. 2020 May 21;19:100192. doi: 10.1016/j.pacs.2020.100192

Correction of wavelength-dependent laser fluence in swept-beam spectroscopic photoacoustic imaging with a hand-held probe

MinWoo Kim a,*, Geng-Shi Jeng b, Matthew O’Donnell a, Ivan Pelivanov a
PMCID: PMC7339128  PMID: 32670789

Abstract

Recently, we demonstrated an integrated photoacoustic (PA) and ultrasound (PAUS) system using a kHz-rate wavelength-tunable laser and a swept-beam delivery approach. It irradiates a medium using a narrow laser beam swept at high repetition rate (∼1 kHz) over the desired imaging area, in contrast to the conventional PA approach using broad-beam illumination at a low repetition rate (10−50 Hz). Here, we present a method to correct the wavelength-dependent fluence distribution and demonstrate its performance in phantom studies using a conventional limited view/bandwidth hand-held US probe. We adopted analytic fluence models, extending diffusion theory for the case of a pencil beam obliquely incident on an optically homogenous turbid medium, and developed a robust method to estimate fluence attenuation in the medium using PA measurements acquired from multiple fiber-irradiation positions swept at a kHz rate. We conducted comprehensive simulation tests and phantom studies using well-known contrast-agents to validate the reliability of the fluence model and its spectral corrections.

Keywords: Photoacoustics, Spectroscopic imaging, Fluence model, Laser fluence correction, Quantitative PA imaging

1. Introduction

Quantitative spectroscopic photoacoustic (PA) imaging relies on the known optical absorption spectrum of target chromophores to estimate their concentrations. The PA signal, however, depends not only on the optical absorption spectrum of a target but also on the wavelength-dependent optical fluence at the same site. A critical challenge for in vivo, PA spectroscopic imaging is predicting light transport noninvasively in a turbid medium [1,2]. The spatial distribution of fluence depends on the optical absorption and scattering properties of that medium, but they are usually unknown in advance. In addition, since light attenuation in a medium is a function of wavelength, the medium induces spectral distortion (sometimes called ‘coloring’) between the nominal PA-measured optical absorption spectrum and the true target’s spectrum [1]. Therefore, for true quantitative spectroscopic (or multi-wavelength) PA imaging that can identify molecular constituents in the medium under study, the optical fluence must be simultaneously estimated and compensated. This requires not only an appropriate fluence model, but also an effective method to estimate the fluence distribution at each wavelength from PA measurements.

Fluence correction methods have been proposed for homogenous absorption and scattering media such that the light distribution and PA measurements can be closely represented by simple formulas or mathematical expressions [[3], [4], [5], [6], [7]]. Unfortunately, very few of them translate into clinical tools because most explicitly depend on prior knowledge of the medium’s optical properties. Indeed, optical constants reported in the literature may vary a few orders of magnitude depending on the measurement technique, tissue condition, and geometry [8]. Background tissue properties may also change dynamically based on tissue blood content and oxygenation level. Optical constants measured even with the same device, the same experimental conditions, and for the same person may vary. Furthermore, these properties may change during medical procedures and interventions.

Many groups have also proposed laser fluence correction techniques without prior knowledge of tissue optical properties [[9], [10], [11], [12], [13], [14], [15], [16], [17], [18], [19], [20]]. Most rely on absorbing structures restored by tomographic array detectors (i.e., fully surrounding the target) or a large planar array detector. Typically, model-based schemes invert background tissue optical properties with approximated light transport models [[9], [10], [11], [12]]. However, they are very sensitive to the geometry of both the background and targets and require full view and bandwidth detection.

Many recent studies have shown the potential of deep learning schemes to access more complex tissue structures and heterogeneous optical properties within a short computation time [[13], [14], [15], [16], [17], [18], [19]]. Their primary challenge is the generation of realistic training data/images. Despite success in simulations, they have not been shown yet to work broadly for in-vitro or in-vivo conditions.

Unfortunately, full view and bandwidth systems used for small animal PA imaging cannot be easily adapted to most clinical applications in humans. Due to limited light penetration, PA systems intended for most clinical applications must use hand-held, limited view array transducers [21], leading to strong artifacts in structural PA images [22]. Diagnostic ultrasound (US) co-registered with PA (e.g. PAUS methods) has many advantages for clinical translation of PA technology, but the limited view/bandwidth of US arrays degrades PA reconstruction. Even with these constraints, PA imaging, particularly spectroscopic imaging, brings a molecular dimension to US. However, there have been no successful demonstrations to date of wavelength-dependent laser fluence correction for these systems that is appropriate for in vivo, let alone ex vivo, applications.

This study focuses on wavelength-dependent laser fluence correction for PA imaging using a limited view/bandwidth diagnostic US probe [21] appropriate for real-time monitoring of interventional procedures. The main target is identifying small and sparse absorbers such as labelled drugs and contrast agents or quantifying functional changes in microvessels in quasi-homogenous optically turbid tissue. In this geometry, fluence correction should be constrained and physics-based because limited view and narrow bandwidth lose some low-frequency components during reconstruction (beamforming) of light-absorbing structures [23]. Note that laser fluence estimation without the help of other tools and at near real-time rates is highly desirable.

In Ref. [24], point source tissue illumination from different positions along the tissue surface was proposed to calculate the laser fluence distribution within the medium and then apply the estimated fluence for PA spectral decomposition. However, the authors did not show how to integrate this method into real-time PA scanners since broad beam illumination is most commonly used in PA imaging systems [24]. Leveraging the results of this study, we recently introduced a real-time interleaved photoacoustic-ultrasound (PAUS) fast-sweep scanner [25], where unlike previous delivery systems coupling laser pulses into all fibers in a bundle simultaneously, light is coupled into individual fibers sequentially (see Fig. 1), but at a very high rate.

Fig. 1.

Fig. 1

Real-time integrated photoacoustic and ultrasound (PAUS) system used in these studies. The ultrasound system programmably controls the laser, motor controller, and US transducer. The motor controller synchronizes emission with the centers of 20 fibers in the bundle, delivering a trigger to the US system when properly aligned for each fiber. The US system then externally triggers the compact laser, transmitting a pulse at about a 1 kHz repetition rate with a wavelength switchable from pulse-to-pulse over the range from 700 nm to 900 nm. With absolute position control, a precise rate is not needed for external laser triggering, ensuring maximal light delivery to each fiber. Motor speed variations only slightly alter the overall frame rate of 50 Hz. A total of 20 fibers are arranged on two sides of the linear array US transducer, as shown in the zoomed front view in the bottom right corner. The system mixes laser beam and focused US transmissions for interleaved PA spectroscopic imaging and US B-mode imaging.

A unique diode-pumped wavelength tunable (700 nm - 900 nm) laser emitting about 1 mJ pulses at 1000 Hz, with wavelength switching in less than 1 ms for any arbitrary wavelength order, was designed especially for the fast-sweep PAUS scanner. To maximize exposure, we illuminate with a narrow (∼ 1 mm in diameter) beam and switch it from fiber-to-fiber at 1000 Hz, resulting in one loop around the US probe per single-wavelength frame in only 20 ms (Fig. 1). The next loop uses another wavelength without delay; the procedure repeats over all wavelengths. That is, instead of illuminating with a broad beam, we use fast-scanning (or fast-sweep) over the same illumination area.

