Skip to main content
EJNMMI Physics logoLink to EJNMMI Physics
. 2020 Dec 7;7:72. doi: 10.1186/s40658-020-00343-6

Impact of the dosimetry approach on the resulting 90Y radioembolization planned absorbed doses based on 99mTc-MAA SPECT-CT: is there agreement between dosimetry methods?

Verónica Morán 1, Elena Prieto 2,3, Lidia Sancho 4, Macarena Rodríguez-Fraile 2,5, Leticia Soria 1, Arantxa Zubiria 6, Josep M Martí-Climent 2,3,
PMCID: PMC7721939  PMID: 33284389

Abstract

Background

Prior radioembolization, a simulation using 99mTc-macroaggregated albumin as 90Y-microspheres surrogate is performed. Gamma scintigraphy images (planar, SPECT, or SPECT-CT) are acquired to evaluate intrahepatic 90Y-microspheres distribution and detect possible extrahepatic and lung shunting. These images may be used for pre-treatment dosimetry evaluation to calculate the 90Y activity that would get an optimal tumor response while sparing healthy tissues.

Several dosimetry methods are available, but there is still no consensus on the best methodology to calculate absorbed doses. The goal of this study was to retrospectively evaluate the impact of using different dosimetry approaches on the resulting 90Y-radioembolization pre-treatment absorbed dose evaluation based on 99mTc-MAA images.

Methods

Absorbed doses within volumes of interest resulting from partition model (PM) and 3D voxel dosimetry methods (3D-VDM) (dose-point kernel convolution and local deposition method) were evaluated. Additionally, a new “Multi-tumor Partition Model” (MTPM) was developed. The differences among dosimetry approaches were evaluated in terms of mean absorbed dose and dose volume histograms within the volumes of interest.

Results

Differences in mean absorbed dose among dosimetry methods are higher in tumor volumes than in non-tumoral ones. The differences between MTPM and both 3D-VDM were substantially lower than those observed between PM and any 3D-VDM. A poor correlation and concordance were found between PM and the other studied dosimetry approaches.

DVH obtained from either 3D-VDM are pretty similar in both healthy liver and individual tumors. Although no relevant global differences, in terms of absorbed dose in Gy, between both 3D-VDM were found, important voxel-by-voxel differences have been observed.

Conclusions

Significant differences among the studied dosimetry approaches for 90Y-radioembolization treatments exist. Differences do not yield a substantial impact in treatment planning for healthy tissue but they do for tumoral liver.

An individual segmentation and evaluation of the tumors is essential. In patients with multiple tumors, the application of PM is not optimal and the 3D-VDM or the new MTPM are suggested instead. If a 3D-VDM method is not available, MTPM is the best option. Furthermore, both 3D-VDM approaches may be indistinctly used.

Supplementary Information

The online version contains supplementary material available at 10.1186/s40658-020-00343-6.

Keywords: 90Y-Microspheres, 99mTc-MAA, Radioembolization, Predictive dosimetry, Partition model, Multi-tumor partition model, 3D voxel dosimetry, Local deposition method, Dose point kernel

Background

90Y-Radioembolization (RE) is an established treatment modality for patients with unresectable hepatic malignancies [14]. 90Y-loaded microspheres are injected selectively into the hepatic arteries that supply the tumors and permanently trapped in capillaries, delivering a high radiation absorbed dose to the tumor, while limiting the dose to non-tumoral liver parenchyma [5].

Prior treatment administration, a simulation is performed in order to assess 90Y-micospheres intrahepatic distribution and detect possible extrahepatic and lung shunting. For that purpose, 99mTc-macroaggregated albumin (MAA) are selectively infused through the most appropriate hepatic arteries—selected after hepatic arteriography—to simulate intra- and extrahepatic 90Y-micospheres distribution. Within an hour after the 99mTc-MAA administration, gamma scintigraphy is acquired and SPECT or SPECT-CT scans are recommended [6, 7] to allow a proper evaluation of intra and extrahepatic distributions, as well as to perform dosimetry evaluation to determine the most adequate 90Y activity that maximize tumor response while maintaining radiation exposure to surrounding healthy tissues within acceptable safety limits [8]. Thus, a treatment planning based on a proper dosimetry estimation has an important role in the optimization of the RE outcomes [9].

To date, several dosimetry approaches have been used for the calculation of 90Y-RE absorbed doses. Currently, for 90Y-resin microspheres (SIR-Spheres®; Sirtex Medical Limited, Australia), the most widely used method is PM [10], a multi-compartmental method based on the Medical internal Radiation Dose (MIRD) approach [11]. Mean absorbed doses (Dmean) are calculated for 3 compartments (aggregated tumor, liver, and lung) assuming uniform distribution within each compartment. However, it must be taken into consideration that in most cases, microspheres distribution in tumoral and non-tumoral liver is not uniform. 3D-VDM methods, in contrast to PM, take into account inhomogeneities due to different intrahepatic distribution of the microspheres not only among individual tumors composing the aggregated tumor but also within each tumor.

In this work, a new multi-compartmental method called MTPM, which allows to calculate Dmean within each individual tumor, was developed and implemented. This dosimetry approach considers individual tumors as different compartments in order to take into account the heterogeneity among them.

The aim of this study was to retrospectively investigate and compare different dosimetry methods applied to 90Y RE pre-treatment planning, including the original MTPM approach. For that purpose, absorbed doses within volumes of interest (VOIs) resulting from multi-compartmental methods (PM and MTPM) were compared to those resulting from 3D-VDM methods (dose-point kernel (DPK) convolution and local deposition method (LDM)). The differences among the dosimetry approaches were evaluated within different VOIs in terms of Dmean and dose volume histograms (DVH).

Methods

Patient characteristics

Fourteen patients with hepatic malignancies who underwent RE with 90Y-resin microspheres in our institution from 2013 to 2015 were retrospectively evaluated. The inclusion criteria were availability of a contrast-enhanced CT or MRI within 4 weeks prior to treatment, lesions that could be unequivocally delineated, and similar positioning of the catheter both in the simulation with 99mTc-MAA and in the therapeutic 90Y-microespheres administration. No other clinical or demographic data was taken into account for the patient selection because it is not required to achieve the principal aim of this study: to compare dosimetry methods.

(99mTc-MAA) protocol scan and activity planning

Once 99mTc-MAA were injected trough the selected arteries during hepatic arteriography, planar and SPECT-CT images were acquired in a Symbia T2 (Siemens Medical Solutions, Erlangen, Germany) with a dual-head variable-angle gammacamera and a two-slice spiral CT scanner. A low-energy high-resolution (LEHR) collimator was used with an energy window centered at 140 keV and 15% wide.

For planar imaging, anterior and posterior images of the abdomen and the thorax (10-min acquisition) were taken in a 128 × 128 matrix. No zoom was applied.

For SPECT acquisition, 128 images (20 s per projection) were acquired over 360° using a 128 × 128 matrix with a pixel size of 4.8 × 4.8 mm2. Images were reconstructed using a Flash 3D algorithm (8 iterations, 4 subsets, 8.4 mm FWHM Gaussian post-filter), an iterative algorithm considering a 3D collimator beam modeling, CT-based attenuation correction, and energy window-based scatter correction. The scan parameters for CT were 130 kV, 25 mAs, and 5-mm slices. Both SPECT and CT images were fused using an Esoft 2000 application package (Siemens Medical Solution, Erlangen, Germany).

As previously published by Gil-Alzugaray et al. [12], in our center, the administered 90Y activity was planned by means of PM for lobar and segmental treatments and by body surface area model for whole liver treatments. This methods were applied according to the microspheres’ manufacturer recommended guidelines [13].

