Dose reduction does not impact precision of CT-RSA in tibial components in total knee arthroplasty: a randomized controlled trial

Fredrik Bru 1, Lars H W Engseth 2,3, Are Hugo Pripp 4,5, Anselm Schulz 3,6, Tommy Frøseth Aae 1,7,8, Vigdis Schnell Husby 9, Otto Schnell Husby 1,7, Stephan M Röhrl 2,3, Frank-David Øhrn 1,7,✉
PMCID: PMC13261921  PMID: 41840645

Abstract

Purpose

Radiostereometric analysis (RSA) has been the gold standard for implant migration analysis for decades. Our group previously demonstrated in a porcine cadaver model that CT-RSA precision was non-inferior by dose reduction. The aim of this study was to investigate whether this finding applied in a clinical setting.

Methods

This study is part of the ClessTKA trial comparing a novel 3D printed uncemented medially stabilized total knee arthroplasty (TKA) with Tritanium TKA. Fifty patients were recruited (GMK Sphere n = 26, Tritanium n = 24). All surgeries were performed with mechanical alignment between January and June 2023. CT acquisitions were performed postoperatively within 2 days (standard dose 0.05 mSv, n = 49) and at 1 year (low dose 0.01 mSv, n = 47) using a GE Revolution scanner. CT-RSA analyses of tibial components were performed with Sectra CTMA software by a certified assessor. The primary endpoint was difference in precision (mean) of maximum total point motion (MTPM) between dose levels, with an equivalence interval of 0–0.1 mm.

Results

Mean difference (95% CI) in MTPM under zero motion circumstances between standard and low dose was 0.007 mm (− 0.018 to 0.032). No clinically meaningful differences were found for migration or rotations. Centre-of-mass analysis showed one significant but clinically irrelevant posterior translation (0.013 mm).

Conclusion

The findings confirm that reducing the effective radiation dose from 0.05 mSv to 0.01 mSv does not compromise the precision of CT-RSA for tibial components in TKA. The results further support the use of low-dose CT-RSA protocols in clinical studies, enabling substantial reduction in patient radiation exposure without loss of precision.

Trial registration

Clinicaltrials.gov identifier NCT05651009. Initial release of the study was November 23, 2022. At that time, the software CTMA (Sectra) did not provide the very important Maximum Total Point Motion (MTPM), this was introduced in a later update of the CTMA. Hence there was change in the original protocol retrospectively.

Introduction

Total knee arthroplasty (TKA) is a well-established treatment for end-stage knee osteoarthritis. A subset of patients experiences persistent postoperative pain and reduced function [1, 2]. Early detection of excessive implant migration is vital, as it is a strong predictor of long-term loosening and revision surgery [3]. For more than four decades, radiostereometric analysis (RSA) has been the gold standard for measuring implant migration [4–6]. Computed tomography-based RSA (CT-RSA) has emerged as a promising alternative to conventional RSA, offering comparable precision in migration measurements for hip, shoulder and knee implants [7–10]. Unlike RSA, CT-RSA does not require tantalum markers, specialized imaging setups, computer aided design or reversed engineered models. Furthermore, it can be performed using standard CT scanners available in most hospitals. A key limitation, however, has been concerns about the higher radiation dose associated with repeated CT imaging compared to RSA, given the established dose-response relationship between ionizing radiation and cancer risk [11].

In a recent porcine cadaver study on tibial implants, we demonstrated that CT-RSA performed with a 75% reduction in radiation dose maintained precision comparable to that of a standard-dose protocol [12]. The present study aims to validate these findings in vivo by evaluating the precision of low-dose CT-RSA in patients undergoing primary TKA. We hypothesized that low-dose CT-RSA maintained clinically equivalent precision to standard-dose CT-RSA, thereby supporting its use in clinical migration studies while minimizing patient radiation exposure.

Methods and materials

Study design and setting

This study was conducted as part of the double-blinded randomized controlled trial “Comparison of the in vivo stability of 2 cementless TKA designs using CT micromotion analysis - A randomized controlled trial” (ClessTKA), investigating implant migration in TKA [13]. The trial compared a novel 3D-printed uncemented medially stabilized TKA (GMK Sphere 3D metal, Medacta International, Switzerland) with a well-documented 3D-printed uncemented cruciate-retaining TKA (Triathlon Tritanium CR, Stryker, Mahwah, USA) [14, 15]. The sample size calculation was thus based on the migration analyses and the expected difference in migration between the implants, rather than on the precision of the measurement method used. The surgeries were performed at Kristiansund Hospital, Møre and Romsdal Hospital Trust, Norway, between January and June 2023. All CT acquisitions were performed at the same location between January 2023 and June 2024. A user representative that was not part of the study was appointed and had the opportunity to comment on the protocol.