In our PAUS system, ten fibers are uniformly spaced along each elevational edge of the US array (e.g. 20 fibers in total, Figs. 1,2). Every laser shot from a single fiber is followed by a sub-image reconstruction, i.e. 20 sub-images in all, which are then coherently summed to form the full PA frame. The kHz rate enables 50 full ‘loops’ of the laser beam around the probe per second, resulting in a 50 Hz PA frame rate. For stable spectral decomposition, 10 wavelengths (i.e., 700, 715−875 nm every 20 nm) form the spectroscopic sequence. More details on the system can be found in Ref [25].

Here we describe in detail how to use partial PA images from every fiber to estimate laser fluence. Indeed, when light emerges from different fibers, it propagates different distances to a target and, therefore, the amplitude of a partial PA image reconstructed from a single fiber illumination will depend on that fiber’s position around the US probe (see Fig. 2). For every fiber light source, we adopted analytical fluence models extending diffusion theory for narrow beam (pencil beam) illumination obliquely incident on a semi-infinite homogeneous scattering medium [26]. Using this model, we explore methods to extract optical parameters to assess the fluence distribution within the medium using data acquired at different fiber positions and wavelengths. In particular, estimated parameters include the effective light attenuation coefficient μeff, and reduced light scattering μs' coefficients of a turbid medium.

Fig. 2.

Fig. 2

Schematic diagram of laser sources and bulk media. Each fiber is mounted on the side of a linear-array transducer in an ambient medium (Medium II), and irradiation at the tip of the fiber into a scattering medium (Medium I) is modeled as a pencil beam. The center position of the face between the transducer and Medium I is (0,0,0). The image plane is (y = 0). The y-directional position of every irradiation point is either 5.68 mm or -5.68 mm, and the tilt angle of the laser beam is θ=35°. The pencil beam can be represented as two isotropic point sources satisfying the boundary condition that light propagation from z<zb into z>zb is approximately 0. The extrapolated boundary z=zb is determined by the refractive indexes of the two media and the transducer. The sign of imaginary sources indicates their polarity. Source positions are determined by the transport mean free path lt.

Building on the basic analytic model, we then explore robust methods to estimate these parameters from noisy data. Simulations were conducted to clarify quantitative errors arising from adopted optical fluence models, variations in medium properties, and measurement noise levels. Finally, we demonstrate the reliability of the correction methods via phantom studies.

2. Model

2.1. Photoacoustic signal

Acoustic pressure obtained for the k th optical fiber, optical wavelength λj at discrete position ri can be denoted as

pj,kri=Γμ¯ajriΦj,kri, (1)

where k{1,2,,20} and j{0,1,,9}, Γ is the Gru¨nesien parameter, μ¯a is the optical absorption coefficient of a target and Φ is the light fluence. We assume here that the parameter Γr governing PA efficiency is constant over space {r|r=x,y,z,rΩ} and wavelength. This assumption is not critical for relative spectroscopic measurements, but is important for absolute concentration estimates of specific chromophores producing the PA signal. The fluence distribution Φj,kr depends on the optical properties of the medium. Absorption arising from imaged targets μ¯a also disturbs the fluence, but the effect is insignificant if the target is small [27]. We thus assume that the acoustic pressure (PA signal) is linearly proportional to the absorption coefficient of the target.

The absorption spectrum is given as

μ¯ajri=l=1Lαl(j)Clri, (2)

where L is the number of chromophore (absorber) types and Cl and αl(j) are the concentration and absorption coefficient of the l th chromophore type, respectively. The ultimate goal of quantitative spectroscopic PA imaging is to estimate the relative, or absolute, concentration of a particular chromophore of interest at each position within the imaging field from measurements given the known absorption spectrum αl(j) of that chromophore. Thus, correct modeling and computation of the wavelength-dependent optical fluence distribution Φj,kri at every source position is required to guarantee accurate quantification of the target chromophore.

2.2. Optical fluence model

Light illumination in a medium can be simply expressed using the diffusion approximation,

Qr=μa(r)Φr-D(r)Φr, (3)

where Q denotes the source, μa denotes the absorption coefficient within a medium, and Φ denotes the fluence [26]. The parameter D is the diffusion coefficient given as

D=13μs',μs'=μs1-g, (4)

where μs is the scattering coefficient of the medium, μs' is the reduced scattering coefficient and g is the anisotropy factor. This expression is derived from the radiative transfer equation (RTE) for high scattering (μsμa) and a nearly isotropic medium [26]. We assume that biological tissue can be regarded as ‘macro-homogeneous’. If the scale of optical micro-heterogeneities is smaller than a few photon transport mean free paths, light fluence will be smooth and can be characterized by a macro-homogeneous medium model [28]. For a homogeneous medium Dr=D,μar=μa,μs'r=μs' and a point source (Qr=δr-r'), the solution to the diffusion equation using the Green’s function is given as

Φr=14πDr-r'exp-μeffr-r', (5)

where μeff=3μaμs' [[29], [30], [31]] is the effective attenuation coefficient and r' is the spatial position of the source.

The solution can be modified for our acquisition environment in which a pencil beam is obliquely incident on a semi-infinite scattering medium. As illustrated in Fig. 2, the beam can be converted into two isotropic point sources mirror-symmetric about the extrapolated boundary z=zb due to the refractive-index mismatch between media [26]. The source below the boundary helps describe the transition from ballistic to scattering regimes, and its symmetrical source is added to satisfy a boundary condition described in Appendix A1. Note that every fiber is mounted on the transducer and covered by a thick acrylic holder and protection glass (BK-7 optical glass). Thus, the main contributor to the mismatch with the scattering medium is the solid transducer rather than the ambient medium (air). Appendix A1 describes the detailed computation of zb using the refractive indexes n of the media.

The relative positions of the point sources with respect to the boundary are determined by the transport mean free path lt=1μs' [26]. For the k th fiber whose tip position is rk'=(xk',yk',zk') and incident angle θ, the fluence solution can be expressed as

Φkr=α1exp-μeffr-rk+4πDr-rk+-exp-μeffr-rk-4πDr-rk-, (6)

where α1 is a scalar, rk+=(xk',yk'-ltsinθ,zk'+ltcosθ) and rk-=(xk',yk'-ltsinθ,-zk'-ltcosθ-2zb). Note that the unknown parameters are α1, μs' and μeff. Other parameters can be derived from them or known initially.

If the positions of two imaginary isotropic sources in Fig.2 are close to each other due to a very small lt (e.g. very high μs'), the fluence can be simplified to an asymptotic expression

Φkr=α2zk(1+μeffr-rk)r-rk3exp-μeffr-rk, (7)

where α2 is a scalar. The derivation is presented in Appendix (B). Here, unknown parameters are only α2 and μeff, i.e. μeff is the only parameter defining the relative distribution of laser fluence in the medium. Held et al. employed this model [22].

In this paper, analytic fluence expressions in Eqs. (6) and (7) are called Model I and II, respectively. Thus, when Model II is quite accurate, it makes fluence estimation from experimental data very simple and stable. We will explore below how well both models compare to Monte-Carlo simulations, and under what conditions these models can be used for laser fluence assessment from swept-beam PA measurements.

