oceanic and Atmospheric Administration (NOAA). We report similar peak absorption coefficients for both isotopologues, and thus similarly precise fits of the individual mole fractions χ=kBTcSTpανdν, where c is the speed of light in a vacuum, kB is the Boltzmann constant, T is the sample temperature, p is the total sample pressure, S(T) is the unweighted temperature-dependent transition intensity (i.e., not weighted by natural terrestrial isotopic relative abundances), and α(ν) is the measured absorption coefficient as a function of frequency ν. Assuming that all stable isotopes of C (13C, 12C) and O (18O, 17O, 16O) are randomly distributed amongst the corresponding isotopologues of CO2, then R(13C/12C)sample = χ(13C16O2)/χ(12C16O2). Taking the ratio of measured mole fractions of the two CO2 isotopologues leads to the working equation

RC13/C12sample=arsr, [1]

where the subscript r indicates the (13C16O2)/(12C16O2) ratio of parameters, aανdν are the measured peak areas for each line, and sr is the line intensity ratio for the transition pair. By concurrent measurements of a for both 13C16O2 and 12C16O2 at similar signal-to-noise ratios, we also eliminated biases associated with the relative frequency axis as well as with the choice of line shape profile.

Substituting Eq. [1] into the definition δ 13C = R(13C/12C)sample/R(13C/12C)VPDB − 1 and solving for R(13C/12C)VPDB gives the VPDB reference value as

RC13/C12VPDB=ar/sr1+δC13, [2]

where all quantities on the right-hand-side are based on values with quantifiable uncertainties.

We emphasize that the line intensities used to evaluate R(13C/12C)sample in Eq. [1] depend upon invariant molecular constants and a temperature-dependent function describing Boltzmann population statistics (Supplementary Information, Section S3). As already stated, the temperature dependence of sr(T) was effectively eliminated by the choice of 13C16O2 and 12C16O2 transition pairs (Supplementary Information, Section S1). For 13C16O2, we used intensities calculated here from an updated, fully ab initio dipole moment surface (DMS) and a semi-empirical potential energy surface (PES). This choice was justified by benchmarking the underlying quantum-chemistry theory against accurate experiments (Supplementary Information, Section S2).

AIR-IS measurements of R(13C/12C)sample for five CO2-in-air samples are plotted in Fig. 2A vs. δ 13C values assigned by IRMS (Supplementary Information, Section S4). In Fig. 2A, the fitted black line reveals a high degree of linearity between the two techniques (r2 = 0.997), and fitted residuals are plotted in Fig. 2B. Note that the CO2-in-air samples were chosen to cover a broad range of δ 13C values that naturally occur across a wide variety of terrestrial and marine carbon sources23, as illustrated in Fig. 2C. For example, AIR-IS R(13C/12C)sample values are accurate enough to unambiguously differentiate between C3 and C4 plant photosynthetic pathways.

Fig. 2. AIR-IS measurements of R(13C/12C)sample vs. IRMS assignments of δ 13C.

Fig. 2.

A. AIR-IS values of R(13C/12C)sample were measured for five CO2-in-air samples with δ 13C VPDB value assignments by IRMS. The light-blue shaded region is the interpolated area bounded by the range of measured R(13C/12C)sample values, while the mean and standard deviation for all five samples are plotted as gray circles. The black line represents a linear regression using the expression R(13C/12C)sample = R(13C/12C)VPDB × [(δ 13C) + 1]. B. Fitted residuals (O–C), along with their corresponding standard deviations. C. Expanded range of δ 13C values for some terrestrial and marine carbon sources23.

With precise values of δ 13C assigned to our CO2-in-air samples traceability to VPDB, we used Eq. [2] to calculate R(13C/12C)VPDB from each individual spectroscopic measurement of R(13C/12C)sample. A scatter plot of R(13C/12C)VPDB versus an arbitrary measurement number is shown in Fig. 3A. From the ensemble statistics of R(13C/12C)VPDB measured using samples with a variety of δ 13C values, we assessed long-term reproducibility associated with daily variations in the spectrum signal-to-noise ratio. The scatter in the data points plotted in Fig. 3A included variability associated with spectroscopic interferences, as well as potential optical interference effects (e.g., spurious etalons) and daily variations in laser power and linewidth which slightly affected optical cavity throughput and scan speed. The data was fitted to a normal distribution function (Fig. 3B) with a mean value of R(13C/12C)VPDB = 0.011125 and a standard error of the mean of 1.7 × 106. Therefore, our Type A evaluation of the long-term reproducibility gave a relative standard uncertainty of 0.15 × 10−3.

