An active electronic, high-density epidural paddle array for chronic spinal cord neuromodulation

Samuel R Parker 1, Jonathan S Calvert 1, Radu Darie 1, Jaeson Jang 2,3, Lakshmi Narasimhan Govindarajan 2,3,4, Keith Angelino 5, Girish Chitnis 5, Yohannes Iyassu 5, Elias Shaaya 6, Jared S Fridley 6, Thomas Serre 2,3, David A Borton 1,3,6,7,8,*, Bryan L McLaughlin 5,8
PMCID: PMC11920892  PMID: 40104941

Abstract

Objective. Epidural electrical stimulation (EES) has shown promise as both a clinical therapy and research tool for studying nervous system function. However, available clinical EES paddles are limited to using a small number of contacts due to the burden of wires necessary to connect each contact to the therapeutic delivery device, limiting the treatment area or density of epidural electrode arrays. We aimed to eliminate this burden using advanced on-paddle electronics. Approach. We developed a smart EES paddle with a 60-electrode programmable array, addressable using an active electronic multiplexer embedded within the electrode paddle body. The electronics are sealed in novel, ultra-low profile hermetic packaging. We conducted extensive reliability testing on the novel array, including a battery of ISO 10993-1 biocompatibility tests and determination of the hermetic package leak rate. We then evaluated the EES device in vivo, placed on the epidural surface of the ovine lumbosacral spinal cord for 15 months. Main results. The active paddle array performed nominally when implanted in sheep for over 15 months and no device-related malfunctions were observed. The onboard multiplexer enabled bespoke electrode arrangements across, and within, experimental sessions. We identified stereotyped responses to stimulation in lower extremity musculature, and examined local field potential responses to EES using high-density recording bipoles. Finally, spatial electrode encoding enabled machine learning models to accurately perform EES parameter inference for unseen stimulation electrodes, reducing the need for extensive training data in future deep models. Significance. We report the development and chronic large animal in vivo evaluation of a high-density EES paddle array containing active electronics. Our results provide a foundation for more advanced computation and processing to be integrated directly into devices implanted at the neural interface, opening new avenues for the study of nervous system function and new therapies to treat neural injury and dysfunction.

Keywords: neuromodulation, epidural electrical stimulation, implanted device, hermetic electronics, spinal electrophysiology, high density array

1. Introduction

Epidural electrical stimulation (EES) of the spinal cord has been used for the treatment of neuropathic pain for almost six decades (Nahm 2020), with over 50 000 procedures performed in the United States each year (Krog et al 2023). In therapeutic or research applications, EES leads are surgically inserted into the epidural potential space on the dorsal aspect of the spinal cord and connected to an implantable or external pulse generator (IPG or EPG, respectively) using inline contacts. EES has been used in preclinical research settings to study central, peripheral, autonomic, and sensorimotor function (Musienko et al 2012, Parker et al 2013, Gad et al 2014, Capogrosso et al 2016, Calvert et al 2023), as well as in clinical research settings to restore sensorimotor function following a spinal cord injury or amputation (Chandrasekaran et al 2020, Rowald et al 2022, Lorach et al 2023, Nanivadekar et al 2023). In parallel, epidural spinal recordings obtained through EES leads have been used as control signals for pain therapies (Nijhuis et al 2023), to study voluntary movement control (Burke et al 2021), and to examine somatosensory evoked spinal potentials (SEPs) (Nainzadeh et al 1988, Urasaki et al 1990, ÇiÇek et al 2000, Tsirikos et al 2004, Insola et al 2008, Sala et al 2013, Woodington et al 2024). However, all prior research studies and clinical applications utilize passive leads, requiring a one-to-one connection between electrical contacts on the tissue and the stimulation electronics. In practice, this has put a constraint on the number of contacts used to a maximum of 32 channels, resulting in limited selectivity during stimulation and resolution during recording.

To ensure proper placement of the EES electrodes for activation of targeted spinal circuits, a key step is intraoperative testing during implantation (Falowski 2019). During intraoperative testing, electrical stimuli are delivered through the implanted paddle. The stimuli primarily activate sensory neurons, which then recruit motor neurons via reflex pathways (Capogrosso et al 2013). Using bilateral electromyography (EMG), the response of muscles at the same spinal level as the sensory targets is evaluated, and lead position is adjusted to achieve the desired placement (Shils and Arle 2018). To recruit more specific pools of neurons on the spinal cord, multipolar stimulation is applied in a current steering approach (Chandrasekaran et al 2020, Rowald et al 2022, Lorach et al 2023, Mishra et al 2023, Nanivadekar et al 2023), which requires access to multiple contacts in a confined region of interest. Subsequently, there have been efforts to design an EES paddle optimized for particular applications (for example, locomotor restoration), using neuroimaging data and computational modeling to guide the arrangements of contacts on the paddle (Rowald et al 2022). Additionally, standard EES paddles principally deployed for the management of neuropathic pain concentrate their electrodes on the central structures of the spinal cord, and do not place stimulating contacts over lateral regions (for example, spinal nerve entry points). Activation of lateral structures has resulted in strong motor recruitment of target muscles (Calvert et al 2023), which may be more selective than stimulation more proximal to contralateral motor pools. However, the distribution of afferent entry points is not uniform along the rostrocaudal axis, again highlighting the importance of proper intraoperative alignment when using sparsely distributed electrode contacts.

The necessity for manual stimulation parameter selection is a major barrier to the widespread adoption of EES for functional improvement after spinal cord injury (Solinsky et al 2020). Consequently, there have been efforts to automate the optimization of stimulation parameters using machine learning models (Zhao et al 2021, Govindarajan et al 2022). Although these approaches have shown improvement in speed, prior models were conditioned on each electrode independently. Such an approach is effective for paddles containing low numbers of contacts but requires significant computing power to run many models in parallel for high-count paddles. Additionally, conditioning on each electrode independently does not allow the model to infer what may happen during stimulation on electrodes excluded from the training set. As the number of stimulation contacts in a paddle increases, the time to collect training data scales linearly. Instead, if neural networks can infer the consequences of stimulation at stimulation contacts not included in the training dataset, collecting training data from all contacts may not be necessary to rapidly program EES on many-contact paddles.

To address both placement limitations and stimulation parameter search optimization, a next-generation EES paddle must (1) span multiple spinal segments to target multiple sensorimotor pools, (2) provide a wide mediolateral span to target lateral structures, (3) enable localized current steering and bipolar re-referencing with densely packed electrodes, and (4) be reconfigurable for patient anatomy. These requirements are incompatible with conventional EES hardware with limited electrode contacts.

Thus, we designed a smart-implant called HD64 to provide 60 electrodes of epidural current steering for EES across a 14.5 mm mediolateral span (almost 2x wider than commercial EES paddles) and 2.5 vertebral segment span (40 mm). Our smart-implant is only 2 mm thick, and sports a high-density electrode array integrated with a hermetic electronic multiplexing package and a 24:64 reconfigurable multiplexer application-specific integrated circuit (ASIC)—thus breaking the one-to-one wiring constraint of percutaneous EES hardware. The programmable smart-paddle enables the spatial layout of stimulating electrodes to be software-controlled dynamically, and is powered by a ±5 V AC power driver ASIC with fail-safe AC power leakage-detection circuits. The leakage detection circuit acts as a ground-fault detection system, powering down the ASIC if power flows to ground through an alternative path (and maintaining normal operation when power returns down the dedicated AGND and DGND return lines). This ensures the power required for the operation of the active implant remains isolated from the user of the HD64. For future human use of the device, we developed good manufacturing practice (GMP) processes (achieving a manufacturing yield of 85%) and performed ISO 14708 aging testing as well as ISO 10993-1:2021 biocompatibility testing. A comparison between three commercial EES paddles with the work presented here can be found in table 1.

Table 1.

A comparison of the physical characteristics of commercially available epidural electrical stimulation paddles and the paddle developed in this work, the HD64.

5-6-5 (Medtronic 2022) CoverEdge (Boston Scientific 2020) CoverEdgeX (Boston Scientific 2020) HD64 (this work)
Manufacturer Medtronic (Minneapolis, MN) Boston Scientific (Marlborough, MA) Boston Scientific (Marlborough, MA) Micro-Leads Medical (Somerville, MA)
Length 64.2 mm 50 mm 67 mm 57.5 mm
Treatment width 10 mm 8 mm 8 mm 14.5 mm
Thickness 2.0 mm 2.0 mm 2.0 mm 0.2 mm edge, 2.0 middle
Number of contacts 16 32 32 60
Inter-electrode spacing (mediolateral, rostrocaudal) (1.0 mm, 4.5 mm) (1.0 mm, 1.0 mm) (1.0 mm, 3.2 mm) (0.7 mm, 0.9 mm)
Density vs 5-6-5 1 2.1 1.4 2.2
Percutaneous lead-tails 2 4 4 2
Contact surface area 6 mm2 3.4 mm2 3.4 mm2 4.15 mm2
Contains an active multiplexer No No No Yes

The culmination of this work resulted in the long-term evaluation of the smart HD64 paddle for EES using benchtop verification and in-vivo validation in two sheep with up to 2 paddles per animal for 15 months. During this time, our goals were to establish the utility of HD64 to perform high density EES studies and observe any device-related functional issues. We quantified the selectivity of EES-evoked motor responses delivered using dense bipoles, differences in SEPs recorded using high- and low-density bipoles, and we extended the state-of-the-art stimulation parameter inference machine learning models to high-density electrodes (Govindarajan et al 2022).

2. Methods

2.1. Treatment span of sensory-motor EES

In the T11 to L1 vertebral regions, the transverse diameter of the human spinal cord is 8–9.6 mm (Ko et al 2004, Fradet et al 2014), the spinal canal sagittal depth ranges between 15.4 and 19.54 mm, and the spinal canal transverse width spans 16.7–26.5 mm (Laporte et al 2000, Busscher et al 2010). Recent work has highlighted the benefit of activation of lateral structures (Calvert et al 2023), and their potential as a target for locomotor neuroprosthesis following spinal cord injury (Rowald et al 2022). Conventional surgical paddles for pain are limited to dorsal column medio-lateral treatment (<7.5 mm with 4 electrode columns) across two vertebral segments (<49 mm with 8 electrodes rows). Extending the medio-lateral therapeutic span of EES delivery will facilitate accessing the dorsal horn and dorsal roots. Based on the desired treatment area, the paddle requirements are to treat a 13.7 × 43 mm stimulation area. Extending the treatment surface area for EES faces two challenges: (1) the lateral epidural potential volume adjacent to nerve roots is much thinner than at the midline, and (2) performing EES with a sparser electrode array would reduce the targeting resolution, potentially introducing off-target effects or non-specific activation.

2.1.1. Mechanical requirements for surgical introduction of the lead

The paddle geometry was designed to mimic conventional EES paddles at the anatomical midline using a 2 mm thick geometry. Due to the 1 mm lateral epidural thickness, the paddle lateral edges were limited to 0.7 mm. Ten cross-sectional geometries of electrode profiles were developed in groups of three and sequentially evaluated. Each design was modeled using computer-aided design software and molded in silicone to exhibit various longitudinal and lateral flexibility profiles. The profile groups were sent to three neurosurgeons to evaluate the preferred paddle design (criteria for their evaluation included steerability, axial stiffness, and bendability) while accessing the epidural space through a small laminectomy. The resulting profile (figure 1(l)) was the final profile for two key reasons: (1) the lateral wings remained flexible and could bend through a conventional-width laminectomy, and (2) a relatively rigid paddle was strongly preferred as it enabled physicians to advance the paddle rostrally using forceps.

Figure 1.

Figure 1.