Participants

50 patients scheduled for primary TKA due to end-stage osteoarthritis (Kellgren & Lawrence 3–4) (Table 1) were enrolled in the ClessTKA, unless they met the exclusion criteria of the study (Table 2) [13]. Patients were randomized into a study group (GMK Sphere, n = 26) and a control group (Triathlon Tritanium CR, n = 24) using the EFORSK software (Unit for Applied Clinical Research, NTNU, Norway) with stratification by sex and varying block sizes. Randomization was performed no more than 1 week before surgery by the project leader (FDØ) and concealed until allocation. FDØ had no access to, and could thus not affect the randomization algorithm in any way. All procedures were performed with concomitant patella resurfacing, using mechanical alignment and without tourniquet, by two experienced orthopaedic surgeons (FDØ/OSH) consistently working together. All complications were accounted for.

Table 1.

Demographics of the patients included in the study

Variables (n = 50) Tritanium (n = 24) GMK Sphere (n = 26)
Age (years, mean, 95% CI) 64.2 (61.1–67.3) 66.8 (64.5–69.2)
HKA (°, 95% CI) 175.4 (172.7–178.1.7.1) 176.1 (174.0–178.2.0.2)
BMI (kg/m2, 95% CI) 29.4 (27.8–31.0) 28.3 (26.8–29.9)
Left/right (n) 12/12 11/15
Male/female (n) 10/14 11/15
Implant (n) 24 26

CI=Confidence Interval, HKA = Hip-Knee-Ankle angle, BMI=Body Mass Index

Table 2.

Exclusion criteria of the ClessTKA study

Exclusion criteria
Preoperative severe deformity (HKA ≥ 5° valgus or > 15° varus) on full-length weight-bearing radiograph
Preoperative flexion contracture > 15°

Age < 50 or > 75 years at time of surgery

Use of walking aids due to other musculoskeletal or neuromuscular problems

Preoperative diagnosis other than OA or avascular necrosis
Revision arthroplasty
BMI > 35
Lateral collateral ligament-deficient knee
Previous knee joint infection

HKA = Hip-Knee-Ankle angle, OA=Osteoarthritis, BMI=Body Mass Index

CT acquisitions

Double CT acquisitions were performed at two postoperative time points using two different effective doses [16]:

  1. Baseline—within 2 days of surgery (standard dose: tube current 160 mA, rotation time 0.5 s, scan length 180 mm, tube voltage 120 kV, effective dose 0.05 mSv, n = 49).

  2. Follow-up—at 1 year postoperatively (low dose: tube current 40 mA, rotation time 0.5 s, scan length 180 mm, tube voltage 120 kV, effective dose 0.01 mSv, n = 47).

All scans were obtained using a GE Revolution CT scanner (GE Healthcare, Chicago, USA). Reconstruction was performed with a Bone Plus metal artefact reduction (MAR) algorithm. Between each double scan, patients moved their knee and the scanner was reset (Fig. 1).

Fig. 1.

Fig. 1

Flow chart of the study. BMI=Body Mass Index, HKA = Hip-Knee-Ankle angle, RA= Rheumatoid Arthritis

CT-RSA analysis

CT-RSA migration measurements were performed using the CT Micromotion Analysis (CTMA) software (Sectra AB, Linköping, Sweden) with Hounsfield Unit (HU) values of 150 for bone and 3000 for metal. The bone was marked selectively, whereas the tibial implants were marked using the double click function of the software (Fig. 2). The standard Digital Imaging and Communication in Medicine (DICOM) coordinate system was used. All multiplanar axes were the same and were set before the analysis. Peripheral points named tip, anterior, posterior, medial and lateral were created (Fig. 3). For each tibial component, the following were recorded:

Fig. 2.

Fig. 2

Double click marking of the tibial implant and selective bone marking of the tibia

Fig. 3.

