Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2023 Nov 9;13:19444. doi: 10.1038/s41598-023-46732-6

The influence of large-diameter multifocal contact lens on ocular surface, visual quality, and visual function for presbyopic adults with dry eye syndromes

Ching-Jen Hsiao 1,2, Hsiao-Ching Tung 3, Chuen‐Lin Tien 4, Yu-Wen Chang 1, Ching-Ying Cheng 1,2,
PMCID: PMC10636056  PMID: 37945680

Abstract

This study investigated the influence of large-diameter multifocal contact lenses on the ocular surface, visual quality, and visual function for presbyopic adults with dry eye syndromes. The study enrolled 40–55-year-old adults with presbyopia and dry eye syndromes (DES). The subjects were randomly assigned to three groups wearing different designs of contact lenses (Proclear, SMR, and Optimum) for 6–8 h a day for two weeks. Ocular surface health, tear quality, visual quality, and visual function were measured before and after lens wear. No significant difference was observed across all three groups for the amount of conjunctival redness, blink frequency (lens on), and stereopsis vision before and after wearing. Although there seemed to be a significant declining trend for corneal staining and limbal redness, non-invasive tear break-up time (TBUT), and lipid layer thickness while lens wear, the measured values were all within the normal range. Vice-versa after lens removal, results also showed significant improvement on lipid layer thickness, blink frequency (lens off), and contact TBUT. A significant improvement was observed in the modulation transfer function (MTF) of the total area ratio after wearing contact lenses. In contrast, the MTF of the high-order aberration area ratio resulting from lens wear was lower than that of the baseline measurement. There are also significant improvements observed for SMR and Optimum regarding near visual acuity, near point of accommodation, and the subjective questionnaire (OSDI and VBP) scores. Although it is difficult to avoid a specific negative impact on the ocular surface and tear film, visual function and visual quality can still be positively improved, especially shown on larger diameter and distance-center designed multifocal contact lenses.

Subject terms: Health care, Medical research

Introduction

Contact lens wear is one effective option for correcting ametropia14. However, the majority of individuals opt for contact lenses to correct their refractive errors before reaching the age of 30 to 34. Beyond this age range, changes in accommodation5 and physiological mechanisms, such as a reduction in the number and size of goblet cells, as well as decreased lacrimal gland secretion, can lead to conjunctival alterations and dryness of the eyes. This, in turn, can lead to discomfort and may even increase the risk of infection when wearing contact lenses6,7.

Dry eyes, a common condition seen in ophthalmology clinics8, are now manifesting at increasingly younger ages912. The frequent use of digital devices often reduces the rate of eye blinking, prolonging the exposure of the eye surface to the air, disrupting the delicate balance between tear evaporation and tear film replenishment13. Long-term users of digital products, approximately 64% to 90% of them, are susceptible to experiencing symptoms associated with Computer Vision Syndrome (CVS), which commonly include headache, shoulder and neck pain, dry eyes, eye-surface congestion, blurred vision, diplopia, and eye fatigue14,15. CVS symptoms are closely related to dry eye symptoms; approximately 75% of people with CVS suffer from dry eyes or dry eye syndromes16,17, high-order aberration, visual fluctuation, glare, poor contrast sensitivity and visual quality18,19.

Many contact lenses have been designed and developed using different materials and optical designs to correct presbyopia and have also been designed based on wearing needs to improve visual quality and visual comfort when wearing contact lenses for extended periods4,2022. According to previous studies, rigid contact lenses (RGP) offer better ocular health due to increased tear exchange and higher oxygen permeability than soft contact lenses (SCL)23. Recent research has been exploring large-diameter RGPs lenses2427, with a particular focus on their impacts on DES, tear film quality, visual quality, visual function, and binocular accommodation.

Large-diameter RGPs, also known as scleral contact lenses, are defined with a diameter > 12.5 mm. It is designed to only have direct contact with the ocular surface at the bulbar conjunctiva and thus create corneal vaults filled with preservative-free (PF) saline solution. The PF saline solution-filled space between corneal surface and the rear surface of the lens provides a stable tear film24,28,29. Due to the breathable material and design, large-diameter RGPs are often used to treat keratoconus (KCN), high astigmatism that cannot be corrected with small-diameter rigid lenses, ocular surface diseases, and simple refractive correction3032. They are also a preferred choice over soft contact lenses for their ability to provide tear storage between the lens and the cornea, offering relief from discomfort associated with dry eye syndrome (DES)25,33.

Large-diameter multifocal contact lenses, whether soft or rigid, are designed for simultaneous vision with minimal movement to achieve a translating effect. These multifocal contact lenses can further be categorized as center-near (CN) for improved near vision with a trade-off of blurrier distant vision, and center-distant (CD) for the vice versa34. The purpose of this study was to analyze the impact of wearing different types of large-diameter multifocal contact lenses on the ocular surface, tear film, and visual quality in presbyopic adults with dry eye syndrome (DES).

Materials and methods

Study design

A mixed two-factor design (3 groups × 4 measurements) experimental study was conducted from November 18, 2020, to May 30, 2021, at Chung Shan Medical University, Taichung, Taiwan. This experimental study protocol was reviewed and approved by the institutional review board of Chung Shan Medical University Hospital (approval no. CS2-20,089), and the study strictly adhered to the research ethics specifications of the Declaration of Helsinki. To ensure accurate measurement results, possible confounding factors such as laboratory brightness, visual target distance, and subjective differences in measurement tools and equipment operators were controlled (STROBE guidelines for reporting the manuscript35,36).

2.2. Research subjects

After the investigators thoroughly described the study content and procedure, subjects voluntarily decided to participate in the study and signed the informed consent form. The Ocular Surface Disease Index (OSDI) and Dry Eye Questionnaire (DEQ-5) were administered for preliminary screening. Subjects with OSDI scores ≥ 13 and 6 < points DEQ-5 scores < 12 were included in the study, it indicated that subjects who met the screening criteria all displayed subjective dry eye symptoms.