The HD64: active electronics hermetically sealed into an epidural electrode paddle array. (a) A functional block diagram of the application specific circuit (ASIC) embedded on the spinal paddle. (b) A photograph of the ASIC die (redistribution layer and solder ball placement shown). (c) An exploded 3D render of the hermetic package. From top to bottom, we show the titanium 6–4 top case, the balled ASIC, a novel alumina feedthrough array, and the feedthrough pins. (d), (e) Photographs of the assembled hermetic feedthrough package. (f) The hermetic package is precision aligned within <10 μm onto the electrode before conductors are thermally micro-bonded followed by silicone injection molding. (g) A histogram of the hermeticity leak rates measured from 171 hermetic packages. The MIL-STD-883K helium leak failure threshold is overlaid as the red dashed line. (h) Left to right: a photograph of the dorsal side of the molded HD64 paddle array, a photograph of the 2 12-contact lead tail connectors, and a photograph of the ventral side of the array, showing the 60-contact grid. (i) Side- and front-on photographs comparing the lower-volume profile of the HD64 compared to a Boston Scientific CoverEdge32 surgical paddle.

2.1.2. Smart-implant multiplexing architecture

Using a fixed number of electrodes (e.g. 16 or 32) spread across a wider medio-lateral span would result in a sparser electrode density and result in a reduced EES targeting resolution. Given the lateral nerve root density, a sparse density was undesirable. Conventional implanted surgical paddles have reached 32 electrodes, but do not exhibit stimulation sites over the nerve roots and require management and connection of 4 lead tails below the skin. Scaling to 64 electrodes using current connector technology would require 8 lead-tails and 64-channel connector on the IPG, which would greatly increase bulk in the body and surgical complexity. Since EES generally needs access to many electrodes but only a subset are used at the same time, we developed a multiplexing architecture to programmably connect electrodes to the pulse generator. Specifically, we sought to develop a scalable manufacturing process to produce paddle electrodes.

2.2. Implant-grade design requirements

HD64 is a smart-paddle containing embedded electronics that is powered and controlled by a pulse generator over a two-lead tail wiring scheme. To realize an active-multiplexed electrode array, we developed multiple new implant technologies including: (1) a high-voltage multiplexer that operates from charge-balanced AC power (figures 1(a) and (b)), with known noise performance (Rachinskiy et al 2022), (2) a high-density and low-cost 93-pin hermetic feedthrough package with >80% yield (figures 1(c)–(e)), (3) precision electrode technology (platinum–iridium with 100 μm lines and 100 μm spaces), and (4) medical-grade micro-bonding processes to permanently bond 85 electrical connections between the hermetic electronic package and electrode array.

For long-term implanted use, we developed requirements based on our clinical inputs for people with spinal cord injury and/or chronic pain in the lower extremities. We established electrical and mechanical safety requirements and testing paradigms in accordance with ISO 14708-3:2017. We further developed a biocompatibility evaluation plan in accordance with ISO 10993-1:2018. The device also had to be manufactured according to GMP and undergo ethylene oxide (ETO) sterilization validation.

2.2.1. Electrical safety: AC-powered implanted satellite devices

In this work, we describe two specific and critical design elements that enable new smart implants including: (1) ultra-low current satellite multiplexer operation using charge-balanced AC power over an implanted lead and pluggable-connector interface, (2) digital charged-balance programmable control of the distal satellite multiplexer. We tested commercial leads and connectors with ±10 V AC power using an accelerated aging paradigm in a sealed saline container. At the end of a 5 year implanted life equivalent, the saline was tested for residual traces of platinum–iridium using inductively coupled mass spectrometry to test for any electromigration. We determined that ±5 V AC through the lead-wire with a 2x voltage multiplier within the hermetic electronics package prevented any risk of electromigration. Additionally, we developed a separate IPG-side driver ASIC chip to deliver low-current ±5 V AC power with charged-balanced differential data lines to program the multiplexer. Board logic-level programming of a satellite multiplexer over a lead wire is not an acceptable risk, as any DC potential on the line can lead to electromigration at the connector. The driver ASIC was designed to deliver a programmable AC-current limited to 10–100 μA and 10–500 kHz to power the multiplexing ASIC (note, this is independent of therapeutic stimulation frequency). Importantly, a novel real-time leakage current detection circuit was developed such that any broken wire or connection could be detected by the ASIC which would then generate an internal digital fault flag.

2.2.2. Biocompatibility evaluation

For translation to future human implanted use, we determined that the following ISO 10991-1 tests must be performed: cytotoxicity, sensitization, irritation, material-mediated pyrogenicity, acute systemic toxicity, subacute toxicity, implantation, genotoxicity, ethylene oxide residuals, and partial chemical characterization. To achieve these endpoints, approximately 300 active-paddle test articles were sent to a certified ISO 10993-1 test house over a 14 month duration. All test articles successfully passed these tests, and no histological abnormalities were identified by the test house following any of the in vivo tests. A summary of the biocompatibility tests performed and their results are included in supplementary table 1.

2.2.3. Precision electrode technology

Precision electrodes were developed with a highly-novel delamination-free process using medical-implant grade silicone (Nusil, Carpentaria, CA), platinum–iridium 90/10 conductors, and a nylon mesh reinforcement layer (figure 1(f)). In this process, the silicone and platinum–iridium are pre-processed using a proprietary laser patterning technique, resulting in 50-100 μm electrode features in 50 μm thick platinum–iridium 90/10. A flexible reinforcement mesh layer was embedded within the silicone to prevent stretching of the silicone to protect the electrode traces from fracture during stretching or repetitive pull cycles. The HD64 paddle consists of three silicone and two metal layers for a total substrate thickness of 400 μm. Using a proprietary process, the silicone layers are chemically fused together to form a seamless and delamination-free construction for long-term implanted operation.

2.2.4. High-voltage 24:64 multiplexing ASIC with AC power and charge-balanced programming

An onboard multiplexer ASIC (figure 1(a)) was developed to support two key functions needed for this architecture: (1) a fail-safe low-voltage and low-current AC powering scheme with on-chip ±9 V compliance voltage multipliers and a bidirectional charge-balanced digital read-write scheme, (2) a 24:64 switch matrix to allow 24 bidirectional wires to connect to any of the 60 high-density electrodes. Fabrication was performed using the X-FAB XH035 18V 350 nm silicon process to construct the ASIC. A polyimide redistribution layer was developed to provide a solder-ball flip-chip interface between the ASIC (figure 1(b)) and hermetic feedthrough (figure 1(c)). The multiplexer is designed to be dynamically programmed using a ±1.8 V charge-balanced, differential digital interface operating from ±5 V AC power. The AC power, 2 programming lines, 2 ground lines, and 19 bidirectional contact connections (24 conductors total) are distributed across 2 12-contact lead tails. Once powered, programming can be used to connect any of 24 input wires to 64 outputs. The multiplexer contains 64 blocks, each containing four switches to ensure every output electrode has a redundant connection to at least four stimulation/recording inputs. Switches were assigned to blocks in such a way as to maximize the number of switches that may become nonfunctional while still enabling sequential programming to raster through the entire 60-electrode array. The ASIC operates from AC power (±5 V AC, 10–500 kHz) and features real-time current-limiting. Charge balance is necessary to prevent corrosion and long-term electro-migration of the conductors. The ASIC uses an on-board rectifier with off-chip capacitors to convert the AC signal to ±9 V DC internally and 3.3 V.

2.2.5. High-density hermetic electronics package

We developed a custom 93-feedthrough ceramic assembly (8 × 8 × 0.75 mm) brazed into a titanium flange. On the interior of the package, the ASIC was flip-chipped to the surface of the feedthrough array (figure 1(c)). A low-outgassing underfill was applied between the ASIC and the ceramic and inspected for voids. A moisture getter was applied to the interior titanium lid, which was then seam-welded to form a hermetic assembly. Surface mount capacitors were soldered to metal pads on the ceramic feedthrough surface. The lid was attached to the feedthrough array and a laser seam-weld was performed in a nitrogen–helium environment. On the exterior of the hermetic packaging, the feedthroughs emerge from the ceramic surface and serve as a surface for 93 subsequent micro-bonds to be performed with the electrode array (figures 1(d) and (e)).

2.2.6. Electrode-hermetic electronics integration

The completed hermetic electronics assembly and high-density electrode assembly were then micro-bonded using a proprietary process. Rigorous machine vision was used to perform automated inspection of the 93 feedthroughs and receiving electrode pad positions. Components with >35Fixturing and vision cameras were used to align the components together until the feedthroughs and receiving pads were overlapping with <20 μm offset (figure 1(f)). Machine vision was used to inspect the tolerances of the electrode, and the feedthrough component and components with >35 μm of tolerance were discarded to ensure reliability of the bonding process. After the precision alignment was performed, a custom-made machine was developed to perform thermal welding between each platinum–iridium pad and each platinum–iridium feedthrough.

2.2.7. Injection molding and lead-tail attachment

The proximal end of the HD64 smart paddle uses two connector tails, each with 12 in-line ring contacts (figure 1(h)), which have decades of long-term implant reliability (Letechipia et al 1991) and are compatible with IPGs. The lead-tails were manufactured and welded to the platinum–iridium receiving pads on the HD64 substrate. A silicone injection molding step was used to create the contoured paddle profile of the HD64, as well as to completely insulate between all of the feedthrough pins. The high-pressure injection molding process with appropriate ports and runners ensured that no air voids were present between feedthrough pins. All fixtures and processes were developed to ensure the silicone flow did not trap bubbles during the assembly process. Following this process, the HD64 was completely assembled and ready for use (figure 1(h)).

2.3. Benchtop evaluation of the HD64

2.3.1. Mechanical testing of the array body and lead tails

To evaluate the performance of the HD64 following the mechanical forces experienced during surgical implantation and in the epidural potential space a set of mechanical tests were conducted. The tests assessed the electrical performance following flexion of independent sections of the paddle-lead tail assembly (as these sections experience different flexion profiles over their lifetime), and assessed the tensile performance of the entire assembly.

To determine the tensile properties of the assembly, 29 HD64 samples (58 lead tails) underwent standardized tensile assessment. All samples were soaked in 0.9 gl−1 ± 10% NaCl solution at 37° ± 2° C. The initial lengths of the sample were measured using a calibrated ruler. The sample under test was then placed into a tensile testing fixture (MTS Systems, Eden Prairie, MN). The array body was placed in the bottom grip, using silicone pneumatic grips, and the proximal ends of the lead tail being assessed was secured in the top grip using a serrated fixture. The 5.5 N tensile load setting of the testing fixture was used to achieve a 5 N tensile force (50 mm min−1 velocity), which was held for 60 s. After testing both lead tails, the sample was removed from the fixture and the final lengths of the samples were measured. Electrical isolation between each conductor and contact site. To pass this assessment, lead tails must exhibit permanent elongation less than 5%, and electrical isolation between all tested nets. The mean permanent elongation observed was All 29 samples measured permanent elongation less than 5% ((3.6 × 10−4 ± 4.3 × 10−4)%, mean ± standard deviation). The distribution of observed permanent elongation is presented in supplementary figure 1(a). Of the 29 samples tested, 1 failed electrical testing (a disconnect between a conductor in the lead tail and the multiplexing ASIC, and two shorts between pairs of lead tail conductors). Such a failure would be easily detectable during system testing prior to implantation of the device, and would simply result in a short delay of surgery while a backup device was located. The low failure rate observed following tensile testing, and low severity impact resulted in an acceptable level of risk.

During surgical implantation, sections of the HD64 experience differing flexion forces. Therefore, independent tests were developed to adequately assess the electrical performance of the HD64 assembly after exposure to these forces. Firstly, to simulate the flexion experienced by the array body during implantation and in the epidural potential space, 29 array bodies were flexed using a custom flexion testing fixture to ±22° for 1500 cycles, with a 0.5 s cycle duration. Following the flexion cycles, the electrical isolation between each conductor and contact site was assessed. In all 29 samples tested, all electrical connections remained nominal, with no open or short circuit conditions created. The second aspect to be tested was the most distal end of the lead tail, which is typically coiled and secured to form a strain relief loop. To assess this section of the lead tail, the paddle bodies were held secure, and the distal section of the lead tails were flexed to ±45° for 1500 cycles, with a 0.5 s cycle duration. Following the flexion cycles, the electrical isolation between each conductor and contact site was assessed. In 3 of the 29 tested samples new short or open circuit conditions were observed, though these were all below the 3 short or open conductor limit afforded by switch matrix redundancy. Finally, the impact of repeated flexion on the resistance of the distal section of the HD64 (simulating the implanted 2 year implanted lifetime required of the device for this study) was assessed. For this test, the distal section of the lead tail was sectioned to be 20 cm in length. With the paddle body secured, the lead tail was flexed to ±90° for 47 000 cycles, with a 0.5 s cycle duration. Following the flexing cycles, the lead tail was cut to separate it from the paddle body, and the 12 conductors in each lead tail was removed from the tail to facilitate resistance testing. The maximum allowable resistance for the 20 cm section is 6 Ω (giving a resistivity of 30 Ωm−1). The distribution of observed resistances is shown in supplementary figure 1(b). All measured conductors exhibited a post-flexion resistance below the 6 Ω criteria, indicating minimal impacts on lead resistance due to flexion over the implanted lifetime of the device. The successful completion of mechanical testing indicated a robust design, which continued to hermetic testing.