The lung shunt fraction (LSF) was calculated by Eq. (1), where Clung and CWL are the geometric mean of total counts (anterior and posterior images) registered within lungs and whole liver, respectively:

LSF%=100ClungClung+CWL 1

Planar images may not be used to determine accurately the tumor to non-tumor liver activity concentration ratio (TNR) [14]; therefore, attenuation-corrected SPECT images were used instead. TNR was calculated by Eq. (2), where CTL and CNLt are the total counts registered within TL and NLt volumes respectively:

TNR=CTL/VTLCNLt/VNLt 2

Contouring

The first step for this retrospective investigation was the anatomic VOIs segmentation. These VOIs were contoured on the CT from 99mTc-MAA SPECT-CT with the aid of a rigidly registered diagnostic scan (contrast enhanced CT or MRI) using a commercial treatment planning software (Pinnacle, Philips Medical System, Anover, MA). A process similar to the one used in external beam radiation therapy was followed. The VOIs were then exported as DICOM-RT structure sets. To avoid inter-operator bias, all VOIs were delineated by a single physician.

For each patient, individual tumors (Ti), the planning target volume (PTV), and the whole liver (WL) were delineated. The PTV refers to the portion of the liver in which it is intended to deliver the radiation dose: one or more segments, one lobe or the whole liver depending on whether the treatment is segmental, lobar, or total. Tumoral liver volume (TL), corresponding to the aggregated tumor volume was generated by summing all the Ti volumes. Target normal liver volume (NLt) was defined by subtracting the TL volume from the PTV volume. Whole normal liver (NLw) was also determined by subtracting the TL volume from the WL. Volumes in mL for individual tumors, aggregated tumoral liver, target normal liver, and whole normal liver (VTi, VTL, VNLt, and VNLw) were calculated for the 14 patients.

Dosimetry assessment

For the purposes of this study, the mean absorbed dose delivered to each compartment (DmeanTi,DmeanTL,DmeanNLt, andDmeanNLw) was estimated according to MIRD formalism [11]. Both multi-compartment dosimetry methods were retrospectively applied to obtain DMean. To implement MTPM, in patients with two individual tumors or more, an Excel-based mean absorbed dose calculator was developed (available in additional file 1).

Additionally, DPK and LDM were applied to calculate a 3D dose map and DVHs. The actual 90Y administered activity and volumes of the contoured VOIs used to determine the absorbed doses were the same for all dosimetry approaches (PM, MTPM, and both 3D-VDM).

For the 90Y dosimetry calculation purposes, an identical 99mTc-MAA and 90Y-microspheres’ biodistributions were assumed, based on previous studies [10, 1519].

Multicompatimental methods

PM was applied to calculate DMeanTL and DMeanNLt according to Eqs. (3) and (4), where A(90Y) is the 90Y-microspheres administered activity, and MTL and MNLtare the masses in kg of the tumoral liver and the target normal liver, respectively. A 1 g/mL tissue density is assumed, and volumes in liters are straight converted in masses in kg.

DMeanNLw was determined by rescaling the DMeanNLt to the NLwvolume, applying (5).

DMeanTLGy=49.67JGBqA(Y)90GBq1LSF100TNRMTLkgTNR+MNLtkg 3
DMeanNLtGy=49.67JGBqA(Y)90GBq1LSF100MTLkgTNR+MNLtkg 4
DMeanNLwGy=DNLtGyMNLtkgMNLwkg 5

In patients with two or more tumors (n), the MTPM method, an (n + 2) compartment partition model, was applied to determine DMeanTiusing Eq. (6), where TNRi is the tumor to normal liver activity concentration ratio for individual tumors calculated by Eq. (7).

DMeanTiGy=49.67JGBqA(Y)90GBq1LSF100TNRiMTLKgTNR+MNLtKg 6
TNRi=CTi/VTiCNLt/VNLt 7

3D-voxel dosimetry

The first step to perform 3D image-based dosimetry using 99mTc-MAA SPECT is to convert, through a calibration factor, the counts registered in each voxel of the reconstructed image to 90Y activity (in MBq). Since 99mTc-MAA administered activity (A(99mTc)) is totally uptaken in the liver with the exception of the fraction that shunts to the lung, the patient-specific calibration factor may be determined as it was described by Chiesa et al. [20].

The 90Y-microspheres activity in a liver voxel at the image acquisition time (Avoxel(90Y)) is directly proportional to the total counts registered within a voxel of 99mTc-MAA SPECT image (Cvoxel(99mTc)). Thus, Avoxel(90Y) may be estimated by means of Eq. (8) where CWL is the total counts of 99mTc registered within the WL volume.

AvoxelY90=Cvoxel(Tc)99mATc99m1LSF100CWLTc99mA(Y)90A(Tc)99m=Cvoxel(Tc)99m1LSF/100A(Y)90CWLTc99m 8

Unlike other internal radionuclide therapy, RE has the advantage of negligible biological clearance following the infusion. Thus, assuming the permanent trapping of microspheres, fitting of time-activity curves is not required, and the total number of disintegrations in a voxel (Ã voxel(9oY)) was calculated as described by Eq. (9), where T1/2(90Y) is the physical 90Y half-life (64.2 h).

A~voxelY90=AvoxelY90eLn2tT12dt=1.443T12Y90.AvoxelY90 9

To convert the cumulative activity in each voxel to a tridimensional 90Y absorbed dose map, two different 3D-VDM approaches were applied: LDM and DPK. For that purpose, a software tool based in MATLAB v.R2016a (The Math Works, Natick, MA) code was developed.

DPK takes into account the high-energy beta particles transport to adjacent voxels. The absorbed dose within the target voxel t DvoxeltY90 was calculated by the convolution of the 3D cumulative activity matrix with a cubic dose kernel, as described in Eq. (10). Where A~voxelsY90 is the time-integrated activity within the source voxel s, and S(voxelt ← voxels) is the well-known S value defined as the absorbed dose to the target voxel t per unit of cumulative activity in the voxel s. The dose kernels used in this work were extracted from Lanconelli database [21].

DvoxeltY90=s=0NA~voxelsY90Svoxeltvoxels 10

LDM assumes that the kinetic energy from each beta particle is deposited within the voxel where the emission occurs. The source voxel s in this case is also the target voxel t. The absorbed dose in each voxel was then determined by Eq. (11), multiplying the cumulative activity within the voxel by a constant scalar factor, which is the S value considering an absorbed fraction equal to 1 in each voxel (Svoxeltvoxelsvoxelt=voxels). S is calculated by means of Eq. (12), where EβY90voxelt=0.9267MeVdisintegration1.60221013JMeVGyKgJ109disintegrationssGBq is the deposited β-energy per disintegration in average, and Mvoxeltis the target voxel mass. For a given cubic voxel size (4.48 mm side), S is 1.603 Gy/GBq.s.

DvoxeltY90=A~voxelsY90×Svoxeltvoxelsvoxelt=voxels 11
Svoxeltvoxelsvoxelt=voxels=EβY90voxeltMvoxeltvoxelt=voxels 12

Dosimetry comparisons and statistical analysis

DmeanNLt and DmeanNLwcalculated by PM, LDM, and DPK methods were compared using a paired Student’s t test or Wilcoxon test in case differences between methods do not meet normal criteria.

DmeanTi and DmeanTLcalculated by PM, MTPM, and both 3D-VDM were also compared using a paired Student’s t test or Wilcoxon test, as corresponds. For MTPM, LDM, and DPK, DmeanTL was calculated for each patient as the average of all DmeanTi. The standard deviation (SD) was also determined. DmeanTi calculated by PM was the same for all individual tumors of the same patient, and equal to DmeanTL, as tumoral liver compartment in PM approach is defined as an aggregated tumor including all Ti.