Fig. 3. AIR-IS value of R(13C/12C)VPDB.

Fig. 3.

A. Scatter plot of all 301 unique AIR-IS measurements of R(13C/12C)VPDB, along with error bars showing relative precision of ≈ 0.2 × 10−3. B. Histogram of values fitted to a normal distribution (black line), with mean value of R(13C/12C)VPDB = 0.011125. C. Comparison of literature values for R(13C/12C)VPDB5,20,2429, including this work. Error bars show combined standard uncertainty.

In addition to long-term reproducibility in ar, we report persistent (systematic) relative standard uncertainties in R(13C/12C)VPDB for the transition intensities, cavity ring-down signal digitizer nonideality, choice of distribution function (Supplementary Information, Section S5), IRMS value assignments, and temperature in Table 1. Adding these components in quadrature gave a combined standard uncertainty in R(13C/12C)VPDB of 43 × 10−6.

Table 1. Uncertainty budget for R(13C/12C)VPDB.

Summary of individual relative standard uncertainties (u) comprising the combined standard uncertainty for R(13C/12C)VPDB of 43 × 10−6.

Symbol Value (10−3) Notes

u 636 3.8 Line intensity, 13C16O2
u 626 0.6 Line intensity, 12C16O2
u ADC 0.6 Digitizer nonideality
uf 0.16 Choice of distribution function
uar 0.15 Standard error in ar
u IRMS 0.06 IRMS δ 13C value assignments
uT 0.04 Temperature-dependence of S

The value of R(13C/12C)VPDB inferred from the unified AIR-IS and IRMS approach reported here is plotted as a gray circle at the top of Fig. 3C and labeled “This work.” The error bar illustrates the combined standard uncertainty in our measurement, dominated by the uncertainty in the 13C16O2 transition intensities. For comparison, several literature values are also plotted in Fig. 3C. The blue diamond corresponds to the most widely cited value of 13RVPDB = 0.0112372 (standard uncertainty, 30 × 10−6), measured by calibrated mass spectrometry of the original Pee Dee belemnite sample performed by Craig in 195720. More recently, in a 2010 technical report24, the International Union of Pure and Applied Chemistry (IUPAC) recommended a value of R(13C/12C)VPDB = 0.011180 (standard uncertainty, 14 × 10−6) based on mass spectrometry measurements by Chang & Li in 1990 (C&L, red triangle)25. Additional experiments performed over the last 20 years5,2629 (open black squares) collectively allude to an even lower absolute value of R(13C/12C)VPDB = 0.011117 (standard uncertainty, 15 × 10−6) show as the black square5, with those literature values deviating by more than −14 ‰ from Craig20.

The AIR-IS result is in good agreement with the 2019 recommendation of Malinovsky et al.5, suggesting from that a reevaluation of the internationally accepted value of R(13C/12C)VPDB is appropriate. Recently, Skrzypek & Dunn30 highlighted an immediate motivation for the adoption of an SI traceable, consensus value of R(13C/12C)VPDB. They report that three different reference values are currently in use by commercial optical stable carbon isotope analyzers, resulting in potential differences between calibrated analyzer δ values of ≈ 0.1 ‰. Therefore, the addition of our independent AIR-IS value to the VPDB discussion will aid in establishing consistency and traceability in stable carbon isotope analysis.

By directly interrogating highly homogenous gas-phase samples of CO2 in air, AIR-IS avoids biases common to the general practice of CO2 extraction from a VPDB traceable carbonate. Furthermore, AIR-IS also eliminates the need for O-isotope calibrations, potentially allowing the technique to accurately quantify R(13C/12C)sample values even in the event of anomalous oxygen isotopic compositions. For example, careful attention must be paid to 17O corrections when analyzing CO2 samples with oxygen from different sources using mass spectrometry (e.g., samples originating from reactions with 17O-enriched stratospheric ozone)24. Oxygen isotope anomalies are also observed in extraterrestrial samples and may explain early solar system evolution31. Specifically, the stable carbon isotopic composition of meteorites with potentially anomalous oxygen compositions reveal 13C-enrichment in the organic molecules carried to early Earth32. These O-isotope anomalies, unless known a priori, present a unique challenge for accurate mass spectrometry where oxygen corrections are explicitly required24,31.