Fig. 3

Peripheral points: tip, anterior, posterior, medial and lateral. Multiplanar reconstruction axis set before each analysis

Primary endpoint Maximum total point motion (MTPM), a vector measured in millimetres representing the point of the implant moving the most between the time points [4].

Secondary endpoints centre-of-mass (COM) and peripheral points translations and rotations in medial-lateral, proximal-distal and anterior-posterior directions.

Analyses were performed by a CTMA-certified author (FDØ) who was not blinded to implant nor dosage level. All other authors, the patient, study nurse and biostatistician remained blinded during data collection and analysis.

Statistical analysis

Analyses were performed using Stata version 17 (StataCorp LLC, College Station, TX, USA). Means and 95% confidence intervals (95% CI) were calculated for the variables for standard- and low-dose protocols, and named the precision. The primary outcome measure was the difference in MTPM precision between standard- and low-dose CT-RSA. An equivalence margin of 0–0.10 mm was predefined [4, 12, 17], with differences exceeding this considered clinically relevant. Linear mixed models were used, with dose level as a fixed factor and patient ID as a random factor. Standard deviations (SD) of the variables of standard- and low-dose protocols were also calculated. A two-sample F test for equality of variances using the Stata code sdtest, was conducted. Statistical significance was defined as p < 0.05.

Results

49 patients completed the baseline (standard dose) and 47 patients the 1-year (low dose) CT acquisitions and were thus included in the precision analysis (Fig. 1). The comparison between standard-dose and low-dose CT-RSA showed a mean difference in MTPM precision of 0.007 mm (95% CI −0.018 to 0.032, p = 0.576), well within the predefined equivalence margin of 0–0.10 mm (Figs. 4 and 5). The mean MTPM was 0.117 (0.10 to 0.12) mm for the standard dose, and 0.124 (0.11 to 0.14) mm for the low dose; the SD of both was 0.12 mm. The SD test gave a p value of 0.452. No statistically significant differences in MTPM precision were found between the two implant groups (GMK Sphere vs. Tritanium), with a mean difference of 0.006 mm (95% CI −0.019 to 0.030). Mean and mean difference of segmental translations and rotations, and peripheral points are shown in Table 3.

Fig. 4.

Fig. 4

MTPM shown in boxplot format. Examination (low-dose and standard-dose) on the x-axis and MTPM (mm) on the y-axis

Fig. 5.

Fig. 5

Difference if MTPM shown in Bland-Altman plot. Mean on the x-axis and difference on the y-axis

Table 3.

Mean and mean difference of segmental translations and rotations, and peripheral points

Variable Mean standard dose (mm) Standard deviation Mean low dose (mm) Standard deviation Mean difference (mm) 95% CI
Segmental translation, mm
Medial 0.00 (−0.01 to 0.01) 0.02 0.00 (−0.01 to 0.01 0.03 −0.001 0.00 to 0.01
Proximal 0.00 (−0.01 to 0.02) 0.05 0.01 (−0.01 to 0.02 0.05 0.002 −0.02 to 0.02
Posterior −0.01 (−0.02 to −0.00) 0.02 0.00 (−0.01 to 0.01) 0.03 0.013 0.00 to 0.02
Segmental rotation, mm
Transversal 0.04 (0.00 to 0.01) 0.13 0.01 (−0.02 to 0.05) 0.13 −0.027 −0.08 to 0.02
Internal −0.01 (−0.03 to 0.01) 0.06 0.00 (−0.02 to 0.02) 0.07 0.012 −0.01 to 0.04
Varus −0.00 (−0.02 to 0.02) 0.06 0.00 (−0.01 to 0.02) 0.07 0.005 −0.02 to 0.03
Translations of the peripheral points, mm
Tip
Medial 0.00 (−0.01 to 0.01) 0.04 −0.00 (−0.01 to 0.01) 0.04 0.04 −0.02 to 0.01
Proximal −0.00 (−0.02 to 0.01) 0.05 0.00 (−0.01 to 0.02) 0.05 0.005 −0.01 to 0.02
Posterior 0.01 (0.00 to 0.03) 0.06 0.01 (−0.01 to 0.02) 0.05 −0.004 −0.03 to 0.02
Medial
Medial 0.00 (−0.01 to 0.01) 0.02 0.00 (−0.01 to 0.05) 0.03 0.000 −0.01 to 0.01
Proximal 0.00 (−0.02 to 0.01) 0.05 0.01 (−0.01 to 0.02) 0.06 0.006 −0.01 to 0.03
Posterior −0.01 (−0.03 to 0.00) 0.04 0.00 (−0.02 to 0.01) 0.06 0.011 −0.01 to 0.03
Lateral
Medial 0.00 (−0.01 to 0.01) 0.02 0.00 (−0.01 to 0.01) 0.03 0.001 −0.01 to 0.01
Proximal 0.00 (−0.02 to 0.02) 0.05 0.00 (−0.02 to 0.02) 0.07 0.000 −0.03 to 0.03
Posterior −0.01 (−0.03 to 0.00) 0.07 0.01 (−0.01 to 0.02) 0.05 0.020 0.00 to 0.04
Anterior
Medial 0.00 (−0.01 to 0.00) 0.03 0.00 (−0.01 to 0.01) 0.03 0.002 −0.01 to 0.01
Proximal −0.01 (−0.03 to 0.01) 0.07 0.00 (−0.02 to 0.02) 0.07 0.013 −0.01 to 0.04
Posterior −0.01 (−0.02 to −0.00) 0.03 0.00 (−0.01 to 0.01) 0.04 0.014 0.00 to 0.03
Posterior
Medial 0.00 (0.00 to 0.01) 0.03 0.00 (−0.01 to 0.01) 0.05 −0.004 −0.02 to 0.01
Proximal 0.02 (0.00 to 0.04) 0.07 0.01 (−0.01 to 0.02) 0.07 −0.013 −0.04 to 0.01
Posterior −0.01 (−0.02 to 0.00) 0.03 0.00 (−0.01 to 0.01) 0.04 0.014 −0.00 to 0.03