The heterogeneity of 90Y-microspheres distribution among the tumors for each patient was evaluated through the TNRi coefficient of variation (COV(TNRi)).

A comparison among the studied dosimetry methods for all VOIs was performed in terms of mean absorbed dose differences (DmeanVOI) in Gy.

The correlation between differences in Dmean between PM and the other studied dosimetry methods (MTPM, LDM, and DPK) and TNR-TNRi differences was evaluated by means of the Spearman’s correlation coefficient (rho).

Dosimetry comparison between DPK and LDM methods was also managed in terms of DVHs. Some metrics were extracted from the DVHs: the minimum dose to 5%, 25%, 50%, 70%, and 95% in the corresponding VOI (D5, D25, D50, D70, and D95, respectively), the percentage of the tumor volume receiving at least 100 Gy (V100) and the percentage of the NLw and NLt volumes receiving at least 20 Gy (V20). A paired Student’s t test or Wilcoxon test, as appropriate, was applied. Absorbed dose differences in Gy were also determined for each VOI.

Bland-Altman analysis was used to evaluate the agreement among the studied dosimetry methods (PM, MTPM, LDM, and DPK), in terms of Dmean, for both tumoral and non-tumoral volumes (NLt, NLw, TL, and Ti). The agreement of DVH between both 3D-VDM methods was also evaluated by means of a Bland-Altman analysis. Pearson correlation (ρ) and Lin concordance (ρc) coefficients were reported.

All analyses were performed with statistical STATA v.15 software (StataCorp, TX, USA). A p value of 0.05 or less was considered statistically significant.

Differences between LDM and DPK methods were also assessed by a voxel by voxel analysis. A voxel based subtraction of the parametric images (in Gy) calculated by both methods was performed, and the calculation of the normalized mean square error (NMSE) between dose absorbed maps obtained applying Eq. (13), as described previously by Pacilio et al. [22], where xi is the ith voxel of the DPK image and pi the ith voxel of the LDM image (used as a reference).

NMSE=100i(xipi)2ipi2 13

Results

Finally, 14 patients were collected according the inclusion criteria. There were 5 patients who received whole liver treatment, 8 who received lobar treatments (7 right and 1 left lobe) and 1 superselective treatment to one hepatic segment. In total, 101 individual tumors were identified and analyzed, with a volume range from 0.6 to 351 mL. Prior to RE, in order to simulate the 90Y-labeled microspheres biodistribution, 161 ± 11 MBq of 99mTc-MAA were administered.

The treatment characteristics and volumes of contoured VOIs are reported in Table 1 for the 14 patients.

Table 1.

Treatment characteristics: treatment approach, number of treated tumors, 90Y administered activity, the percentage of lung shunt fraction, tumor to non-tumor liver activity concentration ratio, and volumes for individual tumors (average ± SD), aggregated tumoral liver, target normal liver, and whole normal liver

Patient Treatment approach Number of tumors 90Y Activity (GBq) LSF (%) TNR VTi (mL) VTL (mL) VNLt (mL) VNLw (mL)
1 Whole-liver 4 1.4 6.1 1.0 143 ± 155 573 1875 1785
2 Lobar 2 1.3 1.4 0.6 2 ± 1 4 789 1235
3 Lobar 1 1.2 10.1 0.8 12 12 969 1509
4 Whole-liver 50 1.8 4.0 1.2 15 ± 33 735 2277 2277
5 Lobar 3 0.6 10.7 1.2 6 ± 1 18 410 1434
6 Lobar 1 0.5 6.6 0.6 93 93 932 1551
7 Whole-liver 9 0.9 1.8 1.7 6 ± 6 50 1176 1176
8 Lobar 1 1.1 9.0 2.3 282 282 982 1577
9 Lobar 2 1 3.9 2.3 13 ± 6 26 753 1240
10 Lobar 4 0.9 2.2 1.5 10 ± 5 42 887 1725
11 Whole-liver 15 1.6 5.3 2.2 12 ± 17 177 1605 1605
12 Segmental 2 1 4.1 1.4 14 ± 1 29 472 1205
13 Lobar 1 1.8 3.6 1.2 130 129 1003 1561
14 Whole-liver 6 1.3 5.8 0.7 82 ± 53 493 1373 1373

An example of 99mTc-MAA SPECT-CT images used to perform the dosimetry calculation is reported in Fig. 1, as well as the delineated VOIs in axial, coronal, and sagittal planes.

Fig. 1.

Fig. 1

Volumes of interest contoured on the 99mTc-MAA SPECT-CT images used for the pre-therapy dosimetry assessment corresponding to patient 12: whole liver volume delimited by red line, target volume delimited by blue line and individual tumors delimited by pink, yellow, and light blue lines

Absorbed dose by normal liver

Box plots summarizing DmeanNLt and DmeanNLw, calculated by means of PM, DPK, and LDM methods, are shown in Fig. 2.

Fig. 2.

Fig. 2

Mean absorbed doses delivered to target normal liver a and whole normal liver b volumes, determined by means of PM and both 3D voxel dosimetry methods (LDM and DPK)

To assess the agreement between the PM and LDM method and also between both 3D-VDM studied dosimetry methods, in terms of Dmean, Bland-Altman plots are shown in Fig. 3 for both NLt and NLw. The comparison between PM and DPK method are shown in additional file 2. Note the different scale according the compared methods. Pearson’s correlation and Lin concordance coefficients for each comparison between dosimetry methods are summarized in Table 2. For both DmeanNLt and DmeanNLw, 1/14 (7.1%) of the points are beyond the ±2 SD lines in all pair of comparisons. The PM method significantly overestimates Dmean with respect to LDM and DPK methods (p < 0.01). The Dmean determined by LDM is also higher than those calculated applying DPK. Maximum DmeanNLw and DmeanNLt between PM and 3D methods were 2.4 Gy and 16.1 Gy respectively, and between both 3D-VDM methods were 0.8 Gy for NLw and 1.3 Gy for NLt. Significant differences were observed in DmeanNLt and DmeanNLwbetween PM and 3D-VDM methods and also between both 3D-VDM methods (p value <0.01).

Fig. 3.

Fig. 3

Bland-Altman diagrams representing the relative dose difference in Gy between dosimetry methods versus the mean absorbed dose values in Gy for both normal target liver a and whole normal liver b. The dashed line represents the average of the differences between the studied dosimetry methods, and the black line is the mean differences of ± 2 SD

Table 2.

Comparison among PM and both 3D dosimetry methods (LDM and DPK) in terms of mean absorbed dose delivered to target and whole normal liver: Bland Altman analysis, Pearson’s correlation, and Lin concordance coefficients

Bland-Altman Correlation and concordance coefficients
DMeanVOI (Gy) Bias [95%CI] (Gy) ρ ρc
Target normal liver
 PM-DPK 4.7 − 4.8; 14.2 0.99 0.94
 PM-LDM 4.2 − 4.9; 13.2 0.99 0.95
 LDM-DPK 0.6 − 0.1; 1.3 1.00 1.00
Whole normal liver
 PM-DPK 1.1 − 0.1; 2.4 1.00 0.99
 PM-LDM 0.7 − 0.3; 1.7 1.00 1.00
 LDM-DPK 0.4 0.0; 0.8 1.00 1.00

DVH curves calculated by means of LDM and DPK dosimetry methods for both NLt and NLw volumes are reported in Fig. 4. DVHs correspond to patient 12, who yielded one of the highest differences between both 3D-VDM methods.

