Classifying Prosthetic Use via Accelerometry in Persons with Trans-Tibial Amputations

Morgan T Redfield 1, John C Cagle 1, Brian J Hafner 1, Joan E Sanders 1
PMCID: PMC4423801  NIHMSID: NIHMS589126  PMID: 24458961

Abstract

Knowledge of how persons with amputation use their prostheses and how this use changes over time may facilitate effective rehabilitation practices and enhance understanding of prosthesis functionality. Perpetual monitoring and classification of prosthesis use may also increase the health and quality of life for prosthetic users. Existing monitoring and classification systems are often limited in that they require the subject to manipulate the sensor (e.g., attach, remove, or reset a sensor), record data over relatively short time periods, and/or classify a limited number of activities and body postures of interest. In this study, a commercially-available three-axis accelerometer (ActiLife ActiGraph GT3X+) was used to characterize the activities and body postures of individuals with trans-tibial amputation. Accelerometers were mounted on prosthetic pylons of ten persons with trans-tibial amputation as they performed a preset routine of actions. Accelerometer data was post-processed using a Binary Decision Tree to identify when the prosthesis was being worn and to classify periods of use as movement (i.e., leg motion like walking or stair climbing), standing (i.e., standing upright with limited leg motion), or sitting (i.e., seated with limited leg motion). Classifications were compared to visual observation by study researchers. The classifier achieved a mean accuracy of 96.6% (SD=3.0%).

Keywords: Accelerometry, Ambulatory Monitoring, Amputees, Artificial Limbs, Activity Monitor

Introduction

Prosthetists, physicians, and prosthetics researchers are challenged to describe how persons with limb loss use their prostheses outside the clinic or laboratory [1]. Performance tests such as the timed up and go test [2] or the six-minute walk [3] can be used to measure mobility of a prosthetic user in a clinic or laboratory [4,5], but information on what prosthesis users do in their daily lives can be difficult to acquire. Characterizing ways that prostheses are used is complicated by the range of situations and environments users encounter. The characterization of prosthesis use could be partially achieved by quantifying prosthetic wear (e.g., donning and doffing) and users’ engagement in locomotor activities (e.g., walking and stair climbing) and fundamental body postures (e.g., standing or sitting). Accurate knowledge of prosthetic use in free-living conditions would enhance prosthetic prescriptions, fitting processes, and measurement of outcomes [6].

Previous methods for measuring prosthetic use outside of a gait laboratory or clinic included self-report surveys and personal activity monitoring devices (e.g., pedometers and step activity monitors). Self-report surveys have been used to quantify frequency and duration of prosthetic use [7]. However, self-report of activity among persons with limb loss has been noted to be unreliable when compared to a step-activity monitor [8]. Pedometers and step-counters have been used to objectively measure step-activity of persons wearing prostheses over extended periods of time [6,813]. While these sensors accurately measure gait activities, they are unable to provide information about body postures that may also be part of a person’s habitual activity [14]. Differentiation of body postures may be clinically important as sitting and standing can affect changes in residual limb volume and alter the fit of a prosthesis [15,16]. Accurate knowledge of how much a prosthetic user sits or stands could thus be useful in determining changes in socket fit throughout the day.

Identification of activities and postures has previously been achieved through classification of data from one or more body-mounted sensors [14,1720]. This technique has been applied to characterize the quality of gait [21], discriminate activity levels [22], and determine body orientations [23] of individuals without amputations. It has also been used on persons with lower limb amputations to quantify step counts [18], estimate ambulation time [24], and describe gait patterns [25]. Algorithms have also been developed to identify locomotion and posture of individuals with an amputation from sensor data obtained over short time periods (i.e., up to several hours) [10,2628].

While these studies demonstrated potential for activity and posture classification based on data from body-mounted sensors, there remain challenges to clinical use such as need for multiple sensors, subject donning requirements, low storage capacities, and short battery lives. Currently available sensors are also often restricted to short-term applications and/or require adherence to specific user protocols. Accordingly, more user-friendly and clinically-relevant solutions are needed to overcome these challenges. Here, we explore the potential for a commercially-available accelerometer and custom signal processing algorithm to identify when prostheses are being worn and to classify periods of use as movement, standing, or sitting. Use of a single sensor mounted to a prosthesis would eliminate the need for the user to attach and remove the sensor, improve wear compliance, and reduce cost. We believe this strategy is a first step toward a prosthesis-integrated sensing system that could be used to collect, process, and convey information of interest to users, practitioners, and researchers.