2.3. Optical fluence estimation

PA measurement can be expressed as

yj,kri=pj,kri+nkri, (8)

where j and k are wavelength and optical fiber indexes, respectively. nk is the system noise containing a DC component bk and zero-mean Gaussian noise n¯k. Control data were recorded by assigning zero laser power at the first wavelength (j = 0) to estimate the noise bias as bk=1ΩiΩy0,kri, where Ω is the cardinality of the entire domain Ω. The unbiased measurement can then be obtained as

y¯j,kri=yj,kri-bk=pj,kri+n¯kri,j1,2,,9, (9)

Note that the unknown parameters needed to estimate pj,kri are μs'(j), μeff(j), and βi,j=Γα1μ¯ajri provided Model I is employed. To enhance estimation efficiency, we selected position indexes as Ω={i|j,ky¯j,kri>τ} where τ denotes the threshold value. Also, we normalized measurements to reduce the degrees of freedom as

yj,kri=y¯j,krik=120y¯j,kriΓμ¯ajriΦj,kri+n¯krik=120Γμ¯ajriΦj,kri=Φj,krik=120Φj,kri+n˘kri, (10)

where iΩ, and yj,kri now depends on two parameters, μs'(j) and μeff(j) for every i, j and k.

The normalized version of fluence Φj,kri=Φj,krik=120ϕj,kri ranges from 0 to 1. The optimal parameters for the j th wavelength can be estimated as

μˆeffj,μˆs'(j)=argmin(μeffj,μs'(j))riΩwik=120yj,kri-Φjkrik=120Φjkri2 (11)

where wi=j,ky¯j,kri is the weight such that a position with higher SNR contributes more to the estimate. Likewise, Model II can be used for estimation, but the search parameter is only μˆeffj for every wavelength index j. The fluence estimate Φˆj,kri can be obtained by substituting optical parameter estimates into either Model I or II.

2.4. Fluence correction of light absorption spectrum

Assume that one type of chromophore is located at a local position of interest and its absorption spectrum is known as αj. The optical absorption spectrum obtained from PA measurements dj=ky¯j,k is distorted due to the wavelength-dependent fluence of the surrounding turbid medium (biological tissue, for example), as shown in Eq. (1). The spectrum can be corrected using the fluence estimate Φˆj,k as cj=k(Φˆjky¯j,k)kΦˆjk2, obtained by solving the least-squares problem, y¯j,k=cjΦˆjk+nk, for every wavelength j. By comparing normalized dj and cj with aj, fluence estimation accuracy can be computed.

3. Numerical simulations

3.1. Simulation parameters

The primary purpose of simulations is to verify the proposed fluence models and associated estimation methods described in Sections 2.2 and 2.3. First, Models I and II were compared to ground truth Monte Carlo simulations. We adopted the medium geometry of Fig. 2 for all simulations. All media were approximately represented as cubical shapes. Their sizes, optical parameters, and refractive indexes are summarized in Table 1. We assumed that one of the fibers transmits light into the scattering medium at the interface. The location and tilt angle of the fiber tip are (0 mm, 5.68 mm, 0 mm) and 35, respectively. The total number of photon packets for Monte Carlo tests is 20 million. The simulation used the open source program, MCX Studio [32].

Table 1.

Range of parameters in numerical simulations.

Cuboid size
(xmm×ymm×zmm)
μa
(cm-1)
μs'
(cm-1)
Refractive index n
Ambient medium (50×50×50) 0.0 0.00 1.00
Scattering medium (50×50×50) 0.01-0.05 5-35 1.33
Transducer medium (30×20×30) 1000.0 1000.0 1.49

We also compared errors in parameter estimation for Model I compared to Model II at a fixed optical wavelength. We generated synthetic data using Model I (Eq. (6)) assuming that Model I is very close to ground truth. The noise power is set to zero to find the contribution of model difference to the error. We assumed a point target located at position {r=(xmm,0mm,zmm)} in a scattering medium whose optical parameters are μa, μs' and μeff. For every (z, μs'), we conducted 100 simulation tests to estimate μeff by randomly changing values of other parameters in the range. A fractional error sample for every test is given as ε=(μeff-μˆeff)/μeff×100 (%), where μˆeff denotes the estimate from Model I. We averaged 100 error samples for every (z, μs') to monitor the mean error pattern ε¯(z, μs') over penetration depth and scattering coefficient.

In the last simulation, we investigated the accuracy of parameter estimation for different wavelengths in a turbid medium where μa and μs' are wavelength dependent. We generated synthetic data using Model I as in the previous simulation, and added white Gaussian noise to mimic experimental conditions and explore reconstruction algorithm stability to noise. The total number of SNR levels was 15, ranging from 20 dB to 50 dB, and the number of test data sets for each SNR was 100. SNR is defined as

SNR=10log10120k=120pk2/σn2, (12)

where σn2 is the noise variance. For every test, we assumed SNR is constant over wavelength. A point target is located at position {ri=(0,0,10mm)} in a scattering medium whose optical parameters are μs'(j), μa(j) and μeff(j) at the j th wavelength.

According to the literature [8,33], the scattering coefficient in brain varies the most over wavelength of all tissue types

μs'λ=40.8λ/500nm-3.089, (13)

providing the most complicated conditions for light absorption spectrum reconstruction using PA signals. We used this scattering function for the test medium in the third simulation but assumed the absorption coefficient μaj(=0.03cm-1) is constant for all wavelengths. This does not reduce the generality of the results because (as we show below) reconstruction error is not highly sensitive to variations in light absorption.

Optical parameters were estimated from noise contaminated data using Eq. (11) based on either Model I or II. Then, the fluence estimate Φˆj,k(ri) was compared with ground truth Φj,k(ri) using the correlation coefficient as

ρ=j(ψˆjψj)jψˆj2jψj2, (14)

where ψˆj=kΦˆj,k(ri)-19j,kΦˆj,k(ri) and ψj=kΦj,k(ri)-19j,kΦj,k(ri). We examined the mean correlation coefficient using 100 samples for every SNR level.

3.2. Simulation results

Fig. 3(a) shows slices of a 3D fluence distribution Φx,y,z from the Monte Carlo simulation. We used the distribution of optical fluence Φ0,0,z along the z axis as a reference to compare with corresponding distributions calculated using Model I and Model II (given by Eqs. (6) and (7) respectively). Fig. 3(c–e) show the distributions obtained for different μs' (2 cm-1, 5 cm-1 and 10 cm-1, respectively). The elevational position y' of the source is fixed at 5.70 mm, which corresponds to the fiber positions in our experimental transducer array (see Fig. 2). For extremely small μs'(=2cm-1), Monte Carlo produces a spike since ballistic transport crosses the image plan. As expected, Model I is very close to Monte Carlo results for diffusive conditions because scattering dominates in the medium. Model II approximates Model I when the measurement point is located at distances from the source and the interface much larger than the transport mean free path lt. The higher the scattering, the closer the agreement between Model II and Monte Carlo, as shown in Fig. 3(d) and (e). We also compared fluence models at different lateral source positions (y' is 2.85 mm, 5.70 mm and 11.40 mm, see Fig. 3(f–h)). Scattering μs' was set to 10 cm-1.