Fig. 4.

Fig. 4

Dose volume histogram curves for both target normal liver a and whole normal liver volumes b corresponding to patient 12. Continuous lines correspond to LDM method and the dashed lines to DPK method

The differences between LDM and DPK methods in DVH in terms of D5, D25, D50, D70, and D95 for the healthy liver volumes (NLt and NLw) are reported in Fig. 5. Higher differences were found for D70 and D95 while D50, D25, and D5 showed lower variations in both compartments. The maximum differences for NLt and NLw were found for D95, with values of 6.34 and 4.71 Gy, respectively.

Fig. 5.

Fig. 5

Boxplot representing absorbed dose differences in Gy between LDM and DPK methods in terms of D5, D25, D50, D70, and D95 delivered to the target normal liver a, whole normal liver b, and individual tumors c

The statistical analysis showed that DVH differences between both 3D-VDM were statistically significant except for the following cases: D25 and D50 for NLt (p value 0.06 and o.17) and D5, D50, and D70 for NLw (p values: 0.26, 0.95, and 0.68).

The mean (range) difference in V20 Gy (evaluated as %) due to applied LDM or DPK methods for both NLt and NLw were − 0.3 (− 1.0, 0.3) % and − 0.2 (− 0.6, 0.3) %, respectively. Significant differences were observed in both compartments (p value < 0.01).

The results of the Bland-Altman analysis to assess differences in DVH in terms of D5, D25, D50, D70, and D95 are summarized in Table 3, for both NLt and NLw. Pearson’s correlation and Lin concordance coefficients are greater than 0.995 for all the comparisons. For both compartments, at most, 1/14 (7.1%) of the points are beyond the ± 2 SD lines for all the studied endpoints.

Table 3.

Comparison among both 3D dosimetry methods (LDM and DPK) in terms of D5, D25, D50, D70, and D95 delivered to target normal liver, whole normal liver, and individual tumors: Bland Altman analysis and Lin concordance coefficients

Bland-Altman Correlation and concordance coefficients
∆D (Gy) Bias [95%CI] (Gy) Points beyond ± 2 SD (%) ρ ρc
Target normal liver
 D5 0.3 − 0.9; 0.4 7.1 0.999 0.999
 D25 − 0.2 − 0.8; 0.4 7.1 1.000 1.000
 D50 0.2 − 0.8; 1.1 0.0 1.000 1.000
 D70 0.4 − 1.0; 1.9 0.0 1.000 0.999
 D95 2.3 0.0; 5.9 7.1 1.000 0.996
Whole normal liver
 D5 0.1 − 0.4; 0.3 7.1 0.999 0.998
 D25 − 0.1 − 0.5; 0.2 7.1 1.000 1.000
 D50 − 0.1 − 0.9; 0.7 7.1 1.000 1.000
 D70 0.1 − 1.2; 1.3 0.0 1.000 1.000
 D95 2.5 0.1; 4.9 7.1 1.000 0.996
Individual tumors
 D5 − 0.3 − 1.6; 1.0 5.0 0.999 0.999
 D25 − 0.1 − 1.7; 1.8 4.0 0.999 0.999
 D50 0.4 − 1.7; 2.5 6.0 1.000 0.999
 D70 1.1 − 2.3; 4.5 4.0 1.000 0.999
 D95 3.1 − 9.0; 15.2 2.0 0.999 0.995

Absorbed dose by tumors

Box plots summarizing DmeanTi and DmeanTL calculated applying PM, MTPM, DPK, and LDM methods are shown in Fig. 6.

Fig. 6.

Fig. 6

Mean absorbed doses delivered to the aggregated tumoral liver a and each individual tumor b volumes, determined by means of multicompartmental (PM and MTPM) and 3D voxel dosimetry methods (LDM and DPK)

For patients with more than one single tumor, DmeanTL determined applying PM is summarized in Table 4, as well as the COV(TNRi). For MTPM, and both 3D-VDM DmeanTL is also showed as average of DmeanTi ± SD. The maximum DmeanTL among the studied methods for PM-MTPM, PM-DPK, PM-LDM, MTPM-DPK, MTPM-LDM, and LDM-DPK were 41.3, 41.5, 40.7, 39.7, 34.1, and 5.6 Gy, respectively. Significant differences were observed among the all dosimetry methods in terms of DmeanTL except for PM-MTPM (p value = 0.32).

Table 4.

Mean absorbed dose delivered to aggregated tumoral liver calculated by means of multicompartmental (PM and MTPM) and 3D voxel dosimetry methods (LDM and DPK), for patients with more than a single lesion. COV(TNRi) is also reported

Patient DmeanTLGy COV(TNRi)
PM MTPM DPK LDM
1 28 26 ± 28 25 ± 27 25 ± 28 109
2 47 50 ± 11 44 ± 9 45 ± 10 22
4 32 33 ± 14 31 ± 13 32 ± 14 44
5 74 76 ± 36 66 ± 31 67 ± 32 47
7 60 56 ± 19 53 ± 17 54 ± 18 33
9 137 177 ± 178 137 ± 138 143 ± 144 100
10 70 75 ± 22 66 ± 20 69 ± 20 29
11 82 41 ± 47 41 ± 46 42 ± 47 113
12 127 129 ± 29 107 ± 26 108 ± 25 23
14 24 29 ± 10 28 ± 10 28 ± 10 36

In Fig. 7, the distribution of the ΔDmeanTifor all dosimetry methods comparisons overall 101 individual tumors is shown. Differences between PM and the other methods (MTPM, LDM, and DPK) were found to be fairly large with more than 30 % of individual tumors with differences exceeding 20 Gy. Differences between MTPM and both 3D-VDM methods were pretty small with 10 and 13 % of the individual tumors with differences higher than 5 Gy, for LDM and DPK methods respectively. The 70% of tumors presented differences between LDM and DPK methods less than 1 Gy.

Fig. 7.

Fig. 7

Distribution of differences in mean dose absorbed within individual tumors for the studied methods compared by pairs

DmeanTi between both multicompartimental methods (PM-MTPM) and between PM and both 3D-VDM (PM-LDM and PM-DPK) were correlated to differences between TNR and TNRi (rho: 0.96, p value: 0.00 for PM-MTPM; rho = 0.97, p value = 0.00 for PM-LDM; and rho = 0.97, p value: 0.00 for PM-DPK).

To assess the agreement between PM and MTPM, PM and LDM, MTPM and LDM and also between both 3D-VDM, in terms of Dmean, Bland-Altman plots are shown in Fig. 8 for Ti and in additional file 3 for TL. Comparisons between both multicompartimental methods (PM and MTPM) and DPK method are presented for Ti in additional file 4. Pearson’s correlation and Lin concordance coefficients for each comparison between dosimetry methods is summarized in Table 5, for both TL and Ti. The maximum ∆DmeanTi among the studied methods were − 166.0 Gy for PM-MTPM, − 97.8 Gy for PM-DPK, − 107.6 Gy for PM-LDM, 68.2 Gy for MTPM-DPK, 58.4 Gy for MTPM-LDM, and 9.8 Gy for LDM-DPK. DmeanTiwere higher between PM and the other methods than between MTPM and both 3D-VDM. The lowest differences were found between both 3D dosimetry methods. Significant differences were observed among the all dosimetry methods in terms of DmeanTi except for PM-MTPM (p value = 0.32) and PM-DPK (p value = 0.06).

Fig. 8.

Fig. 8

Bland-Altman diagrams representing the relative dose difference in Gy between the studied dosimetry methods versus the mean dose values in Gy for each individual tumor by pairs. The dashed line represents the average of the differences between the studied dosimetry methods, and the black line is the mean differences of ± 2 SD

Table 5.

