Skip to main content
eLife logoLink to eLife
. 2026 Aug 24;15:RP111639. doi: 10.7554/eLife.111639

Optimising the tilt increment for in situ cryo-electron tomography

Maarten Willem Tuijtel 1, Tomáš Majtner 1, Beata Turoňová 1,2, Martin Beck 1,2,3,
Editors: Axel T Brunger4, Volker Dötsch5
PMCID: PMC13502964  PMID: 42636038

Abstract

Cryo-electron tomography (cryo-ET) enables high-resolution, three-dimensional imaging of cellular structures in their native, frozen state. However, image quality is limited by a trade-off between angular sampling and radiation damage. Therefore, the choice of the angular increment during data collection is a critical parameter that affects tomogram quality and downstream analyses. Optimising this increment is challenging due to the high demands on microscope time, storage, and computation. In this study, we systematically evaluated tilt increments of 1°, 2°, 3°, 5°, and 10° using lamellae from Dictyostelium discoideum cells. We found that at a constant total electron dose, finer tilt increments (1–3°) produced better-aligned tomograms with higher signal-to-noise ratios and improved outcomes in template matching and subtomogram averaging. A 3° increment emerged as the optimal balance between data quality, alignment accuracy, dose per image, and processing efficiency. This practical recommendation supports both high-throughput and high-resolution structural studies and can guide future cryo-ET data acquisition strategies.

Research organism: Dictyostelium

Introduction

Cryo-electron tomography (cryo-ET) (Förster and Briegel, 2024) is a powerful technique to study virions (Turoňová et al., 2020b), thin bacterial cells (Xue et al., 2022a), isolated organelles, larger eukaryotic cells (Hoffmann et al., 2022; Kreysing et al., 2025; Schiøtz et al., 2024), and even tissue (Glynn et al., 2025; Klumpe et al., 2025) with molecular resolution. It captures the three-dimensional (3D) electrostatic potential of the specimen under scrutiny and enables the structural analysis of macromolecular complexes within their native context. However, it is also limited by a number of technical parameters. One of these is the maximum electron dose that can be applied to biological specimens before irreversible damage occurs, which is about 100–150 e-2 for most eukaryotic cells (Grant and Grigorieff, 2015; Xue et al., 2022b).

For simplicity, the cryo-ET workflow can be described in terms of an experimental and two subsequent computational steps, which are relevant to this study. (i) Multiple two-dimensional (2D) projection images are acquired that capture the respective biological specimen from different angles. (ii) These individual projections, which are also referred to as tilt images, need to be computationally aligned to correct for spatial displacements that occur due to mechanical imperfections of the microscope stage. (iii) From the resulting aligned tilt series of projection images, a three-dimensional (3D) reconstruction is computed.

The maximum electron dose influences all three of these steps. (i) The dose has to be distributed over all tilt images and therefore the angular increment is a key experimental parameter for the data acquisition. (ii) The dose contained in an individual projection determines how well it can be computationally aligned with its neighbouring images. Both (i) the data acquisition and (ii) the alignment of images impact on the quality of (iii) the reconstruction: The smaller the angular increment used during data acquisition, the lower the signal in each individual tilt image, but the finer the angular sampling of the resulting 3D reconstruction. Inversely, the more accurate the alignment of the tilt images, the better the accuracy of the reconstruction. Since both parameters influence the outcome of the cryo-ET workflow in an interdependent way, it is difficult to make predictions about the optimal tilt increment.

These considerations have led to the notion that there is a trade-off between the number of tilts that can be collected and the electron dose that can be allocated per tilt image. This caveat needs to be considered for every cryo-ET data acquisition session, and different tomographic acquisition schemes have been proposed to deal with this phenomenon (Hagen et al., 2017; Sanchez et al., 2020; Saxton et al., 1984; Song et al., 2020; Zheng et al., 2004). Since accumulated radiation dose progressively degrades high-resolution information, this motivated the development of the dose-symmetric tilt scheme, which prioritises acquisition of low-tilt images early to better preserve high-resolution information (Hagen et al., 2017; Turoňová et al., 2020a). However, to the best of our knowledge, this trade-off has never been systematically addressed and empirically validated, possibly due to the extensive experimental and computational effort associated with such investigations. The respective decision regarding acquisition parameters is thus often made on an arbitrary basis.

The alignment is thus a critical step in the cryo-ET workflow, and it is typically performed in two stages. First, coarse alignment is achieved by cross-correlating the entire projection images to correct for large lateral shifts introduced during image acquisition. Subsequently, the images are subdivided into overlapping patches, and local cross-correlations are computed between adjacent projection images to estimate relative shifts. These patch-derived shifts are then refined using a least-squares-fitting procedure to determine optimal alignment parameters (Mastronarde and Held, 2017).

The challenges during the 3D reconstruction can be explained using the central slice theorem (Radon, 1917), which states that the Fourier transform of each 2D projection (tilt image) corresponds to a central slice through the 3D Fourier transform of the final tomogram. This illustrates the sampling challenge that is inherent to cryo-ET. The sample thickness (D) (or the resulting thicknesses of all central slices in Fourier space), and the finite number of 2D projections (n) that are used for the reconstruction, result in regions between the tilts that also remain incompletely sampled beyond a defined spatial frequency. The Crowther criterion conceptualises this phenomenon and provides a means to calculate the maximum attainable resolution (d) based on angular sampling (Crowther et al., 1970; Frangakis, 2024), with

d=πDn.

Similarly, the signal-to-noise ratio (SNR) in 3D reconstructions is described by the dose-fractionation theorem (Hegerl and Hoppe, 1976), which states that the statistical significance of each reconstructed voxel depends solely on the total applied electron dose, irrespective of the number of projection images, provided that perfect alignment between the projections is achieved (McEwen et al., 1995). Thus, when the total dose is kept constant, the SNR in the tomogram depends on the accuracy of the alignment.

Tomograms are often further data mined by locating specific proteins or protein complexes in the crowded cellular environment, using, for example, template matching (TM) approaches (Böhm et al., 2000; Cruz-León et al., 2024), or alternatively, machine-learning based feature recognition (de Teresa-Trueba et al., 2023; Moebel et al., 2021). For in situ structural analysis, the respective particles are extracted, aligned, and averaged, a process called subtomogram averaging (STA) (Castaño-Díez and Zanetti, 2019; Turoňová and Wan, 2024c). These processing routines are critically important to understand the function of large macromolecular assemblies inside cells (Beck and Baumeister, 2016; McCafferty et al., 2024). Therefore, the outcome of TM and STA in terms of accuracy or resolution, respectively, has been used as a metric to score the technical quality of tomograms and the suitability of data acquisition schemes for specific applications (Tuijtel et al., 2024). During STA processing, the alignment of the projections can be refined on a per-particle basis (Bartesaghi et al., 2012; Burt et al., 2024; Chen et al., 2019; Himes and Zhang, 2018; Khavnekar et al., 2021; Tegunov et al., 2021), which is instrumental for achieving high-resolution structures in the sub-4 Å-range (Hoffmann et al., 2022; Tegunov et al., 2021; Xing et al., 2023; Xue et al., 2022b).

According to the above-discussed notion that there should be a trade-off between angular sampling and radiation damage, we reasoned that there should be an optimal balance between angular sampling and tilt image alignment and, thus, tomogram quality. To address this, we acquired data on lamellae of Dictyostelium discoideum cells using various tilt increments, both typical for the field and beyond. We utilised TM and STA as metrics to determine the optimal tilt increment for in situ cryo-ET. Surprisingly, we found that tilt-series alignment improves with smaller angular increments, which, in line with the dose-fractionation theorem, benefits the SNR of the tomograms, despite the decreasing signal in the individual tilt images. Tomographic reconstruction, TM, and STA were successful for all imaging conditions; however, visual tomogram quality, TM accuracy and STA resolution suffered from the use of higher increments. An increment of 3° emerged as the optimal balance between data quality, alignment accuracy, dose per image, and processing efficiency. The results described herein can serve as a guide for the community to choose data acquisition parameters for cryo-ET data collection.

Results

A dataset for benchmarking tilt increments in cryo-electron tomography

To study the effect of the tilt increment, we collected several cryo-ET datasets using SerialEM (Mastronarde, 2005) on lamellae from D. discoideum cells, varying the tilt increment from 1°, 2°, 3°, 5°, and 10° using the dose-symmetric tilt-scheme (Hagen et al., 2017; for details on the sample preparation and microscopy data acquisition, see ‘Methods’). During acquisition, the tilt range (from –60° to +60° relative to the lamella plane) and the total dose (120–130 e-2) were kept constant for each condition. As a consequence, the number of images in a tilt series and the dose per tilt image were varied for each condition (for full details regarding data acquisition parameters, see Table 1). Both data storage requirements and beam time needed per tomogram differed by nearly an order of magnitude (Table 1). After acquisition, tilt series were pre-processed, aligned using patch-tracking, and tomograms were reconstructed in IMOD (Mastronarde and Held, 2017; see ‘Methods’). For most datasets, image acquisition was largely complete, with the exception of the 1° dataset, which showed a markedly higher proportion of missing images at high tilt angles (Figure 1—figure supplement 1). Closer inspection revealed that many of these images were not acquired, as SerialEM applies built-in safeguards (e.g. autofocus inconsistency or insufficient image counts) that can abort a tilt-series branch before completion.

Table 1. Datasets and acquisition parameters used in this study.

Tilt increment Number of tilts Total dose (e-2) Dose/tilt (e-2) Acquisition time (min) File size Tilt series acquired Pixel size (Å)
121 120–130 1.1 65 3.8 Gb 64 1.971
61 120–130 2.0 25 2.0 Gb 113 1.971
3° 41 120–130 3.2 20 1.3 Gb 128 1.971
25 120–130 4.8 14 800 Mb 122 1.971
10° 13 120–130 9.2 9 417 Mb 89 1.971

Reconstructions visually suffer from higher increments

Representative tilt images, slices through reconstructions, and the respective Crowther criteria are visualised in Figure 1, Figure 1—figure supplement 2, and Figure 1—videos 1–10. All tomograms were successfully reconstructed, even with the unusually large angular increment of 10°. However, in reconstructions obtained from the 5° and 10° tilt-increment data, membranes were observed to extend outside of the lamella body (white arrowheads in Figure 1C). Furthermore, a loss in tomogram contrast that was primarily noticeable in the XZ-slices through the volumes was apparent (Figure 1C). In particular, tomograms with 10° increment showed severe artefacts surrounding highly curved membranes, and membranes extending far outside of the lamellae (see Figure 1—figure supplement 3). The Crowther criterion (Figure 1D) extended from 2.3 nm in the 1° data to 30.4 nm in the 10° data, but also depended on the local lamella thickness in the individual tomogram.

Figure 1. Representative tilt images and tomograms at varying tilt increment.

(A) Single projection images of the sample at effective zero-tilt from a tilt series acquired with a tilt increment of 1° (i); 2° (ii); 3° (iii); 5° (iv); and 10° (v). (B) Central XY slices through the tomograms reconstructed from the tilt series shown in (A). (C) Central XZ slices through the tomograms shown in (B). White arrowheads point to membranes that appear to extend outside of the lamella. (D) Power spectra of representative tomograms for all five conditions. Yellow circles represent the Crowther criterion for the tomograms shown in (B) and (C), with the numerical value shown as text in yellow. Scalebar is 200 nm and applies to all relevant panels.