The accurate isotope ratio infrared spectroscopy methodology reported herein is suitable for the possible optical realization of various artifact-based scales, including 14C, 18O, 17O, 15N, 2H, and 34S. For stable carbon isotope ratios, our own efforts to ascertain more accurate values for 13C16O2 transition intensities in the 2 μm wavelength region using gravimetrically prepared, potentially isotopically enriched gas samples with traceability to the kilogram are just beginning. Accurate measurements of analogous 12C16O2 transitions could also reduce uncertainty in the 13C16O2 ab initio predictions (Supplementary Information, Section S2). Those results are expected to reduce our relative combined standard uncertainty associated R(13C/12C)VPDB to ≤ 1 × 10−3. Over time, a consensus determination of R(13C/12C)VPDB through comparisons amongst national metrology institutes and their partners could ultimately obviate the current cumbersome and time-consuming task of artifact-based scale management and long-term scale propagation. Combined with recent advances in the optical detection of 14C with sensitivity below the modern mole fraction of 1.176 pmol/mol33,34, as well as emerging optical techniques for precision clumped isotope analysis35, the possibility exists for finally realizing primary laser-based carbon and oxygen isotope metrology with traceability to the revised, “quantum” SI.

METHODS

Absolute isotope ratio infrared spectroscopy.

We performed frequency agile, rapid-scanning cavity ring-down spectroscopy36 using two distributed feedback diode lasers simultaneously coupled to a single high-finesse enhancement cavity and sample cell. Long-term stability and reproducibility were achieved in part by actively stabilizing the uniform grid of optical frequencies transmitted by the enhancement cavity to a frequency-stabilized HeNe laser37. The enhancement cavity and sample cell comprised two triple-coated high-reflectivity mirrors at opposite ends of a stainless-steel tube. One of the mirrors was mounted to a piezo-electric transducer which was used as a slow actuator to adjust the cavity length to maintain resonance with the HeNe laser. The optical cavity was nominally 75 cm in length, with a free spectral range of 200.07 MHz (standard uncertainty, 30 kHz).

Before introduction of CO2-in-air samples, the sample cell was evacuated under high vacuum for approximately 1 h and then flushed with a continuous flow of high-purity N2 gas for approximately 2 h. The procedure effectively eliminated sample cell memory of the previous CO2-in-air sample, as well as minimized spectroscopic interferences from previously adsorbed molecules within the gas delivery system. CO2-in-air samples were then introduced and maintained at a constant pressure near 8 kPa (as measured by calibrated microbolometer gauges) for a total time of 2 h. After 2 h, the sample was evacuated, and the cavity prepared to receive another sample. Temperature stabilization was enhanced using an insulating box around the AIR-IS sample cell and was monitored during spectral acquisition using calibrated platinum resistance thermometers in good thermal contact with the outside of the enhancement cavity. A comprehensive summary of the NIST technical approach to accurate gas metrology by cavity ring-down spectroscopy can be found in Fleurbaey et al.38.

Digitizer linearity.

Linearity of the analog-to-digital acquisition cards was tested using synthetic exponential decay signals and a metrology-grade reference digitizer as a transfer standard39. The synthetic decays of 1 ms in length were constructed using an arbitrary waveform generator with 14-bit vertical resolution and a sampling rate of 128 MS/s. Their time constants spanned the relevant range for each digitizer; 26 μs to 45 μs for the 2.0 μm digitizer (13C16O2), and 10 μs to 20 μs for the 1.6 μm digitizer (12C16O2). Linear fits of observed time constants (τobs) vs. programmed time constants (τpro) yielded a transformation function between observed and true round-trip losses in the enhancement cavity. The linear transformations were used to estimate the relative shift in measured integrated absorption δa=aobsapro1 for simulated lines at both 2.0 μm and 1.6 μm. The digitizer analysis yielded the following δa, which were then applied to all the measured spectra: δa = 4.7 × 10−3 for 13C16O2 at 2.0 μm, and δa = −4.8 × 10−3 for 12C16O2 at 1.6 μm. We estimated the uncertainty in δa to be equal to 2×0.4×1030.6 × 10−3 by fitting a generalized extreme value distribution function to the observed distribution in τobs over a broad range of synthetic exponential decay signals. We note that the limit in the uncertainty in δa is calculated from the arbitrary waveform generator vertical resolution (14 bits) to be 1/(2Δ+13)0.02 × 10−3. The measured values of δa estimated all potential non-linearities, impedance mismatches, radiofrequency back reflections, dissimilar metallic connectors, and other issues associated with all electronics and cable after the photoreceivers, including the digitizer.