Bland Altman analysis, Pearson’s correlation, and Lin concordance coefficients, of mean absorbed doses delivered to both tumoral liver and individual tumors volumes among the studied dosimetry methods (PM, MTPM, LDM, and DPK)

Bland-Altman Correlation and concordance coefficients
DMeanVOIGy Bias [95%CI] (Gy) ρ ρc
Tumor volume
 PM-MTPM − 0.7 − 32.5; 31.0 0.94 0.92
 PM-DPK 5.6 − 17.2; 28.5 0.95 0.94
 PM-LDM 6.7 − 15.8; 29.2 0.95 0.93
 MTPM-DPK 6.3 − 12.7; 25.4 1.00 0.96
 MTPM-LDM 7.4 − 14.1; 28.9 0.99 0.94
 LDM-DPK 1.1 − 2.0; 4.1 1.00 1.00
Individual tumors
 PM-MTPM 4.7 − 60.1; 69.5 0.55 0.55
 PM-DPK 7.1 − 52.2; 66.4 0.53 0.50
 PM-LDM 7.9 − 50.1; 65.8 0.53 0.50
 MTPM-DPK 2.4 − 10.6; 15.4 0.99 0.98
 MTPM-LDM 3.2 − 11.8; 18.1 0.99 0.97
 LDM-DPK 0.8 − 2.0; 3.5 1.00 1.00

DVH curves calculated by means of LDM and DPK dosimetry methods for two individual tumors are reported in Fig. 9. DVHs correspond to tumors, which yielded the highest differences between both 3D-VDM methods.

Fig. 9.

Fig. 9

Dose volume histogram curves for each individual tumor, corresponding to T1 of patient 9 a and T2 of patient 10 b. LDM dosimetry method is represented by continuous lines and DPK method with the dashed lines

The differences between LDM and DPK methods in DVH for each individual tumor volume in terms of D5, D25, D50, D70, and D95, are reported in Fig. 5. Higher differences were found for D95 while D70, D50, D25 and D5 showed lower variations in both compartments. The maximum difference was 55.3 Gy, corresponding to D95, within T1 of the patient 9 and it is due to a high absorbed dose gradient. The statistical analysis showed that DVH differences between both 3D-VDM methods for Ti were statistically significant except for D25 (p value 0.10).

The mean (range) difference in V100 Gy (%) due to applied LDM or DPK methods was 0.0 (− 1.4, 11.9) %. Significant differences were observed (p value < 0.01).

The results of the Bland-Altman analysis to assess differences in DVH in terms of D5, D25, D50, D70 and D95 are summarized in Table 3, for individual tumor volumes. Pearson’s correlation and Lin concordance coefficients were also reported.

Parametric images in terms of absorbed dose for both 3D-VDM methods, and the differences between them by voxel based subtraction are represented in Fig. 10. The images correspond to a patient who yielded the highest differences between both 3D-VDM methods. Figure 10c shows the voxels where parametric image calculated by applying DPK method take values above parametric image determined by means of LDM method, and Fig. 10d presents the voxels where LDM image take values above DPK image. NMSE between absorbed dose maps obtained with each 3D-VDM method was 0.24%, ranged from 0.12 to 0.78 %.

Fig. 10.

Fig. 10

Central coronal slice of the mean absorbed dose distribution map, for patient 9 determined by means of voxel dosimetry methods: LDM a and DPK b. Voxel based subtraction of parametric images (in Gy): DPK-LDM c and LDM-DPK d.

Discussion

The goal of this study was to evaluate the impact of the dosimetry approach on the resulting 90Y-RE pre-treatment absorbed dose based on 99mTc-MAA images. Several available dosimetry methods were compared on a group of 14 patients overall a total of 101 individual tumors. Two different multi-compartmental dosimetry methods (PM and new MTPM) and two 3D voxel based dosimetry methods (DPK and LDM) were studied. Mean absorbed dose, as well as DVH curves, were compared.

For optimal RE treatment, it is essential to be able to predict the absorbed dose to the tumor, healthy liver and lungs. An accurate prediction of these values can help to optimize patient selection and to truly individualize a safe and effective treatment planning [9]. For that, an optimal and reproducible dosimetry approach is not only important but essential. Currently, several dosimetry methods are available and there is no consensus on the best methodology to calculate the absorbed doses [1, 23].

In clinical practice, PM absorbed doses are calculated for the aggregated tumor compartment, and not for each individual tumor. The limitation of PM is the lack of spatial dose information, since it is assumed a homogeneous uptake of 99mTc-MAA in all individual tumors. The new approach of the MTPM method, developed in this study, takes into account the heterogeneity among the individual tumors but no the non-uniform distribution within a given tumor. Voxel-based dosimetry considers voxel-by-voxel heterogeneities between tumors and within each tumor.

Differences between dosimetry approaches are a critical issue and have been explored by other research groups. Pasciak et al. analyzed the effectiveness of several patient-specific dosimetry techniques by means of 99mTc-MAA simulation images of phantoms, such as Monte Carlo, local energy deposition in patient specific phantoms and kernel transport techniques in water [24]. To the best of our knowledge, this is the first study in which a multicompartmental method is implemented for individual tumors. Besides, it is the first time that a comparison among several multicompartimental (PM and MTPM) and 3D voxel dosimetry (LDM and DPK) methods in terms of absorbed doses within each individual tumor is carried out. Other groups have investigated differences on post-treatment PET and Bremsstrahlung images. Thus, Kao et al. studied a technical comparison of partition model and body surface area method with an emphasis on its clinical implications and discuss about future dosimetry techniques for 90Y-RE [2]. Mikel et al. in a study published in 2016 quantified differences that exist between glass microsphere package insert standard model (assuming tumor and healthy liver as a single compartment), partition model and Monte Carlo [25]. This group, in a different publication, also assessed differences between four different voxel-based dosimetry methods (Monte Carlo, soft tissue kernel with density correction, soft-tissue kernel and local deposition) for tumor, liver and lung absorbed doses based on 90Y-Bremsstrahlung SPECT-CT [26]. Pasciak et al. compared DPK convolution with LDM methods on 90Y-PET-CT images using a NEMA Phantom [27].

Absorbed dose by normal liver

Despite statistical differences were found in mean absorbed dose delivered to the normal liver between all methods, these differences are clinically acceptable. In general, the observed differences were higher between PM and both 3D methods than between LDM and DPK methods. The differences between PM and 3D dosimetry methods were more important within target normal liver than within whole normal liver.

According with Pearson’s correlation and Lin’s concordance coefficients, mean absorbed dose in the healthy liver, calculated by all the dosimetry methods included in this study, were highly correlated and concordant. Besides, only in 1, over the 14 studied patients, the difference is beyond the ± 2 standard deviations lines in the Bland-Altman plots. This concordance, added to the fact that the absolute differences in Dmean have no clinical impact, implies that PM and both 3D voxel dosimetry approaches may be used interchangeably for healthy tissue calculations.

The resulting DVH in healthy liver from both 3D-VDM are greatly similar. The extreme difference was found for D95 corresponding to patients who have received the higher Dmean. A high correlation and concordance were found between them and the resulting differences in the absorbed dose have no clinical impact. According to these results, LDM and DPK methods may be indistinctly used for healthy liver absorbed dose prediction.

Absorbed dose to tumoral liver

In general, the Dmean differences (in Gy) found among the dosimetry methods are higher in tumor volumes (TL and Ti) than in the non-tumoral ones (NLt and NLw), which is explained by the greater heterogeneity of the microspheres and the higher dose gradient within the tumor tissue.