At first, 68 presbyopic adults aged 40–55 years completed the OSDI and DEQ-5 questionnaires. Of these, 60 subjects were enrolled after screening. The inclusion criteria for the study were refractive errors of myopia ≤ 8.00 D, astigmatism ≤ 1.75 D, and near ADD ≥ 0.75 D. Subjects who had undergone at least one eye surgery, eye-related diseases, ocular surface dysfunction, systemic immune-related diseases, long-term usage of eye drops, or could not wear contact lenses were excluded. Of the 60 people included after the initial screening, 18 were subsequently excluded: 5 subjects had refractive errors that did not meet the criteria, 5 were unable to adapt to wearing contact lenses, and 8 did not cooperate with the follow-up schedule. Finally, 42 subjects (12 Proclear Lens: 46.33 ± 3.75; 12 SMR Lens: 46.83 ± 5.06; and 18 Optimum Lens: 45.78 ± 3.17; 12 men: 46.67 ± 5.14 and 30 women: 46.07 ± 3.33) with a mean age of 46.24 ± 3.88 years participated and were randomly assigned to different lens group, If the assigned group is not suitable for the subject, it is handled by dropping them out. One-way analysis of variance (ANOVA) showed no significant differences in age, OSDI, DEQ-5 score, or refractive errors among three lens groups (Table 1).

Table 1.

General information comparison of lens groups (Mean ± Standard Deviation).

Group N Age
M ± SD
OSDI
M ± SD
DEQ-5
M ± SD
Right Sph(D)
M ± SD
Right Cyl(D)
M ± SD
Left Sph(D) M ± SD Left Cyl(D)
M ± SD
Proclear (Soft CL) 12 46.33 ± 3.75 18.33 ± 4.36 9.42 ± 3.18  − 5.33 ± 2.63  − 0.48 ± 0.21  − 5.38 ± 2.28  − 0.52 ± 0.46
SMR (ArtMost, Soft CL) 12 46.83 ± 5.06 17.42 ± 1.78 8.83 ± 1.53  − 6.71 ± 1.22  − 0.67 ± 0.63  − 6.54 ± 1.43  − 0.67 ± 0.56
Optimum (Rigid CL) 18 45.78 ± 3.17 16.44 ± 2.28 9.00 ± 2.38  − 5.31 ± 2.02  − 0.83 ± 0.52  − 5.33 ± 2.02  − 0.86 ± 0.43
ANOVA F and P-value

F = 0.262

p = 0.771

F = 1.532

p = 0.229

F = 1.263

p = 0.294

F = 2.002

p = 0.149

F = 1.828

p = 0.174

F = 1.590

p = 0.217

F = 1.874

p = 0.167

P < 0.05.

The sample size of this study was determined using G*Power analysis, with an effect size of f = 0.8, α = 0.05, power (1-β) = 0.95, and number of groups = 3. The calculated results of the total sample size were 30. The number of participants who completed the study (42) exceeded the sample size required (30); the power appeared to be adequate after recalculation and adjustment (effect size f = 0.8, α = 0.05, power (1 − β) = 0.984).

Research materials

Contact lenses

Three types of multifocal contact lenses were used to compare the dependency variances between different optical or different material designs (Table 2): Lens 1: CooperVision Proclear® 1-day multifocal Lens (Proclear); Lens 2: ArtMost Aspherical progressive soft contact lens (SMR); and Lens 3: Optimum GP rigid gas-permeable contact lenses (Optimum). All the contact lenses used in this study had passed FDA approval for a broad range of indications and contact lens modalities. Table 2 lists the contact lens design and specification parameters. All three contact lenses have multifocal designs; lens material and lens design are enhanced for eye parts prone to dryness. All subjects were asked to wear the contact lenses for 6–8 h a day. To avoid any potential sources of bias, each test was performed by the same optometrist. All participants were invited to undergo baseline (wear multifocal spectacles) examination, as well as fellow up on D1, W1, W2 after wearing contact lenses.

Table 2.

Contact lens design and specification parameters.

Contact lens Optical design Material Basic curve (mm) Diameter (mm) Water content (%) Oxygen permeability(cm2/s) (ml O2/mL × mm Hg)
Proclear (Soft CL) Aspherical Progressive (Near–center) Omafilcon A 8.7 14.2 60% 21 × 10−11
SMR (ArtMost, Soft CL) Aspherical Progressive (Distance–center) Ocufilcon D 9.0 14.4 55% 19.6 × 10−11
Optimum (Rigid CL) Aspherical Progressive (Distance–center) Optimum Extra (Raflufocon D) 6.49–8.43 15.5 100 × 10−11

The process of fitting begins with an initial assessment, which includes evaluating eye health, visual acuity, and specific refractive errors, as well as patient's medical history. Subsequently, a trial lens fitting assessment are performed, which involve checking for signs of decentration, excessive movement, or lens tightness. The central cornea and limbus should not experience compression, and a certain tear thickness should be ensured. Initially, the central area generally needs to maintain a tear thickness of about 300 µm. The initial position of the landing zone should be located at approximately halfway between the corneal edge and the lens edge, and it should not exert excessive pressure on the cornea. After wearing the trial lenses for 30 min, the central tear thickness should still be around 150–200 µm. Only after determining the appropriate trial lenses should a refraction check be performed. Subsequent follow-up appointments, training, education, and regular monitoring were required.