Gas samples and δ 13C value assignments.

Five CO2-in-air samples were measured by AIR-IS. Mentioned in the main text is NIST Standard Reference Material (SRM®) 1720 Northern Continental Air with WMO/NOAA-assigned δ 13C = −8.6 ‰ (cylinder number CC324315 and sample number 1720-A-25)40,41. The remaining four CO2-in-air samples, having nominal CO2 molar fractions approximately equal to atmospheric natural abundances (≈400 μmol/mol), were prepared at NIST. Briefly, the preparation involved mixing (diluting) isotopically distinct pure CO2 (>99.996 %) samples with a balance of synthetic air (N2 = 78.1 %, O2 = 20.89 %, CO2 < 0.3 μmol/mol, N2O < 0.2 nmol/mol) using gravimetric and volumetric standard preparation methods for compressed gas mixtures40,42.

The VPDB-CO2 δ 13C and δ 18O value assignments of these four samples were made at NIST on the parent CO2 by dual inlet isotope ratio mass spectrometry (DI-IRMS)43. The NIST VPDB-CO2 scale realization was achieved using NIST pure CO2 reference materials 8562, 8563, and 856444. A summary of NIST DI-IRMS results and related gas sample information is available as Table S4 of the Supplementary Information. Importantly, the uncertainty of 0.06 ‰ associated with the NIST IRMS value assignments had a minimal impact on the absolute isotope ratio R(13C/12C)VPDB estimated in this work. Further details regarding IRMS value assignment uncertainties can be found in Srivastava & Verkouteren43 and in Section S4 of the Supplementary Information.

Isotope ratio notation and nomenclature.

Here we use notation for isotope ratios recommended by the Commission on Isotopic Abundances and Atomic Weights of the International Union of Pure and Applied Chemistry (IUPAC)45.

Quantum chemistry calculations.

Rotational-vibrational transition intensities were calculated for 13C16O2 using an updated ab initio dipole moment surface (DMS) and semi-empirical potential energy surface (PES). Here we improved upon the ab initio DMS of Polyansky et al.46 by increasing the number of ab initio data points by 50 % (3000 total points). Also, the semi-empirical PES of Huang et al.47, previously used by Polyansky et al.46 without modification, was fitted to known CO2 energy levels using the DVR3D program48 resulting in a standard deviation of 0.02 cm−1. Analogous quantum chemistry methods recently applied to water (H2O) showed that both an increased number of ab initio DMS data points49 and an accurate semi-empirical PES50 will improve accuracy in predicted transition intensities.

For our denser grid of ab initio points we applied the same level of quantum chemistry theory as Polyansky et al.46: an all-electron multireference configuration interaction (MRCI) calculation with the aug-cc-pwCVQZ basis set, inclusive of relativistic corrections determined from separately fitted one-electron mass-velocity-Darwin (MVD1) terms. Calculations used the MOLPRO package51, and the wave functions—required to predict the transition intensities—were solved numerically from the rotational-vibrational Schrödinger equation using the DVR3D program48.

To evaluate uncertainty in the predicted 13C16O2 transition intensities, we also calculated transition intensities for 12C16O2 and compared with highly accurate experiments38,5254. For ground-state rotational quantum numbers of J″ = 12 and J″ = 18 we find that, for five near-infrared vibrational bands of 12C16O2, the standard deviation of a uniform distribution chosen to represent the experimental measurements relative to the predictions of our updated ab initio DMS is equal to 3.8 × 10−3. This value, listed in Table 1 of the main text as the uncertainty in the 13C16O2 transition intensities, is consistent with upper-bound estimates from quantum chemistry methods46,55. Further details, including numerical results from the updated ab initio DMS and their comparison with the experimental literature, can be found in Sections S1 and S2 of the Supplementary Information.