Complications

One patient in the study group underwent revision due to a deep postoperative infection and was therefore excluded from further analysis at 1 year. In the control group, 1 patient required debridement and irrigation for a postoperative haematoma. Postoperative stiffness requiring manipulation under anaesthesia occurred in 1 patient in the study group and in 2 patients in the control group. Minor wound problems and removal of palpable retained sutures from the knee capsule were noted in a few patients in both groups; all resolved without further intervention.

Missing data

Double examinations with the standard dose were performed postoperatively, and those with the low dose at 1 year. The patient who developed infection and was excluded from the study lacked low-dose data. In addition, one patient had missing double examinations postoperatively, and two patients at the 1-year follow-up. Details are presented in Fig. 1.

Discussion

The main finding of this clinical study is that a 75% reduction in effective radiation dose, from 0.05 to 0.01 mSv, does not affect the precision of CT-RSA for tibial components in TKA. These results confirm our previous findings from a porcine cadaver model in a clinical setting, and indicate that low-dose CT-RSA can be implemented without compromising measurement reliability [12, 17].

Our equivalence margin of 0–0.10 mm was based on established thresholds in RSA literature, where increased migration of more than 0.2 mm between 1 and 2 years predicts a higher risk of aseptic loosening [4, 12, 17]. The observed mean difference in MTPM precision (0.007 mm) was far below this margin and statistically non-significant. This supports the clinical feasibility of low-dose CT-RSA for migration monitoring in TKA [12].

The precision values observed in this study are comparable to, or better than, those reported for conventional RSA in tibial components [7]. CT-RSA offers several practical advantages: it does not require tantalum markers, specialized calibration cages, or CAD models, while CT scanners are widely available in hospitals. Furthermore, analysis is less dependent on radiographic positioning and marker visibility, both of which are potential limitations in RSA. The precision described in this study matches the level reported in the only other published study of clinical CT-RSA performed on TKA [18], in which V3MA, not CTMA software, was used [18, 19].

Radiation exposure has been a main argument against the adoption of CT-RSA in clinical research. However, the effective dose in our low-dose protocol (0.01 mSv) is similar to that of a standard RSA examination (0.10–0.15 mSv), especially when accounting for the repeated exposures sometimes required in RSA due to inadequate marker visibility [7]. Given the well-documented dose-response relationship between ionizing radiation and cancer risk, even modest dose reductions are desirable [11].

The only statistically significant difference we observed was a small increase in posterior COM translation at low dose (0.013 mm), which is well within the range of measurement error and not clinically relevant. This finding may reflect normal variability between repeated acquisitions rather than a systematic dose-related effect.