In this study we used commercially-available accelerometers and a custom software algorithm to classify the movements and body postures of persons with trans-tibial amputation. Our hypothesis was that data from a single prosthesis-mounted accelerometer could be used to identify when a prosthetic user was wearing their prosthesis and if they were sitting, standing, or actively moving. An algorithm was designed to identify when the prosthesis was being worn and to classify actions as movement (i.e., regular leg motion like walking or stair climbing, transitioning from one posture to another, or donning or doffing the prosthesis), standing, or sitting.

Methods and Materials

Experimental Design

Persons with trans-tibial amputations were recruited to test the developed classification algorithm in a semi-controlled activity protocol. All subjects were recruited from local prosthetic clinics, peer support groups, and hospitals. Inclusion criteria were ages 18 to 75, trans-tibial amputation that occurred at least two years prior to testing, Medicare Functional Classification Level (MFCL) 2 (limited community ambulator) or higher [29], a healthy residual limb with intact skin, and ability to walk for at least an hour (with rests as necessary). Subjects provided informed consent prior to participation. All procedures were approved by a University of Washington Institutional Review Board (IRB) before study procedures were initiated.

ActiLife ActiGraph (Pensacola, Florida) GT3X+ accelerometers were applied to each subject to measure limb segment accelerations. The GT3X+ has a +/− 6g (gravitational acceleration) dynamic range, 0.00293g resolution, 100Hz maximum sampling rate, up to 31 days of battery life, and up to 40 days of data storage. A dynamic range of +/− 6g has previously been found to be acceptable for quantifying movement patterns during walking [31,32]. The Actigraph accelerometer is packaged in a 4.6×3.3×1.5cm water-resistant enclosure and weighs 19 grams. One accelerometer was attached to the subject’s prosthesis, proximal to the foot. Positioning the accelerometer at this location, instead of one more proximal, ensured the sensor was subjected to high accelerations (i.e., received a strong signal) during leg motions. The sensor was oriented with the positive x-axis along the limb axis and the positive z-axis in the medial-lateral direction (Figure 1). A second accelerometer was affixed to the anterior thigh on the same leg as the prosthesis. It was oriented with the positive x-axis along the limb axis and the positive y-axis facing to the subject’s right. These locations ensured that different postures would be easily differentiated using anterior-posterior acceleration data. A sheet of Tegaderm (3M) was placed on the skin of the thigh. Adhesive backed Velcro was then used to attach the accelerometer to the Tegaderm (Figure 1). The second accelerometer was further secured with an elastic strap to minimize local movements. Sampling rates for both accelerometers were 40Hz. A 40Hz sampling rate was deemed acceptable as most of the energy in gait is concentrated below the resulting Nyquist frequency of 20Hz [33,34]. Using a sample rate of 40Hz rather than a higher rate typically used in gait laboratories, maximized the duration of data collection and still allowed for the identification of relevant gait events. At this sampling rate, GT3X+ accelerometers were capable of recording data for up to 30 days.

Figure 1.

Figure 1

Accelerometer attachments and orientations. One accelerometer was connected securely to the subject’s pylon with the z-axis facing in the medial-lateral direction and the x-axis facing in the long direction. A second accelerometer was attached securely to the subject’s thigh with the y-axis facing to the right and the z-axis facing in the anterior-posterior direction.

Two experiments were performed to assess the accuracy of a novel algorithm designed to classify use of prostheses as 1) movement (e.g., walking, using stairs, or transitioning from one posture to another), 2) standing (i.e., upright standing posture with minimal movement), 3) sitting (i.e., seated posture with minimal movement), or 4) doffed (i.e., prosthesis not being worn) based on data from the pylon-mounted accelerometer. The thigh-mounted accelerometer was used in both experiments to enhance accuracy of activity and posture recognition so as to validate classifications derived by an algorithm run only on the single accelerometer data. Two algorithms were developed to classify movement, posture, and wear. One algorithm made use of data from only the pylon-mounted accelerometer, and the other algorithm made use of data from both accelerometers.

The purpose of the first experiment was to assess the accuracy of single- and dual-accelerometer classifications of activities performed in a laboratory setting compared to visual observation. Subjects were asked to perform a pre-defined activity protocol that included walking over level ground (i.e., an indoor hallway); sitting on office chairs, sofas, or benches; standing; ascending and descending stairs; and doffing and donning the prosthesis. These activities were deemed to be most representative of the activities of clinical interest to prosthetists. Subjects were asked to sit in each type of seat at least three times, stand and walk the hallway at least five times, use the stairs at least once, and doff/don their prosthesis once. Each time the subject performed an activity in the test sequence, they were asked to perform it for at least 60 seconds. Subjects were asked to perform the sequence in a set order, but were asked to engage in each activity, posture, or don/doffing action as they normally would (i.e., no instructions were given for walking speed, sitting posture, etc.). Subjects were visually monitored while they performed the test sequence. A researcher followed and timed each subject with a stopwatch to capture when subjects started and stopped each type of movement, posture, or don/doffing action. This record was used for ground-truth comparisons.