For patients with more than a single tumor, a large heterogeneity in the 99mTc-MAA uptake among the individual tumors was found. The COV(TNRi) was 56 ± 37 %, ranging from 22 to 113%. This heterogeneity was also observed in the large variability of mean absorbed dose in the tumoral liver compartment determined by MTPM, DPK, and LDM methods, expressed as standard deviation (Table 4).

Despite the fact that no statistical differences in DmeanTiwere found for PM-MTPM and PM-LDM, differences in DmeanTi between the PM and the other methods (MTPM, LDM and DPK) were fairly large and may yield a substantial impact in treatment planning. Besides, a poor correlation and concordance between PM and the other studied dosimetry approaches were found. According to these results, this study demonstrates that due to the heterogeneity among tumors, to apply PM method introduces errors in the DmeanTi estimation. Therefore, PM method is not the best approach to evaluate de mean absorbed dose in the tumoral parenchyma as this approach neglects the heterogenenity between individual tumors

The strong correlation observed between differences in Dmean (PM-MTPM, PM-LDM and PM-DPK) and TNR-TNRi differences suggest that the individual segmentation and evaluation of the individual dose of the tumors plays an essential role in a proper dosimetry estimation.

To introduce the original MTMP method in the clinical practice leads to improve the patient selection; since the patient may not be a good candidate for RE if, due heterogeneous 90Y-microspheres distribution, one or more tumors to be treated do not achieve a therapeutic absorbed dose. Moreover, MTMP method could be applied to perform post-treatment dosimetry calculations. For those tumors that have not received a therapeutic absorbed dose, other therapies such as stereotactic body radiation therapy (SBRT), proton therapy, ablation, etc., may be considered.

In specific cases where patients have a single tumor, PM provided acceptable results in Dmean within the tumoral parenchyma; however, MTPM or 3D-VDM reduce uncertainty in the absorbed dose calculations for patients with several tumors. Similar conclusion was showed by Mikell et al. [26] who quantified differences between three dosimetry models used for 90Y RE: PM, glass microsphere package insert standard model and Monte Carlo.

Differences in DmeanTi between MTPM and both 3D-VDM were substantially lower than differences between PM and any 3D-VDM, as MTPM introduces more information regarding the different 99mTc-MAA uptake in each tumor. Although differences between MTPM and both LDM and DPK methods were statistically significant, a strong correlation and concordance were found, and the differences in DmeanTi have no impact in clinical management. Consequently, MTPM or 3D dosimetry approaches may be indistinctly used to calculate Dmean within individual tumor volumes.

The principal advantage of MTPM over 3D-VDM is its easier implementation, because no algorithm or specific software dosimetry is necessary for dose calculation and therefore it has a greater availability in daily clinical practice. Despite applying MTPM is laborious in those cases in which the number of individual tumors is large, the same contours must be delineated to perform 3D voxel dosimetry. Taking this into account, in centers where it is not possible to perform a dosimetry based on 3D voxel methods, to implement MTPM is recommended in order to improve the accuracy of Dmean calculation.

The differences between LDM and DPK methods inDmeanTi and DVH are statistically significant. However, taking into account the uncertainty associated with the dose calculation procedure, these differences are clinically acceptable and have no impact in treatment planning. An almost perfect correlation and concordance were found. Then either 3D dosimetry approaches may be indistinctly used to calculate Dmean within individual tumor volumes.

Although global differences in terms of absorbed dose in Gy and NMSE are small, important voxel-by-voxel differences have been observed (maximum difference: 77 Gy). This is a consequence of the high absorbed dose gradient. Since DPK method takes into account the energy transport to adjacent voxels, it was expected that DPK tended to provide higher absorbed doses than LDM in outer edge of the uptake region, while LDM tended to provide higher absorbed doses in the central area, as the results of this study show (Fig. 10).

The need to compare dosimetry performed on 99mTc-MAA SPECT-CT pre-therapy images to post-therapy images, in order to study the effectiveness of 99mTc-MAA, is pointed out by many studies [2831]. As this study suggest, in some situations, the dosimetry approach has a substantial impact on the resulting absorbed doses; therefore, considering the differences in absorbed doses due to different dosimetry methods is essential not only to compare pre and post dosimetry calculations but also to interpret different clinical studies.

As it is shown in this study, the differences in DmeanNLwbetween PM and both 3D-VDM that we found in Bland-Altman analysis (− 0.3 and 1.7 Gy for DPK; − 0.1 and 2.4 Gy for LDM) are lower than the differences between pre and post-dosimetry comparison (− 7.4 and 9.1 Gy) reported by Richetta et al. [29]. However, the differences in DmeanTibetween PM and the other dosimetry approaches that we found in Bland-Altman analysis (− 60.1 and 69.5 Gy for MTPM, − 52.2 and 66.4 Gy for DPK; − 50.1 and 65.8 Gy for LDM) are comparable to the differences between pre and post-dosimetry comparison (− 79 and 68 Gy) reported by Richetta et al. [29].

Therefore, based on the results of this study, to use the same approach in predictive and post-treatment dosimetry calculation is recommended, in order to be able to make a comparison between them in terms of Dmean within both tumoral and non-tumoral parenchyma. Furthermore, to take into account the differences in Dmean across dosimetry methods is essential in interpreting clinical studies that use different dosimetry approaches.

Limitations and future perspectives

There are other methodological variables, not included in this work, which may be source of differences in the absorbed dose results. Image registration and VOIs segmentation may be a limiting factor for all dosimetry methods, due to large absorbed dose gradients presents in RE treatments, especially important near liver-lung interface [26]. The calibration factor used to convert counts to Bq/cm2 may be determined not only applying a self-calibration factor, as described in this work, but also by means of the evaluation of SPECT system sensitivity [32]. Voxel-based dosimetry may suffer from bias related to acquisition and reconstruction parameters, partial volume effect, etc. [24].

Conclusion

This work shows that significant differences exists among the studied pre-treatment dosimetry approaches (PM, MTPM, LDM, and DPK methods) for 90Y RE treatments. However, these differences do not yield a substantial impact in treatment planning for healthy tissue from a clinical point of view and different dosimetry approaches may be applied indistinctly.

Due to the large heterogeneity found among individual tumors, an individual segmentation and evaluation of the tumors plays an essential role in a proper dosimetry estimation. Therefore, in patients with multiple tumors, to apply PM method is not recommended in tumoral parenchyma since the mean absorbed dose are estimated within the aggregated tumor compartment, and the 3D dosimetry methods or the new MTPM should be applied.

Moreover, when a 3D voxel based dosimetry method is not available, MTPM is the best option to estimate the mean absorbed dose within each tumor. Both LDM and DPK methods may be indistinctly used.

Although this study showed that some methods are interchangeably, to make an optimal comparison between absorbed dose values from different publications or even from the same study, i.e., pre and post-dosimetry, is essential to take into account differences in the absorbed doses caused by the adoption of a different dosimetry approach.