Supplementary Material

Supplementary Material

Acknowledgements

We acknowledge M. E. Kelley and W. R. Miller, Jr. (National Institute of Standards and Technology, NIST) for preparing several CO2-in-air gas samples. Funding was provided by the NIST Greenhouse Gas and Climate Science Program. OLP acknowledges partial support from Quantum Pascal project 18SIB04, which received funding from the EMPIR programme co-financed by the Participating States and the European Union’s Horizon 2020 research and innovations programme. NFZ acknowledges State project 0035-2019-0016.

Footnotes

Competing interests

The authors declare no competing interests.

Data availability

Source data are available for this paper at doi:10.18434/mds2-2369. All other data that support the plots within this paper and other findings of this study are available from the corresponding author upon reasonable request.

References

  • 1.Urey HC, Lowenstam HA, Epstein S & McKinney CR Measurement of paleotemperatures and temperatures of the Upper Cretaceous of England, Denmark, and the Southeastern United States. Geol. Soc. Am. Bull. 62, 399–416 (1951). doi: 10.1130/0016-7606(1951)62[;399:MOPATO];2.0.CO;2 [DOI] [Google Scholar]
  • 2.Coplen TB et al. New guidelines for δ13C measurements. Anal. Chem. 78, 2439–2441 (2006). doi: 10.1021/ac052027c [DOI] [PubMed] [Google Scholar]
  • 3.Assonov S Summary and recommendations from the International Atomic Energy Agency Technical Meeting on the Development of Stable Isotope Reference Products (21–25 November 2016). Rapid Commun. Mass Spectrom. 32, 827–830 (2018). doi: 10.1002/rcm.8102 [DOI] [PubMed] [Google Scholar]
  • 4.Phillips WD The end of artefacts. Nature Phys. 15, 518 (2019). doi: 10.1038/s41567-019-0514-8 [DOI] [Google Scholar]
  • 5.Malinovsky D, Dunn PJH, Holcombe G, Cowen S & Goenaga-Infante H Development and characterization of new glycine certified reference materials for SI-traceable 13C/12C isotope amount ratio measurements. J. Anal. At. Spectrom. 34, 147–159 (2019). doi: 10.1039/c8ja00281a [DOI] [Google Scholar]
  • 6.Kerstel E “Isotope ratio infrared spectrometry” in Handbook of Stable Isotope Analytical Techniques, de Groot PA, Ed. (Elsevier, 2004), pp. 759–787. doi: 10.1016/B978-044451114-0/50036-3 [DOI] [Google Scholar]
  • 7.Brand WA & Coplen TB Stable isotope deltas: tiny, yet robust signatures in nature. Isotopes Environ. Health. Stud. 48, 393–409 (2012). doi: 10.1080/10256016.2012.666977 [DOI] [PubMed] [Google Scholar]
  • 8.Brenna JT, Corso TN, Tobias HJ & Caimi RJ High-precision continuous-flow isotope ratio mass spectrometry. Mass Spectrom. Rev. 16, 227–258 (1997). doi: 10.1002/(SICI)1098-2787(1997)16:5<227::AID-MAS1>3.0.CO;2-J [DOI] [PubMed] [Google Scholar]
  • 9.Flores E, Viallon J, Moussay P, Griffith DWT & Wielgosz RI Calibration strategies for FT-IR and other isotope ratio infrared spectrometer instruments for accurate δ13C and δ18O measurements of CO2 in air. Anal. Chem. 89, 3648–3655 (2017). [DOI] [PubMed] [Google Scholar]
  • 10.Corso TN & Brenna JT High-precision position-specific isotope analysis. Proc. Natl. Acad. Sci. U.S.A. 94, 1049–1053 (1997). doi: 10.1073/pnas.94.4.1049 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Webster CR et al. Isotope ratios of H, C, and O in CO2 and H2O of the Martian atmosphere. Science 341, 260–263 (2013). doi: 10.1126/science.1237961 [DOI] [PubMed] [Google Scholar]
  • 12.Crosson ER et al. Stable isotope ratios using cavity ring-down spectroscopy: determination of 13C/12C for carbon dioxide in human breath. Anal. Chem. 74, 2003–2007 (2002). doi: 10.1021/ac025511d [DOI] [PubMed] [Google Scholar]