Figure 1.

Figure 1—figure supplement 1. Overview of acquired, removed and remaining tilt images for each dataset.

Figure 1—figure supplement 1.

For each dataset, the percentage of tilt images is shown per category: images contributing to the final tomogram (green), images that were removed due to drift, obstruction or other causes (red), and images that were not acquired due to SerialEM’s automated abort criteria (orange).
Figure 1—figure supplement 2. Overview of distinct features in tilt-images and tomogram slices.

Figure 1—figure supplement 2.

(A–E) Columns represent the 1° (A), 2° (B), 3° (C), 5° (D), and 10° (E) datasets. Within each column, the central tilt-image is shown on the left and a slice through the tomogram at the same position is shown on the right. The rows represent different features, from top to bottom: (i) ribosomes, (ii) mitochondria outer membrane, (iii) microtubule, (iv) nuclear pore complex, and (v) VAULT protein. Scalebar is 50 nm and applies to both panels.
Figure 1—figure supplement 3. Artefacts in tomograms acquired with 10° tilt increment.

Figure 1—figure supplement 3.

(A) XY slice of a tomogram, near the lamella surface, reconstructed from a tilt series with a 10° increment. Arrowheads point to artefacts in the reconstruction in the region close to strongly curved mitochondrial membranes. (B) Central XZ slice of the same tomogram shown in (A), reconstructed with extended Z-height. The region where the lamella is positioned is indicated with yellow lines. Membranes that extend far beyond the lamella are indicated with white arrowheads. Scalebar is 200 nm and applies to both panels.
Figure 1—video 1. Representative tilt series acquired with a tilt increment of 1°, as depicted in Figure 1.
Download video file (10.1MB, mp4)
Figure 1—video 2. Representative tilt series acquired with a tilt increment of 2°, as depicted in Figure 1.
Download video file (9.7MB, mp4)
Figure 1—video 3. Representative tilt series acquired with a tilt increment of 3°, as depicted in Figure 1.
Download video file (7.2MB, mp4)
Figure 1—video 4. Representative tilt series acquired with a tilt increment of 5°, as depicted in Figure 1.
Download video file (5.6MB, mp4)
Figure 1—video 5. Representative tilt series acquired with a tilt increment of 10°, as depicted in Figure 1.
Download video file (3MB, mp4)
Figure 1—video 6. Representative tomographic volume of data acquired with a tilt increment of 1°, as depicted in Figure 1.
Download video file (4.6MB, mp4)
Figure 1—video 7. Representative tomographic volume of data acquired with a tilt increment of 2°, as depicted in Figure 1.
Download video file (6.3MB, mp4)
Figure 1—video 8. Representative tomographic volume of data acquired with a tilt increment of 3°, as depicted in Figure 1.
Download video file (6.5MB, mp4)
Figure 1—video 9. Representative tomographic volume of data acquired with a tilt increment of 5°, as depicted in Figure 1.
Download video file (8.3MB, mp4)
Figure 1—video 10. Representative tomographic volume of data acquired with a tilt increment of 10°, as depicted in Figure 1.
Download video file (7.7MB, mp4)

The residual alignment error suffers from higher increments

Next, we quantified the SNR of the tilt images and the tomographic volumes by dividing the square of the mean by the square of the standard deviation of the images (for details, see ‘Methods’). As expected, due to the increased dose per image, the SNR in tilt images increases with the increment (see Figure 2A). However, how the SNR of projection images affects the reconstructions further depends on the alignment accuracy. When analysing tomographic volumes, we found that tomograms from data with a smaller increment displayed higher SNR values (see Figure 2B), whilst showing a similar lamella thickness distribution (see Figure 2—figure supplement 1A). This is also apparent in the XZ-slices shown in Figure 1C. An insightful metric to judge the accuracy of the alignment of projection images is the residual error that is calculated using the Etomo program in IMOD (Mastronarde and Held, 2017), which is the deviation of tilt patches from the alignment fit during tilt-series alignment (Mastronarde and Held, 2017). Interestingly, large differences were observed for this metric when comparing the datasets (Figure 2C). We found that tilt series with a smaller tilt increment could be aligned with higher accuracy. These data show that tilt images with as little as 1.1 e-2 were already aligned with great accuracy.

Figure 2. Assessment of signal-to-noise ratio (SNR), tilt-alignment quality, and CTF fitting parameters at different tilt increments.

(A) SNR measured from projection images of the sample at effective zero-tilt position. SNR was defined as the ratio of the squared mean to the squared standard deviation of each image. (B) SNR measured from the sample volume that represents the lamellae. As the tomogram SNR distribution of particularly the 2° and 3° dataset showed similar SNR distributions, we performed formal significance testing for the data in this panel (see B). Kruskal–Wallis test confirmed significant differences across conditions (H=270.97, p<0.001); pairwise Mann–Whitney U tests with Bonferroni correction revealed all pairs were significantly different except 2° vs. 3° (p=0.093), indicating comparable SNR for these two tilt increments. (C) Residual alignment error after tilt-series alignment in Etomo. (D, E) CTF estimation parameters. (D) Figure of merit, which indicates the confidence level to which the CTF was fitted, and (E) the final resolution to which the CTF was fitted; both estimated with Gctf on projection images of the sample at effective zero-tilt position. (F) Scatter plot showing the final resolution as shown in (E) against the local lamella thickness for all conditions, with coefficients of determination of R2=0.30, 0.38, 0.66, 0.61, and 0.60 for the respective datasets.

Figure 2.

Figure 2—figure supplement 1. Local lamella thickness per condition.

Figure 2—figure supplement 1.

(A) Local lamella thickness for all tomograms used in this study. (B) Local lamella thickness for all tomograms used for the TM and STA analyses. Owing to the poor tomogram reconstruction quality observed for the 10° tilt increment data, lamella thickness could not be determined for each tomogram for this condition.
Figure 2—figure supplement 2. CTF estimation parameters from CTFFIND4.

Figure 2—figure supplement 2.

(A) CTFFIND4 CC score, which indicates the confidence level to which the CTF was fitted. (B) Maximum fitted resolution of the CTF using CTFFIND4. For clarity, only the images of the untilted specimen were used in the analysis.

After estimation of the contrast-transfer-function (CTF) using Gctf (Zhang, 2016), we found that the CTF was fitted with higher confidence to images with higher dose based on the metrics ‘figure of merit’ and ‘CC-score’, as expected. Surprisingly though, the resolution to which the CTF was fitted was similar for all conditions, despite an eightfold increase in dose (see Figure 2E). Similar results were obtained using CTFFIND4 (Rohou and Grigorieff, 2015; Figure 2—figure supplement 2). In cryo-ET, several factors limit the resolution up to which CTF can be reliably estimated. This includes the dose per image and sample thickness determined by the lamellae, as well as tilt-induced effects such as defocus gradients and increased sample thickness. As the data shown in Figure 2D–F pertains to images of the untilted specimen, the tilt-induced effects can be ruled out. Moreover, the results in Figure 2E demonstrate that the observed maximum resolution, at least under the chosen experimental conditions, is not constrained by the per-image-dose. We therefore examined whether the maximum resolution depends on the local lamella thickness in each tomogram. Indeed, the variation in maximum resolution correlates with lamella thickness across all datasets (see Figure 2F).

From these data, we conclude that tilt images can be acquired and aligned for all conditions, but smaller tilt increments result in tomograms that are better aligned and have a higher SNR.

Template matching performance suffers from higher increments

Ribosomes are widely abundant throughout the cytoplasm and, due to their size and electron contrast, have a high degree of processability. We therefore used ribosomes as probes to study the effect of tilt increment on TM and STA, as was done previously to study other effects (Berger et al., 2023; Tuijtel et al., 2024). Firstly, we selected ca. 20 tomograms per condition, based on tomogram content and local lamella thickness (Tuijtel et al., 2024; for more details, see ‘Methods’ and Figure 2—figure supplement 1B). We then performed TM for ribosomes on the selected tomograms with a binning factor of 4 (corresponding to a voxel size of 7.9 Å), and an angular search of 5° using GAPSTOP (Cruz-León et al., 2024; Tuijtel et al., 2024; Wan et al., 2024; see Figure 3—figure supplement 1). For the 1°, 2°, and 3° increment data, we could set peak extraction thresholds such that only true ribosomes were selected, and no 3D classification was necessary to remove, or able to detect junk particles (see Figure 3—figure supplement 2). For 5° and 10° data, however, 3D classification in RELION 3.1 (Zivanov et al., 2018) had to be utilised to separate junk particles from the TM results (see Figure 3—figure supplement 3). For the 10° data, 3D classification did not work reliably with a binning factor of 6 that was sufficient for the 5° data. Instead, we used a binning factor of 2, still with limited success (see Figure 3—figure supplement 3). Finally, in a similar number of tomograms, we identified fewer ribosomes for the 5° and 10° data (see Figure 3—figure supplement 3). To ensure similar data processing strategies for all conditions, extraction thresholds were also lowered for the 1°, 2°, and 3° conditions, and 3D classification was performed to filter out the junk particles (see Figure 3A, Figure 3—figure supplement 3).

Figure 3. The effect of tilt increment on template matching.

(A) Classes after the first round of 3D classification. (B) Maximum F1-score for tomograms used for TM and STA. (C) Area under the precision–recall curves. See corresponding Figure 3—figure supplement 4.

Figure 3.

Figure 3—figure supplement 1. Template matching scores.

Figure 3—figure supplement 1.

(A–E) Volumes with template matching scores were converted to z-scores. For visualisation purposes, here only a subvolume of 500 × 500 × 400 voxels were isolated from representative volumes. Panels (A–E) represent results for 1°, 2°, 3°, 5°, and 10° increments, respectively. A maximum intensity projection was performed along the last dimension (z) and plotted (left). An overlay the scores and the tomogram of the same region (middle) and a histogram of the score volume (right). Scalebar shown in panel (A) is 100 nm and applies to all relevant panels.
Figure 3—figure supplement 2. 3D classification results after high-confidence TM.

Figure 3—figure supplement 2.

Resulting classes after 3D classification with the intention to remove junk particles, performed on subtomograms with binning factor 6, corresponding to a voxel size of 11.8 Å, for tilt increments of 1° (A), 2° (B), and 3° (C). Z-score thresholds of 6, 6, and 5.5 were used for the 1°, 2°, and 3° data, respectively. The total number of particles is indicated above the classes, as well as the number of particles per class below the class. No significant number of junk particles could be detected for these datasets.
Figure 3—figure supplement 3. 3D classification results after TM to remove junk particles.

Figure 3—figure supplement 3.

Resulting classes after 3D classification to remove junk particles, performed on subtomograms with binning factor 6, corresponding to a voxel size of 11.8 Å/px, for tilt increments of 1° (A), 2° (B), 3° (C), 5° (D), and 10° (E). Z-score thresholds of 5 were used for all data, apart from the 10° data, where a threshold value of 3.75 was used. For the 10° dataset, a binning factor of 2 is used, corresponding to a voxel size of 3.9 Å. The total number of starting particles is indicated at the top. The number of particles per class is indicated below each class. Successive classification rounds are visible as rows. Particle classes that were discarded have been marked with a red arrow, particle classes that were taken to the next round of classification are indicated with a downward orange arrow, and particle classes that were accepted as true positives are indicated with a green check mark. The number of accepted particles after 3D classification is indicated below, as well as the percentage with respect to the number of starting particles.
Figure 3—figure supplement 4. F1- and PR-curves for all tomograms.