Measurement

Measurements in the study included: (1) ocular surface; (2) tear quality; (3) visual quality; and (4) visual function. Each examination was performed on the baseline (BL), day 1, week 1, and week 2. In addition to the baseline measurement data (wearing habitual multi-focal spectacles) without participants wearing contact lenses before the experiment, some exams could only be measured after removing the lenses (ocular surface redness, corneal fluorescence staining, and contact TBUT), while others could be performed with or without participants wearing contact lenses, including NI-TBUT, lipid layer, and blink frequency. Visual quality included blinking frequency and modulation transfer function (MTF), as well as the analysis of the subjective questionnaire survey. Blinking frequency was also recorded while wearing and removing contact lenses.

A Topcon VT-10 phoropter (Topcon, Tokyo, Japan), View-M digital visual acuity chart (Quan Chin Industrial Co., Taiwan, 6 M in a bright room), and a TMV near-point card (Brighten Optix Co., Taiwan, 40 cm) were used to measure subjective refraction and visual acuity. Further examinations conducted for this study involved the following four stages: (1) ocular surface evaluation (Topcon SL-D701 slit lamp), including quantifying the redness of the conjunctiva or limbus (Efron Grading Scales for Contact Lens Complications, score 0–4), and corneal fluorescence staining (standard Oxford scale, score 0–5); (2) tear quality, including contact tear-film break-up time (TBUT), non-contact tear-film break-up time (NI-TBUT), and lipid-layer thickness (SBM ICP Tearscope); (3) visual quality examination, including blink frequency, modulation transfer function (Nidek OPD Scan 3, Tokyo, Japan)37, and dry-eye-related questionnaires (Dry Eye Questionnaire, DEQ-5 and Ocular Surface Disease Index, OSDI); (4) visual function measurement, including near visual acuity, stereo visual acuity (Butterfly Stereo Acuity test), and near point of accommodation (Royal Air Force Ruler, RAF; and Push-up method).

Data analysis and statistical analysis

All data were obtained and analyzed using SPSS 22.0 statistical software (IBM, Armonk, NY, USA). A value of p < 0.05 was considered statistically significant. Repeat-measurement analysis, two-way ANOVA and simple main effects analysis were performed. Because the participants were all healthy subjects, there was no significant difference between left and right eyes in terms of refractive errors; the results only showed the left eye parameters.

Institutional review board statement

The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Ethics Committee of Chung Shan Medical University Hospital (Taichung, Taiwan) (Approval Number: CS2-20089). Informed written consent was obtained from all individual participants included in the study.

Informed consent

Patients signed informed written consent regarding the publication of their data or photographs.

Results

Ocular surface

Two-way ANOVA, mixed design, and simple main effects were used to compare the differences between the three lenses and four times of measurement (Table 3 and Fig. 1). The results showed that ocular surface examination with different types of large-diameter multifocal contact lenses might influence the values of corneal fluorescence staining (F = 7.235, p < 0.000), conjunctival redness (F = 3.811, p = 0.012), and limbus redness (F = 3.443, p = 0.042). Simple main effects analysis indicated that all three types of largediameter multifocal contact lenses had negative impacts on the ocular surface, of which the Optimum RGPs lens had the greatest impact on corneal (F = 4.489, p = 0.007) and limbus (F = 3.357, p = 0.026); post hoc analysis showed significant differences between baseline, week 1, and week 2. Additionally, no significant was observed between different materials (soft and rigid) and optical (CN and CD) designs, it appeared that the lens diameter might be the sole factor affecting the ocular surface, the larger the contact lens diameter, the more redness reaction was shown on the cornea and limbus.

Table 3.

Two-way ANOVA, mixed design, and simple main effects analysis on ocular surface measurement.

Two-way ANOVA Main effects Measurement Post hoc( p < 0.05)
F p
Corneal Fluorescence Staining (Left eye) F = 7.235, p = 0.001** Proclear: 2.209 .106
SMR 1.582 .212
Optimum: 4.489 .007 BL > W1, BL > W2
Conjunctive Redness (Left eye)

F = 3.811

p = 0.012*

Proclear: 1.320 .284
SMR 3.163 .079
Optimum: 2.439 .075
Limbus Redness (Left eye)

F = 3.443

p = 0.042*

Proclear: 1.984 .136
SMR 1.375 .268
Optimum: 3.357 .026 BL > W1, BL > W2

* p < .05, ** p < .01, > means “better than”.

Figure 1.

Figure 1

Ocular surface with corneal staining, conjunctival redness, and limbus redness between the three lenses and four times of measurement.

Tear quality

Tear changes when wearing contact lenses (Lens-on, tears on the front surface of the lens)

Two-way ANOVA, mixed design, and simple main effects were used to compare the differences between the three lenses and four times of measurement (Table 4 and Fig. 2). The results showed that the values of non-contact tear breakup time (NI-TBUT) (F = 19.421, p < 0.001) and lipid-layer thickness (F = 5.222, p = 0.002) became significantly worse when wearing contact lenses (Lens-on). Further analysis of the main effects comparison revealed that each type of large diameter contact lenses had significant impacts on NI-TBUT (Proclear: F = 6.177, p = 0.002, SMR: F = 17.132, p < 0.001, Optimum: F = 4.807, p = 0.005) and lipid layer (Proclear: F = 7.118, p = 0.001, SMR: F = 3.778, p = 0.020, Optimum: F = 0.291, p = 0.832), of which Optimum lens did not differ from the baseline, indicating that Optimum lens had little impact on lipid layer while wearing. The above comparison showed that wearing contact lenses affected the NI-TBUT and the lipid-layer thickness on the front surface of the lens with wearing duration. Additionally, no significant was observed between different materials (soft and rigid) and optical (CN and CD) designs, it appeared that the lens diameter might be the main factor affecting the tear quality while lens on. Optimum lens exerts the greatest positive influence and shows a more stable trend compared with the other two lenses.

Table 4.

Two-way ANOVA, mixed design, and simple main effects analysis on tear quality measurement.