2.3.2. Hermeticity of the electronics package

To protect the embedded active electronics from the ionic environment in the epidural space, it is necessary to ensure the hermeticity of the electronics package and feedthrough assembly. Hermeticity model calculations were performed using a 2 year shelf life and a 1 year implanted life. Hermetic tests for electronics (MIL-STD-883 Method 1014.15) have failure limits based on the dew point inside a free volume package. The internal free volume of the HD64 assembly was 0.02 cm3, which corresponds to a failure leakage rate of 5 × 10−9 atm cm3 s−1 of air. This is equivalent to 13.5 × 10−9 atm cm3 s−1 of leakage using helium, which is used to test hermetic components. Using the laser interferometer method, compliant with the standard, the leak rate was measured for 171 hermetic packages. Figure 1(g) shows the distribution of measured leak rate values. Less than 10% of assembled packages had gross leak rates and were sent for subsequent testing. All accepted packages recorded a leak rate at least 15 times lower than the failure threshold set by MIL-STD-883. Based on the empirical leak rates 5 × 10−10, hermeticity calculations suggest the internal free volume of the package will not reach a dew point until far beyond a 2 year shelf life and 2 year implanted use. Using our measured leakage values, the dimensions of the device, and the material properties of our moisture-getter, the calculated hermetic lifespan of the hermetic electronics package is approximately 4 years (Greenhouse, Lowry, and Romenesko 2012, 71). Additional validation may be required prior to implanting the HD64 for longer than this calculated lifespan. We were therefore confident that the hermetic package could continue safely to 15 month in vivo testing.

2.3.3. Control of active-multiplexing

As all 60 contacts are bidirectional (capable of simultaneous stimulation and recording), maintaining knowledge of the state of the HD64 multiplexer was necessary to demultiplex the recorded spinal responses and stimulation information post-hoc. A schematic representation of the connections and devices used in this manuscript is presented in figure 2(a). The HD64 is powered and controlled by an external controller (MB-Controller, Micro-Leads Medical, Somerville, MA), which communicates with a host PC over a serial connection. MATLAB (version 2023b, MathWorks, Natick, MA) functions were provided to connect to, configure, and read from the multiplexer. Stimulation was also controlled using MATLAB. A custom script was written to ensure that the multiplexed stimulation channel was connected to the target electrode contact. Additionally, changes to the multiplexer configuration were logged alongside stimulation information, allowing synchronization with the recorded electrophysiological signals. This enabled recorded multiplexed signals to be split as the multiplexer configuration changed and a sparse 60 × n matrix of demultiplexed signals to be created, where n is the recording length in samples. While the multiplexing ASIC onboard the HD64 is capable of rapidly switching between connection configurations (though this increases the noise floor from 1.11 μVrms using an Intan RHD2164 alone to 2.65 μVrms with the HD64 connected and rapidly switching (Rachinskiy et al 2022)), in this study multiplexer configurations remained static during an experiment.

Figure 2.

Figure 2.

An overview of the in vivo evaluation of the HD64. (a) A schematic representation of the connections and devices used to evaluate the HD64. The host PC controls the collection of spinal electrophysiology, application of transcutaneous electrical nerve stimulation (TENS) and epidural electrical stimulation (EES), and the configuration of the onboard ASIC through the multiplexer controller. In S1 and S2, electromyographic (EMG) signals are collected from implanted, wireless intramuscular EMG telemetry devices or wireless surface EMG devices, respectively. AFE is the analog front end, BF is the biceps femoris, Gas is the gastrocnemius, and EDL is the extensor digitorum longus. (b) A radiograph of sheep 1, showing a single HD64 implanted. (c) A radiograph of sheep 2, showing two HD64s implanted. (d) The violin plots demonstrate that contact impedance remained stable over time, up to 274 d post-implantation.

2.4. In vivo evaluation of the HD64

All surgical and animal handling procedures were completed with approval from the Brown University Institutional Animal Care and Use Committee (IACUC), the Providence VA Medical System IACUC, and in accordance with the National Institutes of Health Guidelines for Animal Research (Guide for the Care and Use of Laboratory Animals). Two sheep (both female, aged 4.19 ± 0.3 years, weighing 92.5 ± 2.5 kg) were used for this study (figures 2(b) and (c)). Recordings were performed with both sheep for 15 months. Animals were kept in separate cages in a controlled environment on a 12 h light/dark cycle with ad libitum access to water and were fed twice daily. The ovine model was chosen for this study as the spine and spinal cord are comparable in size and share many anatomical features with humans, and the use of the ovine model to study the spinal cord has been well established (Marcus et al 1997, Parker et al 2013, 2018, 2020, Wilson et al 2017, Reddy et al 2018). A visual overview of the experimental setup is shown in figure 2(a).

2.4.1. Surgical procedures

The sheep were implanted, as previously reported (Calvert et al 2023). However, S1 was implanted with a single paddle, as the implanting neurosurgeons identified a narrow epidural space at L3 in this animal. Using only the caudal paddle array limits the rostrocaudal span accessible in this sheep, though the target neural structures underlying each array will be assessed independently. To briefly describe our surgical approach, under propofol-based general total intravenous anesthesia, an L4–L6 laminectomy with medial facetectomy was performed. The rostral HD64 paddle (if included) was gently placed, then slid rostrally, after which the caudal paddle was placed. Strain relief loops were made, and then the lead-tails were tunneled to the skin, where they were externalized. Reference and ground electrodes (Cooner AS636 wire, Cooner Wire Company, Chatsworth, CA) were secured epidurally and in the paraspinal muscles, respectively. Strain relief loops were made, and these wires were also tunneled then externalized. Intraoperative testing was used to confirm device functionality. The animals were allowed to recover from anesthesia then returned to their pens.

A second surgical procedure was performed to place intramuscular EMG recording equipment (L03, Data Sciences International, St. Paul, MN) in the lower extremity musculature of S1. The sheep was intubated and placed prone on the operating table. The sheep was placed in a V-shaped foam block to maintain stability throughout the procedure, and a rectangular foam block was placed under the hip to alleviate pressure on the hindlegs (Calvert et al 2023). The legs were hung laterally off of the operating table, and each of the hooves was placed in a sterile surgical glove and wrapped in a sterile bandage for manipulation during the procedure. On each side, the extensor digitorum longus, biceps femoris, and gastrocnemius were identified by palpating anatomic landmarks. Small incisions were made over the bellies of each of the three muscles to be instrumented, and a small subcutaneous pocket was made on the flank. All three channels were tunneled from the subcutaneous pocket to the most proximal muscle. There, a single recording channel and its reference wire were trimmed to length and stripped of insulation. The bare electrodes were inserted into the muscle belly perpendicular to the muscle striations, through 1–2 cm of muscle. The electrodes were secured by suturing to the muscle at the insertion and exit points of the muscle belly. The remaining two channels were tunneled to the next muscle, where the process was repeated before the final electrode was tunneled to the most distal muscle and secured in the same manner. Finally, the telemetry unit was placed in the subcutaneous pocket on the upper rear flank and secured using a suture. Approximately 0.5 g of vancomycin powder was irrigated into the subcutaneous pocket, and the pocket was closed. The process was then repeated on the other side. After brief intraoperative testing, the animal was allowed to recover from anesthesia then returned to its pen. Both sheep recovered full ambulation in less than 6 h.

2.4.2. Post-implantation monitoring

Both sheep were monitored at least twice daily by trained veterinary staff, who assessed changes in bladder function, respiration rate, food and water intake, and gastrointestinal output. Locomotor performance was assessed by a trained animal technician, who administered over 20 min of overground or treadmill-based walking assessments per day. No signs of device-related adverse events were noted.

2.4.3. Recording of EES-evoked motor potentials

At the beginning of the experimental session, the awake sheep was hoisted in a sling (Panepinto, Fort Collins, CO) until clearance between its hooves and the floor was observed. The HD64 was connected, enabling simultaneous stimulation, recording, and control of the multiplexer. Stimulation amplitude ranges were identified for each sheep, ranging from below motor threshold to the maximum comfortable response above motor threshold. Five stimulation amplitudes were selected in the comfortable range, and each stimulation was delivered at four frequencies (10, 25, 50, and 100 Hz).

In monopolar stimulation trials, stimulation was independently applied to each of the 60 contacts. A cathode-leading, 3:1 asymmetrical, charge-balanced waveform was used in all cases. The cathodic-phase pulse width was 167 μs, and the stimulation train duration was 300 ms. The stimulation anode was the implanted reference wire in the paraspinal muscles. The inter-train interval was randomly drawn from a uniform distribution spanning 1–2 s, and stimulus presentation order was randomized.

In bipolar stimulation trials, bipolar pairs were defined such that the spacing between the cathode and anode represented the minimum spacing possible on either the HD64 or Medtronic 5-6-5 paddles. Here, stimulation was provided at 5 amplitude values in the comfortable range for each sheep at a frequency of 10 Hz. This 100 ms inter-pulse-interval was selected to maximize the latency between stimuli so that the neural response to the previous pulse had subsided prior to the delivery of the subsequent stimulation. The stimulation waveform shape, train duration, and inter-train interval were unchanged from the monopolar trials. Stimulation presentation was again randomized.

Data was collected during stimulation across 60 electrode contacts per spinal paddle and 6–8 EMG channels. The spinal, EMG, and stimulation data were synchronized by injecting a known bitstream into the time series data of the spinal and muscular recordings. Stimulation sessions did not exceed three hours, and the sheep were constantly monitored for signs of distress and fed throughout each session. At the conclusion of recording, the animal was returned to their pen.

2.4.4. Recording of EES-evoked spinal potentials

At the beginning of each session, the awake sheep was hoisted in a sling, and the HD64 was connected as described previously. The HD64 was routed to create a stimulating bipole between the caudal-most midline electrodes and a recording bipole, either 7.3 mm, 17.0 mm, 26.7 mm, or 31.6 mm more rostral than the stimulating bipole. As the recording device (A-M Systems Model 1800) consisted of only two recording channels, each of the bipoles was recorded sequentially. For each stimulation event, a cathode-leading, symmetrical EES pulse was delivered to the stimulation bipole. The EES amplitude was set to 2 mA, 4 mA, and 6 mA, while the width of the cathodic phase was kept constant at 25 μs.

2.4.5. Recording TENS-evoked local field potentials on the spinal cord

With the sheep hoisted in the sling apparatus, the bony anatomy of the right-side hind fetlock was palpated, and a ring of wool just proximal to the fetlock was shorn, extending approximately 10 cm proximally. The exposed skin was cleaned with isopropyl alcohol, which was allowed to air dry. 2’ × 2’ TENS (Transcutaneous Electrical Nerve Stimulation) patches (Balego, Minneapolis, MN) were cut to size, placed on the medial and lateral aspects of the cannon bone, then connected to the stimulator device (Model 4100, A-M Systems Inc, Carlsborg, WA). The stimulation monitoring channel was connected to an analog input on the EMG system for synchronization. Stimulation amplitude was set to 25 V, and stimulation pulse width was 250 μs. The interstimulus interval was 500 ms. Spinal local field potentials (LFPs) were recorded from contacts on the HD64. The above process was then repeated for the left hindlimb.