Figure 3—figure supplement 4.

Each line represents a single tomogram used in the analysis. (A–E) F1-curves for tomograms used for TM for 1°, 2°, 3°, 5°, and 10° increments, respectively. On the horizontal axes, the z-score threshold used for peak extraction is plotted and on the vertical axes the F1-score for particles extracted using that threshold. (F–J) Precision–recall curves for the same data.

By comparing particle lists from before and after 3D classification, we calculated F1-scores and precision–recall (PR) curves for each tomogram used for TM (see Figure 3—figure supplement 4 and ‘Methods’). To compare all tomograms from each dataset, we plotted the maximum F1-score and calculated the area under the PR curves for each tomogram (Figure 3B and C). In summary, TM performed equally well on tomograms from the 1°, 2°, and 3° increment data. In contrast, performance was reduced for the 5° and even more so for the 10° data (Figure 3B and C).

Per-particle refinement using STA has an optimal tilt increment

We then proceeded with STA, following the well-established Warp-RELION-M pipeline (Tegunov et al., 2021; Tegunov and Cramer, 2019; Zivanov et al., 2018). Briefly, tilt images and tilt-series alignment files from Etomo were imported into Warp for CTF-estimation, tomogram reconstruction, and subtomogram extraction. Initial particle refinement in RELION 3.1 served as the starting point for subsequent multi-particle refinement in M (see ‘Methods’). We found that RELION 3.1, which treats the subtomograms as static 3D volumes, refined 1°, 2°, and 3° data to roughly the same resolution (10–11 Å), while the 5° and 10° data only averaged up to 19–23 Å (see Figure 4A, Figure 4—figure supplement 1A). After subsequent M refinement, the 3° data showed a markedly higher resolution than all other conditions, averaging to 5.8 Å, compared to 7.0 Å and 7.7 Å for the 1° and 2° data, respectively (see Figure 4B, Figure 4—figure supplement 1B). Furthermore, M improved the resolution for the 5° and 10° data to 7.5 Å and 9.2 Å, respectively, which is very similar to the 1° and 2° data when adjusted for particle numbers (see Figure 4B).

Figure 4. The effect of tilt increment on STA.

(A) Rosenthal–Henderson plots of particle subsets when refined and averaged in RELION 3.1. (B) Rosenthal–Henderson plots of tomogram subsets when refined and averaged in M. Inset: ribosome density maps using all particles, coloured according to the local resolution. (C) Resolution progression for each M refinement, considering all available particles. The grid subdivision shown on the horizontal axis is used for both the Image warp and the Volume warp grids. For more details, see the ‘Methods’ section. (D) Resolution gain for each refinement round, also shown in panel (C).

Figure 4.

Figure 4—figure supplement 1. Fourier shell correlation curves.

Figure 4—figure supplement 1.

(A) FSC curve for all tilt increments after refinement in RELION. (B) FSC curve for all tilt increments after refinement in M. Thresholds at FSC of 0.5 and 0.143 are indicated with dashed horizontal lines.

For each condition, all particles were processed once more in subgroups of different sizes to produce Rosenthal–Henderson plots (Rosenthal and Henderson, 2003), that capture the dependency of resolution on particle number (Figure 4A and B). To capitalise on M’s multi-particle refinement framework, subgroups were defined based on the selection of tomograms rather than an exact number of randomly chosen particles, as was done for the RELION results.

To investigate the results from M further, we plotted the reported resolution through the various rounds of refinement, as shown in Figure 4C and D. We found that the resolution improvements per iteration correlated with the amount of dose in the tilt images.

From our STA-analysis, we conclude that, when subtomograms were treated as independent and static 3D volumes, 1°, 2°, and 3° data averaged to similar resolution. However, the per-particle tilt image refinement from M yielded the highest final resolution for the 3° tilt increment data.

Discussion

Cryo-ET has recently matured into a robust and accessible technique, supported by advances such as improved hardware, fast parallel acquisition (Bouvette et al., 2021; Eisenstein et al., 2023; Khavnekar et al., 2023), and streamlined data processing routines. As a result, it is becoming a mainstream tool for in situ structural biology. Strategies to optimise data acquisition have thus come into focus, and more systematic research has been performed on how parameters such as lamella thickness and acquisition schemes influence the resolution attainable with STA (Berger et al., 2023; Tuijtel et al., 2024; Turoňová et al., 2020a). In this study, we systematically investigated the effects of varying the tilt increment when collecting cryo-ET data, with the total dose and tilt range kept constant. We found that finer increments (1–3°) showed overall similar results in visual tomogram quality, tilt-series alignment, high-confidence TM and STA resolution, whereas coarser increments (5° and 10°) compromised tomogram quality, showed reduced TM accuracy, and subsequently lower STA-resolution. However, after M refinement, the 3° dataset clearly outperformed all other datasets in terms of resolution. For STA, a 3° increment may provide the optimal balance between angular sampling, dose distribution, and computational expense during processing.

To further contextualise these findings, we first examined the quality of the tomograms. Contrary to our expectations, angular sampling, rather than the amount of dose per projection image, was the primary factor that influenced the alignment accuracy as approximated by the residual error. This observation may be explained with differences in alignment strategies used for in vitro and in situ samples. For in vitro studies, gold fiducial beads are typically used for alignment. These beads retain their spherical shape and similarity, regardless of the projection angle or the angular gap between successive images, thereby providing consistent and robust reference points across the tilt series. In contrast, alignment of in situ datasets often relies on patch tracking of endogenous structural features, which are generally irregular in shape. Their projection geometry varies with the projection angle and they therefore lose similarity between successive images more rapidly. As a result, the cross-correlation peaks become broader or less well-defined, ultimately reducing alignment accuracy.

We further found that using a 1–3° increment, despite the reduced SNR in the tilt images, generally led to high contrast in the tomograms. Initially, we reasoned that finer angular sampling would extend the Crowther criterion to higher spatial frequencies and, when combined with improved alignment, would yield tomograms of visibly enhanced quality. However, we could not visually distinguish tomograms coming from 1°, 2°, or a 3° tilt series. Conversely, coarser increments (5° and 10°) compromised tomogram quality and SNR and led to artefacts in the tomograms. Furthermore, whilst the 1° data exhibited the best alignment, it is associated with practical drawbacks. For instance, the acquisition time per tilt series increased more than twofold compared to 2° tilt increments. In addition, a higher number of projections increases demands on data storage and computational resources during pre-processing, including subsequent steps that utilise tilt images or intermediate data. However, differences in storage and processing requirements vanish once actual 3D tomograms or subtomograms are reconstructed or extracted, but re-emerge during subsequent M refinements.

As cryo-ET datasets are becoming increasingly larger, and also publicly available, the focus in the field shifts from data acquisition to accurately and efficiently mining the wealth of information that is present in existing cryo-ET datasets (Ermel et al., 2024; Iudin et al., 2023; Kelley et al., 2026; Last et al., 2025; Peck et al., 2024). TM has shown itself to be a robust and sensitive tool to detect the structural signature of proteins (Cruz-León et al., 2024; Taniguchi et al., 2025), but is computationally very expensive, which has become a major bottleneck of the workflow in practice (Martinez-Sanchez, 2025). Moreover, further particle curation is often necessary to remove false-positive TM picks from the true-positive particles through extensive 3D classification procedures. It is therefore important to understand how the tilt increment influences the TM procedure to enable its optimisation. Here, we found that TM performed equally well for the 1°, 2°, and 3° increment data on ribosomes, and peaks higher than a threshold could be extracted to yield only true positives so that successive 3D classification was not necessary. This was not possible for the 5° and 10° data, however, likely due to a combination of suboptimal tilt-series alignment and reduced SNR in the tomograms. We hypothesise that for smaller templates that comprise more challenging targets, improved tomogram alignment and increased Crowther criteria are crucial for successful TM, thereby favouring the usage of smaller tilt increments, particularly when TM is performed on data with a lower binning factor. This notion is supported by recent work, that showed that the addition of interpolated tilts, generated using cryoTIGER, improved the performance of TM of nucleosomes (Majtner et al., 2025). We anticipate that a well-balanced choice of experimental angular increments and tilt interpolation during post processing may be key to future workflows that will benefit from optimal angular sampling.

In STA, the incomplete angular sampling of individual particles may be less critical compared to the analysis of entire tomograms and TM, due to the large number of particles contributing to the final reconstruction. The initially limited angular sampling can be compensated by the inclusion of particles with slightly different orientations, provided that a sufficient number of particles is available. Furthermore, for datasets with fewer projections, suboptimal initial alignment may be mitigated by the higher SNR of individual tilt images. When analysing STA results in our data, as with TM, we could not find major differences in the 1°, 2°, and 3° datasets when averaged with RELION 3.1, whereas the 5° and 10° datasets yielded a noticeably lower resolution. Using M improved the resolution across all datasets to comparable levels, accounting for reduced particle numbers, and notably rescued the 5° and 10° datasets despite their lower initial resolution. We hypothesise that two factors contribute to the pronounced tilt increment dependency of the resolution gain after M. Firstly, the initial tilt-series alignment was markedly poorer at higher increments, leaving more potential for improvement. Secondly, these images were acquired at a higher dose, and therefore have a higher SNR, which likely facilitates more effective alignment by M.

Remarkably, the 3° dataset reached a higher resolution than all others. Acquiring data with a 3° tilt increment likely strikes an optimal balance for the spatial refinement of projections on a per-particle basis STA refinement: it enables robust initial tilt-series alignment and high SNR in the tomograms to support successful TM, while still delivering sufficient dose per tilt image to facilitate effective M refinement. Furthermore, especially compared to the 1° data, data collection with a 3° increment is more streamlined in terms of acquisition time and data storage.

We note that, while we believe our findings can be applied to cryo-ET in general, the present investigation was confined to a single sample and molecular target. Additionally, although the exact STA resolution may vary, the general trends identified here are unlikely to change when different software packages are used.

Furthermore, we focused our research on in situ data. It is unclear to which extent these results are transferrable to in vitro cryo-ET studies of inherently thin samples. The alignment of those tilt series is usually performed using gold fiducial beads, which provide great precision, as discussed above. Furthermore, reduced tilt ranges may be used to increase signal in the lower tilt images, however, alignment will most likely not be affected.

Ribosomes are abundant, electron dense, and the processing workflow is particularly effective for ribosomes, making them ideal candidates for studies of this nature. However, it remains unclear whether other targets such as smaller particles or particles with a lower abundance, would follow a similar trend, in particular with respect to TM and M refinements. M allows the simultaneous refinement of multiple molecular species within a single dataset. As a result, lower-abundance targets may benefit from alignment driven by nearby ribosomes, provided there is sufficient spatial proximity, and from the advantages of the multi-particle refinement framework. Many in situ cryo-ET studies focus on targets at least partially situated within the cytoplasm, which is filled with relatively dense structures and membranes, thereby aiding tilt-series alignment. Large vacuoles, the extracellular space, or the nucleoplasm, where these contrast-rich structures are absent, may comprise more challenging environments. To what extent tomogram content affects tilt-series alignment remains to be further investigated.