Supplementary Information

Additional file 1:. (335.7KB, xlsm)
Additional file 2:. (44.3KB, jpg)
Additional file 3:. (95.3KB, pptx)
Additional file 4:. (60.7KB, jpg)

Acknowledgements

Not applicable

Abbreviations

3D

Tridimensional

3D-VDM

3D-voxel dosimetry methods

(DMeanVOI)

Mean absorbed dose difference within a VOI

ρ

Pearson correlation coefficient

ρc

Lin concordance coefficient

A(99mTc)

99mTc Administered Activity

A(90Y)

90Y Administered Activity

Avoxel(90Y)

90Y Activity within a voxel

A~voxelY90

9oY Cumulative Activity within a voxel

A~voxelsY90

90Y Cumulative Activity within a source voxel

Clung

Lung total counts

CNLt

Target normal liver total counts

CTi

Individual tumor total counts

CTL

Tumoral liver total counts

Cvoxel(99mTc)

Total counts within a voxel generated by 99mTc

CWL

Whole liver total counts

CWL(99mTc)

Whole liver total counts generated by 99mTc

COV

Coefficient of variation

CT

Computed tomography

D5

Minimum absorbed dose to 5% in the corresponding VOI

D25

Minimum absorbed dose to 25% in the corresponding VOI

D50

Minimum absorbed dose to 50% in the corresponding VOI

D70

Minimum absorbed dose to 70% in the corresponding VOI

D95

Minimum absorbed dose to 95% in the corresponding VOI

Dmean

Mean absorbed dose

DMeanTi

Individual tumor mean absorbed dDose

DMeanTL

Tumoral liver mean absorbed dose

DMeanNLt

Target normal liver mean absorbed dose

DMeanNLw

Whole normal liver mean absorbed dose

DvoxeltY90

90Y Mean absorbed dose within the target voxel

DICOM-RT

Digital Imaging and Communications in Medicine - Radiation Therapy

DPK

Dose Point Kernel

DVH

Dose volume histogram

Eβ(90Y)〉

Deposited β-energy per disintegration in average

FWHM

Full width half maximum

LDM

Local deposition method

LEHR

Low energy high resolution

LSF

Lung shunt fraction

Mvoxel

Voxel mass

MAA

Macroaggregated albumin

MIRD

Medical internal radiation dose

MRI

Magnetic resonance imaging

MTPM

Multi-tumor partition model

n

Number or tumors

NMSE

Normalized mean square error

NLt

Target normal liver

NLw

Whole normal liver

pi

Ith voxel of the LDM parametric image

PM

Partition model

PTV

Planning target volume

RE

Radioembolization

S(voxelt ← voxels)

Absorbed dose within the target voxel per unit of cumulative activity within the source voxel

SPECT

Single photon emission computed tomography

T1/2(90Y)

Physical 90Y half life

Ti

Individual tumors

TL

Tumoral liver

TNR

Tumor to non-tumor ratio

TNRi

Individual tumor to non-tumor ratio

V20

Percentage of corresponding VOI receiving at least 20 Gy

V100

Percentage of the tumor volume receiving at least 100 Gy

VNLt

Target normal liver volume

VNLw

Whole normal liver volume

VTi

Individual tumor volume

VTL

Tumoral liver volume

VOI

Volume of interest

WL

Whole liver

xi

Ith voxel of the DPK parametric image

Authors’ contributions

The contribution of the authors is substantiated below:VM participated in the conceptualization and design of the investigation. Also contributed to the statistical analysis and interpretation of data, developed the software used in this work and was the major contributor in writing the manuscript. EP and JMM-C contributed to the conception and design of the work and substantively revised it. MR-F and LS-R analyzed and interpreted the patient clinical data. LS have substantially revised the work, and ensuring English language is of sufficient quality to be understood. AZ delineated all the contours in order to perform dosimetry. All authors read, advised, and approved the final manuscript

Funding

Not applicable.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Ethics approval and consent to participate

This study has been approved by the institutional review board (Comité Ético de Investigación Clínica Universidad de Navarra).

Consent for publication

Not applicable. Images in Fig. 1 and Fig. 10 are entirely unidentifiable and there are no details on individuals reported within the manuscript.