2.5. Extension of state-of-the-art EES parameter inference neural networks

2.5.1. Model reparameterization and data acquisition

To extend the state-of-the-art EES parameter inference neural network models to 60 electrodes, the input space was reparameterized. The original model accepted a 3D input feature vector: amplitude, frequency, and electrode index (Govindarajan et al 2022). The new model used here now accepts a 4D input vector: amplitude, frequency, electrode mediolateral coordinate, and electrode rostrocaudal coordinate (i.e. inputting an xy location rather than hardcoding an arbitrary electrode number). By defining the electrode position on a continual space, positional relationships between each electrode could be learned. As in the original model, frequency and amplitude values were normalized to the range [0, 1).

When surgically placing the electrodes, it is impractical to ensure the electrode is perfectly aligned with the spinal cord in two dimensions, and is sitting flush with the dorsal surface. Additionally, the paddle may shift slightly during closing. For each sheep, radiographs were collected following the implantation of the HD64 electrodes. By examining the orientation of each electrode contact, an affine transformation was computed to account for skew between the electrode paddles and the radiograph camera (or equivalently, the spinal cord, as the sheep was placed sternally with the radiograph camera perpendicular to the longitudinal plane of the spinal cord). In this transformed space, the relative position of the rostral paddle (if placed) was determined with respect to the caudal paddle. Then, using the known dimensions of the HD64, the coordinates of each electrode contact were calculated. The electrode coordinates were normalized to the same range as the amplitude and frequency (the left-most caudal electrode contact had a coordinate of (0, 0), with increasing values extending rightward and rostrally). This process enabled the contact separation of the HD64 to be correctly determined, even if the HD64 was not sitting perfectly aligned with the anatomy.

The data acquisition phase proceeded as described for recording of monopolar EES-evoked motor responses.

2.5.2. Data handling & evaluation

The data handling and preprocessing steps were performed as previously described (Govindarajan et al 2022). Briefly, the EMG recordings from bilateral extensor digitorum longus, biceps femoris, and gastrocnemius were high-pass filtered at 3 Hz, then a second-order infinite impulse response notch filter with a quality factor of 35 at 60 Hz. EMG envelopes were calculated by computing the moving RMS (root mean squared) value of each signal for a 300 ms window (50% window overlap). The enveloped data was epoched from 100 ms prior to the onset of stimulation to 300 ms after the conclusion of the stimulation train (700 ms epochs). The epoched data was labeled with the stimulation amplitude, frequency, and electrode coordinates determined by the map described previously. This dataset was the 100% density dataset, as it contained every stimulation electrode. The 50% density dataset was created by removing 30 electrode contacts from the 100% density dataset. Finally, the 25% density dataset was created by removing an additional 14 contacts from the 50% density dataset. The remaining data handling steps (outlier removal, unreliable sample removal, subthreshold EMG removal, and EMG summarization) were performed as previously described (Govindarajan et al 2022). Independent models were then trained for each density dataset.

Following the completion of training and inference, stimulation proposals were generated for a set of target EMG responses obtained from testing EES parameter sets (40% of all conditions were held out, as in (Govindarajan et al 2022)). Since these testing data were held out from the training data (the remaining 60% EES parameter conditions), the combination of target EMG responses and their EES parameters has never been shown to the network during the training phase. Proposals to achieve these target EMG responses were generated independently by all three models. If the stimulation contact coordinate inferred for a proposal was not encircled by a stimulation contact, the proposal was snapped to the nearest contact. The proposed stimulation amplitudes were checked to comply with the ranges the researcher found comfortable for each sheep and through software limits. To assess changes in the distribution of evoked motor responses, a set of stimulation parameters applied during the collection phase were repeated at the start of the evaluation phase. Then, each proposal was delivered 10 times, and the proposal order was randomized. After collection, the L1 error between the proposed EMG response and the achieved EMG response was calculated.

2.6. Statistics and data processing

Following the completion of an experimental session, the recorded data were analyzed offline using custom-written code in MATLAB (R2023b) and Python 3.8.17 (using SciKit-Learn version 1.4.1 (Pedregosa et al 2011) and uniform manifold approximation and projection (UMAP) module (McInnes et al 2018)). To compute recruitment curves, the recorded EMG signals were split into 500 ms stimulation-triggered epochs. This epoch length was selected to capture the 300 ms pulse train and residual motor effects. The data was then rectified, and the power was recruited by taking the absolute value and then summing all samples in each epoch. The power values were then normalized to a range of [0, 1] for each muscle by subtracting the minimum activation and dividing by the range of activations. The EMG power (rectified area under the curve, or rAUC) was averaged across the four stimulation trials for each amplitude, frequency, and electrode combination, and the standard deviation was computed. To identify statistical clusters of evoked EMG responses, the single-trial recruitment data was rearranged into a 2D table, where rows were stimulation electrodes, and the recruitment data for each stimulation frequency was horizontally concatenated to form the columns. Raw rAUCs were normalized using the Yeo–Johnson power transformation (Yeo and Johnson 2000), and used as input features for the UMAP algorithm (Gorman 2018, McInnes et al 2018). A two-dimensional embedding was generated, where each point represents the rAUC response pattern across muscles, stimulation amplitudes, and frequencies for a given electrode (four points per electrode, corresponding to the four stimulation repetitions). Spectral clustering (Shi and Malik 2000) was used to identify between 1 and 16 clusters of electrodes in the dataset, representing groups of electrodes with similar rAUC response patterns. A silhouette analysis (Rousseeuw 1987) indicated that the optimal number of clusters was eight. If the four repeats of the same electrode did not fall into the same cluster, that electrode was assigned the cluster label corresponding to the majority of the four labels. The process of epoching, rectifying and integrating the EMG signal was repeated for the bipolar stimulation datasets. Before normalizing the bipolar data, the unnormalized monopolar data was included such that both monopolar and bipolar EMG responses were normalized to the same range. The EMG distributions for each stimulation configuration were compared using a Mann–Whitney U test. This non-parametric test was selected as it does not assume normality of the underlying distributions. Finally, the selectivity indexes were computed for each muscle in all stimulation electrode configurations (Badi et al 2021, Bryson et al 2023). The distributions of selectivity indexes were also compared using a Mann–Whitney U test.

Spinal evoked compound action potentials (ECAPs) and SEPs were first split into 500 μs and 100 ms stimulation-triggered epochs, respectively. The epochs were averaged across 50 stimulation trials and filtered with a low-pass filter at 2000 Hz. The mean and standard deviation of the ECAPs were calculated for each amplitude and recording bipole. The latency of the peak of the response was identified for each of the 50 single trials (Chakravarthy et al 2020). A linear fit of the latency-distance relationship was calculated, and the gradient of this fit was calculated to determine the conduction velocity (CV) (Parker et al 2013, Lam et al 2024). The SEPs were re-referenced by subtracting the inverting electrode recording from the non-inverting electrode recording for each recording bipole (Verma and Romanauski et al 2023). The mean SEP response was calculated for each recording channel. To assess the spatial correlation between recorded channels, the normalized zero-lag cross-correlation coefficient was computed between each recording channel for single trials (Swindale and Spacek 2016). For each channel, the mean pairwise correlation to other channels was calculated over each trial. To assess the uniqueness between channels, the absolute value of the correlation was computed and then subtracted from 1. That is, perfectly correlated channels, with a mean pairwise correlation coefficient of 1, would receive a score of 0, while completely uncorrelated channels would receive a score of 1. The distribution of uniqueness scores was compared using a Kruskal–Wallis test. This non-parametric test was selected as it does not assume normality of the underlying distributions.

To evaluate the performance of the EMG response simulation neural network (the forward model), the L1 loss was computed by subtracting the predicted EMG vector from the evoked EMG vector and then taking the absolute value. The L1 loss was selected as the error term as this function penalizes errors linearly (rather than quadratically, in the case of a mean squared error loss function), which reduces the training effect of outlier data points which may become apparent due to volitional movements made by the sheep. The evaluation process occurred on a held-out dataset of mean EMG responses to 480 stimulation combinations, each with 4 repeats. The distributions were compared using a Kruskal–Wallis test. The random performance of each network was determined by generating a randomized prediction for each muscle from a uniform distribution ranging from 0 to the maximum response in the training dataset. Then, this randomized EMG vector was compared to the target vector. The L1 losses were split for each model by inclusion of the stimulating electrode in the training dataset. The L1 losses were compared within models using the Mann–Whitney U test. Finally, the L1 losses were split for each model by muscle, and their distributions were compared using a Kruskal–Wallis test. To evaluate the performance of the parameter inference neural network (the inverse model), the rAUC was computed for each of the responses to inferred parameters. The mean L1 error between the predicted and achieved EMG rAUC vectors was calculated across muscles. Additionally, the L1 error between the EMG responses to the original EES parameters and the responses to these same parameters sent during the evaluation session (which occurred some hours after the training session) was calculated. The distributions of L1 errors were compared using the Kruskal–Wallis test.

3. Results

3.1. Successful chronic implantation of the HD64

To evaluate the safety of chronic implantation of the HD64 devices, we monitored the condition of the sheep continuously throughout the study and evaluated our ability to control the multiplexer on the active array. We implanted two sheep with 1 (S1) or 2 (S2) HD64 EES paddles which remained implanted for 15 months. The multiplexers were configured 1598 and 4562 times on 34 and 51 unique days, respectively, throughout the study. No device-related malfunctions were observed, and control of the multiplexer remained stable. The impedance from the distal connector end of the lead-tails to the contacts (including the paths through the multiplexer) were measured at several points up to 274 d post implant. Over this time, no significant changes in the distribution of impedances were observed (p > 0.05, Mann–Whitney U test) (figure 2(d)). While the HD64 was implanted in the animals, we observed consistent stimulation to EMG responses and spinal responses to spinal and peripheral stimulation.

3.2. Active multiplexing enables precise motor recruitment

After implantation, we evaluated the benefit of a high-density electrode paddle in generating targeted motor recruitment from EES by recording motor-evoked responses from 6 to 8 bilateral lower extremity EMG sensors. Following EES, the EMG signal was rectified, then the rAUC was calculated and plotted as a function of amplitude (figure 3(a)). Applying monopolar EES to each of the 60 contacts in S1 produced a detectable motor response in at least 1 of 6 muscles. For each electrode contact, the EES amplitude required to reach 33% of the maximum activation for each muscle was identified (figure 3(b)). Contacts marked N.R. could not recruit the muscle to 33% of its maximum activation. Our analysis identified a lower EES amplitude required to recruit ipsilateral muscles than contralateral muscles, consistent with previous literature (Calvert et al 2019). The results for S2 are presented in supplemental figure 2.

Figure 3.

Figure 3.

Clustering analysis of motor outputs evoked by monopolar stimulation. (a) A graphical overview of the electromyography (EMG) preprocessing. A representative EMG trace (black) is rectified (blue trace), then the rectified area under the curve is calculated (blue shaded region). A sample recruitment curve is shown for six muscles. The points are the mean, and the bars show the standard deviation. A blue dashed line indicates the 33% of maximum activation threshold used for panel b. (b) The minimum required epidural electrical stimulation (EES) amplitude required to recruit each of the six instrumented muscles to 33% of their maximum activation. Electrodes marked ‘N.R.’ could not recruit the muscle to 33% of its maximum activation at the amplitude range tested. (c) A 2D projection of the embedded recruitment data. Each point is a stimulation event, and the points are colored by which cluster the stimulating electrode resides in. (d) A diagram of the HD64, with the electrode contacts colored by which cluster the electrode resides in. A scale drawing of the Medtronic 5-6-5 is overlaid to highlight which cluster(s) each of the 5-6-5’s electrodes contact. (e) Representative radar plots (EES frequency = 50 Hz) show the mean responses for each stimulation contact cluster. The clusters exhibit diverse recruitment patterns and relationships with amplitude.

The spatial diversity of EES-evoked motor responses was examined by performing unsupervised clustering. UMAP dimensionality reduction was performed on the recruitment data for each of the 60 electrodes. The low-dimensional embeddings were then clustered by statistical similarity. Eight independent clusters were identified. Electrodes for each cluster exhibited colocalization (figure 3(c), supplemental figure 3) on the HD64 paddle. To simulate the performance of a commercially available EES paddle (5-6-5, Medtronic, Minneapolis, MN), a scale drawing of the commercial paddle was overlaid on the HD64 clusters. While some commercial paddle contacts overlaid a single cluster, many contacts straddled two or more independent clusters, and cluster 5 was not reachable (figure 3(d)). Consequently, the motor responses evoked using the commercial paddle may lack the fidelity available using the HD64. To explore the diversity of the recruitment patterns across clusters, the mean recruitment for each electrode in each cluster was calculated and then normalized using the Yeo–Johnson power transformation (Yeo and Johnson 2000). Figure 3(e) provides an example of the recruitment patterns observed for each cluster at a stimulation frequency of 50 Hz.