Taken together, our findings support the use of a 3° tilt increment for tilt-series acquisition, as it yields tomograms with high SNR, accurate tilt-series alignment, robust performance during TM, and maximised resolution when using multi-particle refinement in M. From our perspective, a 3° increment also represents an economical choice, reducing beam time through shorter acquisition, and lowering data storage and computational demands. Smaller or more challenging targets may still benefit from finer increments.

Methods

Key resources table.

Reagent type (species) or resource Designation Source or reference Identifiers Additional information
Strain, strain background (Dictyostelium discoideum) Ax2-214 dictyBase, Depositor Guenter Gerisch DBS0235534
Chemical compound, drug HL5 medium Formedium HLB0102
Chemical compound, drug Ampicillin Formedium A9518
Chemical compound, drug Geneticin G418 Sigma-Aldrich G5013
Software, algorithm SerialEM (4.0.6, 4.0.10 & 4.0.20) Mastronarde, 2005 RRID:SCR_017293 https://bio3d.colorado.edu/SerialEM/
Software, algorithm IMOD (4.11.5) Kremer et al., 1996 RRID:SCR_003297 https://bio3d.colorado.edu/imod/
Software, algorithm Matlab (2019b) The MathWorks, Inc RRID:SCR_001622 https://www.mathworks.com/products/matlab.html
Software, algorithm Gctf (v1.06 2016-05-22) Zhang, 2016 RRID:SCR_016500 https://github.com/ProteinGod/Gctf
Software, algorithm Ctffind 4.1 Rohou and Grigorieff, 2015 RRID:SCR_016732 https://grigoriefflab.umassmed.edu/ctf_estimation_ctffind_ctftilt
Software, algorithm WARP (1.0.9) Tegunov and Cramer, 2019 RRID:SCR_018071 https://github.com/warpem/warp
Software, algorithm Relion 3.1 Zivanov et al., 2018 RRID:SCR_016274 https://github.com/3dem/relion
Software, algorithm M (1.0.9) Tegunov et al., 2021 https://github.com/warpem/warp
Software, algorithm STOPGAP (0.7.1) Wan, 2023; Wan et al., 2024 https://github.com/wan-lab-vanderbilt/STOPGAP
Software, algorithm GAPSTOP (v0.3) Cruz-León et al., 2024; Wan et al., 2024 https://gitlab.mpcdf.mpg.de/bturo/gapstop_tm; Turoňová, 2025
Software, algorithm ChimeraX (1.6) Meng et al., 2023; Pettersen et al., 2021 RRID:SCR_015872 https://www.cgl.ucsf.edu/chimerax/
Software, algorithm cryoCAT Turoňová, 2024b https://github.com/turonova/cryoCAT
Software, algorithm Adobe Illustrator 2022 Adobe https://www.adobe.com/
Software, algorithm Python (3.9.7) https://www.python.org/ https://www.python.org/downloads/release/python-397/
Software, algorithm SciPy (1.13.1) https://scipy.org/ RRID:SCR_008058 https://github.com/scipy/scipy
Software, algorithm Numpy (1.26.4) https://numpy.org/ RRID:SCR_008633 https://github.com/numpy/numpy
Software, algorithm Matplotlib (3.9.4) https://matplotlib.org/ RRID:SCR_008624 https://github.com/matplotlib/matplotlib
Software, algorithm Seaborn (0.13.2) Waskom, 2021 RRID:SCR_018132 https://seaborn.pydata.org/
Software, algorithm starparser (v1.38) Chabaan https://github.com/sami-chaaban/starparser; Chaaban, 2022a
Other Pelco easiGlow Glow Discharger Cleaning System Ted Pella, Inc easiGlow Used for hydrophilisation of EM grids prior to vitrification
Other Quantifoil R 1/4, 200 Mesh, Au, SiO2 film Quantifoil EM grids and support film
Other Whatman filter paper #1 Whatman WHA1001329 Blotting paper during vitrification
Other EM GP2 Automatic Plunge Freezer Leica Microsystems Leica GP2 Used for plunge-freezing
Other Aquilos dual beam cryo-FIB/SEM Thermo Fisher Scientific Aquilos cryo-FIB Used for the cryo-FIB milling of lamellae
Other Titan Krios G4 (300 kV cryo-transmission electron microscope) Thermo Fisher Scientific Titan Krios Used for cryo-ET data acquisition, see ‘Methods’
Other Falcon 4 direct electron detector with Selectris X imaging filter Thermo Fisher Scientific F4 with selectris X Direct electron detector and energy filter use for data acquisition, see ‘Methods’

Cryo-ET sample preparation

D. discoideum cells (strain Ax2-214) were grown in HL5 medium (Formedium) containing 50 µg/mL ampicillin and 20 µg/mL geneticin G418 (Sigma Aldrich) at ca. 20°C until exponential growth was achieved.

EM grids, R1/4 200 mesh Au grids with SiO2 support film (Quantifoil), were glow discharged for 90 s at 0.38 mbar and 15 mA using an easyGlow instrument (PELCO) immediately before usage. The cells were diluted to a concentration of 3.3 × 105 cells/ml, and a droplet of 100 µL cell suspension was placed on each grid, after which the cells were allowed to attach to the grid for 2–4 hr at room temperature in the dark. The grids were subsequently blotted and vitrified by plunge freezing into liquid ethane using a Leica GP2 plunger. Cryo-FIB milling was performed using an Aquilos cryo-FIB system (Thermo Scientific). Prior to gallium FIB-milling, grids were coated with an organometallic platinum layer using a gas injection system for 10 s and additionally sputter coated with platinum at 1kV and 10 mA current for 20 s. SEM imaging was performed at 10 kV and 13 pA current to guide the milling progress. Milling was performed at 30 kV in a stepwise fashion, where the current was reduced from 500 pA to 30 pA whilst reducing the thickness of the remaining lamella. Finally, lamella polishing was performed with 30 pA current.

Cryo-ET acquisition

Cryo-ET datasets were collected at 300 kV on a Titan Krios G4 microscope equipped with a cold FEG, Selectris X imaging filter, and Falcon 4 direct electron detector, operated in counting mode (all from Thermo Scientific). Medium magnification overview montages of lamellae were acquired with 3.0 nm pixel size to inspect the lamellae and determine positions for tilt series collection. Tilt series were acquired using SerialEM (versions 4.0.6, 4.0.10, and 4.0.20) in low-dose mode as dose-fractionated movies with size 4096 × 4096 at a magnification of 64,000×, corresponding to a calibrated pixel size of 1.971 Å. Dose-fractionation and motion-correction were performed on-the-fly using 10 equally dosed fractions per tilt movie in SerialEM. Tilt series acquisition started from the lamella pretilt of +8°, and a dose symmetric acquisition scheme was applied (Hagen et al., 2017), with tilt increment and exposure as described in Table 1, all with a tilt grouping of 2. The energy filter was set to a slit width of 10 eV, and a nominal defocus of −2.5 to −5 µm was used. The dose rate on the detector was targeted to be ca. 6 e/px/s at a representative spot on the specimen.

Tomogram reconstruction

The tilt series were dose-filtered using custom Matlab scripts (Wan et al., 2017) and manually cleaned based on visual inspection of the tilt images to remove unusable projection images. CTF-estimation was performed on the tilt images using Gctf (Zhang, 2016) and CTFFIND4 (Rohou and Grigorieff, 2015). The dose-filtered tilt series were then aligned through patch-tracking in IMOD 4.11.5 (Kremer et al., 1996) and reconstructed using weighted back projection and fSIRT for visual inspection. During fiducial model generation using tiltxcorr, optimal filter parameters were found for several representative tilt series for each dataset, after which these were applied to all tilt series in the dataset. Specifically, low-frequency rolloff sigma was 0.01 for the 1° and 3° datasets, 0.02 for the 2° dataset and 0.001 for 5° and 10° datasets. High-frequency cutoff radius was 0.1 for the 2° data, and 0.08 for all other datasets. High-frequency rolloff sigma was kept constant at 0.05. During tomogram reconstruction, all images were multiplied with a factor of 250 by the tilt command. To ensure compatibility with the Warp and M software suites, tomograms were then reconstructed again in Warp using the alignment obtained from Etomo. Both standard and deconvolved tomograms were reconstructed using Warp’s standard tomogram reconstruction settings that weigh and correct the data for CTF, dose and tilt. Local lamella thickness was measured manually for each tomogram using IMOD.

Crowther criterion

The Crowther criterion, or the resolution to which the tomogram is completely filled (disregarding the missing wedge), d is given by

d=πDn,

where n is the number of projection images and D is the specimen thickness (Crowther et al., 1970). The conversion of resolution to number of pixels in the Fourier transform was done using the function resolution2pixels from cryoCAT (Turoňová, 2024b). When calculating the numerical values for the Crowther criterion, we simplified by assuming evenly spaced projections over the full tilt range, thereby disregarding the missing wedge.

SNR measurements

The SNR was calculated according to

SNR=μ2/σ2,

with µ the mean of the image and σ the standard deviation of the image (Heymann, 2022). For tilt images, SNR calculations were based on unbinned, unprocessed motion-corrected tilt images of the specimen at effective zero-tilt. In the rare event where this image was removed during pre-processing, the next available tilt image was used instead. For tomogram volumes, voxels outside of the lamella body were masked out before the SNR was calculated.

To quantify differences in SNR across tilt increment conditions, a non-parametric Kruskal–Wallis test was performed as an omnibus test of the null hypothesis that all groups are drawn from the same distribution. Because SNR distributions were not assumed to be normal and sample sizes differed across conditions, non-parametric tests were used throughout. Following the omnibus test, all 10 pairwise comparisons between conditions were assessed using two-sided Mann–Whitney U tests. To control for multiple comparisons, raw p-values were adjusted using the Bonferroni correction (multiplied by the number of comparisons, n=10, capped at 1.0). Statistical significance was defined as a Bonferroni-corrected p-value below 0.05. All analyses were performed in Python using the scipy.stats module.

Template matching and 3D classification

To ensure processing feasibility, ca. 20 tomograms were selected for each condition to be subjected to TM and STA. Tomograms were selected based on the following criteria: local lamella thickness was limited to ca. 180 nm to optimise resolution (Tuijtel et al., 2024); tomograms that predominantly contained organelles that exclude ribosomes, such as mitochondria or the nucleus, were excluded from analysis. From the resulting list of tomograms, 20 were chosen at random. Care was taken that these tomograms formed a representative subset of the original thickness and alignment distributions to avoid bias by tomogram selection.

TM was performed using GAPSTOP (Cruz-León et al., 2024; Turoňová, 2024a), a GPU-accelerated implementation of STOPGAP (Wan et al., 2024). As a template, a previously determined high-resolution ribosome map from D. discoideum was used, from reference (Tuijtel et al., 2024), and low-pass filtered to a resolution of 30 Å. TM was performed on deconvolved tomograms from Warp with a voxel size of 7.9 Å (corresponding to a binning factor of 4) using an angular search of 5°. TM was performed on a high-performance computing cluster using 16 A100 GPUs (Nvidia), running for approximately 1 hr and 20 min per tomogram, independent from the tilt increment. The resulting score maps were converted to z-scores for more generalisable peak extraction across tomograms (Cruz-León et al., 2024). Initially, peaks were thresholded using a carefully selected z-score (6, 6, and 5.5 for the 1°, 2°, and 3° datasets, respectively) to optimise particle extraction. Later, thresholds were lowered to a z-score of 5 for all conditions, except for the 10° dataset, where peaks were extracted using a threshold at z-score 3.75. Peak location and angular information from TM were converted to particle STAR files using the cryoCAT package (Turoňová, 2024b).