In a recent publication from the ClessTKA study using the same patient population, we investigated the precision of CT-RSA on femoral implants [13]. Femoral and tibial components have different geometrical shapes that may affect the precision differently. Because of this, CT-RSA of femoral components have acceptable, though lower precision than the findings in the present study [13]. In TKA, the tibial component is more prone to loosening than the femoral component. Establishing the precision of CT-RSA in tibial components at different dosage levels in a clinical study therefore represent a huge and important step towards proving that CT-RSA is a tool that can be used in migrations studies of TKAs.

Strengths and limitations

A major strength of this study is the randomized controlled design embedded in an ongoing clinical trial, ensuring high methodological quality and generalizability. To our knowledge, this is the first randomized controlled trial assessing the precision of CT-RSA on tibial components. Previous studies are either cadaver studies, or single-group studies with the same dosage level throughout the study [18–20]. All analyses were performed by a CTMA-certified observer (FDØ). Although the observer was not blinded, the patients, study nurse, statistician and first author (FB) were blinded for both implant type and dosage level, thus minimizing bias. Furthermore, the use of two implant designs enabled the assessment of dose effects in different geometries. In addition, our study was performed in accordance with existing recommendations for migration analysis [21–23].

Limitations include the single-centre setting and a relatively small sample size (n = 50), which may have limited the detection of very small differences in precision. However, this is still by far the largest study on CT-RSA on TKA to date [18, 20]. 50 patients is usually considered sufficient for a randomized controlled trial [24]. In addition, it was a precision study with double examinations under the assumption of zero motion at two follow-up points. The accuracy of CT-RSA was not measured. The fact that the double acquisitions for the low-dose and standard-dose CT-RSA protocols were performed at two different time points may introduce potential confounding. Nevertheless, each double examination was conducted within only a few minutes, and the assumption of negligible true implant motion between acquisitions is therefore considered valid, despite the two protocols being performed 12 months apart. This design was chosen to simplify the CT workflow for radiographers and thereby ensure feasibility in routine clinical practice. The published Guideline for RSA and CT-RSA implant migration measurements recommend double acquisitions in clinical studies performed at the 1-year control or earlier to prevent potential dropouts [23]. Thus, this approach is not assumed to compromise the narrow confidence intervals shown in the results, and the results should be interpreted as reliable. The original protocol was changed after the study commenced, because the CTMA software originally did not provide MTPM. In November 2022 when the initial registration in clinicaltrials.gov was performed, we could only assess the maximum translation of the 5 predefined peripheral points (maximum total translation, MTT), and not the MTPM [12, 25]. In a later update of the software, it was possible to calculate the actual MTPM. Finally, while the radiation dose was reduced to 0.01 mSv, which is the lowest value that can be recorded by our CT scanner, further optimization could be explored, including iterative reconstruction techniques and alternative scanning parameters.

Clinical implications

Our findings indicate that low-dose CT-RSA can be adopted in clinical migration studies without loss of precision, and thereby reducing patient radiation exposure. Given its practical advantages and comparable or superior precision to RSA, CT-RSA with optimized low-dose protocols may become the preferred method for implant migration assessment in future clinical trials.

Conclusion

The findings confirm that reducing the effective radiation dose from 0.05 mSv to 0.01 mSv does not compromise the precision of CT-RSA for tibial components in TKA. The results further support the use of low-dose CT-RSA protocols in clinical studies, enabling substantial reduction in patient radiation exposure without loss of precision.

Acknowledgements

We would like to thank research coordinator Myrthle Slettvåg Hoel and the Radiology Department at Kristiansund Hospital, especially radiographer Torbjørn Malm, for their invaluable contribution to this study. We would also express our sincere gratitude to the scrub nurses at the OR of Kristiansund Hospital. Without their efforts, the study would not have been possible.

Author contributions

Fredrik Bru: Writing – original draft, methodology, formal analysis, data curation, review & editing, conceptualization Lars H.W. Engseth: Review & editing, methodology, conceptualization. Are H. Pripp: Review & editing. methodology, formal analysis, data curation, conceptualization. Tommy Frøseth Aae: Review & editing, methodology, conceptualization. Anselm Schulz: Review & editing, methodology, conceptualization. Otto Schnell Husby: Review & editing, methodology, funding acquisition, conceptualization. Vigdis Schnell Husby Review & editing, methodology, conceptualization. Stephan M. Röhrl: Review & editing, methodology, conceptualization. Frank-David Øhrn: Writing – review & editing, methodology, funding acquisition, formal analysis, data curation, original idea, conceptualization, supervision.