To further explore the stimulation fidelity afforded using a high-density electrode paddle compared to a commercially available paddle, we examined the effect of current steering. Our stimulating analog front-ends were equipped with a single current source, enabling us to create a single bipole. When no anodic contact was specified, current returned through the ground wire implanted in the paraspinal muscles. This creates a much more dispersed electric field than if the current return was more proximal to the cathodic contact. By using an anodic stimulation contact, current is simultaneously sourced and sinked from the cathodic and anodic contacts, respectively. Thus, the electric field is ‘steered’ towards the anodic contact, changing the intensity and dispersion of the electric field. EES was delivered using bipolar stimulation pairs spaced using the minimum inter-electrode spacing available on the HD64 (‘narrow’, supp. figures 4(a) and (g)), or the minimum spacing available on the 5-6-5 (‘wide’, supp. figures 4(d) and (g)), and was compared to each other, or a monopolar baseline (supp. figures 4(a) and (d)). For each lower extremity muscle, the rAUC and selectivity of the response to stimulation was calculated, as described previously. Using a Mann–Whitney U test, a significant difference was identified between the narrow and wide bipolar stimulations for all stimulation amplitudes (p < 0.05), though never in all muscles, indicative of a difference in the selectivity of activation (supp. figure 4(h)). Comparing the selectivity of the narrow and wide bipolar fields, a significant difference was identified in at least one muscle for all stimulation amplitudes. At the maximum EES amplitude tested (1500 μA) 66% of instrumented muscles demonstrated significantly different response selectivity (supp. figure 4(i)). Other bipolar arrangements are presented in supplemental figure 5.

3.3. Active multiplexing enables localized referencing

To evaluate the benefits of dense recording bipoles, we began examining epidural spinal responses to spinal and peripheral stimulation. The experimental configuration for this section is shown in figure 4(a). To examine the propagation of ECAPs on the spinal cord using the HD64, a EES bipole was created on the caudal midline of the paddle. Four bipolar recording pairs were established along the midline of the paddle, with centers distanced 7.3 mm, 17.0 mm, 26.7 mm, and 31.6 mm rostrally from the center of the stimulating bipole respectively (figure 4(b)). Single cathode-leading, biphasic, symmetrical EES pulses were delivered with amplitudes 2 mA, 4 mA, and 6 mA. The pulse width remained constant at 25 μs. Note that this pulse width is shorter than the stimulation waveform used to study EES-evoked motor responses. This short pulse width and symmetrical aspect ratio was selected to minimize the effect of the stimulation artifact in the temporal window of interest immediately following stimulation. Decreasing the pulse width (and aspect ratio) mandated correspondingly increasing the stimulation amplitude range used. The responses recorded by each bipole are presented in figure 4(c). The elapsed time between stimulation and the peak of the ECAP was identified at each bipole for each of the 50 trials (figure 4(d)). Using the distance between the stimulating and recording bipoles and the time of peak response, the CV of the evoked response was calculated. For stimulation amplitudes 2 mA, 4 mA, and 6 mA, the conduction velocities were 114.9 m s−1, 114.6 m s−1, and 111.8 m s−1, respectively. These conduction velocities are within the range identified in previous work (Capogrosso et al 2013, Parker et al 2020). Additionally, the conduction velocities presented less than 2.5% deviation as stimulation amplitude increased, consistent with activating similar distributions of neural fibers. The presence of such an orthodromically propagating response is a fundamental demonstration of the HD64’s ability to faithfully convey potentials of neurogenic origin.

Figure 4.

Figure 4.

Assessing spinal responses detected by monopolar, wide bipolar, and narrow bipolar recordings. (a) A graphical overview of the spinal local field potential (LFP) recording, epidural electrical stimulation (EES) and transcutaneous electrical nerve stimulation (TENS) setup. (b) A map of the HD64, showing the stimulation cathode (red) and anode (blue). The recording bipoles and their distances from stimulation are shown. (c) The mean and standard deviation of the spinal evoked compound action potential (ECAP) response recorded using various bipolar pairs along the midline of the spinal cord, for three stimulation amplitudes. The peak of the 4 mA stimulation response is shown with a black arrow. Note that peak latency increases with separation from the stimulation bipole. (d) The peak latencies of 50 single trials, plotted as a function of separation from the stimulating bipole, for each stimulation amplitude. The samples are colored by their recording bipole. A linear fit is plotted in blue, with the conduction velocity (CV) inset.

We examined the SEPs in response to peripheral application of TENS using three recording strategies, each containing six recording channels (supp. figure 6(a)). The monopolar recording strategy used six consecutive HD64 contacts, with a recording reference wire implanted in the epidural space. The bipolar HD64 arrangement created six consecutive bipolar recording pairs using the minimum inter-electrode spacing available on the HD64. Finally, the bipolar 5-6-5 arrangement simulated the best-case bipolar arrangement using a commercial spinal paddle by creating six bipolar recording pairs with an interelectrode distance similar to that found on the Medtronic 5-6-5 (a to-scale drawing of this paddle is presented in blue in supp. figure 6(a)). Supplementary figure 6(b) shows the trial average (50 trials) SEP in response to 25 V TENS applied to the right fetlock for all three arrangements. While the SEP is clearly prominent in the monopolar arrangement, the responses recorded by all channels are highly correlated. To identify the degree of correlation in recorded SEPs between channels, the mean pairwise zero-lag cross-correlation coefficient was calculated for each channel on single trials. The effect of arbitrary bipole orientation was removed by taking the absolute value of the correlation coefficient. A statistically significant difference between the distribution of correlations was identified for all recording arrangements (p < 0.05, Kruskal–Wallis test) (supp. figure 6(c)). Spectral comparisons are made in supplemental figure 7.

3.4. Continuous encoding of electrode position enables inference over electrode space

Increasing the number of stimulation contacts increases the solution space of automated parameter inference neural networks. We leveraged the dense grid of contacts on the HD64 to enable spatial encoding of EES parameters during training and inference. The reparameterized neural network models were successfully trained to infer evoked EMG power following EES for all three electrode densities (figure 5(a)). The cumulative L1 loss between the predicted and actual EMG power for the held-out dataset is shown at the top of figure 5(b) for each instance of the ‘forward’ EMG prediction model. Significant differences were found between the 25% model and all other models (p < 0.05, Kruskal–Wallis test), but not between the 50% and 100% models (p > 0.05, Kruskal–Wallis test). Examining the L1 prediction error for each model, a significantly higher error was produced when predicting responses to stimulation on unseen electrodes for the 25% model (p < 0.05, Mann–Whitney U test). This was not present for the 50% model, indicating the 50% model’s ability to accurately infer EMG responses for unseen stimulation sites (p > 0.05, Mann–Whitney U test) (figure 5(b) bottom). In all cases, the prediction error was less than random chance (indicated by blue dots in figure 5(b)). Decomposing the L1 loss by muscle reveals the same result identified previously, where the only significant difference in L1 distribution occurs between the 25% model and the remaining models (p < 0.05, Kruskal–Wallis test), except for the left gastrocnemius, where prediction loss followed the same distribution for all models (p > 0.05, Kruskal–Wallis test) (supp. figure 8).

Figure 5.

Figure 5.

Evaluating the performance of the forward neural network model to predict sensorimotor computations. In all boxplots the horizontal bar is the median, the boxes extend from the 25th to the 75th percentile, and the blue dots represent a ‘random’ baseline performance. Statistical significance was calculated using a Kruskal–Wallis non-parametric test, with a maximum p-value of 0.05. (a) A map of the electrode contacts included in the training dataset for each model. Included electrodes are shaded blue. (b) (top) The distribution of L1 errors between the ground truth and simulated electromyography (EMG) for a held out dataset of 480 stimulation parameter combinations for the three models. (bottom) The data in (top) split by inclusion of the stimulating electrode in the training dataset. (c) An example target normalized electromyography (EMG) vector is shown at the top. The posterior densities over are shown for 3 models. Higher likelihood regions are indicated in lighter colors. (d) The in vivo evaluation procedure. The posterior densities of each model are sampled to produce proposed stimulation parameter combinations for each target. These parameters are applied in vivo, and the evoked EMG vector is calculated. The L1-error between the target EMG vector and the evoked EMG vector is then computed. (e) The distribution of mean L1 errors between the target EMG vector and the evoked EMG vector for each of the three models, and between EMG vectors evoked using identical stimuli in the training and evaluation sessions (‘ground truth’, or GT).

Following successful amortization of spinal sensorimotor computations by the forward models, we performed inference using the inverse models to identify EES parameters predicted to evoke target EMG responses. The inverse models produced joint posterior density functions over stimulation parameters and electrode spaces for 32 target EMG vectors. Figure 5(c) compares the posterior densities across the three inverse models for a single EMG target. Posteriors for additional EMG targets are provided in supplemental figure 9. Here, the 25% model predicts low likelihood across all electrode- and parameter-space. As the number of included electrode positions increases, the model localizes an optimal stimulation location, indicated by the focal region of high likelihood. To evaluate the performance of the models, we selected target EMG vectors from the held-out dataset for evaluation in vivo. The workflow and representative results are outlined in figure 5(d). The stimulation parameter and electrode combination that produced the highest likelihood is identified and then delivered using the stimulator device. The evoked EMG responses from the six instrumented muscles are recorded and then compared offline to the target EMG vector. This process was repeated for all three models. We then compared the L1 error between the target EMG response and the evoked EMG response across models (figure 5(e)). A nonparametric Kruskal–Wallis test revealed that the underlying distributions of L1 errors between targets and the responses were not significantly different between models (p > 0.05). This indicates that adequate training data to infer stimulation parameters to faithfully reproduce target EMG activity could be collected using a sparse subset (25%) of the available stimulation contacts. Further, no significant differences were observed between the re-sent stimulation parameter (the ‘ground truth’) distribution and the proposal distributions. This indicates that the parameter inference performance approached the soft upper bound given by the natural variations in EMG recorded across time.

4. Discussion

In this work, we presented a novel, high-density, active electronic EES paddle, the HD64. We evaluated the hermeticity of the active electronics assembly, and found leak rates 15 times smaller than the threshold defined by MIL-STD-883K. Following ISO 10993-1:2018 biocompatibility testing, we confidently progressed to in vivo testing. Using the onboard electronics, we quantified the improvements in the diversity and selectivity of motor responses evoked using the HD64 compared to wider bipolar configurations. Further, we demonstrated the capabilities of the HD64 to provide superior referencing of somatosensory evoked potentials following peripheral nerve stimulation, significantly reducing the mean spatial correlation compared to bipoles with similar spacing to commercial paddles. Finally, we extended the current state-of-the-art stimulation parameter inference model to encode sensorimotor-EES relations, enabling increased spatial resolution of inferred EES parameters and utilizing 2.5x higher channel count compared to prior preclinical work and clinical standards.

The HD64 meets all 4 criteria we established for a next-generation EES paddle. The rostrocaudal length of the treatment area enables delivery of stimulation to 2 vertebral segments per array. The length of the HD64 is within the bounds of existing paddle arrays, which are used commercially to target multiple spinal segments for pain modulation. The mediolateral span is sufficiently large to target lateral structures, such as the dorsal roots. This wider span is facilitated by the tapered cross section of the HD64, enabling the edges of the device to flex as needed in the epidural potential space. We have demonstrated localized current steering, by establishing cathodic and anodic currents on neighboring electrodes, and leveraged increased recording resolution to uncover spatial differences in SEPs. Finally, the reprogrammability enabled by the onboard multiplexer has been used to reposition the stimulating contact in response to optimal stimulation suggestions made by neural network models.