The purpose of the second experiment was to assess differences between single- and dual-accelerometer classifications of activities performed in free-living settings. In the second experiment, two subjects were asked to wear both accelerometers for two days outside of the laboratory. Subjects were asked to go about their days normally. When they doffed their prosthesis, they were to remove the thigh mounted accelerometer and set it on a flat surface. Use of Velcro and elastic band for accelerometer attachment allowed the subject to easily reattach the thigh mounted accelerometer.

Classification Algorithm

Raw acceleration data, such as those shown in Figure 2, obtained from the GT3X+ accelerometers were post-processed using custom algorithms written using MathWorks Matlab 7.12.0 (Natick, Massachusetts) software. No filtering was performed on the data. Data were buffered into short windows with an overlap of 50%. Window length was experimentally determined as described below and set to 45 samples (i.e., 1.125s).

Figure 2.

Figure 2

Plots of signal magnitude area and pylon acceleration signals for different activities and postures. If SMA was higher than the upper threshold then the subject was considered engaged in movement. If the SMA was between the thresholds, the subject was considered stationary. If the SMA was below the lower threshold for over 320 seconds then the prosthesis was considered doffed. The above plot shows SMA in dB to accurately show the difference between the thresholds. The lower threshold, set to 0.01g corresponds to −40dB and the upper threshold of 0.1g corresponds to −20dB.

A binary decision tree (BDT) algorithm [17] was designed to classify the windowed data. Data from all three axes of the pylon-mounted accelerometer were used to determine if the subject was active, stationary (i.e., sitting or standing), or had doffed the prosthesis. Determination of posture was performed using only the anterior-posterior data of one accelerometer. For the algorithm that used data from only the pylon-mounted accelerometer, data from the anterior-posterior axis of that accelerometer was used. For the algorithm that used data from both accelerometers, data from the anterior-posterior axis of the thigh-mounted accelerometer was used.

The BDT (Figure 3) used signal magnitude area (SMA) to determine if the prosthesis was moving or stationary within each window. SMA was calculated by subtracting the mean (μ) of the window from each of the accelerometer axes, integrating the absolute value of the result over a full window, and dividing by the window size (T). This method has previously been used to detect activity levels [20,23]. SMA was evaluated using the following equation:

SMA=1T0T[X(t)μX+Y(t)μY+Z(t)μZ]dt

The developed algorithm required several subject-specific parameters for calibration. First, the accelerometers’ locations (i.e., left or right leg) were required to correctly orient the pylon-mounted accelerometer’s anterior-posterior axis; the axial direction did not change and the y-axis of the accelerometer rotated with leg change to remain pointing in the lateral direction. Second, the pylon accelerometer’s inclination while the prosthesis was doffed and standing upright with the foot on the floor was required. This doffed position served as a reference to differentiate sitting and standing postures. This strategy was effective because the anterior-posterior inclination angle (with respect to the vertical axis) was found to be greater than the doffed reference angle for sitting and less than it for standing (Figure 4).

Figure 3.

Figure 3

The binary decision tree algorithm used for activity and posture classification. Thigh acceleration signals (dashed line) were considered only in cases where two accelerometers were used

Figure 4.

Figure 4

Pylon and thigh acceleration signals over a sixty second period when the subject was sitting, had doffed their prosthesis and place the foot flat on floor in the reference position, or was standing.

Two activity thresholds were used to guide classifications. The lower and upper activity thresholds were experimentally determined via a sensitivity analysis using the laboratory-based experiment data as described below. These thresholds were set to 0.01g and 0.1g, respectively (Figure 2). When SMA was below the lower threshold, the subject was deemed either to be stationary or to have doffed their prosthesis. When SMA remained below the lower threshold for more than 320s, the prosthesis was considered doffed. The 320s parameter was chosen based on the observation that subjects in stationary postures during the first experiment were not completely immobile for the full length of time they were in the posture, but further research will be needed to validate it. Otherwise, the prosthesis was assumed to be donned and windows were classified as a stationary posture (i.e., standing or sitting). When SMA was between the lower and upper thresholds, the accelerometer data from that window were averaged to find the inclination [35]. Inclination was then compared to the subject’s reference inclination to determine if the subject was sitting, standing, or doffed (Figure 2). If the prosthesis was oriented in a way that did not correspond to one of those postures, indicated by the inclination being outside of a range that could be obtained by a sitting or standing individual, the window was classified as unknown. Lastly, when SMA exceeded the upper threshold, the subject was considered to be engaged in movement.