Funding

Open access funding provided by NTNU Norwegian University of Science and Technology (incl St. Olavs Hospital - Trondheim University Hospital). The study was investigator-initiated and funded by Møre and Romsdal Hospital Trust and Medacta International. NorTrials MU (St. Olavs Hospital, Trondheim, Norway) served as the contractual partner to safeguard investigator independence and ensure transparency in line with the recommendations of the International Committee of Medical Journal Editors (ICMJE) [26]. The authors declare no conflicts of interest relevant to this work.

Data availability

The datasets generated and/or analysed during the current study are not publicly available but are available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

Service user involvement was incorporated into the study design through the appointment of a service user representative who was not enrolled as a study participant. The study received ethical approval from the Regional Committee for Medical and Health Research Ethics in Central Norway (reference number 520246) as well as from the Data Access Committee of Møre and Romsdal Hospital Trust. The trial was registered at ClinicalTrials.gov (NCT05651009) and conducted in accordance with the CONSORT guidelines [27] and the Helsinki Declaration [28]. Prior to inclusion in the study, all participants were provided with both written and oral information and gave their written informed consent.

Consent for publication

No identifying images or other personal or clinical details of participants are presented that may compromise anonymity. Prior to submittance, alle authors have read the manuscript and approved the publication.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Baker PN, van der Meulen JH, Lewsey J, Gregg PJ. The role of pain and function in determining patient satisfaction after total knee replacement. Data from the National Joint Registry for England and Wales. J Bone Joint Surg Br. 2007;89(7):893–900. [DOI] [PubMed] [Google Scholar]
  • 2.Beswick AD, Wylde V, Gooberman-Hill R, Blom A, Dieppe P. What proportion of patients report long-term pain after total hip or knee replacement for osteoarthritis? A systematic review of prospective studies in unselected patients. BMJ Open. 2012;2(1):e000435. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Gunaratne R, Pratt DN, Banda J, Fick DP, Khan RJK, Robertson BW. Patient dissatisfaction following total knee arthroplasty: a systematic review of the literature. J Arthroplasty. 2017;32(12):3854–60. [DOI] [PubMed] [Google Scholar]
  • 4.Ryd L, Albrektsson BE, Carlsson L, Dansgard F, Herberts P, Lindstrand A, et al. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J bone joint Surg Br volume. 1995;77(3):377–83. [PubMed] [Google Scholar]
  • 5.Ryd L. Roentgen stereophotogrammetric analysis of prosthetic fixation in the hip and knee joint. Clin Orthop Relat Res. 1992;276:56–65. [PubMed] [Google Scholar]
  • 6.Selvik G. Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand Suppl. 1989;232:1–51. [PubMed] [Google Scholar]
  • 7.Engseth LHW, Schulz A, Pripp AH, Röhrl SMH, Øhrn FD. CT-based migration analysis is more precise than radiostereometric analysis for tibial implants: a phantom study on a porcine cadaver. Acta Orthop. 2023;94:207–14. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Brodén C, Giles JW, Popat R, Fetherston S, Olivecrona H, Sandberg O, et al. Accuracy and precision of a CT method for assessing migration in shoulder arthroplasty: an experimental study. Acta Radiol. 2020;61(6):776–82. [DOI] [PubMed] [Google Scholar]
  • 9.Brodén C, Reilly P, Khanna M, Popat R, Olivecrona H, Griffiths D, et al. CT-based micromotion analysis method can assess early implant migration and development of radiolucent lines in cemented glenoid components: a clinical feasibility study. Acta Orthop. 2022;93:277–83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Brodén C, Sandberg O, Sköldenberg O, Stigbrand H, Hänni M, Giles JW, et al. Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups: a clinical study of 24 patients. Acta Orthop. 2020;91(3):260–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, Hanson R, et al. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology. 2009;251(1):175–84. [DOI] [PubMed] [Google Scholar]