When considering EES-enabled neurorehabilitation, the HD64 presents a rostrocaudal span comparable to that of commercial pain modulation arrays. This span limits the number of motor pools available for targeting during spatiotemporal motor-restorative EES, compared to larger clinical electrode arrays (e.g. Medtronic 5-6-5, Boston Scientific CoverEdge X). However, a stacked approach utilizing 2 HD64 arrays enables the benefits afforded by high channel counts to be maintained while targeting the necessary span of motor pools and possessing a clinically-relevant number of lead tails (for example, 2 HD64s present a longer rostrocaudal span, approximately 4 times the available stimulation contacts, and increased contact density compared to the Boston Scientific CoverEdgeX, while maintaining the same number of lead tails). An alternative approach to address an increased rostrocaudal span would be to increase the number of contacts as the size of the array increases. This approach would require 1 surgical entrance to the epidural potential space, compared to 2 in the stacked approach. However, using a long, contiguous paddle constrains the rostrocaudal span to the length of the electrode, which may not be appropriate for all patient anatomies. In the stacked approach, the rostrocaudal separation between the two paddles can be chosen to ensure optimal placement of HD64 contacts over motor pools (and space between paddles in areas without motor pools). Thus, using the stacked 2-paddle approach shown in S2, future EES for neurorehabilitation research leveraging the HD64 can access the required span of motor pools using highly selective stimulation.

Looking forward, our work here is a fundamental demonstration of active electronics embedded into neural interface electrodes themselves, and the resultant improvements in stimulation targeting and recording resolution. While our work leverages multiplexing to achieve a high channel count, other architectures may use the same hermetic approach to integrate digital signal processing, neural network inference, or stimulation control directly into electrodes. We expect others to continue the path towards highly functional integrated neural interfaces to restore lost function following trauma or disease.

4.1. Active electronics can be chronically hermetically sealed in an EES paddle

Our novel paddle contains an active multiplexing chip hermetically sealed on the body of the EES paddle, enabling communication with 60 bidirectional contacts through two 12-contact percutaneous lead tails. The increased contact density combinatoriality increases the number of possible stimulation configurations available for research and therapeutic applications. In contrast to this reconfigurability, current clinically available EES paddles sacrifice spatial resolution in favor of a general, ‘one-size-fits-most’ approach. In research studies, prior work has suggested a library of paddles of different sizes to suit person-to-person variability in anatomy (Rowald et al 2022). Instead, the reconfigurability of the HD64 allows for localized concentration of stimulating electrodes to activate a specific area of the spinal cord, enabling focal stimulation with current steering to prevent off-target effects without sacrificing the ability to simultaneously stimulate other distal regions.

As the first example of an EES paddle including active electronics, we were particularly interested in assessing the ability of the encapsulant to maintain patency during chronic implantation. We found that the HD64 remained functional for at least 15 months after implantation and did not display symptoms of fluid ingress. Communication between the externalized stimulation and recording hardware and the onboard multiplexer remained robust throughout the study. This initial result is promising for the future of epidural medical devices containing active electronics. With current technologies, chips of similar dimensions to the present device are capable of significant computational power. Our results may open the door for diverse advances in on-paddle processing, including signal processing or closed-loop stimulation parameter control, as has been hypothesized in prior literature (Zhang et al 2022). While additional testing and regulatory approval is required before such technologies can be translated into clinical applications, this demonstration represents a promising first step.

4.2. Onboard active multiplexing enables chronic stimulation-evoked responses

Our study benefits from decades of prior research showing the selective activation of motor neuron pools using EES applied to the lumbosacral spinal cord. Specifically, EES has been utilized heavily to restore walking ability following a spinal cord injury (Capogrosso et al 2016, Angeli et al 2018, Gill et al 2018). Other studies have used EES to establish somatotopic maps of stimulation location to recruit lower limb extremities (Hofstoetter et al 2021). Extending the mediolateral span through which EES can be delivered, while maintaining a high contact-density, enabled us to specifically target lateral structures of the spinal cord. Stimulation at these lateral sites generated stereotyped EMG responses statistically different from those achieved by medial stimulation, as determined by our spectral clustering analysis. These differences highlight the importance of lateral stimulation sites for enhancing the capability of locomotor-restorative EES. Such sites are not present on stimulation paddle arrays designed for chronic pain management.

The statistically similar clusters of electrodes identified in figure 3(d) are not as clearly demarcated as hypothesized. Examining the recruitment curves in figure 3(e) provides some additional clarity. Clusters 0 and 1 evoke strong, bilateral activation, consistent with midline stimulation. Cluster 2 evokes a strong, right-side biased response, while cluster 3 evokes a weaker, left-side biased response. Cluster 5 evokes a weak, right-side specific response, consistent with right-side lateral stimulation, with the remaining clusters evoking weak, non-specific responses. Overall, contacts on the caudal 5 rows of contacts evoked the strongest responses (as seen in figure 3(b)). From prior literature, the stronger EMG activity in this region indicates the approximate location of a dorsal root entry zone (Hofstoetter et al 2021). In this region, the mediolateral organization identified in prior literature is maintained (Capogrosso et al 2016, Rowald et al 2022, Calvert et al 2023, Lorach et al 2023). In S1, IM EMG recordings were taken, however S2 used surface EMG. Due to the proximity of instrumented muscles on the sheep leg, and the lower resolution of surface recordings, we did not observe selectivity between proximal and distal muscles as the stimulation site was moved rostrocaudally. This is reflected in supplementary figure 3. Instead, we observe the mediolateral striation described in prior literature (Capogrosso et al 2016, Rowald et al 2022, Calvert et al 2023, Lorach et al 2023).

Recording EES-evoked spinal responses, or ECAPs, can provide quantitative data regarding EES paddle status. ECAPs have been used as a control signal for pain-modulating EES (Mekhail et al 2020), and we have previously suggested their utility in the identification of neural anatomy (Calvert et al 2023). Using bipolar recording configurations on the HD64, we observed an orthodromically propagating response to EES. The conduction velocities of these responses are consistent with the published ranges for Type Ia and Ib axons, which have been shown to be recruited by EES in computational models (Capogrosso et al 2013, Parker et al 2020). Further, our work demonstrates the formation of dense recording bipoles on the ovine spinal cord, and uses these recording bipoles to examine EES- and TENS-evoked spinal responses. This is timely for the field of neural interfaces, with the recent advent of high-density neural probes (Jun et al 2017, Steinmetz et al 2021, Chamanzar et al 2023) and electrocorticography grids (Rachinskiy et al 2022, Palopoli-Trojani et al 2024). Using conventional EES paddles, contacts are sparsely distributed to ensure coverage of a large area using a small number of electrodes. This sets a high minimum limit on contact separation.

When considering EES or sensing, we suggest that the active multiplexing on the HD64 enables dense bipoles to be created at arbitrary points of interest. Importantly, contact separation can be non-uniform (for example, tight spacing in regions of high spatial variability and wider spacing in less volatile regions) or change with time or desired function. These ‘activity-dependent’ paddle arrangements allow for the capabilities of the stimulation system to be adjusted to suit the needs of patients when completing various activities of daily living (for example, greater trunk control during sitting vs lower extremity motor control during locomotion in patients with motor dysfunction).

4.3. Spatial encoding of EES enables parameter inference over arbitrary channels

Stimulation parameter inference remains an obstacle to widespread clinical implementation of EES (Solinsky et al 2020). Our previous parameter selection automation work (Govindarajan et al 2022), and other subsequent literature (Bonizzato et al 2023) treat stimulating electrodes as discrete variables. As such, spatial relationships (such as those uncovered in figure 3(d)) cannot be leveraged by the model. In this work, we instead evaluated a novel electrode encoding scheme that accounts for spatial relationships between electrode contacts. This approach is extensible to very high channel counts and allows the network to identify optimal stimulation locations not included in the training dataset. Using the results shown here, active electronics conducting neural network inference may be hermetically sealed into the paddle array body, enabling stimulation parameter inference to be conducted without additional hardware or communication requirements.

As the number of available monopolar stimulation channels increases, the time required to collect training data from every contact scales linearly. Unaddressed, this is a barrier to clinical translation of parameter selection automation techniques and high-density EES paddles. Here, our work shows that exhaustive sampling of every stimulation contact is not required if the electrode encoding scheme is continuous. No significant decrease in EMG response prediction performance is observed for electrode inclusion rates as low as 50%, and stimulation parameter inference performance did not significantly vary across inclusion rates tested. The development of efficient machine learning tools, such as the one presented in this manuscript, greatly reduces the amount of training data required. Increasing efficiency and decreasing data collection efforts will aid in scaling EES for clinical use, which could yield meaningful benefits during spinal rehabilitation in patients with neural dysfunction.

4.4. Study limitations and implications for future research

Our study introduces a high-density smart EES paddle with an integrated multiplexer and demonstrates the utility of increased electrode density. However, several limitations should be considered. While the characterization presented in this manuscript empirically supports the use of high-density electrode paddles in these applications, a more thorough assessment may be possible with the addition of computational models. Previous advances in electrode design were informed by finite element models of 15 healthy volunteers (Rowald et al 2022). Such models could quantify recruitment of off-target motor neuron pools after applying stimulation at bipoles of various sizes, however the work presented here utilizes an ovine model. The spinal cord of the sheep is similar to that of humans in terms of gross anatomy and size (critically, the transverse width of the epidural space in the lumbar region is 17.5 ± 1.2 mm (Wilke et al 1997) compared to 17.87 ± 1.47 mm in humans (Lee et al 2022)) and has been used as a model for the study of spinal electrophysiology (Parker et al 2013, 2020, Chakravarthy et al 2020, Calvert et al 2023), but the results of this manuscript may not be generalizable to human anatomy. The mediolateral span of the HD64 is approximately 4 mm wider than commercial EES paddles traditionally used for the treatment of neuropathic pain, however the HD64 presents a tapered profile. The HD64 maintains its full 2 mm thickness for a narrower mediolateral span than commercial SCS leads. Examining prior studies of spinal cord and spinal canal dimensions, the transverse diameter of the spinal cord in the T11 to L1 vertebral regions is between 8 and 9.6 mm (Ko et al 2004, Fradet et al 2014). In the same region, the spinal canal sagittal depth is between 15.4 and 19.54 mm, while the canal transverse width is 16.7–26.5 mm (Laporte et al 2000, Busscher et al 2010). Further, the spinal cord exhibits positive anteroposterior eccentricity in this region (that is, the centroid of the spinal cord is more anterior than the centroid of the spinal canal (Fradet et al 2014)). A scale diagram showing the HD64 implanted in this region is depicted in supplementary figure 10. The previously published dimensions and tapered profile do not indicate substantial difficulty avoiding neural structures during implantation of the HD64 in the examined vertebral levels. Future studies are required to evaluate the therapeutic potential of high-density EES electrodes or to quantify the effect of stimulation focality on behavioral values such as sensation. One current barrier to clinical translation of the HD64 is the lack of existing implantable pulse generators (IPGs) capable of generating the power and communication signals necessary to control the onboard multiplexer. Additionally, the inclusion of active electronics in implanted electrodes increases the power draw of the implantable system, and must be considered when selecting an appropriate IPG battery size. An example rechargeable IPG contains a 200 mAh battery (Saluda Medical 2022), which is sufficient to deliver therapy for between 24 and 168 h (The Advanced Spine Center 2022). This equates to a typical current draw during therapy of between 1.19 and 8.3 mA. The HD64 consumes only 20 μA in steady-state operation and 100 μA when reconfiguring connections, representing a worst-case increase in current draw of less than 5%. Therefore, we predict the impacts on recharge interval or IPG battery capacity to be minimal. While the hermeticity evaluations performed in this manuscript exceeded our requirements for a 15 month implant duration, additional evaluation of the package may be required for more chronic implants. Our neural network models demonstrate sufficient accuracy to match EMG patterns, however, these experiments were conducted with the sheep at rest and elevated in a sling. A more functional task may instead consider applying a stimulation sequence to match a target kinematic trajectory. Additionally, our neural network parameter inference is limited to monopolar stimulation patterns and, therefore, cannot take advantage of the improved selectivity identified by bipolar stimulation in this manuscript. Bipolar stimulation dramatically increases the number of electrode configurations the network must evaluate; the additional complexity of which is a promising target for future research.