When data from only the pylon-mounted accelerometer was used for classification, sitting or standing was calculated based on the inclination of the prosthesis as determined by anterior-posterior accelerometer measurements. When data from both the pylon-mounted accelerometer and the thigh-mounted accelerometer were used for classification, the pylon-mounted accelerometer data was used to determine movement, and data from the anterior-posterior axis of the thigh-mounted accelerometer was used to determine posture (Figure 4).

A sensitivity analysis was performed on the three experimentally-determined parameters of the classification algorithm. Window size, upper activity threshold, and lower activity threshold were varied to determine the optimal values described above and to assess the sensitivity of the results to changes in those parameters.

Accuracy Evaluation

Classification accuracy for each experiment was calculated by comparing (on a window-by-window basis) the predicted activity or body posture to that which was recorded by the researcher during the experiment. The accuracy was computed as the ratio of the number correctly classified windows to the total number of windows that were evaluated. Accuracy was calculated for both single- and dual- accelerometer classification algorithms. Confusion matrices, a method of describing which postures and activities were misclassified and as what posture or activity they were misclassified, were calculated for each subject [36]. A mean confusion matrix was also calculated using classifications from all subjects. For the second experiment, accuracy in differentiating sit and stand postures and activities was quantified by comparing the prediction of the BDT algorithm using only data from the pylon-mounted accelerometer to that from both accelerometers.

Results

Eight subjects (six male and two female) were recruited to participate in the first (i.e., lab-based test sequence) experiment and two subjects (both male) were recruited to participate in the second (i.e., multiple day validation) experiment (Table 1). All subjects had a unilateral trans-tibial amputation. Subjects’ mean age was 53.0 (SD=11.6), mean weight was 90.4 kg (SD=11.6 kg), mean height was 178.0 cm (SD=7.2 cm), and mean time since amputation was 19.6 years (SD=13.9 years). Subjects were classified as MFCL-2 (n=3), MFCL-3 (n=3), and MFCL-4 (n=4) by the study prosthetist based on interview and clinical evaluation.

Table 1.

Subject demographics.

Subject Limb
side
gender age etiology years
since
amputation
residuum
length
(cm)
weight
(kg)
height
(cm)
MFCL
1 r m 69 trauma 47 15 99.5 182.9 3
2 r m 58 trauma 5 17 80.9 177.8 4
3 l f 56 trauma 9 17 100.0 167.6 2
4 r m 31 Infection 2 16 88.7 170.2 3
5 l m 49 tumor tumor 12 19 108.5 177.8 2
6 l m 49 trauma 22 10 100.0 182.9 4
7 l m 36 trauma 2 23 98.2 190.5 3
8 r f 65 trauma 58 10 78.4 167.0 2
9 r m 65 trauma 11 17 72.7 175.3 4
10 r m 52 trauma 28 19 77.3 188.0 4
Mean - - 53 - 19.6 16.2 90.4 178.0 3.10
SD - - 12 - 18.4 3.7 11.7 7.8 0.83

Window lengths from 20 to 80 samples were tested to determine the effect of window length on classification accuracy. The sensitivity of both classification algorithms to window length was similar. Both algorithms had a maximum classification accuracy at approximately 45 samples per window (Figure 5).

Figure 5.

Figure 5

Sensitivity of classification accuracy to window length. The classification accuracy reached a maximum at 45 samples per window for both algorithms that were tested.

Accuracies for classifications derived using lower activity thresholds between 0.001 and 0.02g and upper activity thresholds between 0.01 and 0.2 g were computed. It was found that if the activity thresholds were low, accuracy decreased substantially because stationary postures (e.g., standing or sitting) were classified as active use (Figure 6). As thresholds increased from their optimal value, accuracy dropped off due to periods of activity being misclassified as sitting, standing, or doffed. Maximum accuracy was achieved for the pylon data classification algorithm using a 0.01 g lower threshold and a 0.1 g upper threshold. Maximum accuracy was achieved for the pylon and thigh data classification algorithm using a 0.008 g lower threshold and a 0.1 g upper threshold. These optimal thresholds were used in all subsequent analyses.

Figure 6.

Figure 6

Sensitivity of classification accuracy to activity thresholds. Classification accuracy reached a maximum when the lower threshold was 0.01g and the upper threshold was 0.1g. Accuracy decreases significantly if lower thresholds are chosen, but higher thresholds result in smaller losses in accuracy.