Subtomograms were extracted using Warp at a binning factor of 6 (corresponding to a voxel size of 11.8 Å) and subjected to multiple rounds of 3D classification in RELION 3.1 (Zivanov et al., 2018) until a clean particle list was obtained.

Template matching evaluation

To ascertain the quality of the TM procedure, we calculated the precision, recall, and F1-scores. Firstly, ribosome positions confirmed by 3D classification were considered the ground truth (GT). Then, TM peaks were extracted at various z-score thresholds, and these particle locations were compared to the GT. For each extraction, the overlap of GT-particles and extracted particles are true positives (TP). The false positives (FP) are locations that were extracted, but not in the GT. Then, the precision (P) is calculated as

P=TPTP+FP.

Similarly, the recall (R) is given by

R=TPTP+FN,

where FN are the false negatives, meaning the particles in the GT list that were not extracted when using that extraction threshold. Precision–recall curves were plotted by calculating both the precision and recall values for all extraction thresholds described above.

The F1-score is the harmonic mean of precision and recall:

F1score=2×P×RP+R.

Ribosome particle refinement

After 3D classification, a 3D refinement was carried out on the particles selected as true positive particles using RELION's 3D Refine job type. These refined particles were re-extracted in Warp using a binning factor of 2 (corresponding to a voxel size of 3.9 Å), refined again, after which subtomograms were extracted in an unbinned fashion, refined in RELION once more, and the results served as inputs for M (version 1.0.9) (Tegunov et al., 2021), starting from the on-the-fly frame-aligned projection images. M refinements were carried out as shown in Table 2. When convergence was reached, no further refinements were carried out.

Table 2. Parameters of M refinement rounds.

Geometry Tilt series CTF
M refinement Image warp grid Particle poses Stage angles Volume warp grid Defocus Tilt series acquired
Round 1 3 × 3 3 × 3 × 2 × 10 - -
Round 2 6 × 6 6 × 6 × 8 ×10 - -
Round 3 6 × 6 6 × 6 × 8 × 10
Round 4 10 × 10 10 × 10 × 10 × 10

Rosenthal–Henderson plots

To produce Rosenthal–Henderson plots (Rosenthal and Henderson, 2003), particle subgroups were randomly selected in two ways. For RELION refinement, the ‘extract_random’ function from starparser was used (Chaaban, 2022b) on the unrefined and unbinned STAR file exported from Warp to produce particle subsets of 5000, 2500, 1000, and 500 particles. These particle subgroups were then refined independently using RELION as before, and post-processed using a solvent mask.

For M refinements, particles were pseudo-randomly selected based on whole tomograms using a custom Python script (see Source code 1). Care was taken to first process the smallest particle subset in M, and then continue with the subsets containing more particles to avoid cross-interference of the previous refinement rounds.

Acknowledgements

This work was funded by the Max Planck Society, has been made possible in part by grant number 2021-234666 from the Chan Zuckerberg Initiative DAF, an advised fund of Silicon Valley Community Foundation, and by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation)—SFB1507 Project No. 450648163 (P17). We thank Patrick C Hoffmann for help with the cell culture. We thank the Central Electron Microscopy Facility of the Max Planck Institute of Biophysics for providing microscope access and support, and are especially grateful to Mark Linder and Sonja Welsch for technical support. We thank Jan Philipp Kreysing, Sergio Cruz-León, and Gerhard Hummer for useful discussions; Stefanie Böhm and Sonja Welsch for critical reading of the manuscript. We thank Iskander Khusainov, Özkan Yildiz, Juan F Castillo Hernandez, Andre Schwarz, Erin Schuman, and the Max Planck Computing and Data Facility for support with scientific computing.

Funding Statement

The funders had no role in study design, data collection and interpretation, or the decision to submit the work for publication. Open access funding provided by Max Planck Society.

Contributor Information

Martin Beck, Email: Martin.Beck@biophys.mpg.de.

Axel T Brunger, Stanford University School of Medicine, Howard Hughes Medical Institute, United States.

Volker Dötsch, Goethe University Frankfurt, Germany.

Funding Information

This paper was supported by the following grants:

  • Max-Planck-Gesellschaft to Beata Turoňová, Martin Beck.

  • Chan Zuckerberg Initiative 2021-234666 to Beata Turoňová, Martin Beck.

  • Deutsche Forschungsgemeinschaft SFB1507 Project No. 450648163 to Martin Beck.

Additional information

Competing interests

No competing interests declared.

Author contributions

Conceptualization, Data curation, Software, Formal analysis, Validation, Investigation, Visualization, Methodology, Writing – original draft, Project administration, Writing – review and editing.

Software, Writing – review and editing.

Conceptualization, Supervision, Funding acquisition, Methodology, Writing – review and editing.

Conceptualization, Resources, Data curation, Software, Formal analysis, Supervision, Funding acquisition, Validation, Investigation, Visualization, Methodology, Writing – original draft, Project administration, Writing – review and editing.

Additional files

MDAR checklist
Source code 1. Jupyter notebook used to pseudo-randomly select particles based on whole tomograms for M refinements.
elife-111639-code1.zip (10.7KB, zip)

Data availability

Relevant cryo-ET density maps generated in this study have been deposited in the EM Data Bank (EMDB) with the following access code: EMD- 57021. All raw data have been deposited to the Electron Microscopy Public Image Archive (EMPIAR) database (accessing code EMPIAR-13391), including acquisition metadata, information, tilt-series alignment data, all raw TM locations, and particle locations after 3D classification and refined particle motive lists. All other data needed to evaluate the conclusions in the study are present in the paper.

The following datasets were generated:

Tuijtel MW, Beck M. 2026. Cryo-ET dataset on lamellae of Dictyostelium discoideum cells acquired with various tilt-increments. EMPIAR. EMPIAR-13391