  • 13.Zare RN et al. High-precision optical measurements of 13C/12C isotope ratios in organic compounds at natural abundance. Proc. Natl. Acad. Sci. U.S.A. 106, 10928–10932 (2009). doi: 10.1073/pnas.0904230106 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Long DA, Okumura M, Miller CE & Hodges JT Frequency-stabilized cavity ring-down spectroscopy measurements of carbon dioxide isotope ratios. Appl. Phys. B 105, 471–477 (2011). doi: 10.1007/s00340-011-4518-z [DOI] [Google Scholar]
  • 15.Castrillo A et al. Amount-ratio determinations of water isotopologues by dual-laser absorption spectroscopy. Phys. Rev. A 86, 052515 (2012). doi: 10.1103/PhysRevA.86.052515 [DOI] [Google Scholar]
  • 16.Kiseleva M, Mandon J, Persijn S & Harren FJM Line strength measurements and relative isotope ratio 13C/12C measurements in carbon dioxide using cavity ring down spectroscopy. J. Quant. Spectrosc. Radiat. Transfer 204, 152–158 (2018). doi: 10.1016/j.jqsrt.2017.09.021 [DOI] [Google Scholar]
  • 17.Stoltmann T, Casado M, Daëron M, Landais A & Kassi S Direct, precision measurements of isotopologue abundance ratios in CO2 using molecular absorption spectroscopy: application to Δ 17O. Anal. Chem. 89, 10129–10132 (2017). doi: 10.1021/acs.analchem.7b02853 [DOI] [PubMed] [Google Scholar]
  • 18.Kääriäinen T, Richmond CA & Manninen A Determining biogenic content of biogas by measuring stable isotopologues 12CH4, 13CH4, and CH3D with mid-infrared direct absorption laser spectrometer. Sensors 18, 496 (2018). doi: 10.3390/s18020496 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Abe M et al. Dual wavelength 3.2-μm source for isotope ratio measurements of 13CH4/12CH4. Opt. Express 23, 21786–21797 (2015). doi: 10.1364/OE.23.021786 [DOI] [PubMed] [Google Scholar]
  • 20.Craig H Isotopic standards for carbon and oxygen and correction factors for mass-spectrometric analysis of carbon dioxide. Geochim. Cosmochim. Acta 12, 133–149 (1957). doi: 10.1016/0016-7037(57)90024-8 [DOI] [Google Scholar]
  • 21.Assonov S, Groening M, Fajgelj A, Hélie J-F & Hillaire-Marcel C Preparation and characterization of IAEA-603, a new primary reference material aimed at the VPDB scale realization for δ13C and δ18O determination. Rapid Commun. Mass Spectrom. 34, e8867 (2020). 10.1002/rcm.8867 [DOI] [PubMed] [Google Scholar]
  • 22.18th WMO/IAEA Meeting on Carbon Dioxide, Other Greenhouse Gases and Related Tracers Measurement Techniques (GGMT-2015), La Jolla, CA, USA, 13–17 September 2015, GAW Report No. 229. https://library.wmo.int/opac/doc_num.php?explnum_id=3074 [Google Scholar]
  • 23.Coplen TB & Shrestha Y Isotope-abundance variations and atomic weights of selected elements: 2016 (IUPAC Technical Report). Pure Appl. Chem. 88, 1203–1224 (2016). doi: 10.1515/pac-2016-0302 [DOI] [Google Scholar]
  • 24.Brand WA, Assonov SS & Coplen TB Correction for the 17O interference in δ(13C) measurements when analyzing CO2 with stable isotope mass spectrometry (IUPAC Technical Report). Pure Appl. Chem. 82, 1719–1733 (2010). doi: 10.1351/PAC-REP-09-01-05 [DOI] [Google Scholar]
  • 25.Chang TL & Li W-J A calibrated measurement of the atomic weight of carbon. Chin. Sci. Bull. 35, 290–296 (1990). doi: 10.1360/sb1990-35-4-290 [DOI] [Google Scholar]
  • 26.Nørgaard JV et al. The International Measurement Evaluation Program, IMEP-8: carbon and oxygen isotope ratios in CO2. Anal. Bioanal. Chem. 374, 1147–1154 (2002). doi: 10.1007/s00216-002-1572-8 [DOI] [PubMed] [Google Scholar]
  • 27.Ruße K, Valkiers S & Taylor PDP Synthetic isotope mixtures for the calibration of isotope amount ratio measurements of carbon. Int. J. Mass. Spec. 235, 255–262 (2004). doi: 10.1016/j.ijms.2004.05.007 [DOI] [Google Scholar]