Fig. 3.

Fig. 3

(a) Slices (planes) of 3D light fluence Φ(x,y,z) in a scattering medium. The magnitude distribution is obtained by Monte Carlo (MC) simulation. (b) Position (axial) zmax of the maximum optical fluence location along line (0,0,z) when a fiber source is located at different lateral positions (0,y',0). The graphs represent zmax over y' for several scattering (e.g. different μs') conditions. (c-e) Distribution of optical fluence for different μs', but fixed source position y'. (f-h) Distribution of optical fluence for different source position y' but fixed μs'. The light absorption coefficient of the medium is μa= 0.03cm-1.

The position zmax of maximum optical fluence along the z axis increases with increasing distance y' between the image plane and light source, and Model II converges quickly to Model I soon after zmax. Fig. 3(b) summarizes the shift in zmax as a function of source position y' for different μs' typical of biological tissue [8]. For the source position in our system (y'=5.7mm), zmax barely changes for μs' varying from 5-20 cm-1. This means that simplified Model II converges to MC and Model I at very similar depths over a wide range of tissue scattering. For our transducer configuration, zmax is around 3 mm. This fact is very important for practical fluence assessment because μs' is not known a priori.

Based on these initial simulations, it is clear that Model I can be used as ground truth if the light source is located more than a distance lt from the light source. We now focus on the difference in parameter estimation error (bias) between Models I and II in more detail. The fractional error for Model II is defined as (ΦI-ΦII)/ΦI×100 (%), where ΦI and ΦII denote Model I and Model II, respectively. Fig. 4(a) shows the error over z/lt for several μs' over the tissue range. The error is under 10 % when z/lt is larger than ∼10. Fig. 4(b) shows the error for several assumed μa values. Note that the μa variation rarely changes the error. Fig. 4(c) illustrates the fractional error ε¯(z, μs') for 5<μs'<35cm-1 and 0<z<40mm when Model II is used to estimate μeff. Fig. 4(d) displays pixel values of (c) over the z/lt (= zμs') axis.

Fig. 4.

Fig. 4

(a) Fractional errors (relative percentage changes) of Model II with respect to Model I over z/lt when the reduced scattering coefficient μs' is 5, 10, 20 and 30cm-1. The common light absorption coefficient μa is 0.03cm-1. (b) Fractional error over z/lt when the μa value is 0.01, 0.02, 0.03, 0.04 and 0.05cm-1. The common reduced scattering coefficient μs' is 10cm-1. (c) Mean fractional error ε¯(z, μs') of μeff estimates with respect to ground truth over 2D domains, axial depth z, and reduced scattering coefficient μs'. (d) Mean fractional error ε¯(z, μs') over 1D axis z/lt = zμs'. Pixel values of (c) are represented by dots on this graph.

Fig. 5 presents optical parameter estimates using Models I and II when data are contaminated by noise. We used Eq. (13) to set ground-truth scattering coefficient variations over wavelength (715−875nm). Fig. 5(a–c) show estimates when the SNR is the highest (50 dB). Fig. 5(d) shows the correlation between the estimated optical fluence spectrum and ground truth over a wide range of SNR. Model I estimation error (bias and deviation) is only a function of measurement noise. Estimates of μeff are unbiased at high SNR while those of μs' are biased and their errors are high even at this high SNR level. However, despite large error in μs' estimated by Model I, the correlation between resultant fluence estimates and ground truth is close to 1. As shown in Fig. 5(d), the lower SNR level causes lower correlation because both estimation deviation and bias are higher. In Model II, estimation error is caused by model discrepancy as well as measurement noise. Note that Model II underestimates μeff at high SNR, but the ratio of estimate μˆeff to ground truth μeff remains almost constant over the wavelength range. Thus, it provides competitive correlation between fluence estimate and ground truth despite the bias. The lower SNR, of course, leads to lower estimation performance (see Fig. 5(d)).

Fig. 5.

Fig. 5

(a-c) Optical parameter estimates and fluence estimates over wavelength. The reduced scattering coefficient μs' and the effective attenuation coefficient μeff of the medium vary with wavelength. The measurement SNR is 50 dB. (a) Shows estimates of μ^eff using Model I and Model II, and (b) shows estimated μ^s' using Model I. (c) Shows kΦ^j,k(ri) estimates at position ri using Model I and Model II. The markers (o) and () indicate estimates using Model I and II, respectively. The dotted line denotes ground truth kΦj,k(ri). (d) Shows the strength of association between ground truth kΦj,k(ri) and its estimate kΦ^j,k(ri) over the SNR range. The markers (o) and () indicate correlation coefficients using Model I and II, respectively.

4. Experiments

Phantom experiments produce realistic PA measurements to help validate the proposed methods. We conducted two studies.

4.1. Phantom study I

In the first study, we explored the accuracy of optical fluence estimates in a turbid medium using a human hair as the absorbing target. Each hair was aligned parallel to the transducer elevational direction (y-axis) so that it appears as a point target in the z-x image plane (see Fig. 6(a)). The target was positioned in a cubic tank (open on top) filled with an optical scattering medium, and a transducer was positioned at the medium surface. The tank was sufficiently large so that boundary effects from all faces except the top interface between media can be neglected. The optical medium was an intralipid solution (20 % IV fat emulsion, Fresenius Kabi, Deerfield, USA) with homogeneous scattering. We diluted the original 20 % emulsion to 0.5 %–4 % ones to control the scattering coefficient.

Fig. 6.

Fig. 6

Experimental assessment of turbid medium optical properties using the fast-sweep concept. (a) Measurement diagram with a human hair as absorbing target immersed in a turbid medium (intralipid solution of different concentrations as indicated in top right corners of panels). Optical properties (reduced scattering coefficient μs' and effective attenuation coefficient μeff) of the medium were determined using Model I (orange) and Model II (blue - μeff only). The point and error bar indicate estimation mean and standard deviation. A total of 8 samples (8 datasets) were used for statistics. The dotted line (μeff) and the shaded region between dotted lines (μs') indicate the estimated value and range, respectively, for the results of Ref. [24].

Radio frequency ultrasound data were recorded using the Verasonics system and a 128 element linear-array transducer (LA 15/128-1633, Vermon S.A. Tours, France). The transducer center frequency is 15 MHz and the 3-dB bandwidth is 11−19 MHz. The array element pitch is 0.1 mm. The laser pulse energy was between 0.4 mJ and 0.5 mJ at the tissue surface, depending on the wavelength. A little (4% portion of it) was taken from the main path by a beam splitter and recorded for every laser pulse with a photodetector located before the fiber bundle. Thus, the energy of every laser pulse was measured to compensate its dependence on wavelength and take into account pulse-to-pulse energy variations (∼ 8%). The number of axial samples and scan-lines were adjusted to the field of view (25.2 mm × 12.7 mm). One PA data set consists of 2048 samples × 128 channels × 20 fibers × 10 wavelengths × 40 frames acquired at a 1 kHz repetition rate, representing a total acquisition period of 8 s. We averaged every data set over frames to enhance SNR and processed PA signals using conventional delay-and-sum beamforming to obtain an image (512 × 128) for every fiber and wavelength. A Hilbert transform was then applied to obtain the smooth envelope of axial image samples. The enveloped image was used as yj,k in Eq. (8) for the k th fiber and j th wavelength.