Tuijtel MW, Beck M. 2026. Optimal tilt-increment for cryo-ET. EMD. EMD-57021

References

  1. Bartesaghi A, Lecumberry F, Sapiro G, Subramaniam S. Protein secondary structure determination by constrained single-particle cryo-electron tomography. Structure. 2012;20:2003–2013. doi: 10.1016/j.str.2012.10.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Beck M, Baumeister W. Cryo-electron tomography: can it reveal the molecular sociology of cells in atomic detail? Trends in Cell Biology. 2016;26:825–837. doi: 10.1016/j.tcb.2016.08.006. [DOI] [PubMed] [Google Scholar]
  3. Berger C, Dumoux M, Glen T, Yee NB-Y, Mitchels JM, Patáková Z, Darrow MC, Naismith JH, Grange M. Plasma FIB milling for the determination of structures in situ. Nature Communications. 2023;14:629. doi: 10.1038/s41467-023-36372-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Böhm J, Frangakis AS, Hegerl R, Nickell S, Typke D, Baumeister W. Toward detecting and identifying macromolecules in a cellular context: Template matching applied to electron tomograms. PNAS. 2000;97:14245–14250. doi: 10.1073/pnas.230282097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Bouvette J, Liu H-F, Du X, Zhou Y, Sikkema AP, da Fonseca Rezende E Mello J, Klemm BP, Huang R, Schaaper RM, Borgnia MJ, Bartesaghi A. Beam image-shift accelerated data acquisition for near-atomic resolution single-particle cryo-electron tomography. Nature Communications. 2021;12:1957. doi: 10.1038/s41467-021-22251-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Burt A, Toader B, Warshamanage R, von Kügelgen A, Pyle E, Zivanov J, Kimanius D, Bharat TAM, Scheres SHW. An image processing pipeline for electron cryo-tomography in RELION-5. FEBS Open Bio. 2024;14:1788–1804. doi: 10.1002/2211-5463.13873. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Castaño-Díez D, Zanetti G. In situ structure determination by subtomogram averaging. Biophysical and Computational Methods - Part B. 2019;58:68–75. doi: 10.1016/j.sbi.2019.05.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Chaaban S. Starparser. v1.38GitHub. 2022a https://github.com/sami-chaaban/starparser
  9. Chaaban S. Starparser. v1.38Zenodo. 2022b doi: 10.5281/zenodo.6792794. [DOI]
  10. Chen M, Bell JM, Shi X, Sun SY, Wang Z, Ludtke SJ. A complete data processing workflow for cryo-ET and subtomogram averaging. Nature Methods. 2019;16:1161–1168. doi: 10.1038/s41592-019-0591-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Crowther RA, DeRosier DJ, Klug A. The reconstruction of a three-dimensional structure from projections and its application to electron microscopy. Proceedings of the Royal Society of London. A. Mathematical and Physical Sciences. 1970;317:319–340. doi: 10.1098/rspa.1970.0119. [DOI] [Google Scholar]
  12. Cruz-León S, Majtner T, Hoffmann PC, Kreysing JP, Kehl S, Tuijtel MW, Schaefer SL, Geißler K, Beck M, Turoňová B, Hummer G. High-confidence 3D template matching for cryo-electron tomography. Nature Communications. 2024;15:3992. doi: 10.1038/s41467-024-47839-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. de Teresa-Trueba I, Goetz SK, Mattausch A, Stojanovska F, Zimmerli CE, Toro-Nahuelpan M, Cheng DWC, Tollervey F, Pape C, Beck M, Diz-Muñoz A, Kreshuk A, Mahamid J, Zaugg JB. Convolutional networks for supervised mining of molecular patterns within cellular context. Nature Methods. 2023;20:284–294. doi: 10.1038/s41592-022-01746-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. Eisenstein F, Yanagisawa H, Kashihara H, Kikkawa M, Tsukita S, Danev R. Parallel cryo electron tomography on in situ lamellae. Nature Methods. 2023;20:131–138. doi: 10.1038/s41592-022-01690-1. [DOI] [PubMed] [Google Scholar]
  15. Ermel U, Cheng A, Ni JX, Gadling J, Venkatakrishnan M, Evans K, Asuncion J, Sweet A, Pourroy J, Wang ZS, Khandwala K, Nelson B, McCarthy D, Wang EM, Agarwal R, Carragher B. A data portal for providing standardized annotations for cryo-electron tomography. Nature Methods. 2024;21:2200–2202. doi: 10.1038/s41592-024-02477-2. [DOI] [PubMed] [Google Scholar]
  16. Förster F, Briegel A. Cryo-Electron Tomography: Structural Biology in Situ. Springer International Publishing; 2024. [DOI] [Google Scholar]
  17. Frangakis AS. In: Cryo-Electron Tomography. Förster F, Briegel A, editors. Springer International Publishing; 2024. Principles of tomographic reconstruction; pp. 53–83. [DOI] [Google Scholar]
  18. Glynn C, Smith JLR, Case M, Csöndör R, Katsini A, Sanita ME, Glen TS, Pennington A, Grange M. A generalizable and targeted molecular biopsy approach for in situ cryogenic electron tomography of vitreous brain tissue. Cell Reports Methods. 2025;5:101080. doi: 10.1016/j.crmeth.2025.101080. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Grant T, Grigorieff N. Measuring the optimal exposure for single particle cryo-EM using a 2.6 Å reconstruction of rotavirus VP6. eLife. 2015;4:e06980. doi: 10.7554/eLife.06980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Hagen WJH, Wan W, Briggs JAG. Implementation of a cryo-electron tomography tilt-scheme optimized for high resolution subtomogram averaging. Journal of Structural Biology. 2017;197:191–198. doi: 10.1016/j.jsb.2016.06.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Hegerl R, Hoppe W. Influence of electron noise on three-dimensional image reconstruction. Zeitschrift Für Naturforschung A. 1976;31:1717–1721. doi: 10.1515/zna-1976-1241. [DOI] [Google Scholar]
  22. Heymann JB. The progressive spectral signal-to-noise ratio of cryo-electron micrograph movies as a tool to assess quality and radiation damage. Computer Methods and Programs in Biomedicine. 2022;220:106799. doi: 10.1016/j.cmpb.2022.106799. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Himes BA, Zhang P. emClarity: software for high-resolution cryo-electron tomography and subtomogram averaging. Nature Methods. 2018;15:955–961. doi: 10.1038/s41592-018-0167-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Hoffmann PC, Kreysing JP, Khusainov I, Tuijtel MW, Welsch S, Beck M. Structures of the eukaryotic ribosome and its translational states in situ. Nature Communications. 2022;13:7435. doi: 10.1038/s41467-022-34997-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Iudin A, Korir PK, Somasundharam S, Weyand S, Cattavitello C, Fonseca N, Salih O, Kleywegt GJ, Patwardhan A. EMPIAR: the electron microscopy public image archive. Nucleic Acids Research. 2023;51:D1503–D1511. doi: 10.1093/nar/gkac1062. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Kelley R, Khavnekar S, Righetto RD, Heebner J, Obr M, Zhang X, Chakraborty S, Tagiltsev G, Michael AK, van Dorst S, Waltz F, McCafferty CL, Lamm L, Zufferey S, Van der Stappen P, van den Hoek H, Wietrzynski W, Harar P, Wan W, Briggs JAG, Plitzko JM, Engel BD, Kotecha A. Toward community-driven visual proteomics with large-scale cryo-electron tomography of Chlamydomonas reinhardtii. Molecular Cell. 2026;86:213–230. doi: 10.1016/j.molcel.2025.11.029. [DOI] [PubMed] [Google Scholar]
  27. Khavnekar S, Wan W, Erdmann P, Plitzko J. STOPGAP_refine: Tilt series refinement for high-resolution subtomogram averaging. Microscopy and Microanalysis. 2021;27:3240. doi: 10.1017/S143192762101117X. [DOI] [Google Scholar]
  28. Khavnekar S, Wan W, Majumder P, Wietrzynski W, Erdmann PS, Plitzko JM. Multishot tomography for high-resolution in situ subtomogram averaging. Journal of Structural Biology. 2023;215:107911. doi: 10.1016/j.jsb.2022.107911. [DOI] [PubMed] [Google Scholar]
  29. Klumpe S, Senti KA, Beck F, Sachweh J, Hampoelz B, Ronchi P, Oorschot V, Brandstetter M, Yeroslaviz A, Briggs JAG, Brennecke J, Beck M, Plitzko JM. In-cell structure and snapshots of copia retrotransposons in intact tissue by cryo-ET. Cell. 2025;188:2094–2110. doi: 10.1016/j.cell.2025.02.003. [DOI] [PubMed] [Google Scholar]
  30. Kremer JR, Mastronarde DN, McIntosh JR. Computer visualization of three-dimensional image data using IMOD. Journal of Structural Biology. 1996;116:71–76. doi: 10.1006/jsbi.1996.0013. [DOI] [PubMed] [Google Scholar]
  31. Kreysing JP, Heidari M, Zila V, Cruz-León S, Obarska-Kosinska A, Laketa V, Rohleder L, Welsch S, Köfinger J, Turoňová B, Hummer G, Kräusslich HG, Beck M. Passage of the HIV capsid cracks the nuclear pore. Cell. 2025;188:930–943. doi: 10.1016/j.cell.2024.12.008. [DOI] [PubMed] [Google Scholar]
  32. Last MGF, Voortman LM, Sharp TH. Scaling data analyses in cellular cryoET using comprehensive segmentation. bioRxiv. 2025 doi: 10.1101/2025.01.16.633326. [DOI]
  33. Majtner T, Kreysing JP, Tuijtel MW, Cruz-León S, Liu J, Hummer G, Beck M, Turoňová B. cryoTIGER: deep-learning based tilt interpolation generator for enhanced reconstruction in cryo electron tomography. Communications Biology. 2025;8:1443. doi: 10.1038/s42003-025-08961-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Martinez-Sanchez A. Template matching and machine learning for cryo-electron tomography. Current Opinion in Structural Biology. 2025;93:103058. doi: 10.1016/j.sbi.2025.103058. [DOI] [PubMed] [Google Scholar]
  35. Mastronarde DN. Automated electron microscope tomography using robust prediction of specimen movements. Journal of Structural Biology. 2005;152:36–51. doi: 10.1016/j.jsb.2005.07.007. [DOI] [PubMed] [Google Scholar]
  36. Mastronarde DN, Held SR. Automated tilt series alignment and tomographic reconstruction in IMOD. Journal of Structural Biology. 2017;197:102–113. doi: 10.1016/j.jsb.2016.07.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. McCafferty CL, Klumpe S, Amaro RE, Kukulski W, Collinson L, Engel BD. Integrating cellular electron microscopy with multimodal data to explore biology across space and time. Cell. 2024;187:563–584. doi: 10.1016/j.cell.2024.01.005. [DOI] [PubMed] [Google Scholar]
  38. McEwen BF, Downing KH, Glaeser RM. The relevance of dose-fractionation in tomography of radiation-sensitive specimens. Ultramicroscopy. 1995;60:357–373. doi: 10.1016/0304-3991(95)00082-8. [DOI] [PubMed] [Google Scholar]
  39. Meng EC, Goddard TD, Pettersen EF, Couch GS, Pearson ZJ, Morris JH, Ferrin TE. UCSF ChimeraX: tools for structure building and analysis. Protein Science. 2023;32:e4792. doi: 10.1002/pro.4792. [DOI] [PMC free article] [PubMed] [Google Scholar]
  40. Moebel E, Martinez-Sanchez A, Lamm L, Righetto RD, Wietrzynski W, Albert S, Larivière D, Fourmentin E, Pfeffer S, Ortiz J, Baumeister W, Peng T, Engel BD, Kervrann C. Deep learning improves macromolecule identification in 3D cellular cryo-electron tomograms. Nature Methods. 2021;18:1386–1394. doi: 10.1038/s41592-021-01275-4. [DOI] [PubMed] [Google Scholar]
  41. Peck A, Yu Y, Schwartz J, Cheng A, Ermel UH, Kandel S, Kimanius D, Montabana E, Serwas D, Siems H, Wang F, Zhao Z, Zheng S, Haury M, Agard D, Potter C, Carragher B, Harrington K, Paraan M. Annotating CryoET volumes: a machine learning challenge. bioRxiv. 2024 doi: 10.1101/2024.11.04.621686. [DOI]
  42. Pettersen EF, Goddard TD, Huang CC, Meng EC, Couch GS, Croll TI, Morris JH, Ferrin TE. UCSF ChimeraX: structure visualization for researchers, educators, and developers. Protein Science. 2021;30:70–82. doi: 10.1002/pro.3943. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Radon J. Über die Bestimmung von Funktionen durch ihre Integralwerte längs gewisser Mannigfaltigkeiten. Berichte Über Die Verhandlungen Der Sächsische Akademie Der Wissenschaften. 1917;69:262–277. [Google Scholar]
  44. Rohou A, Grigorieff N. CTFFIND4: fast and accurate defocus estimation from electron micrographs. Journal of Structural Biology. 2015;192:216–221. doi: 10.1016/j.jsb.2015.08.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  45. Rosenthal PB, Henderson R. Optimal determination of particle orientation, absolute hand, and contrast loss in single-particle electron cryomicroscopy. Journal of Molecular Biology. 2003;333:721–745. doi: 10.1016/j.jmb.2003.07.013. [DOI] [PubMed] [Google Scholar]
  46. Sanchez RM, Zhang Y, Chen W, Dietrich L, Kudryashev M. Subnanometer-resolution structure determination in situ by hybrid subtomogram averaging - single particle cryo-EM. Nature Communications. 2020;11:3709. doi: 10.1038/s41467-020-17466-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Saxton WO, Baumeister W, Hahn M. Three-dimensional reconstruction of imperfect two-dimensional crystals. Ultramicroscopy. 1984;13:57–70. doi: 10.1016/0304-3991(84)90057-3. [DOI] [PubMed] [Google Scholar]
  48. Schiøtz OH, Kaiser CJO, Klumpe S, Morado DR, Poege M, Schneider J, Beck F, Klebl DP, Thompson C, Plitzko JM. Serial Lift-Out: sampling the molecular anatomy of whole organisms. Nature Methods. 2024;21:1684–1692. doi: 10.1038/s41592-023-02113-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  49. Song K, Shang Z, Fu X, Lou X, Grigorieff N, Nicastro D. In situ structure determination at nanometer resolution using TYGRESS. Nature Methods. 2020;17:201–208. doi: 10.1038/s41592-019-0651-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  50. Taniguchi R, Orniacki C, Kreysing JP, Zila V, Zimmerli CE, Böhm S, Turoňová B, Kräusslich HG, Doye V, Beck M. Nuclear pores safeguard the integrity of the nuclear envelope. Nature Cell Biology. 2025;27:762–775. doi: 10.1038/s41556-025-01648-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. Tegunov D, Cramer P. Real-time cryo-electron microscopy data preprocessing with Warp. Nature Methods. 2019;16:1146–1152. doi: 10.1038/s41592-019-0580-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. Tegunov D, Xue L, Dienemann C, Cramer P, Mahamid J. Multi-particle cryo-EM refinement with M visualizes ribosome-antibiotic complex at 3.5 Å in cells. Nature Methods. 2021;18:186–193. doi: 10.1038/s41592-020-01054-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  53. Tuijtel MW, Cruz-León S, Kreysing JP, Welsch S, Hummer G, Beck M, Turoňová B. Thinner is not always better: optimizing cryo-lamellae for subtomogram averaging. Science Advances. 2024;10:eadk6285. doi: 10.1126/sciadv.adk6285. [DOI] [PMC free article] [PubMed] [Google Scholar]
  54. Turoňová B, Hagen WJH, Obr M, Mosalaganti S, Beugelink JW, Zimmerli CE, Kräusslich HG, Beck M. Benchmarking tomographic acquisition schemes for high-resolution structural biology. Nature Communications. 2020a;11:876. doi: 10.1038/s41467-020-14535-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  55. Turoňová B, Sikora M, Schürmann C, Hagen WJH, Welsch S, Blanc FEC, Bülow S, Gecht M, Bagola K, Hörner C, Zandbergen G, Landry J, Azevedo NTD, Mosalaganti S, Schwarz A, Covino R, Mühlebach MD, Hummer G, Krijnse Locker J, Beck M. In situ structural analysis of SARS-CoV-2 spike reveals flexibility mediated by three hinges. Science. 2020b;370:203–208. doi: 10.1126/science.abd5223. [DOI] [PMC free article] [PubMed] [Google Scholar]
  56. Turoňová B. GAPStop(TM)—GPU accelerated python-base stopgap for template matching. v0.3Zenodo. 2024a doi: 10.5281/zenodo.10822455. [DOI]
  57. Turoňová B. Turonova/cryocat. v0.6.1GitHub. 2024b https://github.com/turonova/cryoCAT
  58. Turoňová B, Wan W. In: Cryo-Electron Tomography. Förster F, Briegel A, editors. Springer International Publishing; 2024c. Subtomogram averaging; pp. 219–256. [DOI] [Google Scholar]
  59. Turoňová B. Gapstop_tm. 22d461edGitLab. 2025 https://gitlab.mpcdf.mpg.de/bturo/gapstop_tm
  60. Wan W, Kolesnikova L, Clarke M, Koehler A, Noda T, Becker S, Briggs JAG. Structure and assembly of the Ebola virus nucleocapsid. Nature. 2017;551:394–397. doi: 10.1038/nature24490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  61. Wan W. STOPGAP. v0.7.1GitHub. 2023 https://github.com/wan-lab-vanderbilt/STOPGAP
  62. Wan W, Khavnekar S, Wagner J. STOPGAP: an open-source package for template matching, subtomogram alignment and classification. Acta Crystallographica. Section D, Structural Biology. 2024;80:336–349. doi: 10.1107/S205979832400295X. [DOI] [PMC free article] [PubMed] [Google Scholar]
  63. Waskom ML. seaborn: statistical data visualization. Journal of Open Source Software. 2021;6:3021. doi: 10.21105/joss.03021. [DOI] [Google Scholar]
  64. Xing H, Taniguchi R, Khusainov I, Kreysing JP, Welsch S, Turoňová B, Beck M. Translation dynamics in human cells visualized at high resolution reveal cancer drug action. Science. 2023;381:70–75. doi: 10.1126/science.adh1411. [DOI] [PubMed] [Google Scholar]
  65. Xue L, Lenz S, Zimmermann-Kogadeeva M, Tegunov D, Cramer P, Bork P, Rappsilber J, Mahamid J. Visualizing translation dynamics at atomic detail inside a bacterial cell. Nature. 2022a;610:205–211. doi: 10.1038/s41586-022-05255-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  66. Xue H, Zhang M, Liu J, Wang J, Ren G. Cryo-electron tomography related radiation-damage parameters for individual-molecule 3D structure determination. Frontiers in Chemistry. 2022b;10:889203. doi: 10.3389/fchem.2022.889203. [DOI] [PMC free article] [PubMed] [Google Scholar]
  67. Zhang K. Gctf: Real-time CTF determination and correction. Journal of Structural Biology. 2016;193:1–12. doi: 10.1016/j.jsb.2015.11.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  68. Zheng QS, Braunfeld MB, Sedat JW, Agard DA. An improved strategy for automated electron microscopic tomography. Journal of Structural Biology. 2004;147:91–101. doi: 10.1016/j.jsb.2004.02.005. [DOI] [PubMed] [Google Scholar]
  69. Zivanov J, Nakane T, Forsberg BO, Kimanius D, Hagen WJ, Lindahl E, Scheres SH. New tools for automated high-resolution cryo-EM structure determination in RELION-3. eLife. 2018;7:e42166. doi: 10.7554/eLife.42166. [DOI] [PMC free article] [PubMed] [Google Scholar]