  • 28.Valkiers S et al. Preparation of synthetic isotope mixtures for the calibration of carbon and oxygen isotope ratio measurements (in carbon dioxide) to the SI. Int. J. Mass. Spec. 264, 10–21 (2007). doi: 10.1016/j.ijms.2007.03.012 [DOI] [Google Scholar]
  • 29.Dunn PJH, Malinovsky D & Goenaga-Infante H Calibration strategies for the determination of stable carbon absolute isotope ratios in a glycine candidate reference material by element analyser-isotope ratio mass spectrometry. Anal. Bioanal. Chem. 407, 3169–3180 (2015). doi: 10.1007/s00216-014-7926-1 [DOI] [PubMed] [Google Scholar]
  • 30.Skrzypek G & Dunn PJH Absolute isotope ratios defining scales used in isotope ratio mass spectrometers and optical isotope instruments. Rapid Commun. Mass Spectrom. 34, e8890 (2020). doi: 10.1002/rcm.8890 [DOI] [PubMed] [Google Scholar]
  • 31.Yurimoto H et al. “Origin and evolution of oxygen-isotopic compositions of the solar system” in Protostars and Planets Reipurth V, Jewitt B, D & Keil K, Eds. (The University of Arizona Press, 2007), pp. 849–862. [Google Scholar]
  • 32.Furukawa Y et al. Extraterrestrial ribose and other sugars in primitive meteorites. Proc. Natl. Acad. Sci. 116, 24440–24445 (2019). doi: 10.1073/pnas.1907169116 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Galli I et al. Spectroscopic detection of radiocarbon dioxide at parts-per-quadrillion sensitivity. Optica 3, 385–388 (2016). doi: 10.1364/OPTICA.3.000385 [DOI] [Google Scholar]
  • 34.Fleisher AJ, Long DA, Liu Q, Gameson L & Hodges JT Optical measurement of radiocarbon below unity fraction modern by linear absorption spectroscopy. J. Phys. Chem. Lett 8, 4550–4556 (2017). doi: 10.1021/acs.jpclett.7b02105 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Prokhorov I, Kluge T & Janssen C Optical clumped isotope thermometry of carbon dioxide. Sci. Rep. 9, 4765 (2019). doi: 10.1038/s41598-019-40750-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Truong GW et al. Frequency-agile, rapid scanning spectroscopy. Nat. Photon. 7, 532–534 (2013). doi: 10.1038/nphoton.2013.98 [DOI] [Google Scholar]
  • 37.Hodges JT, Layer HP, Miller WW & Scace G Frequency-stabilized single-mode cavity ring-down apparatus for high-resolution absorption spectroscopy. Rev. Sci. Instrum. 75, 849–863 (2004). doi: 10.1063/1.1666984 [DOI] [Google Scholar]
  • 38.Fleurbaey H, Yi H, Adkins EM, Fleisher AJ & Hodges JT Cavity ring-down spectroscopy of CO2 near λ = 2.06 μm: Accurate transition intensities for the Orbiting Carbon Observatory-2 (OCO-2) “strong band.” J. Quant. Spectrosc. Radiat. Transfer 252, 107104 (2020). doi: 10.1016/j.jqsrt.2020.107104 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Fleisher AJ et al. Twenty-five-fold reduction in measurement uncertainty for a molecular line intensity. Phys. Rev. Lett. 123, 043001 (2019). doi: 10.1103/PhysRevLett.123.043001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Rhoderick GC et al. Development of a Northern Continental Air Standard Reference Material. Anal. Chem. 88, 3376–3385 (2016). doi: 10.1021/acs.analchem.6b00123 [DOI] [PubMed] [Google Scholar]
  • 41.NOAA Central Calibration Laboratory reference gas calibration results for NIST Standard Reference Material 1720 Northern Continental Air serial number CC324315 available at https://www.esrl.noaa.gov/gmd/ccl/refgas.html (last accessed on 17 February 2021).
  • 42.Milton MJT, Guenther F, Miller WR & Brown AS Validation of gravimetric values and uncertainties of independently prepared primary standard gas mixtures. Metrologia 43, L7–L10 (2006). doi: 10.1088/0026-1394/43/3/N01 [DOI] [Google Scholar]