5. Conclusions

In summary, we successfully designed and chronically implanted a high-density smart EES paddle, the HD64, which is the first to integrate active electronics into a hermetic package on the spinal cord. Our hermetic assembly was tested to exceed standards for active hermetic electronics. During chronic in vivo implantation, no device-related malfunctions, nor symptoms of moisture ingress were observed, enabling demonstration of fundamental stimulation-response characteristics in the ovine spinal cord. These results highlight the translational potential of high-density EES paddles, and provide a foundation for more advanced computation and processing to be integrated directly into neural interfaces.

Acknowledgment

We would like to thank Aaron Gregoire and the Center for Animal Resources and Education (CARE) staff at Brown University for their assistance with animal care, as well as Alison Gorbatov and Krista Colling from the Center for Neurorestoration and Neurotechnology (CfNN) for their assistance with the Providence VA Medical System regulatory work. We would also like to thank John Murphy for his engineering expertise in designing modifications to the sling apparatus. We would like to thank Moses Goddard for his surgical time and expertise in implanting the EMG system. Finally, we would like to thank the reviewers of this manuscript for their careful reading and valuable enhancements. Figures 2(a), 4(a), 5(d), supp. 3(a) and supp. 10(b) were created with BioRender.com.

Data availability statement

The data that support the findings of this study are openly available at the following URL/DOI: https://doi.org/10.5281/zenodo.14713531.

Financial support

This work was sponsored in part by the Defense Advanced Research Projects Agency (DARPA) BTO under the auspices of Drs. Alfred Emondi, Jean-Paul Chretien, and Pedro Irazoqui (through the Space and Naval Warfare Systems Center, Pacific DARPA Contracts Management Office) Grant/Contract Numbers D15AP00112 and D19AC00015 to DAB, and by the Merit Review Award #I01RX002835 from the United States Department of Veterans Affairs (VA), Rehabilitation Research and Development Service to DAB. This work was supported in part by the Center for Neurorestoration and Neurotechnology (N2864-C) from the United States (U.S.) Department of Veterans Affairs, Rehabilitation research and Development Service, Providence, RI. The contents of this manuscript do not represent the views of the U.S. Department of Veterans Affairs or the United States Government. Computing resources were provided by the Brown University Center for Computation and Visualization (CCV) using the Carney Condo, supported by National Institute of Health (NIH) Office of the Director grant S10OD025181 (Jerome Sanes). SRP was supported by the 2020 Susan and Isaac Wakil Foundation John Monash Scholarship. JSC was supported by the National Institute for Neurological Disorders and Stroke (NINDS) T32 Postdoctoral Training Program in Recovery and Restoration of Central Nervous System Health and Function (5T32NS100663-04) under the guidance of DAB. JJ was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (RS-2023-00246822). LNG was supported by the 2021-2022 Carney Institute for Brain Science Graduate Fellowship under the guidance of TS. BM, YI, KA, and GC were supported in part by the National Institute for Neurological Disorders and Stroke (NINDS) grant 3U44NS115111.

Authorship statement

SRP: Conceptualization, Methodology, Software, Formal analysis, Investigation, Data Curation, Writing—original draft, Visualization. JSC: Conceptualization, Methodology, Software, Formal analysis, Investigation, Data Curation, Writing—review & editing. RD: Conceptualization, Methodology, Software, Formal analysis, Investigation, Data Curation, Writing—review & editing. JJ: Conceptualization, Methodology, Software, Investigation, Writing—review & editing. LNG: Conceptualization, Methodology, Software, Investigation, Writing—review & editing. KA: Conceptualization, Methodology, Validation, Writing—review & editing. GC: Conceptualization, Methodology, Validation, Writing—review & editing. YI: Conceptualization, Methodology, Validation, Supervision, Writing—review & editing. ES: Conceptualization, Investigation, Writing—review & editing. JSF: Conceptualization, Investigation, Supervision, Funding acquisition, Writing—review & editing. TS: Conceptualization, Supervision, Funding acquisition, Writing—review & editing. DAB: Conceptualization, Methodology, Investigation, Resources, Writing—original draft, Writing—review & editing, Supervision, Project administration, Funding acquisition. BM: Conceptualization, Methodology, Investigation, Resources, Writing—original draft, Writing—review & editing, Supervision, Project administration, Funding acquisition.

DAB and BM contributed equally to this manuscript. Final approval was given by all authors.

Conflict of interest

SRP, JSC, RD, and DAB have patents pending regarding the recording of spinal electrophysiological signals during spinal cord stimulation (PCT-US2022-034450: ‘A novel method to modulate nervous system activation based on one or more spinal field potentials’). BM, KA, GC have patents and pending patents related to high-resolution spinal cord stimulator devices (US-11 116 964-B2: ‘Multi-electrode array with unitary body’, 11 027 122: ‘Spinal cord stimulation method to treat lateral neural tissues’, US-11 395 924-B2: ‘Implantable devices with welded multi-contact electrodes and continuous conductive elements’). Micro-Leads Medical is a commercial company developing precision neuromodulation therapeutic devices and BM, KA, GC, and YI are shareholders or have stock options in Micro-Leads.