eLife Assessment

Axel T Brunger 1

This convincing contribution addresses a question of practical importance: when collecting tilt-series data, what is the optimal angular step size between successive tilt images? The work provides valuable practical insights into cryo-ET data acquisition by demonstrating that balancing two competing demands—sufficient dose per individual tilt image and fine angular sampling—is essential to achieve high-quality tomographic reconstructions. They demonstrate that tilt-series acquired with finer increments (1-3°) yield superior alignment accuracy and improved template-matching performance.

Reviewer #1 (Public review):

Anonymous

This work addresses a question of practical importance that had never been systematically analysed in the cryo-ET field: when collecting tilt-series data, what is the optimal angular step size between successive tilt images? Due to the upper limit in electron exposure (100 - 150 e⁻/Ų), this question is important, since finer angular sampling improves attainable reconstruction resolution (Crowther criterion) but reduces the signal-to-noise ratio of each individual image, potentially compromising both image quality and the ability to computationally align successive frames. To address this, the authors designed a thorough benchmarking study comparing five tilt increments (1°, 2°, 3°, 5°, and 10°) while keeping the total dose and tilt range constant. They evaluated the consequences at every stage of the cryo-ET workflow - from raw image quality and tilt-series alignment, through template matching for ribosome detection, to high-resolution subtomogram averaging - with the goal of providing the community with an evidence-based recommendation for data acquisition.

The manuscript is well written, and the experimental design is carefully thought out. The work provides valuable practical insights into cryo-ET data acquisition by demonstrating that balancing two competing demands - sufficient dose per individual tilt image and fine angular sampling - is essential to achieve high-quality tomographic reconstructions. The identification of a practical optimum at 3° tilt increment is the key contribution of the work. It will be interesting to see in the future whether this optimum shifts for smaller molecular targets, and how emerging tilt interpolation strategies such as cryoTIGER may interact with the choice of experimental angular increment.

Comments on revised version.

Well done! I really like the manuscript and from my point of view it's an excellent piece of work and super useful for the community. Thank you so much for the meticulous work!

Reviewer #2 (Public review):

Anonymous

The determination of macromolecular structures directly within their native cellular environment is becoming increasingly routine, making standardized data collection strategies essential. In this manuscript, Tuijtel et al. provide a timely and valuable contribution by benchmarking key acquisition parameters and establishing practical guidelines for in situ cryo-electron tomography (cryo-ET). Critically, the authors present a systematic framework for optimizing data collection to achieve the highest attainable resolution.

Using Dictyostelium cells as a model system, the authors generate multiple datasets at a constant total dose while varying the tilt increment. They demonstrate that tilt-series acquired with finer increments (1-3°) yield superior alignment accuracy and improved template-matching performance, resulting in higher-quality reconstructions than those collected with coarser increments (5° or above). Furthermore, the authors show that for subtomogram averaging, a 3° tilt increment outperforms all other conditions tested, particularly after per-particle refinement as implemented in M.

Comments on revised version.

The authors have addressed all my concerns, and I have no further issues.

eLife. 2026 Aug 24;15:RP111639. doi: 10.7554/eLife.111639.3.sa3

Author response

Maarten Willem Tuijtel 1, Tomáš Majtner 2, Beata Turoňová 3, Martin Beck 4

The following is the authors’ response to the original reviews.

Public Reviews:

Reviewer #1 (Public review):

This work addresses a question of practical importance that had never been systematically analysed in the cryo-ET field: when collecting tilt-series data, what is the optimal angular step size between successive tilt images? Due to the upper limit in electron exposure (100 - 150 e-2), this question is important, since finer angular sampling improves attainable reconstruction resolution (Crowther criterion) but reduces the signal-to-noise ratio of each individual image, potentially compromising both image quality and the ability to computationally align successive frames. To address this, the authors designed a thorough benchmarking study comparing five tilt increments (1°, 2°, 3°, 5°, and 10°) while keeping the total dose and tilt range constant. They evaluated the consequences at every stage of the cryo-ET workflow - from raw image quality and tilt-series alignment, through template matching for ribosome detection, to high-resolution subtomogram averaging - with the goal of providing the community with an evidence-based recommendation for data acquisition.

The manuscript is well written, and the experimental design is carefully thought out. The work provides valuable practical insights into cryo-ET data acquisition by demonstrating that balancing two competing demands - sufficient dose per individual tilt image and fine angular sampling - is essential to achieve high-quality tomographic reconstructions. The identification of a practical optimum at 3° tilt increment is the key contribution of the work. It will be interesting to see in the future whether this optimum shifts for smaller molecular targets, and how emerging tilt interpolation strategies such as cryoTIGER may interact with the choice of experimental angular increment.

The conclusions of this paper are mostly well supported by data, but some aspects of data analysis need to be clarified and/or extended, including:

(1) Line 109: The authors state that the tilt range was kept at ± 60° relative to the lamella plane. Assuming a typical lamella pre-tilt of ~10°, the absolute stage tilt would approach its mechanical limit. Two clarifications would be appreciated: (a) What was the average pre-tilt across all lamellae? (b) How many dark tilt images, if any, were excluded during tomogram reconstruction?

We thank the reviewer for asking for further clarification. For all our datasets, the pre-tilt of the stage was +8° with the lamella untilted under the e-beam, resulting in a tilt range of -52° to + 68°, thereby not reaching the mechanical limit, which is 70° for our microscope stage.

Regarding “dark tilt images”, for most datasets, we did not need to remove many tilt images. However, we now noticed notably more absence of images from higher tilt values for the 1° dataset (see SFig 1). When analysing further, we noticed that for this dataset, we did not actively remove many images prior to tomogram reconstruction, but rather that they were not acquired in the first place by SerialEM. During acquisition, SerialEM performs various safeguarding checks that can abort the acquisition of a tilt series (or of a single branch). As this seems predominantly a problem for the 1° tilt-increment dataset, we have decided to add this to the manuscript as follows, including the figure as new SFig 1.

In the main text:

“For most datasets, image acquisition was largely complete, with the exception of the 1° dataset, which showed a markedly higher proportion of missing images at high tilt angles (SFig. 1). Closer inspection revealed that many of these images were not acquired, as SerialEM applies built-in safeguards (e.g. autofocus inconsistency or insufficient image counts) that can abort a tilt-series branch before completion.”

(2) Line 148: "When analysing tomographic volumes, we found that tomograms from data with a smaller increment displayed higher SNR values (see Fig. 2B)." It would be helpful to specify which comparisons are statistically meaningful (e.g. Mann-Whitney U test?). While the difference between 1° and 2° appears pronounced, the differences between 2°, 3°, and 5° seem minimal. From my point of view, reporting the mean SNR values +/- standard deviations for each condition would already indicate some significance. Furthermore, since SNR is expected to depend on lamella thickness, it should be clarified whether the average lamella thickness is comparable across the five datasets.

We have now calculated the mean and standard deviation of the tomogram SNR, as follows:

Author response table 1.

mean=0.0612 std=0.0259 n=63
mean=0.0406 std=0.0091 n=119
mean=0.0381 std=0.0135 n=119
mean=0.0322 std=0.0117 n=122
10° mean=0.0145 std=0.0055 n=89

Furthermore, we performed a statistical significance test. Kruskal-Wallis test confirmed significant differences in SNR across tilt increments (H=270.97, p<0.001). Pairwise Mann-Whitney U tests with Bonferroni correction revealed significant differences between all pairs except 2° and 3° (p=0.093), suggesting these two conditions indeed yield comparable SNR.

Lastly, we have now added data regarding local lamella thickness for all tilt-series used in the study, as displayed in SFig. 4.