In experiment 1, overall classification accuracy for each subject ranged from 90.1% to 99.6% when using data from only the pylon-mounted accelerometer (Table 2). Mean classification accuracy was 96.6% (SD=3.0%). The most commonly misclassified body posture was sitting, which was typically misclassified as standing. The confusion matrix (Table 3) shows the percentage of actions that were classified correctly or misclassified as other activities.

Table 2.

Classification accuracies for all subjects during the lab-based test sequence. Subject accuracies ranged from 90%-99%.

Subject 1 2 3 4 5 6 7 8 Mean SD
Ankle Only 0.996 0.968 0.901 0.991 0.962 0.975 0.993 0.940 0.966 0.030
Ankle and Thigh 0.996 0.997 0.920 0.994 0.987 0.998 0.993 0.996 0.985 0.025

Table 3.

The confusion matrix for activities in the first experiment shows that there were few unclassifiable activities when using only the pylon-mounted monitor. The most commonly misclassified activities were sit and stand. The rows of the confusion matrix represent the visually identified activity or posture. The columns represent the classifications generated by the binary decision tree (BDT) algorithm.

known\predicted doffed sit stand active
doffed 0.987 0.005 0.005 0.003
sit 0.017 0.919 0.054 0.010
stand 0.003 0.025 0.967 0.005
active 0.000 0.001 0.002 0.997

The algorithm that used data from both accelerometers had classification accuracy equal to or better than the algorithm that used only data from the pylon-mounted accelerometer. Average classification accuracy using both accelerometers was 98.5% (SD=2.5%), 1.9% greater than that obtained using a single accelerometer (Table 2).

In experiment 2, mean agreement between classification with one and both accelerometers was 90.3% (SD=5.2%). Classification of activities for one subject (Subject 9) was 84.7% while for the other subject (Subject 10) it was 95.8%.

Discussion

We demonstrated that a prosthesis-mounted ActiLife ActiGraph GT3X+ three-axis accelerometer can be used to identify the postures and movement of persons with trans-tibial amputation. We also showed that differentiation of sit and stand postures is possible using knowledge of the prosthesis doffed with the foot flat on the floor. A custom algorithm classified data gathered from users as they performed actions in a free-living situation. In a laboratory-based experiment, use of a pylon-mounted accelerometer alone or in conjunction with a thigh-mounted accelerometer allowed for classification of movement and posture with accuracy greater than 90.1%, which may be deemed acceptable for clinical and/or scientific applications.

Although the developed algorithm accurately classified movements and postures in general (Table 2), it was challenged in select situations. For example, the algorithm commonly misclassified sitting as standing when subjects were seated on a high bench. Although sitting accuracy was above 98.2% for low chairs, subjects often oriented their prosthesis at a right angle to the ground in higher chairs rather than extending their prosthetic leg forward as they did in a lower seat. As such, standing may be overestimated (and sitting underestimated) if users sit with their knees flexed at 90 degrees. Use of a second accelerometer may eliminate this issue, but may introduce other problems, such as compliance or improper attachment and removal. It may be possible to reduce the number of single-accelerometer misclassifications by using a probabilistic model to account for transitions between activities.

The classification algorithm achieved a comparatively low accuracy of 90.1% for one test subject in experiment 1. This subject wore a brace on her contra-lateral limb, which may have been a factor in the observed low classification accuracy. Further research may be required to investigate the effects of braces and other walking aids (e.g., canes and walkers) on the classification algorithm.

This movement and posture classification algorithm was capable of identifying the specified behaviors with accuracies of 94.0% with only a single accelerometer for all but one prosthesis user (Table 2). It also appears that using a second accelerometer may not be necessary for accurate classification, as it increased classification accuracy by an average of only about 1.9% during the laboratory based experiment. Agreement between single accelerometer classification and dual accelerometer classification in the second (free living) experiment was 84.7-95.8%. The large difference in results between the two subjects is most likely due to the subjects standing or sitting in different ways, but more research must be done to investigate.

This activity monitoring method may therefore avoid the need for multiple monitors as used in other studies [18,23,27,37]. The developed system appears to meet or exceed classification accuracies reported in related studies on able-bodied and elderly populations [17,18,23,37]. Our system also used fewer sensors than other systems [14,19,37]. Our classification system is designed to detect sitting and standing postures as well as movement, unlike currently available prosthetic monitors such as the StepWatch3 (Orthocare Innovations, Mountlake Terrace, WA) and the Patient Activity Monitor (Reykjavik, Iceland) [18].

