Abstract
A germ-free isolator must have a perfect bacterial filter. This paper describes a new, relatively inexpensive, stainless-steel filter frame, which is easily and quickly assembled and protects the enclosed filter material at all times. Resistance to the flow of air was less than 4 inches of water at an airflow of 30 ft3/min through the filter frame with 204 inches2 of surface area and four, one-half inch thick pieces of fiberglass filter material. This filter performed satisfactorily in our gnotobiotic laboratory and was found to be consistently 100% efficient in removing an aerosol containing Serratia marcescens from an air stream under a variety of operating conditions.
Full text
PDFImages in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- HUMPHREY A. E. Air sterilization. Adv Appl Microbiol. 1960;2:301–311. doi: 10.1016/s0065-2164(08)70132-1. [DOI] [PubMed] [Google Scholar]
- MAY K. R., HARPER G. J. The efficiency of various liquid impinger samplers in bacterial aerosols. Br J Ind Med. 1957 Oct;14(4):287–297. doi: 10.1136/oem.14.4.287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- TREXLER P. C., REYNOLDS L. I. Flexible film apparatus for the rearing and use of germfree animals. Appl Microbiol. 1957 Nov;5(6):406–412. doi: 10.1128/am.5.6.406-412.1957. [DOI] [PMC free article] [PubMed] [Google Scholar]