Fig. 6(b) illustrates optical parameter estimates over the range of laser wavelengths (715−875 nm) for the 0.5 % intralipid solution as the scattering medium. Each marker and error bar indicate mean and standard deviation of 8 samples, respectively. The reference spectrum employed here was obtained from Ref. [24]. Since this spectrum was measured for a 1.2 % intralipid concentration in [24], it was scaled here to compare with the 0.5 % solution and plotted as dotted lines (Fig. 6(c)). The expected reduced scattering coefficient shown on the right is presented as a shaded region since reference papers [24,[34], [35], [36]] give different measurement values. The scattering coefficient μs' is under 5 cm-1 over the measured wavelength range, close to the lower edge of medium scattering used in numerical simulations. As shown in Fig. 6(b,c), estimated μs', and μeff match very closely with those reported in the literature under similar measurement conditions. Due to the small μs', Model II provides a less accurate spectral estimate.

Estimates of effective attenuation coefficient for three higher intralipid concentrations (0.5 %, 1% and 2%) are presented in Fig. 6(b), (d) and (f) respectively. As expected, both Models I and II provide very accurate assessment of μeff, whereas μs' estimates from Model I are clearly not accurate for these higher scattering media (Fig. 6(e,g)). However, as shown in the numerical simulations, one parameter, μeff, is enough when Model II is valid. Thus, inaccuracies in the reconstructed μs' do not significantly affect the accuracy of laser fluence estimates.

4.2. Phantom study II

The purpose of the second study is to (i) perform optical fluence estimation in the medium, and (ii) apply optical fluence corrections to spectroscopic PA measurements. For this test, the container held three cylindrical tubes aligned parallel to the y-axis (see Fig. 7(a)). We injected a nanoparticle solution (gold nanorods (GNR), width (11.4 nm), length (44.8 nm), mass concentration (2.2 mg/mL), longitudinal peak (776 nm), NanoHybrids Inc. Austin, USA) [25] and a black ink solution (Higgines Black Magic Ink, Chartpark Inc., Leeds, USA) as absorbers in Tubes I and III, respectively, where the absorption spectra of the solutions are well-known. Also, we injected water into Tube II as a control. The maximum absorption coefficients of the nanoparticle solution and black ink solution are roughly 35 cm-1 and 82 cm-1, respectively, over the wavelength range (715 nm to 875 nm). The container was filled with a 1% intralipid solution as the scattering medium. We additionally added customized Prussian blue ink [25] to the solution to increase spectral distortion at the expense of high attenuation (low data SNR).

Fig. 7.

Fig. 7

Spectroscopic PA imaging of absorbing targets in a turbid medium. (a) Examples of PA images displayed on a log (dB) scale at 3 different wavelengths (715 nm, 795 nm and 875 nm) for Phantom Study II. The first, second and third tubes were filled with GNR solution, water and black ink, respectively. The scattering medium was a solution of intralipid and Prussian blue ink. The pixel value of the image is associated with the PA pressure. Dotted circles in the image indicate tube cross-sections. (b) and (c) columns show parameter estimates using nanoparticle signals in Tube I and black signals in Tube III, respectively, and spectrum correction results using the estimates. First and second rows represent estimates of effective light attenuation and reduced scattering coefficients in a turbid medium, respectively. Points marked (o) and () denote estimates using Model I and Model II, respectively. The dotted line is obtained by smoothing estimates over wavelength. Third and fourth rows show absorption spectra of the nanoparticle and black ink, respectively. Magenta curves with points marked (◇) correspond to measured PA absorption spectra, i.e. without fluence correction. Orange and blue curves with points marked (o) and () respectively correspond to ink and GNR absorption spectra obtained after applying fluence correction using Model I and Model II, respectively. Dotted lines indicate the reference target spectra measured independently with optical spectrophotometry.

We used a total of 8 datasets and averaged them to improve SNR. Fig. 7(a) presents PA images at three particular wavelengths, where every pixel value is proportional to the photoacoustic signal magnitude presented on a log scale. Due to the limited view and bandwidth of the transducer, the signal only appears at the top and bottom of the tube. The signal beneath the tubes is from a reverberant wave in the tube. The region near Tube II has a weak signal because the tube material itself weakly absorbs light over this wavelength range. The ranges of SNR in Tube I and Tube III are 27.8–39.9 dB and 41.1–47.5 dB, respectively, over the wavelength range.

Fig. 7(b) and (c) show estimation results using signals in Tube I and Tube III for laser fluence assessment, respectively. In other words, we demonstrate here that any target, independent of its absorption properties (for example, for a solution GNR or black ink), effectively absorbing light over the spectral range probed by the system can be used for laser fluence estimation in the swept-beam concept. In practice, such targets can be blood vessels, injected contrast agents or labelled drugs, or other absorbing targets.

The first and second row in columns (b) and (c) of Fig. 7 show effective light attenuation and reduced scattering coefficient estimates, respectively. We smoothed all estimates over wavelength to reduce bias and applied them for fluence compensation. Note that the signals in Tube III provide more stable estimates than those in Tube I due to higher SNR, as predicted in numerical simulations above. The third and fourth row compare raw PA spectra and spectra after fluence correction for nanoparticles and black ink, respectively. Corrected spectra more closely reflect true absorption spectra and approach the ground truth. As predicted in simulations, high SNR increases correction accuracy. Note that laser fluence estimation was performed here using only PA signals from a single target, yet reconstruction accuracy is already reasonable. If a larger number of points within the PA image are used simultaneously for fluence correction, then more accurate corrections can be obtained without the need for signal averaging.

5. Discussion and conclusions

In simulations, we investigated optical parameter estimation using measurements on chromophores acting as targets in a turbid medium. Although we did not fully validate over a wide range of absorption target types, simulations and phantom studies presented here yield consistent estimation results. We can clearly infer that estimation using more sparse targets can lead to lower bias and variance at any noise level if their signal strengths are similar. Also, when multiple targets are used for estimation, targets located near image edges will contribute more to estimation results than those near the center because they provide a wider range of distances r-rk' between target and source positions.

We approximated the narrow beam as an ideal pencil beam. Even though every fiber produces a diverging beam corresponding to their NA = 0.22, we found through MC simulations that the fractional error in the fluence with respect to the true fluence (e.g. for the diverging beam of the same diameter at the medium surface) is less than 3% in the image domain. This minor discrepancy does not affect estimation performance.

The primary advantage of Model I is that it more closely matches the results of MC simulations when the source-target distance exceeds the photon transport mean free path. Thus, estimates of μeff based on this model are unbiased and accurate at zero noise levels, as shown in Fig. 6. A disadvantage of this model is the additional parameter, μs', that must be simultaneously estimated. A two-parameter search often produces higher estimation errors under real, not ideal, conditions, especially when experimental measurements are contaminated by noise. Thus, using constraints for narrowing the search range of μs' based on prior knowledge would increase the performance under low SNR conditions.