Competing interests

The authors declare that they have no competing interests.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Tafti BA, Padia SA. Dosimetry of Y-90 Microspheres Utilizing Tc-99 m SPECT and Y-90 PET. Semin Nucl Med. 2019;49:211–217. doi: 10.1053/j.semnuclmed.2019.01.005. [DOI] [PubMed] [Google Scholar]
  • 2.Kao YH, Tan EH, Ng CE, Goh SW. Clinical implications of the body surface area method versus partition model dosimetry for yttrium-90 radioembolization using resin microspheres: a technical review. Ann Nucl Med. 2011;25:455–461. doi: 10.1007/s12149-011-0499-6. [DOI] [PubMed] [Google Scholar]
  • 3.Salem R, Lewandowski RJ, Mulcahy MF, Riaz A, Ryu RK, Ibrahim S, et al. Radioembolization for Hepatocellular Carcinoma Using Yttrium-90 Microspheres: A Comprehensive Report of Long-term Outcomes. Gastroenterology. 2010;138(1):52–64. doi: 10.1053/j.gastro.2009.09.006. [DOI] [PubMed] [Google Scholar]
  • 4.Sangro B, Bilbao JI, Boan J, Martinez-Cuesta A, Benito A, Rodriguez J, et al. Radioembolization using 90Y-resin microspheres for patients with advanced hepatocellular carcinoma. Int J Radiat Oncol Biol Phys. 2006;66:792–800. doi: 10.1016/j.ijrobp.2006.05.065. [DOI] [PubMed] [Google Scholar]
  • 5.Kennedy A, Coldwell D, Sangro B, Wasan H, Salem R. Radioembolization for the treatment of liver tumors: general principles. Am J Clin Oncol. 2012;35:91–99. doi: 10.1097/COC.0b013e3181f47583. [DOI] [PubMed] [Google Scholar]
  • 6.Rodríguez-Lago I, Carretero C, Herráiz M, Subtil JC, Betés M, Rodríguez-Fraile M, et al. Long-term follow-up study of gastroduodenal lesions after radioembolization of hepatic tumors. World J Gastroenterol. 2013;19:2935–2940. doi: 10.3748/wjg.v19.i19.2935. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ahmadzadehfar H, Sabet A, Biermann K, Muckle M, Brockmann H, Kuhl C, et al. The Significance of 99mTc-MAA SPECT/CT Liver Perfusion Imaging in Treatment Planning for 90Y-Microsphere Selective Internal Radiation Treatment. J Nucl Med. 2010;51:1206–1212. doi: 10.2967/jnumed.109.074559. [DOI] [PubMed] [Google Scholar]
  • 8.Garin E, Lenoir L, Rolland Y, Edeline J, Mesbah H, Laffont S, et al. Dosimetry based on 99mTc-macroaggregated albumin SPECT/CT accurately predicts tumor response and survival in hepatocellular carcinoma patients treated with 90Y-loaded glass microspheres: preliminary results. J Nucl Med. 2012;53:255–263. doi: 10.2967/jnumed.111.094235. [DOI] [PubMed] [Google Scholar]
  • 9.Smits MLJ, Elschot M, Sze DY, Kao YH, Nijsen JFW, Iagaru AH, et al. Radioembolization Dosimetry: The Road Ahead. Cardiovasc Intervent Radiol. 2015;38:261–269. doi: 10.1007/s00270-014-1042-7. [DOI] [PubMed] [Google Scholar]
  • 10.Ho S, Lau WY, Leung TWT, Chan M, Ngar YK, Johnson PJ, et al. Partition model for estimating radiation doses from yttrium-90 microspheres in treating hepatic tumours. Eur J Nucl Med. 1996;23:947–952. doi: 10.1007/BF01084369. [DOI] [PubMed] [Google Scholar]
  • 11.Loevinger R, Budinger TF, Watson EE, S of NM (1953-, Committee. MIRD.MIRD primer for absorbed dose calculations. New York: Society of Nuclear Medicine; 1991.
  • 12.Gil-alzugaray B, Chopitea A, Mercedes I, Bilbao JI, Rodriguez-fraile M, Rodriguez J, et al. Prognostic factors and prevention of radioembolization-induced liver disease. Hepatology. 2013;57:1078–1087. doi: 10.1002/hep.26191. [DOI] [PubMed] [Google Scholar]
  • 13.Sirtex Medical Limited. Sirtex Package Insert. 2017;57:3–5 Available from: https://www.sirtex.com/eu/clinicians/package-insert/..
  • 14.Giammarile F, Bodei L, Chiesa C, Flux G, Forrer F, Kraeber-Bodere F, et al. EANM procedure guideline for the treatment of liver cancer and liver metastases with intra-arterial radioactive compounds. Eur J Nucl Med Mol Imaging. 2011;38:1393–1406. doi: 10.1007/s00259-011-1812-2. [DOI] [PubMed] [Google Scholar]
  • 15.Kao YH, Steinberg JD, Tay YS, Lim GK, Yan J, Townsend DW, et al. Post-radioembolization yttrium-90 PET/CT - part 2: dose-response and tumor predictive dosimetry for resin microspheres. EJNMMI Res. 2013;3(1):57. doi: 10.1186/2191-219X-3-57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Garin E, Lenoir L, Rolland Y, Laffont S, Pracht M, Mesbah H, et al. Effectiveness of quantitative MAA SPECT/CT for the definition of vascularized hepatic volume and dosimetric approach: phantom validation and clinical preliminary results in patients with complex hepatic vascularization treated with yttrium-90-labeled micr. Nucl Med Commun. 2011;32:1245–1255. doi: 10.1097/MNM.0b013e32834a716b. [DOI] [PubMed] [Google Scholar]
  • 17.Flamen P, Vanderlinden B, Delatte P, Ghanem G, Ameye L, Van Den Eynde M, et al. Multimodality imaging can predict the metabolic response of unresectable colorectal liver metastases to radioembolization therapy with Yttrium-90 labeled resin microspheres. Phys Med Biol. 2008;53:6591. doi: 10.1088/0031-9155/53/22/019. [DOI] [PubMed] [Google Scholar]
  • 18.Machac J, Muzinic M, Dacosta M, Zhang Z, Ph D, Heiba S. Quantitative comparison of yttrium-90 ( Y ) -microspheres and technetium-99 m ( Tc ) -macroaggregated albumin SPECT images for planning 90 Y. Ther Liver Cancer. 2010;3:253–262. doi: 10.1177/153303461000900304. [DOI] [PubMed] [Google Scholar]
  • 19.Pacilio M, Amato E, Lanconelli N, Basile C, Torres LA, Botta F, et al. Differences in 3D dose distributions due to calculation method of voxel S-values and the influence of image blurring in SPECT. Physics in medicine and biology. 2015;60:1945–64. doi: 10.1088/0031-9155/60/5/1945. [DOI] [PubMed] [Google Scholar]
  • 20.Chiesa C, Mira M, Maccauro M, Spreafico C, Romito R, Morosi C, et al. Radioembolization of hepatocarcinoma with 90Y glass microspheres: development of an individualized treatment planning strategy based on dosimetry and radiobiology. Eur J Nucl Med Mol Imaging. 2015;42:1718–1738. doi: 10.1007/s00259-015-3068-8. [DOI] [PubMed] [Google Scholar]
  • 21.Lanconelli N, Pacilio M, Lo MS, Botta F, Di DA, LAT A, et al. A free database of radionuclide voxel S values for the dosimetry of nonuniform activity distributions. Physics in medicine and biology. 2012;57:517–33. [DOI] [PubMed]
  • 22.Pacilio M, Ferrari M, Chiesa C, Lorenzon L, Mira M, Botta F, et al. Impact of SPECT corrections on 3D-dosimetry for liver transarterial radioembolization using the patient relative calibration methodology. Med Phys. 2016;43:4053–4064. doi: 10.1118/1.4953203. [DOI] [PubMed] [Google Scholar]
  • 23.Tong AKT, Kao YH, Too C, Chin KFW, Ng DCE, Chow PKH. Yttrium-90 hepatic radioembolization: Clinical review and current techniques in interventional radiology and personalized dosimetry. Br J Radiol. 2016;89(1062):20150943. doi: 10.1259/bjr.20150943. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Pasciak AS, Erwin WD. Effect of Voxel Size and Computation Method on Tc-99 m MAA SPECT/CT-Based Dose Estimation for Y-90 Microsphere Therapy. IEEE Trans Med Imaging. 2009;28:1754–1758. doi: 10.1109/TMI.2009.2022753. [DOI] [PubMed] [Google Scholar]
  • 25.Mikell JK, Mahvash A, Siman W, Baladandayuthapani V, Mourtada F, Kappadath SC. Selective Internal Radiation Therapy With Yttrium-90 Glass Microspheres: Biases and Uncertainties in Absorbed Dose Calculations Between Clinical Dosimetry Models. Int J Radiat Oncol Biol Phys. 2016;96:888–896. doi: 10.1016/j.ijrobp.2016.07.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Mikell JK, Mahvash A, Siman W, Mourtada F, Kappadath SC. Comparing voxel-based absorbed dosimetry methods in tumors, liver, lung, and at the liver-lung interface for 90Y microsphere selective internal radiation therapy. EJNMMI Phys. 2015;2:1–14. doi: 10.1186/s40658-015-0119-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Pasciak AS, Bourgeois AC, Bradley YC. A comparison of techniques for 90Y PET/CT image-based dosimetry following radioembolization with resin microspheres. Front Oncol. 2014;4:121. [DOI] [PMC free article] [PubMed]
  • 28.Kafrouni M, Allimant C, Fourcade M, Vauclin S, Guiu B, Mariano-goulart D, et al. Analysis of differences between 99 m Tc-MAA dosimetry for hepatocellular carcinoma selective internal radiation therapy. EJNMMI Res. 2019;9:62. doi: 10.1186/s13550-019-0533-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Richetta E, Pasquino M, Poli M, Cutaia C, Valero C, Tabone M, et al. Physica Medica PET-CT post therapy dosimetry in radioembolization with resin 90 Y microspheres : Comparison with pre-treatment SPECT-CT 99 m Tc-MAA results. Phys Med. 2019;64:16–23. doi: 10.1016/j.ejmp.2019.05.025. [DOI] [PubMed] [Google Scholar]
  • 30.Jadoul A, Hustinx R. Comparative dosimetry between 99 m Tc-MAA SPECT / CT and 90 Y PET / CT in primary and metastatic liver tumors. Eur J Nucl Med Mol Imaging. 2019;47:828–37. [DOI] [PubMed]
  • 31.Gnesin S, Canetti L, Adib S, Cherbuin N, Silva-Monteiro M, Bize P, et al. Partition model based 99mTc-MAA SPECT/CT predictive dosimetry compared to 90Y TOF PET/CT post-treatment dosimetry in radioembolisation of hepatocellular carcinoma: A quantitative agreement comparison. J Nucl Med. 2016;57(11):1672–8. doi: 10.2967/jnumed.116.173104. [DOI] [PubMed] [Google Scholar]
  • 32.Lassmann M, Chiesa C, Flux G, Bardiès M. EANM Dosimetry Committee guidance document. Good Pract Clin Dosimetry Report.Eur J Nucl Med Mol Imaging. 2011:192–200. [DOI] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Additional file 1:. (335.7KB, xlsm)
Additional file 2:. (44.3KB, jpg)
Additional file 3:. (95.3KB, pptx)
Additional file 4:. (60.7KB, jpg)

Data Availability Statement

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.


Articles from EJNMMI Physics are provided here courtesy of Springer-Verlag

RESOURCES