Two-way ANOVA Main effects Measurement Post hoc( p < 0.05)
F p

NI-TBUT

(Lens on_Left eye)

F = 19.421 p < 0.001** Proclear: 6.177 .002 BL > W1
SMR 17.132  < .001 BL > D1, BL > W1, BL > W2
Optimum: 4.807 .005 BL > D1

Lipid Layer

(Lens on_Left eye)

F = 5.222

p = 0.002**

Proclear: 7.118 .001 BL > W1, BL > W2
SMR 3.778 .020 BL > W2, D1 > W2
Optimum: 0.291 .832

NI-TBUT

(Lens off_Left eye)

F = 7.965 p < 0.001** Proclear: 2.836 .094
SMR 11.786  < .001 BL > W1, BL > W2, D1 > W2
Optimum: 2.017 .123

Lipid Layer

(Lens off_Left eye)

F = 3.811

p = 0.012*

Proclear: 1.759 .174
SMR 8.423  < .001 W2 > D1
Optimum: 7.043  < .001 D1 > BL

Contact-TBUT

(Lens off_Left eye)

F = 3.443

p = 0.042*

Proclear: 0.298 .827
SMR 2.491 .050 BL > W2
Optimum: 4.860 .005 D1 > W2

NI-TBUT

(Lens on_Left eye)

F = 19.421 p < 0.001** Proclear: 6.177 .002 BL > W1
SMR 17.132  < .001 BL > D1, BL > W1, BL > W2
Optimum: 4.807 .005 BL > D1

Lipid Layer

(Lens on_Left eye)

F = 5.222

p = 0.002**

Proclear: 7.118 .001 BL > W1, BL > W2
SMR 3.778 .020 BL > W2, D1 > W2
Optimum: 0.291 .832

NI-TBUT

(Lens off_Left eye)

F = 7.965 p < 0.001** Proclear: 2.836 .094
SMR 11.786  < .001 BL > W1, BL > W2, D1 > W2
Optimum: 2.017 .123

Lipid Layer

(Lens off_Left eye)

F = 3.811

p = 0.012*

Proclear: 1.759 .174
SMR 8.423  < .001 W2 > D1
Optimum: 7.043  < .001 D1 > BL

Contact-TBUT

(Lens off_Left eye)

F = 3.443

p = 0.042*

Proclear: 0.298 .827
SMR 2.491 .050 BL > W2
Optimum: 4.860 .005 D1 > W2

NI-TBUT

(Lens on_Left eye)

F = 19.421 p < 0.001** Proclear: 6.177 .002 BL > W1
SMR 17.132  < .001 BL > D1, BL > W1, BL > W2
Optimum: 4.807 .005 BL > D1

Lipid Layer

(Lens on_Left eye)

F = 5.222

p = 0.002**

Proclear: 7.118 .001 BL > W1, BL > W2
SMR 3.778 .020 BL > W2, D1 > W2
Optimum: 0.291 .832

NI-TBUT

(Lens off_Left eye)

F = 7.965 p < 0.001** Proclear: 2.836 .094
SMR 11.786  < .001 BL > W1, BL > W2, D1 > W2
Optimum: 2.017 .123

Lipid Layer

(Lens off_Left eye)

F = 3.811

p = 0.012*

Proclear: 1.759 .174
SMR 8.423  < .001 W2 > D1
Optimum: 7.043  < .001 D1 > BL

Contact-TBUT

(Lens off_Left eye)

F = 3.443

p = 0.042*

Proclear: 0.298 .827
SMR 2.491 .050 BL > W2
Optimum: 4.860 .005 D1 > W2

* p < .05, ** p < .01, > means “better than”.

Figure 2.

Figure 2

Tear changes of NI-TBUT, lipid-layer thickness, and Contact-TBUT between the three lenses and four times of measurement.

Tear changes after removing contact lenses (Lens-off, tears on the ocular surface).

Like tears on the front surface, analysis on the ocular surface indicated that NI-TBUT (F = 7.965, p < 0.001), lipid-layer thickness (F = 3.811, p = 0.012), and Contact-TBUT (F = 3.443, p = 0.042) were significantly different while removing contact lenses. Main effects comparison analysis of the three lens groups revealed that although the performance of tear breakup time for the three lens groups was still worse after the contact lenses were removed (NI-TBUT: Proclear: F = 2.836, p = 0.094; SMR: F = 11.786, p < 0.001; Optimum: F = 2.017, p = 0.123; Contact-TBUT: Proclear: F = 0.298, p = 0.827, SMR: F = 2.491, p = 0.050, Optimum: F = 4.860, p = 0.005), especially for the SMR lenses. What was surprising that the positive performance of lipid-layer thickness after removing contact lenses (Proclear: F = 1.759, p = 0.174, SMR: F = 8.423, p < 0.001, Optimum: F = 7.043, p < 0.001), especially shown on the SMR and Optimum lenses. This might suggest that larger-diameter or different material design (soft and rigid, F = 7.985, p < 0.001) contact lenses are beneficial to lipid-layer thickness.

Visual quality

Blinking frequency

Table 5 and Fig. 3 showed that the blinking frequency while wearing contact lenses was relatively stable, and there was no significant difference compared to that when wearing spectacles (F = 1.088, p = 0.357). Even so, there were positive trends in the performance of Proclear and Optimum lenses (both were soft lens); however, upon removing contact lenses (F = 5.318, p = 0.002), the blinking frequency of Optimum (F = 7.729, p < 0.001) lenses increased significantly, indicating that the comfort of wearing large-diameter rigid contact lenses was relatively satisfactory.

Table 5.

Two-way ANOVA, mixed design, and simple main effects analysis on visual quality measurement.