Model II is an approximate version of Model I where μs' can be considered high. In reality, this approximation works when a target is located far (i.e. at distances much larger than lt) from all light sources and interfaces. As shown in Fig. 3, Fig. 4, within the range of typical light scattering of biological tissue, Model II can be used for targets located at depths deeper than 5−7 mm from the medium surface, i.e. at depths where the wavelength dependence of laser fluence starts affecting spectroscopic measurements. As depth z increases, Model II converges to Model I very quickly. Note, that the ratio of μˆeff to μeff remains mostly constant as μeff varies over wavelength, as shown in Fig. 5, even though we used the highest variation of μs' (for brain tissue) from all biological tissue reported in the literature. Model II is a single parameter model that provides fast and stable laser fluence estimates for the range where it is valid.

In our signal model, we assumed that the only factor creating PA intensity variation over fibers is the fluence. However, during data acquisition using our system, every fiber delivered slightly different laser pulse energy due to positioning deviations and pulse-to-pulse laser emission fluctuations. Thus, we took into account pulse energy variations with simultaneous measurements of laser pulse energy for every laser pulse to prevent additional estimation biases. Phantom studies validated the overall performance of spectral corrections based on optical parameter estimation. Fluence-corrected spectra closely match ground-truth spectra, as shown in Fig. 6, Fig. 7. Estimation variance depends on SNR at each wavelength. Thus, estimates based on the GNR target are less accurate because of relatively small signal amplitude at the edges (715 nm and 875 nm wavelengths) of the GNA absorption spectrum.

In our studies, we neglected shadowing (laser fluence at one absorber shadowed by another absorber) because absorbers are located sparsely over the imaging field. However, in clinical applications, the presence of large-scale absorbers such as arterial vessels can induce unavoidable fluence attenuation in the entire image field. We conducted simple MC simulations to explore this effect (Fig. 8(a, b)) assuming a cylindrical-shaped vessel close to a point target in a medium (μeff = 0.9 cm-1). When the vessel diameter is 2.4 mm, our estimation method using noise-free measurements from the target position yields μeff = 1.57 cm-1.

Fig. 8.

Fig. 8

Simulation to explore the contribution of a large cylindrical absorber (μ¯a = 10 cm-1) on fluence variations and estimates of optical properties. It is aligned parallel to the y-axis in a turbid medium (μeff = 0.9 cm-1). (a) 2-D Fluence distribution Φk(x,0,z) for a fiber index k (1 to 10). The center of the cross-section of the absorber is located at (-3.6mm,0mm,5mm). (b) Signal magnitude for a target, indicated as a red dot in panel (a), at (6mm,0mm,8mm) over distance between the k th source and the target when the medium either has the source or does not have the source. The diameter of the source is either 0.8, 1.6 or 2.4 mm. (c) Fluence Φk(6,0,z) over depth z. The red and cyan solid lines denote ground-truth and estimate, respectively, when the absorber diameter is 2.4 mm. The black dotted line denotes the best fit to the ground-truth (ranging from 5 to 10 mm) using our analytical model. The black sold line denotes fluence when the absorber is not present.

Note that our primary purpose is estimating true fluence rather than estimating true optical parameters of the medium. As shown in Fig. 8(c), the actual in-depth fluence attenuation, calculated through the target (as indicated in the upper left panel of Fig. 8(a)), is much closer to our estimate than to the fluence without the vessel. Indeed, the best fit to the ground-truth using the analytical fluence model (blue dotted line) yields μeff = 1.41 cm-1 (using the depth range 5−10 mm). Thus, our estimates are not perfect, but they account for large absorbers and provide accurate compensation of fluence variations within the medium. Our future studies will develop more detailed algorithms to further improve reconstruction accuracy.

Our system emits around 1 mJ pulses at 1 kHz. This complies with the maximum permissible exposure (MPE) for skin [37] over the wavelength range (700-900nm) given a small standoff between fiber and skin. During data acquisition, the maximum laser pulse energy at the fiber output was less than 1 mJ due to ∼ 60 % efficiency of light coupling into optical fibers. Coupling efficiency can be improved and will be a focus of future studies.

In Phantom Study II, we used signal averaging to increase SNR because of the high optical attenuation of the medium. Since most human soft tissues have lower attenuation [8,33], our compensation method should require much lower averaging to obtain comparable results for in vivo studies. Also, as we have shown, any absorber independent of its absorption spectrum can be used for fluence reconstruction. Therefore, the larger the number of image points participating in fluence estimation, the better the accuracy and stability, and fewer averages will be required. In addition, our ongoing research on noise reduction using deep-learning may potentially reduce the need for averaging [23].

In conclusion, our study shows that one of the fundamental problems of PA imaging, i.e. decoupling the local light absorption spectrum from wavelength-dependent optical fluence, may be solved automatically using a swept-beam imaging concept at least when the medium under study can be considered macro-homogeneous. We adopted an analytic laser fluence model for a homogenous scattering medium and estimated wavelength-dependent fluence variations within the image plane from PA measurements without knowledge of optical properties and without large computational costs. In phantom studies, we have shown that fluence correction considerably improves the accuracy of measured chromophore spectra. Real-time PAUS systems providing fluence-compensated spectroscopic PA imaging have the potential to enable molecular imaging for many clinical applications, such as interventional procedure guidance using molecularly labelled therapeutic agents and procedure validation based on spectroscopic confirmation of modifications in microvascular networks.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgments

We would like to greatly thank Professor Martin Frenz at University of Bern for several helpful discussions about the general principles of laser fluence estimation. This work was partially supported by NIH grant HL-125339, GE Healthcare, and the Department of Bioengineering at University of Washington.

Biographies

graphic file with name fx1.jpg

MinWoo Kim received the M.S. degree from the Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, South Korea, in 2009, and the Ph. D. degree from the Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, USA, in 2018. He was a research engineer with Samsung Medison Company, Ltd, Korea, from 2009 to 2012, where he specialized in ultrasonic imaging equipment. He is currently working as a senior fellow with the Department of Bioengineering, University of Washington, USA. His current research interests include photoacoustic imaging, spectroscopic imaging, ultrasound vascular imaging, and image reconstruction using novel signal processing and machine learning.

graphic file with name fx2.jpg

Geng-Shi Jeng received the B.S. degree in electronics engineering from National Chiao Tung University in 1996, and the M.S. and Ph.D. degrees in electrical engineering from National Taiwan University in 1998 and 2005, respectively. After graduation, he started his career in the Industrial Technology Research Institute (ITRI) and moved onto Coretronic Display Solution Corporation as the Director. In 2011, he joined S-Sharp Corporation and then served as the CTO, where he has developed high-frequency preclinical ultrasound imaging systems, research purposed array imaging systems, and ultrasound reflection tomography. From 2017 to 2019, he was a Visiting Scientist in Bioengineering, University of Washington. Now he is an Assistant Professor of Electronics Engineering with National Chiao Tung University. His research interests include photoacoustics, real-time ultrasound systems, beamforming, and elastography.