References

  1. Angeli C A, Boakye M, Morton R A, Vogt J, Benton K, Chen Y, Ferreira C K, Harkema S J. Recovery of over-ground walking after chronic motor complete spinal cord injury. New Engl. J. Med. 2018;379:1244–50. doi: 10.1056/NEJMoa1803588. [DOI] [PubMed] [Google Scholar]
  2. Badi M, et al. Intrafascicular peripheral nerve stimulation produces fine functional hand movements in primates. Sci. Trans. Med. 2021;13:eabg6463. doi: 10.1126/scitranslmed.abg6463. [DOI] [PubMed] [Google Scholar]
  3. Bonizzato M, et al. Autonomous optimization of neuroprosthetic stimulation parameters that drive the motor cortex and spinal cord outputs in rats and monkeys. Cell Rep. Med. 2023;4:101008. doi: 10.1016/j.xcrm.2023.101008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Boston Scientific Surgical leads directions for use. 2020. (available at: www.bostonscientific.com/content/dam/elabeling/nm/92395562-01_B_Surgical_Leads_DFU_en-US_s.pdf)
  5. Bryson N, Lombardi L, Hawthorn R, Fei J, Keesey R, Peiffer J D, Seáñez I. Enhanced selectivity of transcutaneous spinal cord stimulation by multielectrode configuration. J. Neural Eng. 2023;20:046015. doi: 10.1088/1741-2552/ace552. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Burke J F, et al. Epidural spinal electrogram provides direct spinal recordings in awake human participants. Front. Hum. Neurosci. 2021;15:721076. doi: 10.3389/fnhum.2021.721076. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Busscher I, Ploegmakers J J W, Verkerke G J, Veldhuizen A G. Comparative anatomical dimensions of the complete human and porcine spine. Eur. Spine J. 2010;19:1104–14. doi: 10.1007/s00586-010-1326-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Calvert J S, Darie R, Parker S R, Shaaya E, Syed S, McLaughlin B L, Fridley J S, Borton D A. Spatiotemporal distribution of electrically evoked spinal compound action potentials during spinal cord stimulation. Neuromodulation. 2023;26:961–74. doi: 10.1016/j.neurom.2022.03.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Calvert J S, Manson G A, Grahn P J, Sayenko D G. Preferential activation of spinal sensorimotor networks via lateralized transcutaneous spinal stimulation in neurologically intact humans. J. Neurophysiol. 2019;122:2111–8. doi: 10.1152/jn.00454.2019. [DOI] [PubMed] [Google Scholar]
  10. Capogrosso M, et al. A brain-spine interface alleviating gait deficits after spinal cord injury in primates. Nature. 2016;539:284–8. doi: 10.1038/nature20118. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Capogrosso M, Wenger N, Raspopovic S, Musienko P, Beauparlant J, Bassi Luciani L, Courtine G, Micera S. A computational model for epidural electrical stimulation of spinal sensorimotor circuits. J. Neurosci. 2013;33:19326–40. doi: 10.1523/JNEUROSCI.1688-13.2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Chakravarthy K, Bink H, Dinsmoor D. Sensing evoked compound action potentials from the spinal cord: novel preclinical and clinical considerations for the pain management researcher and clinician. J. Pain Res. 2020;13:3269–79. doi: 10.2147/JPR.S289098. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Chamanzar M, et al. Fully flexible high-density implantable neural probes for electrophysiology recording and controlled neurochemical modulation. Research Square. 2023 doi: 10.21203/rs.3.rs-2720944/v1. [DOI] [PMC free article] [PubMed]
  14. Chandrasekaran S, Nanivadekar A C, McKernan G, Helm E R, Boninger M L, Collinger J L, Gaunt R A, Fisher L E. Sensory restoration by epidural stimulation of the lateral spinal cord in upper-limb amputees. eLife. 2020;9:e54349. doi: 10.7554/eLife.54349. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. ÇiÇek S, Attar A, Tuna H, Keçik Y, Egemen N. Effects of different doses of epidural midazolam on spinal somatosensory evoked potentials. Acta Neurochir. 2000;142:921–7. doi: 10.1007/s007010070079. [DOI] [PubMed] [Google Scholar]
  16. Falowski S M. An observational case series of spinal cord stimulation waveforms visualized on intraoperative neuromonitoring. Neuromodulation. 2019;22:219–28. doi: 10.1111/ner.12781. [DOI] [PubMed] [Google Scholar]
  17. Fradet L, Arnoux P-J, Ranjeva J-P, Petit Y, Callot V. Morphometrics of the entire human spinal cord and spinal canal measured from in vivo high-resolution anatomical magnetic resonance imaging. Spine. 2014;39:E262–69. doi: 10.1097/brs.0000000000000125. [DOI] [PubMed] [Google Scholar]
  18. Gad P N, Roy R R, Zhong H, Lu D C, Gerasimenko Y P, Edgerton V R. Initiation of bladder voiding with epidural stimulation in paralyzed, step trained rats. PLoS One. 2014;9:e108184. doi: 10.1371/journal.pone.0108184. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Gill M L, et al. Neuromodulation of lumbosacral spinal networks enables independent stepping after complete paraplegia. Nat. Med. 2018;24:1677–82. doi: 10.1038/s41591-018-0175-7. [DOI] [PubMed] [Google Scholar]
  20. Gorman K. How to use UMAP—Umap 0.5 documentation. 2018. (available at: https://umap-learn.readthedocs.io/en/latest/basic_usage.html)
  21. Govindarajan L N, Calvert J S, Parker S R, Jung M, Darie R, Miranda P, Shaaya E, Borton D A, Serre T. Fast inference of spinal neuromodulation for motor control using amortized neural networks. J. Neural Eng. 2022;19:056037. doi: 10.1088/1741-2552/ac9646. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Greenhouse H, et al. Hermeticity of Electronic Packages. Elsevier; 2012. The flow of gases into sealed packages; pp. 63–106. [DOI] [Google Scholar]
  23. Hofstoetter U S, Perret I, Bayart A, Lackner P, Binder H, Freundl B, Minassian K. Spinal motor mapping by epidural stimulation of lumbosacral posterior roots in humans. iScience. 2021;24:101930. doi: 10.1016/j.isci.2020.101930. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Insola A, Padua L, Mazzone P, Valeriani M. Unmasking of presynaptic and postsynaptic high-frequency oscillations in epidural cervical somatosensory evoked potentials during voluntary movement. Clin. Neurophysiol. 2008;119:237–45. doi: 10.1016/j.clinph.2007.09.132. [DOI] [PubMed] [Google Scholar]
  25. Jun J J, et al. Fully integrated silicon probes for high-density recording of neural activity. Nature. 2017;551:232–6. doi: 10.1038/nature24636. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Ko H-Y, Park J H, Shin Y B, Baek S Y. Gross quantitative measurements of spinal cord segments in human. Spinal Cord. 2004;42:35–40. doi: 10.1038/sj.sc.3101538. [DOI] [PubMed] [Google Scholar]
  27. Krog L, et al. Cervical spinal cord stimulation: a review. Curr. Pain Headache Rep. 2023;28:239–49. doi: 10.1007/s11916-023-01200-8. [DOI] [PubMed] [Google Scholar]
  28. Lam D V, et al. The role of spinal cord neuroanatomy in the variances of epidural spinal recordings. Bioelectron. Med. 2024;10:1. doi: 10.1186/s42234-024-00149-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Laporte S, Mitton D, Ismael B, de Fouchecour M, Lasseau J P, Lavaste F, Skalli W. Quantitative morphometric study of thoracic spine a preliminary parameters statistical analysis. Eur. J. Orthop. Surg. Traumatol. 2000;10:85–91. doi: 10.1007/bf02803102. [DOI] [Google Scholar]
  30. Lee S Y, Kim I Y, Jeong K W, Ryu T, Kwak S K, Jung J Y. Comparison between the coronal diameters of the cervical spinal canal and spinal cord measured using computed tomography and magnetic resonance imaging in Korean Patients. Korean J. Anesthesiol. 2022;75:323–30. doi: 10.4097/kja.22006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Letechipia J E, Peckham P H, Gazdik M, Smith B. In-line lead connector for use with implanted neuroprosthesis. IEEE Trans. Bio-Med. Eng. 1991;38:707–9. doi: 10.1109/10.83572. [DOI] [PubMed] [Google Scholar]
  32. Lorach H, et al. Walking naturally after spinal cord injury using a brain-spine interface. Nature. 2023;618:126–33. doi: 10.1038/s41586-023-06094-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  33. Marcus M A, Bruyninckx F L, Vertommen J D, Wouters P F, Van Aken H. Spinal somatosensory evoked potentials after epidural isoproterenol in awake sheep. Can. J. Anaesth. J. Canad. D’anesthesie. 1997;44:85–89. doi: 10.1007/BF03014330. [DOI] [PubMed] [Google Scholar]
  34. McInnes L, Healy J, Saul N, Großberger L. UMAP: uniform manifold approximation and projection. J. Open Source Softw. 2018;3:861. doi: 10.21105/joss.00861. [DOI] [Google Scholar]
  35. Medtronic . UC202000365eEN. Medtronic; 2022. Product performance report: summary of data from the Medtronic post-market registry. (available at: www.medtronic.com/content/dam/medtronic-wide/public/united-states/products/neurological/product-performance-report-spinal-cord-stimulation-systems.pdf) [Google Scholar]
  36. Mekhail N, et al. Long-term safety and efficacy of closed-loop spinal cord stimulation to treat chronic back and leg pain (evoke): a double-blind, randomised, controlled trial. Lancet Neurol. 2020;19:123–34. doi: 10.1016/S1474-4422(19)30414-4. [DOI] [PubMed] [Google Scholar]
  37. Mishra L N, Kulkarni G, Gadgil M. A novel current steering method for targeted spinal cord stimulation. Front. Pain Res. 2023;4:1028368. doi: 10.3389/fpain.2023.1028368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Musienko P, Courtine G, Tibbs J E, Kilimnik V, Savochin A, Garfinkel A, Roy R R, Edgerton V R, Gerasimenko Y. Somatosensory control of balance during locomotion in decerebrated cat. J. Neurophysiol. 2012;107:2072–82. doi: 10.1152/jn.00730.2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. Nahm F S. From the torpedo fish to the spinal cord stimulator. Korean J. Pain. 2020;33:97–98. doi: 10.3344/kjp.2020.33.2.97. [DOI] [PMC free article] [PubMed] [Google Scholar]
  40. Nainzadeh N K, Neuwirth M G, Bernstein R, Cohen L S. Neurophysiology and Standards of Spinal Cord Monitoring. Springer; 1988. Direct recording of spinal evoked potentials to peripheral nerve stimulation by a specially modified electrode; pp. 234–44. [DOI] [Google Scholar]
  41. Nanivadekar A C, et al. Restoration of sensory feedback from the foot and reduction of phantom limb pain via closed-loop spinal cord stimulation. Nat. Biomed. Eng. 2023;8:992-1003. doi: 10.1038/s41551-023-01153-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Nijhuis H J A, Hofsté W-J, Krabbenbos I P, Dietz B E, Mugan D, Huygen F. First report on real-world outcomes with evoked compound action potential (ECAP)-controlled closed-loop spinal cord stimulation for treatment of chronic pain. Pain Ther. 2023;12:1221–33. doi: 10.1007/s40122-023-00540-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Palopoli-Trojani K, Trumpis M, Chiang C-H, Wang C, Williams A J, Evans C L, Turner D A, Viventi J, Hoffmann U. High-density cortical μECoG arrays concurrently track spreading depolarizations and long-term evolution of stroke in awake rats. Commun. Biol. 2024;7:263. doi: 10.1038/s42003-024-05932-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. Parker J L, Karantonis D M, Single P S, Obradovic M, Laird J, Gorman R B, Ladd L A, Cousins M J. Electrically evoked compound action potentials recorded from the sheep spinal cord. Neuromodulation. 2013;16:295–303. doi: 10.1111/ner.12053. [DOI] [PubMed] [Google Scholar]
  45. Parker J L, Laird-Wah J, Cousins M J. Comparison of a simple model of dorsal column axons with the electrically evoked compound action potential. Bioelectron. Med. 2018;1:117–30. doi: 10.2217/bem-2017-0006. [DOI] [Google Scholar]
  46. Parker J L, Obradovic M, Hesam Shariati N, Gorman R B, Karantonis D M, Single P S, Laird-Wah J, Bickerstaff M, Cousins M J. Evoked compound action potentials reveal spinal cord dorsal column neuroanatomy: ECAPs and spinal neuroanatomy. Neuromodulation. 2020;23:82–95. doi: 10.1111/ner.12968. [DOI] [PubMed] [Google Scholar]
  47. Pedregosa F, et al. Scikit-learn: machine learning in python. J. Mach. Learn. Res. 2011;12:2825–30. doi: 10.5555/1953048.2078195. [DOI] [Google Scholar]
  48. Rachinskiy I, Wong L, Chiang C-H, Wang C, Trumpis M, Ogren J I, Hu Z, McLaughlin B, Viventi J. High-density, actively multiplexed μECoG array on reinforced silicone substrate. Front. Nanotechnol. 2022;4:837328. doi: 10.3389/fnano.2022.837328. [DOI] [PMC free article] [PubMed] [Google Scholar]
  49. Reddy C G, Miller J, Abode-Iyamah K, Safayi S, Wilson S, Dalm B, Fredericks D, Gillies G, Howard III M A, Brennan T. Ovine model of neuropathic pain for assessing mechanisms of spinal cord stimulation therapy via dorsal horn recordings, von Frey filaments, and gait analysis. J. Pain Res. 2018;11:1147–62. doi: 10.2147/JPR.S139843. [DOI] [PMC free article] [PubMed] [Google Scholar]
  50. Rousseeuw P J. Silhouettes: a graphical aid to the interpretation and validation of cluster analysis. J. Comput. Appl. Math. 1987;20:53–65. doi: 10.1016/0377-0427(87)90125-7. [DOI] [Google Scholar]
  51. Rowald A, et al. Activity-dependent spinal cord neuromodulation rapidly restores trunk and leg motor functions after complete paralysis. Nat. Med. 2022;28:260–71. doi: 10.1038/s41591-021-01663-5. [DOI] [PubMed] [Google Scholar]
  52. Sala F, Squintani G, Tramontano V, Arcaro C, Faccioli F, Mazza C. Intraoperative neurophysiology in tethered cord surgery: techniques and results. Child’s Nerv. Syst. 2013;29:1611–24. doi: 10.1007/s00381-013-2188-3. [DOI] [PubMed] [Google Scholar]
  53. Saluda Medical . Saluda Medical; 2022. Evoke ® SCS system user manual CLIN-UMAN-005176. (available at: www.accessdata.fda.gov/cdrh_docs/pdf19/P190002C.pdf) [Google Scholar]
  54. Shi J, Malik J. Normalized cuts and image segmentation. IEEE Trans. Pattern Anal. Mach. Intell. 2000;22:888–905. doi: 10.1109/34.868688. [DOI] [Google Scholar]
  55. Shils J L, Arle J E. Neuromonitoring for spinal cord stimulation lead placement under general anesthesia. J. Clin. Neurol. 2018;14:444–53. doi: 10.3988/jcn.2018.14.4.444. [DOI] [PMC free article] [PubMed] [Google Scholar]
  56. Solinsky R, Specker-Sullivan L, Wexler A. Current barriers and ethical considerations for clinical implementation of epidural stimulation for functional improvement after spinal cord injury. J. Spinal Cord Med. 2020;43:653–6. doi: 10.1080/10790268.2019.1666240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  57. Steinmetz N A, et al. Neuropixels 2.0: a miniaturized high-density probe for stable, long-term brain recordings. Science. 2021;372:eabf4588. doi: 10.1126/science.abf4588. [DOI] [PMC free article] [PubMed] [Google Scholar]
  58. Swindale N V, Spacek M A. Verification of multichannel electrode array integrity by use of cross-channel correlations. J. Neurosci. Methods. 2016;263:95–102. doi: 10.1016/j.jneumeth.2016.02.009. [DOI] [PubMed] [Google Scholar]
  59. The Advanced Spine Center Life-changing spinal cord stimulation | advanced spine blog. 2022. (available at: https://theadvancedspinecenter.com/living-with-a-spinal-cord-stimulator/) (Accessed 6 December 2024)
  60. Tsirikos A I, Aderinto J, Tucker S K, Noordeen H H. Spinal cord monitoring using intraoperative somatosensory evoked potentials for spinal trauma. J. Spinal Disorders Tech. 2004;17:385–94. doi: 10.1097/01.bsd.0000095825.98982.1a. [DOI] [PubMed] [Google Scholar]
  61. Urasaki E, Wada S-I, Kadoya C, Tokimura T, Yokota A, Matsuoka S, Fukumura A, Hamada S. Skin and epidural recording of spinal somatosensory evoked potentials following median nerve stimulation: correlation between the absence of spinal N13 and impaired pain sense. J. Neurol. 1990;237:410–5. doi: 10.1007/BF00314731. [DOI] [PubMed] [Google Scholar]
  62. Verma N, et al. Characterization and applications of evoked responses during epidural electrical stimulation. Bioelectron. Med. 2023;9:5. doi: 10.1186/s42234-023-00106-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  63. Wilke H J, Kettler A, Wenger K H, Claes L E. Anatomy of the sheep spine and its comparison to the human spine. Anat. Rec. 1997;247:542–55. doi: 10.1002/(SICI)1097-0185(199704)247:4&#x0003c;542::AID-AR13&#x0003e;3.0.CO;2-P. [DOI] [PubMed] [Google Scholar]
  64. Wilson S, et al. An ovine model of spinal cord injury. J. Spinal Cord Med. 2017;40:346–60. doi: 10.1080/10790268.2016.1222475. [DOI] [PMC free article] [PubMed] [Google Scholar]
  65. Woodington B J, Lei J, Carnicer-Lombarte A, Güemes-González A, Naegele T E, Hilton S, El-Hadwe S, Trivedi R A, Malliaras G G, Barone D G. Flexible circumferential bioelectronics to enable 360-degree recording and stimulation of the spinal cord. Sci. Adv. 2024;10:eadl1230. doi: 10.1126/sciadv.adl1230. [DOI] [PMC free article] [PubMed] [Google Scholar]
  66. Yeo I-K, Johnson R A. A new family of power transformations to improve normality or symmetry. Biometrika. 2000;87:954–9. doi: 10.1093/biomet/87.4.954. [DOI] [Google Scholar]
  67. Zhao Z, et al. Optimization of spinal cord stimulation using Bayesian preference learning and its validation. IEEE Trans. Neural Syst. Rehabil. Eng. 2021;29:1987–97. doi: 10.1109/TNSRE.2021.3113636. [DOI] [PubMed] [Google Scholar]
  68. Zheng Z, Savolainen O W, Constandinou T G. Algorithm and hardware considerations for real-time neural signal on-implant processing. J. Neural Eng. 2022;19:016029. doi: 10.1088/1741-2552/ac5268. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are openly available at the following URL/DOI: https://doi.org/10.5281/zenodo.14713531.