  • 12.Øhrn FD, Engseth LHW, Pripp AH, Röhrl SMH, Schulz A. Dose reduction does not impact the precision of CT-based RSA in tibial implants: a diagnostic accuracy study on precision in a porcine cadaver. Acta Orthop. 2023;94:550–44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Øhrn FD, Engseth LHW, Pripp AH, Frøseth Aae T, Schulz A, Schnell Husby O, et al. Computed tomography-based radiostereometric analysis achieves sufficient precision for femoral implants in total knee arthroplasty: a randomized controlled trial. J Orthop Surg Res. 2025;20(1):1043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Hasan S, Hamersveld K, Mheen P-v, Kaptein BL, Nelissen RGHH, Toksvig-Larsen S. Migration of a novel 3D-printed cementless versus a cemented total knee arthroplasty: two-year results of a randomized controlled trial using radiostereometric analysis. Bone Joint J. 2020;102-B(8):1016–24. [DOI] [PubMed] [Google Scholar]
  • 15.van der Lelij TJN, van de Marang- Mheen PJ, Kaptein BL, Toksvig-Larsen S, Nelissen R. Continued stabilization of a cementless 3D-printed total knee arthroplasty: five-year results of a randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2023. 10.2106/JBJS.23.00221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Saltybaeva N, Jafari ME, Hupfer M, Kalender WA. Estimates of effective dose for CT scans of the lower extremities. Radiology. 2014;273(1):153–9. [DOI] [PubMed] [Google Scholar]
  • 17.Øhrn F-D, Engseth LHW, Pripp AH, Schulz A, Röhrl SMH. Precision of computer tomography based RSA on femoral implants in total knee arthroplasty: a porcine cadaver study. Clin Biomech. 2025;123:106456. [DOI] [PubMed] [Google Scholar]
  • 18.Hext RA, Kaptein BL, Howard JL, Lanting BA, Teeter MG. Inducible displacement of cementless total knee arthroplasty components with conventional and weight-bearing CT-based radiostereometric analysis. J Orthop Res. 2024. 10.1002/jor.26017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.de Laat NN, Koster LA, Robertson JE, Nelissen RGHH, Kaptein BL. Clinical validation and excellent interobserver agreement of volumetric matching micromotion analysis (V3MA) in total knee arthroplasty. J Orthop Res. 2026. 10.1002/jor.70049. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Engseth LHW, Øhrn FD, Schulz A, Röhrl SM. CT-based migration analysis in a tibial implant compared to Radiostereometric analysis – 1-year results of a single-group implant safety study. In press. 10.1302/0301-620X.107B2.BJJ-2024-0356.R2. The bone & joint journal. 2025. [DOI] [PubMed]
  • 21.Valstar ER, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76(4):563–72. [DOI] [PubMed] [Google Scholar]
  • 22.Sandberg OH, Kärrholm J, Olivecrona H, Röhrl SM, Sköldenberg OG, Brodén C. Computed tomography-based radiostereometric analysis in orthopedic research: practical guidelines. Acta Orthop. 2023;94:373–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Kaptein BL, Pijls B, Koster L, Kärrholm J, Hull M, Niesen A, et al. Guideline for RSA and CT-RSA implant migration measurements: an update of standardizations and recommendations. Acta Orthop. 2024;95:256–67. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Derbyshire B, Prescott RJ, Porter ML. Notes on the use and interpretation of radiostereometric analysis. Acta Orthop. 2009;80(1):124–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Øhrn FD, Engseth LHW, Pripp AH, Schulz A, Röhrl SMH. Precision of computer tomography based RSA on femoral implants in total knee arthroplasty: a porcine cadaver study. Clin Biomech (Bristol). 2025;123:106456. [DOI] [PubMed] [Google Scholar]
  • 26.No authors listed. International Committee of Medical Journal Editors. Available at https://www.icmje.org/ (Date assessed October 29 2025).
  • 27.No authors listed. Consort Transparent Reporting of Trials. Available at http://www.consort-spirit.org (Date assessed October 29 2025).
  • 28.No authors listed. World Medical Association Declaration of Helsinki-Ethical principles for medical research involving human subjects: https://www.wma.net/policies-post/wma-declaration-of-helsinki-ethical-principles-for-medical-research-involving-human-subjects/ (date assessed DEcember 29 2025).

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets generated and/or analysed during the current study are not publicly available but are available from the corresponding author on reasonable request.


Articles from Journal of Orthopaedic Surgery and Research are provided here courtesy of BMC

RESOURCES