Additional research is needed before this classification strategy can be recommended for clinical use. The system must be validated more thoroughly while prosthesis users move through their free-living environments. The choice of classification parameters, including the upper and lower SMA thresholds and the 320s stationary period, must also be validated. One potential limitation in use of the developed posture and activity classification strategy is the unknown effect of riding in a motorized vehicle. Further research is required to isolate or account for external accelerations to which a user may be subjected while riding in vehicles. Additionally, postures (e.g., lying down) or specific activities (e.g., stair climbing) of interest to researchers should be explored to ensure they can be appropriately classified with this system.

Clinical uses of this technology (e.g., rehabilitation training, componentry evaluation, etc.) should be explored to determine if the accuracy of which this system is capable is sufficient for such applications. The long battery life and large storage capacity of the ActiGraph GT3X+ accelerometer suggest that this sensor may be suitable for long-term data collection. Anticipated advances in battery and storage technologies will also likely extend the length of time such sensors can be used to monitor subjects. Such research will help to determine if prosthesis-integrated monitoring systems can enhance clinical care and improve quality of life for prosthesis users.

Conclusion

A classification algorithm was developed to identify periods of prosthesis use and to discriminate activities and body postures of individuals with a lower-limb amputation. Periods of movement, standing, sitting or a doffed prosthesis were accurately classified in a laboratory based experiment more than 92.0% of the time when using data from two body-mounted accelerometers and more than 90.1% of the time when using data from a single accelerometer. These data suggest that activities and body postures can be well classified using this classification algorithm and data from a single, commercially-available accelerometer. More research is required to validate the system in situations where the activity duration and type are not controlled. The authors believe that the information derived from this system will provide valuable clinical insight into how persons with trans-tibial amputation use their prostheses in their free-living environments and that such evidence can be used to facilitate prosthetic treatment and rehabilitation of persons with limb loss. It may also be useful for automatic feedback control to adjust prosthesis mechanisms based on activity and posture.

Acknowledgements

The authors would like to acknowledge Kate Allyn for her assistance with data collection.

The project described was supported by Award Number EEC-1028725 from the National Science Foundation and Award number 1R01HD069387 from the National Institutes of Health. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Science Foundation or of the National Institutes of Health.

The described research protocol was approved by the University of Washington Human Subjects Division.

Funding Sources National Science Foundation Award Number EEC-1028725 National Institutes of Health Award number 1R01HD069387

