Abstract
Platinum electrodes with surface area ratios of four to one were used to detect and enumerate a variety of gram-positive and gram-negative organisms. Linear relationships were established between inoculum size and detection time. End points for platinum electrodes were similar to those obtained with a platinum-reference electrode combination. Shape of the overall response curves and length of detection times for gram-positive organisms were markedly different than those for the majority of gram-negative species. Platinum electrodes are better than the platinum-reference electrode combination because of cost, ease of handling, and clearer definition of the end point.
Full text
PDF![683](https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b3e4/243121/d9e337fdd983/aem00214-0063.png)
![684](https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b3e4/243121/ea5c2c4bc8a6/aem00214-0064.png)
![685](https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b3e4/243121/1794f2321b0d/aem00214-0065.png)
![686](https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b3e4/243121/898a614f7763/aem00214-0066.png)
![687](https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b3e4/243121/50dce758faf3/aem00214-0067.png)
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Lamb V. A., Dalton H. P., Wilkins J. R. Electrochemical method for the early detection of urinary-tract infections. Am J Clin Pathol. 1976 Jul;66(1):91–95. doi: 10.1093/ajcp/66.1.91. [DOI] [PubMed] [Google Scholar]