Two-way
ANOVA
main effects Measurement Post hoc( p < 0.05)
F p
Blinking Frequency Lens on

F = 1.088

p = 0.357

Proclear: 1.899 .149
SMR 0.421 .739
Optimum: 2.198 .100
Blinking Frequency Lens off

F = 5.318

p = 0.002**

Proclear: 0.895 .454
SMR 3.050 .057
Optimum: 7.729  < .001 W1 > BL , W1 > D1

MTF Total Area Ratio

(Left eye)

F = 78.958

p < 0.001**

Proclear: 71.318  < .001 D1 > BL, W1 > BL, W2 > BL
SMR 13.752  < .001 D1 > BL, W1 > BL, W2 > BL
Optimum: 20.766  < .001 D1 > BL, W1 > BL, W2 > BL

MTF Total HO Ratio

(Left eye)

F = 39.467

p < 0.001**

Proclear: 13.440  < .001 BL > D1, BL > W1, BL > W2
SMR 19.702  < .001 BL > D1, BL > W1, BL > W2
Optimum: 20.211  < .001 BL > D1, BL > W1, BL > W2

OSDI Score

(BL, W1, W2)

F = 3.443

p = 0.042*

Proclear: 0.207 .814
SMR 8.117 .002 W2 > BL, W2 > W1
Optimum: 3.567 .041 W1 > BL,

VBP Score

(BL, W1, W2)

F = 19.421 p < 0.001** Proclear: 1.045 .369
SMR 1.233 .311
Optimum: 3.132 .047 W1 > BL, W2 > BL

* p < .05, ** p < .01, > means “better than”.

Figure 3.

Figure 3

Visual quality between the three lenses and four times of measurement.

Modulation transfer function

The higher the contrast of the modulation transfer function (MTF), the better the resolution. After the correction of the total area ratio by the contact lens, compared with unaided vision, the resolution of the three contact lenses improved significantly (F = 78.958, p < 0.001). Main effects comparison analysis revealed that day 1, week 1, and week 2 performed better than baseline in each lens group (Table 5 and Fig. 3, Proclear: F = 71.318, p < 0.001; SMR: F = 13.752, p < 0.001; Optimum: F = 20.766, p < 0.001).

Additionally, the performance of the MTF total higher order (HO) ratio was significantly worse (F = 39.467, p < 0.001). Main effects comparison analysis revealed that the performance was worse on day 1, week 1, and week 2 compared with baseline in each lens group (Table 5 and Fig. 3, Proclear: F = 13.440, p < 0.001, SMR: F = 19.702, p < 0.001, Optimum: F = 20.211, p < 0.001). This might be related to the design of the lens, as Poclear lens is CN and had a smaller effect on the MTF total HO ratio compared with the effect of the distance-center design of the SMR and Optimum lenses which are CD; however, performance on day 1, week 1, and week 2 was significantly worse in each lens group compared with the asymptotic multifocal design.

Questionnaire

The OSDI questionnaires were performed at three different times: baseline, after contact lenses were worn for 1 week, and after contact lenses were worn for 2 weeks. There was significant variation among the time points (F = 3.443, p = 0.042), with subjects reporting higher satisfaction with SMR (F = 8.117, p = 0.002) and Optimum (F = 3.567, p = 0.041) after 1 or 2 weeks of use. Additionally, people with dry eye or discomfort reported increased comfort after wearing the lenses. Visual Behavior Performance scores also significantly improved (F = 19.421, p < 0.001), particularly in association with the Optimum lens (F = 3.132, p = 0.047) after 1 or 2 weeks of use. This suggests that large-diameter contact lenses might be helpful for near work, comfort, perception, and balance.

Visual function

The two-way ANOVA analysis of near visual acuity (F = 0.461, p = 0.637) and stereopsis (F = 0.440, p = 0.648) did not show any significant difference between the three lens groups or the four times points of measurement (Table 6 and Fig. 4). However, there was an effect on near visual acuity. The Optimum lenses performed better than spectacles for correcting near visual acuity (F = 3.847, p = 0.015). Additionally, Proclear lens performed significant worse than baseline both on the right and left eyes in terms of the near point of accommodation (right eye: F = 3.358, p = 0.021; left eye: F = 5.377, p = 0.002). Nevertheless, there was no significant difference between SMR and Optimum, and there was even a positive trend after baseline in terms of the near point of accommodation.

Table 6.

Two-way ANOVA, mixed design, and simple main effects analysis on visual function examination.

Two-way
ANOVA
Main effects Measurement Post hoc ( p < 0.05)
F p
Near Visual Acuity

F = 0.461

p = 0.637

Proclear: 1.656 .196
SMR .844 .480
Optimum: 3.847 .015 D1 > BL
Stereopsis

F = 0.440

p = 0.648

Proclear: 3.633 .076
SMR 2.904 .092
Optimum: 0.486 .694
Near Point of Accommodation (D)—Right eye

F = 3.358

p = 0.021*

Proclear: 12.255  < .001 BL > D1, BL > W1, BL > W2
SMR 1.048 .384
Optimum: 3.283 .078

Near Point of Accommodation (D)—

Left eye

F = 5.377

p = 0.002**

Proclear: 13.436  < .001 BL > D1, BL > W1, BL > W2
SMR 2.810 .081
Optimum: 3.007 .074

* p < .05, ** p < .01, > means “better than”.

Figure 4.

Figure 4

Visual Function between the three lenses and four times of measurement.

Discussion

Analysis of the ocular surface condition after wearing large-diameter multifocal contact lenses for 2 weeks revealed that although the redness of the limbus and corneal fluorescent staining worsened compared to baseline measurements, but the values were both within the clinically normal and acceptable standard38. For people with cornea damage, previous research has indicated that wearing large-diameter rigid contact lenses can significantly reduce corneal fluorescent staining29. The present study did not demonstrate any significant improvements in the state of the ocular surfaces, which may be related to the health of the initial eye itself. For patients who wear contact lenses to prevent ocular surface damage, it may be better to assess any effect after utilizing those lenses for extended period of time.