We incorporated this in the manuscript as follows.

In the main text:

“When analysing tomographic volumes, we found that tomograms from data with a smaller increment displayed higher SNR values (see Fig. 2B and Supplementary Note), whilst showing a similar lamella thickness distribution (see SFig. 4A).”

and:

“Firstly, we selected ca. 20 tomograms per condition, based on tomogram content and local lamella thickness [31] (for more details, see Methods and SFig. 4B).”

As a supplementary note:

“As the tomogram SNR distribution of particularly the 2° and 3° dataset showed similar SNR distributions, we performed formal significance testing for the data in this panel (see Fig. 2B). Kruskal-Wallis test confirmed significant differences across conditions (H=270.97, p<0.001); pairwise Mann-Whitney U tests with Bonferroni correction revealed all pairs were significantly different except 2° vs. 3° (p=0.093), indicating comparable SNR for these two tilt increments.”

And, adding the test in the Methods:

“To quantify differences in signal-to-noise ratio (SNR) across tilt increment conditions, a non-parametric Kruskal-Wallis test was performed as an omnibus test of the null hypothesis that all groups are drawn from the same distribution. Because SNR distributions were not assumed to be normal, and sample sizes differed across conditions, non-parametric tests were used throughout. Following the omnibus test, all 10 pairwise comparisons between conditions were assessed using two-sided Mann-Whitney U tests. To control for multiple comparisons, raw p-values were adjusted using the Bonferroni correction (multiplied by the number of comparisons, n=10, capped at 1.0). Statistical significance was defined as a Bonferroni-corrected p-value below 0.05. All analyses were performed in Python using the scipy.stats module.”

(3) Line 167: "Indeed, the variation in maximum resolution correlates with lamella thickness across all datasets (see Fig. 2F)." The reported R2 values of 0.30 (1°), 0.38 (2°), 0.66 (3°), 0.61 (5°), and 0.60 (10°) reveal a notably weak linear relationship for the finer tilt increments. It is also difficult to assess whether the lamella thickness distributions are comparable across conditions from the current figures - visually, the 1° dataset appears to be based on thinner lamellae, while the 10° dataset appears to include thicker samples. A histogram of lamella thickness distributions for each condition, provided as supplementary material, would greatly aid interpretation. Given this thickness dependency, reporting mean +/- standard deviation of lamella thickness per condition is highly appreciated.

We have added the full lamella thickness distribution per dataset now in SFig. 4.

The apparent weaker relationship between resolution fit and local lamella thickness for the 1 dataset seems to be largely apparent to the few very thin data points in this data (for more clarity, see the same data plotted separately in Author response image 1). We speculate that this is due to even less signal in these very thin and very low-dose images.

Author response image 1.

Author response image 1.

(4) Figure 4: It should be specified which tomogram subsets were used for the Rosenthal-Henderson analysis, whether lamella thickness was taken into account in the subset selection, and whether ribosomes too close to the lamella edges were excluded. Finally, linear fits should be displayed across the full x-axis range for all tilt increments to facilitate direct visual comparison.

We have described the process of tomogram subset selection in detail in the Methods section Template matching and 3D classification. To further add clarity, we have incorporated the local lamella distribution plots for the full data, as well as specifically for the tomograms subjected to TM and STA in SFig. 4.

Regarding the linear fits, we respectfully disagree with this suggestion. Displaying the linear fits only over the range used for their calculation avoids implying that the linear relationship extends beyond the measured data, and in our view produces a clearer figure.

(5) General: Were ribosomes located at the lamella edges excluded from the analysis? As demonstrated in the authors' own prior work (Tuijtel et al., Science Advances, 2024), Ga-FIB milling induces structural damage at the lamella surfaces. To exclude the influence on the STA results, particles near the lamella edges should be removed prior to analysis, and the criteria for this exclusion should be stated explicitly.

We have not excluded any ribosomes from close to the surface. As we still treated all data the same for each condition, we anticipate that the results of the comparison reported here still hold true.

The aim of the authors was to provide the cryo-ET community with an evidence-based recommendation for the choice of tilt increment, and they largely succeeded in this goal. The identification of 3° as a practical optimum - balancing sufficient dose per tilt image for effective per-particle refinement with fine enough angular sampling for accurate tilt-series alignment - is well supported by the data and consistent across the multiple quality metrics employed. The conclusion that coarser increments (5° and 10°) compromise tomogram quality, template matching accuracy, and STA resolution is robust and clearly demonstrated. However, the conclusion rests entirely on a single biological system using ribosomes as the sole molecular target, which are exceptionally favourable due to their abundance, size, and electron contrast. Whether the identified optimum holds for smaller, lower-abundance, or lower-contrast targets remains an open question.

In future, it would be particularly interesting to test whether emerging tilt interpolation strategies, such as cryoTIGER, which is particularly intriguing, can effectively compensate for coarser experimental angular sampling in post-processing. Here, the optimal experimental increment may shift, and the interaction between these two approaches represents a promising direction for future work. More broadly, as cryo-ET datasets grow larger and public repositories expand, the practical tradeoffs between acquisition time, data storage, and structural quality identified here will become increasingly relevant to the field.

We agree with the reviewer and thank them for this positive assessment. An interesting note to the use of cryoTIGER in particular is that it uses already aligned tilt-series as an input, and it therefore is unlikely to overcome severe alignment issues associated with large tilt-increments.

Reviewer #2 (Public review):

The determination of macromolecular structures directly within their native cellular environment is becoming increasingly routine, making standardized data collection strategies essential. In this manuscript, Tuijtel et al. provide a timely and valuable contribution by benchmarking key acquisition parameters and establishing practical guidelines for in situ cryo-electron tomography (cryo-ET). Critically, the authors present a systematic framework for optimizing data collection to achieve the highest attainable resolution.

Using Dictyostelium cells as a model system, the authors generate multiple datasets at a constant total dose while varying the tilt increment. They demonstrate that tilt-series acquired with finer increments (1-3°) yield superior alignment accuracy and improved template-matching performance, resulting in higher-quality reconstructions than those collected with coarser increments (5° or above). Furthermore, the authors show that for subtomogram averaging, a 3° tilt increment outperforms all other conditions tested, particularly after per-particle refinement as implemented in M.

Overall, the manuscript is clearly written, and the conclusions are well supported by the data presented. I have no major concerns. There are some minor points that the authors should address, including:

(1) The phrase "electron optical density distribution" (line 31, Introduction) should be revised to "electrostatic potential" or "Coulomb potential distribution," which more accurately reflects what is measured in cryo-EM/ET.

We thank the reviewer for this correction and have adjusted it in the text:

“It captures the 3-dimensional (3D) electrostatic potential of the specimen under scrutiny and enables the structural analysis of macromolecular complexes within their native context.”

(2) The authors state that the maximum tolerable electron dose is approximately 100-150 e-2 (line 34, Introduction). This is an oversimplification, as bacterial specimens, for example, have been shown to tolerate doses of 200 e-2 or higher (see Breigel et al., PNAS, 2009; https://www.pnas.org/doi/10.1073/pnas.0905181106#T1). The statement should be revised to reflect this variability.

We adjusted this statement to now read:

“One of these is the maximum electron dose that can be applied to biological specimens before irreversible damage occurs, which is about 100-150 e-/Å2 for most eukaryotic cells.”

(3) Lines 56-57: The authors do not cite their own prior work benchmarking tilt-series acquisition strategies on in vitro samples. This earlier study provides important context and should be referenced and briefly discussed.

We assume the reviewer meant this study: Turonova et al., Nat. Comms. (2020). We have now added and discussed this reference as follows:

“Since accumulated radiation dose progressively degrades high-resolution information, this motivated the development of the dose-symmetric tilt scheme, which prioritizes acquisition of low-tilt images early to better preserve high-resolution information [11, 16].”

Recommendations for the authors:

Reviewer #1 (Recommendations for the authors):

(1) Line 159: "Surprisingly though, the resolution to which the CTF was fitted was similar for all conditions, despite an 8-fold increase in dose (see Fig. 2B, E)." The reference to Figure 2B at this point is unclear.

We thank the reviewer for pointing this out, we have removed the reference to panel B.

(2) Line 162: "As the data shown in Fig. 2D-F pertains to images of the untilted specimen, ..." For clarity, this should also be stated explicitly in the figure caption.

We have added this to the figure legend; “both estimated with Gctf on projection images of the sample at effective zero-tilt position.”

(3) Figure 3: These are compelling results, and the 3D classification outcomes provide an excellent visual representation of the quantitative data shown in Figure 3. Including (some or all) of the initial five classes in the figure would further strengthen this already convincing presentation. Additionally, applying a uniform extraction threshold (e.g., z-score of 3.75 or 5) across all tilt increments would facilitate a more direct comparison. But this is really a minor remark, the authors and the editors may judge if the current presentation is already sufficient.

We have adjusted Figure 3 according to the reviewer’s recommendation:

We referenced this in the main text as:

“To ensure similar data processing strategies for all conditions, extraction thresholds were also lowered for the 1, 2 and 3° conditions, and 3D classification was performed to filter out the junk particles (see Fig. 3 A and SFig. 8). “

In order to directly compare the particle extraction, we have already carried out such a uniform extraction threshold, with a z-score threshold of 5 (apart for the 10° data, where this was not possible). This extraction was then used for the 3D classification that led to the TM analysis and further STA investigations.

Reviewer #2 (Recommendations for the authors):

(1) Supplementary Figure 1: To improve accessibility for a broader readership, the authors should annotate or highlight the key organelles and protein complexes visible in the tomographic slices.

We thank the reviewer for this suggestion, but adding arrowheads made this figure too crowded in our opinion. Furthermore, for most of the panels, the mentioned features of interest is centred in the image panel, which should make identification straightforward.

(2) 'In situ' should be in italics throughout the text.

We have changed this.

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Data Citations

    1. Tuijtel MW, Beck M. 2026. Cryo-ET dataset on lamellae of Dictyostelium discoideum cells acquired with various tilt-increments. EMPIAR. EMPIAR-13391
    2. Tuijtel MW, Beck M. 2026. Optimal tilt-increment for cryo-ET. EMD. EMD-57021 [DOI] [PMC free article] [PubMed]

    Supplementary Materials

    MDAR checklist
    Source code 1. Jupyter notebook used to pseudo-randomly select particles based on whole tomograms for M refinements.
    elife-111639-code1.zip (10.7KB, zip)

    Data Availability Statement

    Relevant cryo-ET density maps generated in this study have been deposited in the EM Data Bank (EMDB) with the following access code: EMD- 57021. All raw data have been deposited to the Electron Microscopy Public Image Archive (EMPIAR) database (accessing code EMPIAR-13391), including acquisition metadata, information, tilt-series alignment data, all raw TM locations, and particle locations after 3D classification and refined particle motive lists. All other data needed to evaluate the conclusions in the study are present in the paper.

    The following datasets were generated:

    Tuijtel MW, Beck M. 2026. Cryo-ET dataset on lamellae of Dictyostelium discoideum cells acquired with various tilt-increments. EMPIAR. EMPIAR-13391

    Tuijtel MW, Beck M. 2026. Optimal tilt-increment for cryo-ET. EMD. EMD-57021


    Articles from eLife are provided here courtesy of eLife Sciences Publications, Ltd

    RESOURCES