graphic file with name fx3.jpg

Dr. O’Donnell has worked at General Electric CRD, the University of Michigan, where he was Chair of the BME Department from 1999-2006, and the University of Washington (UW), where he was the Frank and Julie Jungers Dean of Engineering from 2006-2012. He is now Professor of Bioengineering at UW. His most recent research has focused on elasticity imaging, optoacoustic devices, photoacoustic contrast agents, laser ultrasound systems, and catheter-based devices. He is a fellow of the IEEE and AIMBE and is an elected member of the Washington State Academy of Sciences and the National Academy of Engineering.

graphic file with name fx4.jpg

Dr. Ivan Pelivanov have worked in the areas of photoacoustics (PA) and laser-ultrasonics (LU) since 1994 when the fields were still very new, well before their recent boom in both the optics and ultrasound communities. He addressed a very diverse spectrum of problems, both fundamental and applied, from NDE to biomedical diagnostics and imaging with a combination of light and ultrasound. A large part of his research has focused on new imaging approaches, techniques and devices among which: ultra-wideband PVDF transducers of different shapes and sizes for PA signal reception, fully non-contact laser-ultrasound systems for industrial NDE applications, non-contact Optical Coherence Elastography for in-vivo volumetric mapping of soft tissue elastic moduli, and different PA systems, including the most recent fast-swept PAUS.

Appendix A

A.1 Refractive-index-mismatched boundary

Assume that the boundary between medium I and II is refractive-index-mismatched and one source to medium I is located at the interface. Outgoing light toward II from I is partially reflected at the boundary due to the mismatch. This can be mathematically expressed as

sn>0Lr,ssndΩ=sn<0R(sn)Lr,ssndΩ, (15)

where L and R denote the radiance and Fresnel reflection, respectively [26]. The vectors r, s and n denote the position on the boundary, unit direction vector, and unit normal vector pointing toward medium I, respectively. dΩ denotes a differential solid angle element. Using a spherical harmonics expansion, the radiance can be approximated as

Lr,s=14πΦr+34πJrs, (16)

where J is the current density, or the energy flow per unit area. The Fresnel reflection is given as

Rsn=12nrcosθ'-cosθnrcosθ'+cosθ2+nrcosθ-cosθ'nrcosθ+cosθ'2,if0θθc1,otherwise (17)

where θ=cos-1(sn) is the angle of incidence, θ'=sin-1(nrsinθ) is the angle of refraction, θc=sin-1(nr-1) is the critical angle, and nr is the ratio of the refractive index of medium I to that of medium II. Substituting Eq. (16) into Eq. (15) yields

14Φr-12J(r)n=14RΦΦr-12RJJrn, (18)

where RΦ=0π/22sinθcosθRcosθdθ and RJ=0π/23sinθcosθ2Rcosθdθ. Substituting Fick’s law, Jr=-DΦ(r), into Eq. (18) results in

Φr-2D1+RJ1-RΦΦ(r)z=0, (19)

A Taylor series expansion to first order leads to the fluence Φz at z=-zb=-2D1+RJ1-RΦ approaching zero. The face z=-zb is called the extrapolated boundary. Details can be found in the literature [26].

A.2 Simplified diffusion equation

Let exp-μeffr-rk'r-rk' be fx,y,z for convenience. If the two imaginary sources are close, Eq. (6) simplifies to

Φkrlimlt03f(x,y+ltsinθ,z-ltcosθ)4πlt-3f(x,y+ltsinθ,z+lt(cosθ+κ))4πlt=34πfysinθ-fzcosθ-34πfysinθ+fz(cosθ+κ)=-34π(2cosθ+κ)f(x,y,z)z=34π2cosθ+κzk1+μeffr-rk'r-rk'3exp-μeffr-rk', (20)

where lt=3D and κ=4(1+RJ)3(1+RΦ). In Eq. (7), the constant α2 is 34π(2cosθ+κ).

A.3 Computation time

The largest computational burden in the total process is beamforming/reconstruction. We implemented a pixel-based delay-and sum algorithm. A sparse matrix performs mixer, low-pass filter, interpolation filter and phase rotator operations to transform one sub-frame data (2048 × 128) to one sub-frame image (512 × 128). The beamforming computation with the sparse matrix used a GPU processor (NVIDIA GeForce RTX 2080 Ti). When the receive f-number is a constant value of 0.1 with depth as the aperture opens, and decreases with depth once the full aperture is opened, the computation time is about 0.8 ms, less than the data acquisition time of 1 ms.