Measurements of tear quality can be conducted in two stages: the front surface of the contact lens and the front surface of the cornea39. The tears on the front surface of both the contact lens and the cornea are affected by lenses usage. After 1 week of wearing the contact lens, both the NI-TBUT and lipid-layer thickness significantly decreased. The correlation between the two variables may be used to evaluate DES caused by contact lenses4042. Previous studies have also pointed out that the hydrophobic area of the lens will increase tear evaporation and reduce tear breakup time on the non-moisturizing lens surface. In addition, the results may be due to the electrostatic effect between the lipid layer and the contact lenses, resulting in oil deposition on the lens, which decreases the lipid-layer thickness and increases tear evaporation40. affecting the integrity of tear coverage.

One limiting factor of this study was the duration being only 2 weeks of data collection for each subject. Given the limited time frame, it is impossible to conclude whether prolonged wear of the lens beyond the 2-week period will have continuous negative impact on the tear film quality the front of the contact lens and the front surface of the cornea. However, analysis of tear changes showed that among the three lens groups, Optimum exerts higher stability of tear films, which may be related to the design and rigid lens material. Further investigation is necessary to be conclusive.

Blink frequency is a major factor affecting the stability of the tear film. Each complete blink will allow even distribution of the tear film covering the ocular surface to increase stability and concurrently improve visual quality43. The analysis results of blink frequency showed no significant change amont the three groups while wearing the lenses. However, there is significant improvement for the Optimum group after lens removal. Therefore, it is speculated the factors responsible for the change in tear balance are not affected by blink frequency, but rather conservation of ocular surface with adequate tear support.

The simulated value of the MTF of the total aberration and measurement results with contact lenses have better visual quality than the eye without wearing contact lenses at the beginning. This result shows that the difference is possibly due to the correction of ametropia44. In addition, material, design, and adaptation of contact lenses also affect changes in aberrations. Previous studies compared the performance of contrast sensitivity of spectacles, soft contact lenses, and rigid contact lenses. Although there was no significant difference among the three types of lenses, the contrast sensitivity of the rigid contact lenses was better than that of the soft contact lenses and spectacles. In addition, the rigid contact lens design with a large diameter can correct most of the corneal astigmatism and high-order aberrations29,45.

The subjective questionnaire showed little difference in the scores of the OSDI before and after wearing contact lenses. On the other hand, the visual behavior performance scale significantly improved after 1 week of lens wear. Previous studieshave proposed that there is a significant proportional relationship between the ocular surface disease index scale and the convergence fatigue symptom survey (CISS) showing benefits to enhance basic binocular visual function test to the evaluation of DES46. Although this portion is not entirely conclusive from the results of this subjective questionnaire, it still has reference value for the subjective comfort of wearing contact lenses. Visual function includes eye accommodation, stereo vision, and near visual acuity. The test results showed that after two weeks of lens wear, there was little difference in stereo vision compared to habitual spectacles correction. CD lens could induce better near visual acuity and accommodation; on the contrary, CN lens perform worse in accommodation34,4751.

Conclusions

Each lens design has its advantages and disadvantages. Patients can choose different modalities of presbyopia-correcting contact lenses based on their visual needs and adaptability. Although there may be specific negative impacts on the ocular surface and tear film with contact lens wear, one can still expect improvement in eye health, visual quality, and overall quality of life with proper lens fitting, wearing modality, and lens care. More studies need to be performed to explore a variety of contact lens designs that may be suitable for people with different needs and circumstances to provide more references on contact lens fitting, reduce complaints and discomfort from contact lens wear, and provide presbyopic adults with more choices to achieve comfortable vision and enhance quality of life.

Author contributions

Conceptualization, C.-J.H., H.C.T., Y.-W.C. C.-L.T., and C.-Y.C.; investigation, data curation, and formal analysis, Y.-W.C. and C.-Y.C.; writing—original draft preparation, C.-J.H., Y.-W.C., and C.-Y.C.; writing—review and editing, C.-Y.C, C.-L.T., and H.-C.T. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by Chung Shan Medical University Research Project (FCU/CSMU-111-002) in Taiwan.