Abbreviations

BDT

Binary Decision Tree

g

Gravitational acceleration

SMA

Signal Magnitude Area

MFCL

Medicare Functional Classification Level

SD

Standard Deviation

References

  • 1.Condie E, Scott H, Treweek S. Lower limb prosthetic outcome measures: a review of the literature 1995 to 2005. Journal of Prosthetics and Orthotics. 2006;18(6):13–45. [Google Scholar]
  • 2.Podsiadlo D, Richardson S. The timed“ Up & Go”: a test of basic functional mobility for frail elderly persons. Journal of the American geriatrics Society. 1991;39(2):142–148. doi: 10.1111/j.1532-5415.1991.tb01616.x. [DOI] [PubMed] [Google Scholar]
  • 3.Enright PL. The six-minute walk test. Respiratory care. 2003 Aug;48(8):783–5. [PubMed] [Google Scholar]
  • 4.Lin S-J, Bose NH. Six-minute walk test in persons with transtibial amputation. Archives of physical medicine and rehabilitation. the American Congress of Rehabilitation Medicine and the American Academy of Physical Medicine and Rehabilitation. 2008 Dec;89(12):2354–9. doi: 10.1016/j.apmr.2008.05.021. [DOI] [PubMed] [Google Scholar]
  • 5.Schoppen T, Boonstra A, Groothoff JW, De Vries J, Göeken LN, Eisma WH. The Timed “up and go” test: reliability and validity in persons with unilateral lower limb amputation. Archives of physical medicine and rehabilitation. 1999 Jul;80(7):825–8. doi: 10.1016/s0003-9993(99)90234-4. [DOI] [PubMed] [Google Scholar]
  • 6.Dudek NL. Ambulation monitoring of transtibial amputation subjects with patient activity monitor versus pedometer. Journal of Rehabilitation Research and Development. 2008 Dec 1;45(4):577–586. doi: 10.1682/jrrd.2007.05.0069. [DOI] [PubMed] [Google Scholar]
  • 7.Pezzin LE, Dillingham TR, MacKenzie EJ, Ephraim P, Rossbach P. Use and satisfaction with prosthetic limb devices and related services. Archives of Physical Medicine and Rehabilitation. 2004 May;85(5):723–729. doi: 10.1016/j.apmr.2003.06.002. [DOI] [PubMed] [Google Scholar]
  • 8.Stepien JM, Cavenett S, Taylor L, Crotty M. Activity levels among lower-limb amputees: self-report versus step activity monitor. Archives of Physical Medicine and Rehabilitation. 2007 Jul;88(7):896–900. doi: 10.1016/j.apmr.2007.03.016. [DOI] [PubMed] [Google Scholar]
  • 9.Parker K, Kirby RL, Adderson J, Thompson K. Archives of Physical Medicine and Rehabilitation. 4. Vol. 91. Elsevier Inc.; Apr, 2010. Ambulation of people with lower-limb amputations: relationship between capacity and performance measures; pp. 543–9. [DOI] [PubMed] [Google Scholar]
  • 10.Ramstrand N, Nilsson K. Validation of a patient activity monitor to quantify ambulatory activity in an amputee population. Prosthetics and Orthotics International. 2007;31(2):157–166. doi: 10.1080/03093640600988617. [DOI] [PubMed] [Google Scholar]
  • 11.Hafner BJ, Willingham LL, Buell NC, Allyn KJ, Smith DG. Evaluation of function, performance, and preference as transfemoral amputees transition from mechanical to microprocessor control of the prosthetic knee. Archives of Physical Medicine and Rehabilitation. 2007 Feb;88(2):207–17. doi: 10.1016/j.apmr.2006.10.030. [DOI] [PubMed] [Google Scholar]
  • 12.Bussmann JBJ, Culhane KM, Horemans HLD, Lyons GM, Stam HJ. Validity of the prosthetic activity monitor to assess the duration and spatio-temporal characteristics of prosthetic walking. IEEE Transactions on Neural Systems and Rehabilitation Engineering : A Publication of the IEEE Engineering in Medicine and Biology Society. 2004 Dec;12(4):379–86. doi: 10.1109/TNSRE.2004.840495. [DOI] [PubMed] [Google Scholar]
  • 13.Chou TGR, Webster JB, Shahrebani M, Roberts TL, Bloebaum RD. Characterization of Step Count Accuracy of Actigraph Activity Monitor in Persons With Lower Limb Amputation. Journal of Prosthetics and Orthotics. 2009 Oct;21(4):208–214. [Google Scholar]
  • 14.Frossard L, Stevenson N, Sullivan J, Uden M, Pearcy M. Categorization of Activities of Daily Living of Lower Limb Amputees During Short-Term Use of a Portable Kinetic Recording System: A Preliminary Study. Journal of Prosthetics and Orthotics. 2011 Jan;23(1):2–11. [Google Scholar]
  • 15.Sanders JE, Fatone S. Residual limb volume change: Systematic review of measurement and management. Journal of Rehabilitation Research and Development. 2011;48(8):949. doi: 10.1682/jrrd.2010.09.0189. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Sanders JE, Harrison DS, Allyn KJ, Myers TR. Clinical utility of in-socket residual limb volume change measurement: case study results. Prosthetics and Orthotics International. 2009 Dec;33(4):378–90. doi: 10.3109/03093640903214067. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Mathie MJ, Celler B. Classification of basic daily movements using a triaxial accelerometer. Medical & Biological Engineering & Computing. 2004;42(5):679–687. doi: 10.1007/BF02347551. [DOI] [PubMed] [Google Scholar]