References

  • 1.Cox B.T., Laufer J.G., Beard P.C., Arridge S.R. Quantitative spectroscopic photoacoustic imaging: a review. J. Biomed. Opt. 2012;17(6) doi: 10.1117/1.JBO.17.6.061202. [DOI] [PubMed] [Google Scholar]
  • 2.Cox B.T., Laufer J.G., Beard P.C. The challenges for quantitative photoacoustic imaging. Photons Plus Ultrasound: Imaging Sens. 2009;7177 [Google Scholar]
  • 3.Liu Y.-H., Xu Y., Liao L.-D., Chan K.C., Thakor N.V. A handheld real-time photoacoustic imaging system for animal neurological disease models: from simulation to realization. Sensors. 2018;18(11):4081. doi: 10.3390/s18114081. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Zhou X., Akhlaghi N., Wear K.A., Garra B.S., Pfefer T.J., Vogt W.C. Evaluation of fluence correction algorithms in multispectral photoacoustic imaging. Photoacoustics. 2020 doi: 10.1016/j.pacs.2020.100181. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Rajian J.R., Carson P.L., Wang X. Quantitative photoacoustic measurement of tissue optical absorption spectrum aided by an optical contrast agent. Opt. Express. 2009;17(6):4879–4889. doi: 10.1364/oe.17.004879. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Zhao L., Yang M., Jiang Y., Li C. Optical fluence compensation for handheld photoacoustic probe: an in vivo human study case. J. Innov. Opt. Health Sci. 2017;10(04) [Google Scholar]
  • 7.Kim S., Chen Y.-S., Luke G.P., Emelianov S.Y. In vivo three-dimensional spectroscopic photoacoustic imaging for monitoring nanoparticle delivery. Biomed. Opt. Express. 2011;2(9):2540–2550. doi: 10.1364/BOE.2.002540. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Jacques S.L. Optical properties of biological tissues: a review. Phys. Med. Biol. 2013;58(11):R37. doi: 10.1088/0031-9155/58/11/R37. [DOI] [PubMed] [Google Scholar]
  • 9.Brochu F.M., Brunker J., Joseph J., Tomaszewski M.R., Morscher S., Bohndiek S.E. Towards quantitative evaluation of tissue absorption coefficients using light fluence correction in optoacoustic tomography. IEEE Trans. Med. Imaging. 2016;36(1):322–331. doi: 10.1109/TMI.2016.2607199. [DOI] [PubMed] [Google Scholar]
  • 10.Buchmann J., Kaplan B., Powell S., Prohaska S., Laufer J. Quantitative PA tomography of high resolution 3-D images: experimental validation in a tissue phantom. Photoacoustics. 2020 doi: 10.1016/j.pacs.2019.100157. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Buchmann J., Kaplan B.A., Powell S., Prohaska S., Laufer J. Three-dimensional quantitative photoacoustic tomography using an adjoint radiance Monte Carlo model and gradient descent. J. Biomed. Opt. 2019;24(6) doi: 10.1117/1.JBO.24.6.066001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Fonseca M., Malone E., Lucka F., Ellwood R., An L., Arridge S., Beard P., Cox B. Three-dimensional photoacoustic imaging and inversion for accurate quantification of chromophore distributions. Photons Plus Ultrasound: Imaging Sens. 2017;10064 [Google Scholar]
  • 13.Gröhl J., Kirchner T., Adler T., Maier-Hein L. 2019. Estimation of Blood Oxygenation with Learned Spectral Decoloring for Quantitative Photoacoustic Imaging (LSD-qPAI) arXiv preprint arXiv:1902.05839. [Google Scholar]
  • 14.Chen T., Lu T., Song S., Miao S., Gao F., Li J. Vol. 11240. 2020. p. 112403V. (A Deep Learning Method Based on U-Net for Quantitative Photoacoustic Imaging). [Google Scholar]
  • 15.Luke G.P., Hoffer-Hawlik K., Van Namen A.C., Shang R. 2019. O-Net: A Convolutional Neural Network for Quantitative Photoacoustic Image Segmentation and Oximetry. arXiv preprint arXiv:1911.01935. [Google Scholar]
  • 16.Yang C., Lan H., Zhong H., Gao F. Quantitative photoacoustic blood oxygenation imaging using deep residual and recurrent neural network. 2019 IEEE 16th International Symposium on Biomedical Imaging. 2019:741–744. [Google Scholar]
  • 17.Yang C., Gao F. Eda-net: dense aggregation of deep and shallow information achieves quantitative photoacoustic blood oxygenation imaging deep in human breast. International Conference on Medical Image Computing and Computer-Assisted Intervention. 2019:246–254. [Google Scholar]
  • 18.Kirchner T., Gröhl J., Maier-Hein L. Context encoding enables machine learning-based quantitative photoacoustics. J. Biomed. Opt. 2018;23(5) doi: 10.1117/1.JBO.23.5.056008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Cai C., Deng K., Ma C., Luo J. End-to-end deep neural network for optical inversion in quantitative photoacoustic imaging. Opt. Lett. 2018;43(12):2752–2755. doi: 10.1364/OL.43.002752. [DOI] [PubMed] [Google Scholar]
  • 20.Tzoumas S., Nunes A., Olefir I., Stangl S., Symvoulidis P., Glasl S., Bayer C., Multhoff G., Ntziachristos V. Eigenspectra optoacoustic tomography achieves quantitative blood oxygenation imaging deep in tissues. Nat. Commun. 2016;7:12121. doi: 10.1038/ncomms12121. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Schellenberg M.W., Hunt H.K. Hand-held optoacoustic imaging: a review. Photoacoustics. 2018;11:14–27. doi: 10.1016/j.pacs.2018.07.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Deán-Ben X.L., Merčep E., Razansky D. Hybrid-array-based optoacoustic and ultrasound (OPUS) imaging of biological tissues. Appl. Phys. Lett. 2017;110(20) [Google Scholar]
  • 23.Kim M., Jeng G.-S., Pelivanov I., O’Donnell M. Deep-learning image reconstruction for real-time photoacoustic system. IEEE Transcation on Medical Imaging. 2020 doi: 10.1109/TMI.2020.2993835. https://ieeexplore.ieee.org/document/9091172 Early access. Available: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Held K.G., Jaeger M., Rička J., Frenz M., Akarçay H.G. Multiple irradiation sensing of the optical effective attenuation coefficient for spectral correction in handheld OA imaging. Photoacoustics. 2016;4(2):70–80. doi: 10.1016/j.pacs.2016.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Jeng G.-S., Li M.-L., Kim M., Yoon S.J., Pitre J.J., Li D.S., Pelivanov I., O’Donnell M. Real-time spectroscopic photoacoustic/ultrasound (PAUS) scanning with simultaneous fluence compensation and motion correction for quantitative molecular imaging. bioRxiv. 2019 doi: 10.1101/2019.12.20.885251. Available: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Wang L.V., Wu H.-i. John Wiley & Sons; 2012. Biomedical Optics: Principles and Imaging. [Google Scholar]
  • 27.Sivaramakrishnan M., Maslov K., Zhang H.F., Stoica G., Wang L.V. Limitations of quantitative photoacoustic measurements of blood oxygenation in small vessels. Phys. Med. Biol. 2007;52(5):1349. doi: 10.1088/0031-9155/52/5/010. [DOI] [PubMed] [Google Scholar]
  • 28.Pelivanov M., Belov S.A., Solomatin V.S., Khokhlova T.D., Karabutov A.A. Direct measurement of the spatial distribution of laser radiation intensity in biological tissues in-vitro by the laser optoacoustic method. Quantum Elec. 2007;36(12):1089–1096. [Google Scholar]
  • 29.Grashin P.S., Karabutov A.A., Oraevsky A.A., Pelivanov I.M., Podymova N.Y., Savateeva E.V., Solomatin V.S. Distribution of the laser radiation intensity in turbid media: Monte Carlo simulations, theoretical analysis, and results of optoacoustic measurements. Quantum Elec. 2002;32(10):868. [Google Scholar]
  • 30.Karabutov A.A., Pelivanov I.M., Podymova Ny.B., Skipetrov S.E. Determination of the optical characteristics of turbid media by the laser optoacoustic method. Quantum Elec. 1999;29(12):1054. [Google Scholar]
  • 31.Glasstone S., Edlund M.C. Van Nostrand Co., Inc.; New York: 1952. The Elements of Nuclear Reactor Theory, D. [Google Scholar]
  • 32.Fang Q., Boas D.A. Monte Carlo simulation of photon migration in 3D turbid media accelerated by graphics processing units. Opt. Express. 2009;17(22):20178–20190. doi: 10.1364/OE.17.020178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Sandell J.L., Zhu T.C. A review of in‐vivo optical properties of human tissues and its impact on PDT. J. Biophotonics. 2011;4(11-12):773–787. doi: 10.1002/jbio.201100062. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Van Staveren H.J., Moes C.J.M., van Marie J., Prahl S.A., Van Gemert M.J.C. Light scattering in lntralipid-10% in the wavelength range of 400–1100 nm. Appl. Opt. 1991;30(31):4507–4514. doi: 10.1364/AO.30.004507. [DOI] [PubMed] [Google Scholar]
  • 35.Flock S.T., Jacques S.L., Wilson B.C., Star W.M., van Gemert M.J.C. Optical properties of Intralipid: a phantom medium for light propagation studies. Lasers Surg. Med. 1992;12(5):510–519. doi: 10.1002/lsm.1900120510. [DOI] [PubMed] [Google Scholar]
  • 36.Michels R., Foschum F., Kienle A. Optical properties of fat emulsions. Opt. Express. 2008;16(8):5907–5925. doi: 10.1364/oe.16.005907. [DOI] [PubMed] [Google Scholar]
  • 37.ANSI Standard Z136 . Laser Institute of America; Orlando, FL, USA: 2000. 1-2000: For Safe Use of Lasers. [Google Scholar]

Articles from Photoacoustics are provided here courtesy of Elsevier