Data availability

All data generated or analyzed during this study are included in this published article and its supplementary information files. Correspondence and requests for materials should be addressed to C.-Y.C.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Flaxman SR, et al. Global causes of blindness and distance vision impairment 1990–2020: A systematic review and meta-analysis. Lancet Glob. Health. 2017;5(12):e1221–e1234. doi: 10.1016/S2214-109X(17)30393-5. [DOI] [PubMed] [Google Scholar]
  • 2.Heus P, Verbeek JH, Tikka C. Optical correction of refractive error for preventing and treating eye symptoms in computer users. Cochrane Database Syst. Rev. 2018;4(4):Cd009877. doi: 10.1002/14651858.CD009877.pub2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Mohd-Ali B, Tan XL. Patterns of use and knowledge about contact lens wear amongst teenagers in rural areas in Malaysia. Int. J. Environ. Res. Public Health. 2019;16(24):5161. doi: 10.3390/ijerph16245161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Musgrave CSA, Fang F. Contact lens materials: A materials science perspective. Materials. 2019;12(2):261. doi: 10.3390/ma12020261. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Pamela, A. & Lowe, O. Priorities for presbyopes: Maintain comfort and continuity. August 15, 2016; Available from: https://www.reviewofoptometry.com/article/priorities-for-presbyopes-maintain-comfort-and-continuity.
  • 6.Moss SE. Prevalence of and risk factors for dry eye syndrome. Arch. Ophthalmol. 2000;118(9):1264. doi: 10.1001/archopht.118.9.1264. [DOI] [PubMed] [Google Scholar]
  • 7.Bennett ES. Contact lens correction of presbyopia. Clin. Exp. Optom. 2008;91(3):265–278. doi: 10.1111/j.1444-0938.2007.00242.x. [DOI] [PubMed] [Google Scholar]
  • 8.Chang WB, et al. An analysis of drug expenses in the National Health Insurance Database for dry eye syndrome patients in 2008. Chang Gung Med. J. 2012;8(3):18–26. [Google Scholar]
  • 9.Moon JH, Kim KW, Moon NJ. Smartphone use is a risk factor for pediatric dry eye disease according to region and age: A case control study. BMC Ophthalmol. 2016;16(1):1–7. doi: 10.1186/s12886-016-0364-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Farrand KF, et al. Prevalence of diagnosed dry eye disease in the United States among adults aged 18 years and older. Am. J. Ophthalmol. 2017;182:90–98. doi: 10.1016/j.ajo.2017.06.033. [DOI] [PubMed] [Google Scholar]
  • 11.Dana R, et al. Estimated prevalence and incidence of dry eye disease based on coding analysis of a large, all-age United States Health Care System. Am. J. Ophthalmol. 2019;202:47–54. doi: 10.1016/j.ajo.2019.01.026. [DOI] [PubMed] [Google Scholar]
  • 12.Mai ELC, et al. Population-based study on the epidemiology of dry eye disease and its association with presbyopia and other risk factors. Int. Ophthalmol. 2019;39(12):2731–2739. doi: 10.1007/s10792-019-01117-5. [DOI] [PubMed] [Google Scholar]
  • 13.Jaiswal S, et al. Ocular and visual discomfort associated with smartphones, tablets and computers: What we do and do not know. Clin. Exp. Optom. 2019;102(5):463–477. doi: 10.1111/cxo.12851. [DOI] [PubMed] [Google Scholar]
  • 14.Rosenfield M. Computer vision syndrome: a review of ocular causes and potential treatments. Ophthalmic. Physiol. Opt. 2011;31(5):502–515. doi: 10.1111/j.1475-1313.2011.00834.x. [DOI] [PubMed] [Google Scholar]
  • 15.Sheppard AL, Wolffsohn JS. Digital eye strain: Prevalence, measurement and amelioration. BMJ Open Ophthalmol. 2018;3(1):e000146. doi: 10.1136/bmjophth-2018-000146. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Kaido M, et al. Severe symptoms of short tear break-up time dry eye are associated with accommodative microfluctuations. Clin. Ophthalmol. 2017;11:861–869. doi: 10.2147/OPTH.S128939. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Messmer, E. M. The pathophysiology, diagnosis, and treatment of dry eye disease. Deutsches Aerzteblatt Online (2015). [DOI] [PMC free article] [PubMed]
  • 18.D'Souza S, et al. Study of tear film optics and its impact on quality of vision. Indian J. Ophthalmol. 2020;68(12):2899–2902. doi: 10.4103/ijo.IJO_2629_20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Pucker AD, Tichenor AA. A review of contact lens dropout. Clin. Optom. 2020;12:85–94. doi: 10.2147/OPTO.S198637. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Bausch + Lomb Ultra Contact lenses. Available from: http://www.bausch.com.tw/zh-tw/our-products/contact-lenses/lenses-for-short-sighted-long-sighted/ultra-soft-contact-lenses/.
  • 21.Alcon Dailies Total 1 Available from: https://alcon-vc.com.tw/dailiestotal1/.
  • 22.Poclear Contact lenses. Available from: https://coopervision.com.tw/contact-lenses/%E5%AF%B6%E6%99%B4%E7%B3%BB%E5%88%97/%E5%AF%B6%E6%99%B4%E6%97%A5%E6%8B%8B.
  • 23.Efron N. Obituary–rigid contact lenses. Cont. Lens Anterior Eye. 2010;33(5):245–252. doi: 10.1016/j.clae.2010.06.009. [DOI] [PubMed] [Google Scholar]
  • 24.Carracedo G, et al. Post-lens tear turbidity and visual quality after Optimum lens wear. Clin. Exp. Optom. 2017;100(6):577–582. doi: 10.1111/cxo.12512. [DOI] [PubMed] [Google Scholar]
  • 25.Alipour F, Kheirkhah A, Jabarvand Behrouz M. Use of mini optimum contact lenses in moderate to severe dry eye. Cont. Lens Anterior Eye. 2012;35(6):272–276. doi: 10.1016/j.clae.2012.07.006. [DOI] [PubMed] [Google Scholar]
  • 26.Schornack MM. Optimum lenses: A literature review. Eye Contact Lens. 2015;41(1):3–11. doi: 10.1097/ICL.0000000000000083. [DOI] [PubMed] [Google Scholar]
  • 27.Shorter E, et al. Optimum lenses in the management of corneal irregularity and ocular surface disease. Eye Contact Lens. 2018;44(6):372–378. doi: 10.1097/ICL.0000000000000436. [DOI] [PubMed] [Google Scholar]