  • 43.Srivastava A & Verkouteren RM Metrology for stable isotope reference materials: 13C/12C and 18O/16O isotope ratio value assignment of pure carbon dioxide gas samples on the Vienna PeeDee Belemnite-CO2 scale using dual-inlet mass spectrometry. Anal. Bioanal. Chem. 410, 4153–4163 (2018). doi: 10.1007/s00216-018-1064-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Verkouteren RM & Klinedinst DB Value assignment and uncertainty estimation of selected light stable isotope reference materials: RMs 8543–8545, RMs 8562–8564, and RM 8566. NIST Special; Publication 260–149 (2004). https://www.nist.gov/document-10350 [Google Scholar]
  • 45.Coplen TB Guidelines and recommended terms for expression of stable-isotope-ratio and gas-ratio measurements results. Rapid Commun. Mass Spectrom. 25, 2538–2560 (2011). doi: 10.1002/rcm.5129 [DOI] [PubMed] [Google Scholar]
  • 46.Polyansky OL et al. High-accuracy CO2 line intensities determined from theory and experiment. Phys. Rev. Lett. 114, 243001 (2015). doi: 10.1103/PhysRevLett.114.243001 [DOI] [PubMed] [Google Scholar]
  • 47.Huang X, Schwenke DW, Tashkun SA & Lee TJ An isotopic-independent highly accurate potential energy surface for CO2 isotopologues and an initial 12C16O2 infrared line list. J. Chem. Phys. 136, 124311 (2012). doi: 10.1063/1.3697540 [DOI] [PubMed] [Google Scholar]
  • 48.Tennyson J et al. DVR3D: a program suite for the calculation of rotation-vibration spectra of triatomic molecules. Comput. Phys. Commun. 163, 85–116 (2004). doi: 10.1016/j.cpc.2003.10.003 [DOI] [Google Scholar]
  • 49.Conway EK, Kyuberis AA, Polyansky OL, Tennyson J & Zobov NF A highly accurate ab initio dipole moment surface for the ground electronic state of water vapour for spectra extending into the ultraviolet. J. Chem. Phys. 149, 084307 (2018). doi: 10.1063/1.5043545 [DOI] [PubMed] [Google Scholar]
  • 50.Mizus II et al. High-accuracy water potential energy surface for the calculation of infrared spectra. Phil. Trans. R. Soc. 376, 20170149 (2018). doi: 10.1098/rsta.2017.0149 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Werner H-J, Knowles PJ, Knizia G, Manby FR & Schütz M Molpro: a general-purpose quantum chemistry program package. Wiley Interdiscip. Rev.: Comput. Mol. Sci. 2, 242–253 (2012). doi: 10.1002/wcms.82 [DOI] [Google Scholar]
  • 52.Long DA et al. High-accuracy near-infrared carbon dioxide intensity measurements to support remote sensing. Geophys. Res. Lett. 47, e2019GL086344. doi: 10.1029/2019GL086344 [DOI] [Google Scholar]
  • 53.Wübbeler G, Víquez GJP, Jousten K, Werhahn O & Elster C Comparison and assessment of procedures for calculating the R(12) line strength of the ν1 + 2ν2 + ν3 band of CO2. J. Chem. Phys. 135, 204304 (2011). doi: 10.1063/1.3662134 [DOI] [PubMed] [Google Scholar]
  • 54.Yi H, Liu Q, Gameson L, Fleisher AJ & Hodges JT High-accuracy 12C16O2 line intensities in the 2 μm wavelength region measured by frequency-stabilized cavity ring-down spectroscopy. J. Quant. Spectrosc. Radiat. Transfer 206, 367–377 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Lodi L & Tennyson J Line lists for H218O and H217O based on empirical line positions and ab initio intensities. J. Quant. Spectrosc. Radiat. Transfer 113, 850–858 (2012). doi: 10.1016/j.jqsrt.2012.02.023 [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material

Data Availability Statement

Source data are available for this paper at doi:10.18434/mds2-2369. All other data that support the plots within this paper and other findings of this study are available from the corresponding author upon reasonable request.

RESOURCES