  • 18.Godfrey A, Conway R, Meagher D, OLaighin G. Direct measurement of human movement by accelerometry. Medical Engineering & Physics. 2008 Dec;30(10):1364–86. doi: 10.1016/j.medengphy.2008.09.005. [DOI] [PubMed] [Google Scholar]
  • 19.Lester J, Choudhury T. A hybrid discriminative/generative approach for modeling human activities. In: Leslie Pack Kaelbling AS, editor. International Joint Conference on Artificial Intelligence. Edinburgh, Scotland, UK: 2005. [Google Scholar]
  • 20.Mathie MJ, Coster a CF, Lovell NH, Celler BG. Detection of daily physical activities using a triaxial accelerometer. Medical & Biological Engineering & Computing. 2003 May;41(3):296–301. doi: 10.1007/BF02348434. [DOI] [PubMed] [Google Scholar]
  • 21.Atallah L, Aziz O, Yang G, Lo B. Detecting walking gait impairment with an ear-worn sensor. 6th International Workshop on Wearable and Implantable Body Sensor Networks; Berkeley, California. 2009. pp. 175–180. [Google Scholar]
  • 22.Hendrick P, Bell ML, Bagge PJ, Milosavljevic S. Can accelerometry be used to discriminate levels of activity? Ergonomics. 2009 Aug;52(8):1019–25. doi: 10.1080/00140130902846464. [DOI] [PubMed] [Google Scholar]
  • 23.Curone D, Bertolotti GM, Cristiani A, Secco EL, Magenes G. A real-time and self-calibrating algorithm based on triaxial accelerometer signals for the detection of human posture and activity. IEEE Transactions on Information Technology in Biomedicine : a Publication of the IEEE Engineering in Medicine and Biology Society. 2010 Jul;14(4):1098–105. doi: 10.1109/TITB.2010.2050696. [DOI] [PubMed] [Google Scholar]
  • 24.Stam H. A device for long term ambulatory monitoring in trans-tibial amputees. Prosthetics and Orthotics International. 1995;19:53–55. doi: 10.3109/03093649509078232. [DOI] [PubMed] [Google Scholar]
  • 25.Selles RW, Formanoy M a G, Bussmann JBJ, Janssens PJ, Stam HJ. Automated estimation of initial and terminal contact timing using accelerometers; development and validation in transtibial amputees and controls. IEEE Transactions on Neural Systems and Rehabilitation Engineering : A Publication of the IEEE Engineering in Medicine and Biology Society. 2005 Mar;13(1):81–8. doi: 10.1109/TNSRE.2004.843176. [DOI] [PubMed] [Google Scholar]
  • 26.Bussmann JB, Schrauwen HJ, Stam HJ. Daily physical activity and heart rate response in people with a unilateral traumatic transtibial amputation. Archives of Physical Medicine and Rehabilitation. 2008 Mar;89(3):430–4. doi: 10.1016/j.apmr.2007.11.012. [DOI] [PubMed] [Google Scholar]
  • 27.Bussmann HB, Reuvekamp PJ, Veltink PH, Martens WL, Stam HJ. Validity and reliability of measurements obtained with an “activity monitor” in people with and without a transtibial amputation. Physical Therapy. 1998 Sep;78(9):989–98. doi: 10.1093/ptj/78.9.989. [DOI] [PubMed] [Google Scholar]
  • 28.Ooi S, Osman N. Ambulatory Function Monitor for Amputees. 5th International Conference on Biomedical Engineering; Kuala Lumur, Malaysia. 2011. pp. 732–734. [Google Scholar]
  • 29.CGS Administrators L . Juisdiction C Supplier Manual. Nashville, TN: 2011. pp. 4–5. [Google Scholar]
  • 30.Santos-Lozano A, Marín PJ, Torres-Luque G, Ruiz JR, Lucía A, Garatachea N. [cited 2013 Feb 3];Technical variability of the GT3X accelerometer. Medical engineering & physics [Internet] Institute of Physics and Engineering in Medicine. 2012 Jul;34(6):787–90. doi: 10.1016/j.medengphy.2012.02.005. Available from: http://www.ncbi.nlm.nih.gov/pubmed/22417978. [DOI] [PubMed] [Google Scholar]
  • 31.Ledoux WR, Hillstrom HJ. Acceleration of the calcaneus at heel strike in neutrally aligned and pes planus feet. Clinical Biomechanics (Bristol, Avon) 2001 Aug;16(7):608–13. doi: 10.1016/s0268-0033(01)00041-9. [DOI] [PubMed] [Google Scholar]
  • 32.Kavanagh JJ, Menz HB. Accelerometry: a technique for quantifying movement patterns during walking. Gait & posture. 2008 Jul;28(1):1–15. doi: 10.1016/j.gaitpost.2007.10.010. [DOI] [PubMed] [Google Scholar]
  • 33.Antonsson E, Mann R. The frequency content of gait. Journal of Biomechanics. 1985;18(1):39–47. doi: 10.1016/0021-9290(85)90043-0. [DOI] [PubMed] [Google Scholar]
  • 34.Angeloni C, Riley PO, Krebs DE. Frequency content of whole body gait kinematic data. IEEE Transactions on Rehabilitation Engineering. 1994 Mar;2(1):40–46. [Google Scholar]
  • 35.Kwapisz J, Weiss G, Moore S. Activity recognition using cell phone accelerometers. ACM SIGKDD Explorations. 2011;12(2):74–82. [Google Scholar]
  • 36.Schuldt C, Laptev I, Caputo B. Recognizing Human Actions : A Local SVM Approach. Proceedings of the 17th International Conference on Pattern Recognition.2004. pp. 32–36. [Google Scholar]
  • 37.Najafi B, Aminian K, Loew F. An ambulatory system for physical activity monitoring in elderly. 1st Annual International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine & Biology; Lyon, France. 2000. pp. 562–566. [Google Scholar]

RESOURCES