  • 28.Lafosse E, et al. Comparison of the influence of corneo-Optimum and Optimum lenses on ocular surface and tear film metrics in a presbyopic population. Contact Lens Anterior Eye. 2018;41(1):122–127. doi: 10.1016/j.clae.2017.09.014. [DOI] [PubMed] [Google Scholar]
  • 29.Serramito M, et al. Corneal surface wettability and tear film stability before and after Optimum lens wear. Contact Lens Anterior Eye. 2019;42(5):520–525. doi: 10.1016/j.clae.2019.04.001. [DOI] [PubMed] [Google Scholar]
  • 30.Harthan J, et al. Optimum lens prescription and management practices: The SCOPE Study. Eye Contact Lens. 2018;44(Suppl 1):S228–S232. doi: 10.1097/ICL.0000000000000387. [DOI] [PubMed] [Google Scholar]
  • 31.Erdinest, N. et al. Quadrant asymmetric design contact lens for visual rehabilitation after eye trauma. Case Rep. Ophthalmol., 330–336 (2021). [DOI] [PMC free article] [PubMed]
  • 32.Uysal, B. S., Yaman, D., Kalkan Akcay, E., Kilicarslan, A., Sarac, O. & Cagil, N. Evaluation of corneal topography, tear film function and conjunctival impression cytology after long-term Optimum contact lens wear in keratoconus patients. Semin. Ophthalmol. 27, 1–7 (2021). [DOI] [PubMed]
  • 33.Bavinger JC, DeLoss K, Mian SI. Optimum lens use in dry eye syndrome. Curr. Opin. Ophthalmol. 2015;26(4):319–324. doi: 10.1097/ICU.0000000000000171. [DOI] [PubMed] [Google Scholar]
  • 34.Michaud L, Bennett ES, Woo SL, Reeder R, Morgan BW, Dinardo A, Harthan JS. Clinical evaluation of large diameter rigid-gas permeable versus soft toric contact lenses for the correction of refractive astigmatism. A multicentre study. Eye Contact Lens. 2018;44(3):164–169. doi: 10.1097/ICL.0000000000000323. [DOI] [PubMed] [Google Scholar]
  • 35.Sunjic-Alic A, Zebenholzer K, Gall W. Reporting of studies conducted on Austrian claims data. Stud. Health Technol. Inform. 2021;279:62–69. doi: 10.3233/SHTI210090. [DOI] [PubMed] [Google Scholar]
  • 36.Cuschieri ST. STROBE guidelines. Saudi J. Anaesth. 2019;13(1):S31–S34. doi: 10.4103/sja.SJA_543_18. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Xu Z, et al. Comparison of the clinical performance of refractive rotationally asymmetric multifocal IOLs with other types of IOLs: A meta-analysis. J. Ophthalmol. 2018 doi: 10.1155/2018/4728258. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Murphy PJ, et al. How red is a white eye? Clinical grading of normal conjunctival hyperaemia. Eye. 2007;21(5):633–638. doi: 10.1038/sj.eye.6702295. [DOI] [PubMed] [Google Scholar]
  • 39.Müller C, et al. Subjective comparison of pre-lens tear film stability of daily disposable contact lenses using ring mire projection. Clin. Optom. (Auckl.) 2020;12:17–26. doi: 10.2147/OPTO.S235167. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Nichols JJ, Sinnott LT. Tear film, contact lens, and patient-related factors associated with contact lens–related dry rye. Investig. Opthalmol. Vis. Sci. 2006;47(4):1319. doi: 10.1167/iovs.05-1392. [DOI] [PubMed] [Google Scholar]
  • 41.Rohit A, Wilcoxon M, Stapleton F. Tear lipid layer and contact lens comfort: A review. Eye Contact Lens. 2013;39(3):247–253. doi: 10.1097/ICL.0b013e31828af164. [DOI] [PubMed] [Google Scholar]
  • 42.Korb DR, Greiner JV, Glonek T. Tear film lipid layer formation: Implications for contact lens wear. Optom. Vis. Sci. 1996;73(3):189–192. doi: 10.1097/00006324-199603000-00011. [DOI] [PubMed] [Google Scholar]
  • 43.Braun RJ, et al. Dynamics and function of the tear film in relation to the blink cycle. Prog. Retina Eye Res. 2015;45:132–164. doi: 10.1016/j.preteyeres.2014.11.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.García-Marqués JV, Macedo-De-Araújo RJ, Cervino A, García-Lázaro S, McAlinden C, González-Méijome JM. Comparison of short-term light disturbance, optical and visual performance outcomes between a myopia control contact lens and a single-vision contact lens. Ophthal. Physiol. Opt. 2020;40(6):718–727. doi: 10.1111/opo.12729. [DOI] [PubMed] [Google Scholar]
  • 45.Jinabhai A, Radhakrishnan H, Tromans C, O’Donnell C. Visual performance and optical quality with soft lenses in keratoconus patients. Ophthal. Physiol. Opt. 2012;32(2):100–116. doi: 10.1111/j.1475-1313.2011.00889.x. [DOI] [PubMed] [Google Scholar]
  • 46.Rueff EM, King-Smith PE, Bailey MD. Can binocular vision disorders contribute to contact lens discomfort? Optom. Vis. Sci. 2015;92(9):e214–e221. doi: 10.1097/OPX.0000000000000671. [DOI] [PubMed] [Google Scholar]
  • 47.Rajagopalan ASMB, Edward S. Visual performance of subjects wearing presbyopic contact lenses. Optom. Vis. Sci. 2006;83(8):611–615. doi: 10.1097/01.opx.0000232185.00091.45. [DOI] [PubMed] [Google Scholar]
  • 48.Balamurali Vasudevan MF. Sara Gaib Objective and subjective visual performance of multifocal contact lenses: Pilot study. Contact Lens Anterior Eye. 2014;37(3):168–174. doi: 10.1016/j.clae.2013.09.013. [DOI] [PubMed] [Google Scholar]
  • 49.Montés-Micó R, Madrid-Costa D, Radhakrishnan H, Charman WN, Ferrer-Blasco T. Accommodative functions with multifocal contact lenses: A pilot study. Optom. Vis. Sci. 2011;88(8):998–1004. doi: 10.1097/OPX.0b013e31821c0ed8. [DOI] [PubMed] [Google Scholar]
  • 50.Madrid-Costa D, Ruiz-Alcocer J, Radhakrishnan H, Ferrer-Blasco T, Montes-Mico R. Changes in accommodative responses with multifocal contact lenses: A pilot study. Optom. Vis. Sci. 2011;88(11):1309–1316. doi: 10.1097/OPX.0b013e31822be35a. [DOI] [PubMed] [Google Scholar]
  • 51.Efron, N. Contact Lens Practice, 3rd Edition. 2016/11/27: Elsevier.

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All data generated or analyzed during this study are included in this published article and its supplementary information files. Correspondence and requests for materials should be addressed to C.-Y.C.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES