Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2021 Sep 14;16(9):e0257468. doi: 10.1371/journal.pone.0257468

Can sterilization of disposable face masks be an alternative for imported face masks? A nationwide field study including 19 sterilization departments and 471 imported brand types during COVID-19 shortages

B van Straten 1,*, P D Robertson 1, H Oussoren 2, S Pereira Espindola 3, E Ghanbari 3, J Dankelman 1, S Picken 3, T Horeman 1
Editor: Mahendra Singh Dhaka4
PMCID: PMC8439445  PMID: 34520503

Abstract

Background

Face masks, also referred to as half masks, are essential to protect healthcare professionals working in close contact with patients with COVID-19-related symptoms. Because of the Corona material shortages, healthcare institutions sought an approach to reuse face masks or to purchase new, imported masks. The filter quality of these masks remained unclear. Therefore, the aim of this study was to assess the quality of sterilized and imported FFP2/KN95 face masks.

Methods

A 48-minute steam sterilization process of single-use FFP2/KN95 face masks with a 15 minute holding time at 121°C was developed, validated and implemented in the Central Sterilization Departments (CSSD) of 19 different hospitals. Masks sterilized by steam and H2O2 plasma as well as new, imported masks were tested for particle filtration efficiency (PFE) and pressure drop in a custom-made test setup.

Results

The results of 84 masks tested on the PFE dry particle test setup showed differences of 2.3±2% (mean±SD). Test data showed that the mean PFE values of 444 sterilized FFP2 face masks from the 19 CSSDs were 90±11% (mean±SD), and those of 474 new, imported KN95/FFP2 face masks were 83±16% (mean±SD). Differences in PFE of masks received from different sterilization departments were found.

Conclusion

Face masks can be reprocessed with 121 °C steam or H2O2 plasma sterilization with a minimal reduction in PFE. PFE comparison between filter material of sterilized masks and new, imported masks indicates that the filter material of most reprocessed masks of high quality brands can outperform new, imported face masks of unknown brands. Although the PFE of tested face masks from different sterilization departments remained efficient, using different types of sterilization equipment, can result in different PFE outcomes.

Introduction

After the outbreak of COVID-19, this respiratory disease has spread at a rapid pace [1, 2]. Adequate face masks are essential to protect healthcare professionals. In many hospitals shortages of personal protection equipment occurred due to increased demand [3]. In the search for alternative sources, hospitals started to consider reusing their single-use face masks by sterilizing them [4].

Face masks, also referred to as half masks, are used to protect individuals against airborne particles during aerosol generating procedures. Three classes of particle filtering face piece (FFPs) are described in European Norm (EN) 149:2001+ A1:2009 [5]. The most commonly used masks in relation to COVID-19 are the Class 2 FFP2 masks. These are considered to be equivalent to the American N95 mask [6], conforming to the standards of the National Institute for Occupational Safety and Health (NIOSH) 42 CFR 84 mask [7], and the Chinese KN95 mask complying to the Guobiao (GB) 2626–2006 standard [8]. The filter efficiency of smaller particles is a crucial element. The European Norm requires a minimum filter efficiency of 94%, whereas NIOSH [7] and GB [8] require 95%.

Testing filter material of face mask

EN 149:2001+A1:2009 [5] and more specifically NEN-EN 13274–7:2019, part 7 describe a test setup to determine the particle filtration efficiency (PFE) of face masks. It consists of a flow tube, a flow generator, a NaCl particle generator and two particle measurement devices and generates flows up to 120 l/min with NaCl particles of 0.1 to 10 μm. Unfortunately, this setup is costly to build. Therefore, in the first 2 months of COVID-19 only 2 systems were operational in the Netherlands that could be used for testing new, imported face masks. The cost for testing one face mask was approximately 1,500 Euro with a waiting list of up to 4 weeks. A new quick testing method was needed.

Potential reprocessing methods

Several studies have shown the effects of different sterilization methods, including gamma sterilization, plasma sterilization, steam and dry heat sterilization, microwaves, washing machines and UV–C light, as methods to decontaminate face masks for reuse [913]. These studies suggest that gamma and steam sterilization conducted at 134°C damage the microstructure of the filter material [9].

Washing machines and microwaves have a low capacity, and microwaves do not create a uniform heat distribution and require a steam bag [1012]. Some studies suggest that the high concentration of liquid H2O2 in plasma sterilization (approx. 60%), and its strongly charged ionized vapor may neutralize the electrostatic charge of the filter media [11, 12]. Moreover, the sterilization efficacy would likely be affected by the presence of moisture (e.g. exhaled breath) in worn masks, as water is a polar molecule. Finally, the capacity per run remains low due to the vacuum-driven process [13]. The evaporation of moisture may restrict the sterilizer’s ability to pull deep vacuum. UV treatment of face masks seems to have potential but requires preparation time as face masks need to be unfolded in such a way that UV light reaches all of the mask material [1013]. UV-B sterilization was not considered as this method is not yet commonly used and not readily available at hospital sterilization departments. Steam sterilization at 121°C could be an option since studies have shown the effectiveness at 121°C in inactivating the Coronavirus [14, 15].

Pilot studies, that included ATCC 12228 bacterial testing, have been conducted to determine whether 121°C sterilization was a safe and effective method to deactivate the Corona virus. The protocols and results were made available to hospitals via the repository of the Delft University of Technology after demonstrating that sterilization of face masks was possible up to 5 times for high-quality face masks [9, 16]. Although proven efficient, the potential of this new 121°C sterilization method was not explored. Moreover, a study where many different brands of face masks were processed at different CSSD’s, with comparisons between new, imported masks and sterilized masks, did not exist. Therefore the aim of this study is to find the best alternative for high quality face masks in times of shortage by assessing the quality of sterilized and imported FFP2/KN95 face mask filter materials.

The following research questions were defined:

  1. Can FFP2 masks be reprocessed using 121°C steam or H2O2 plasma sterilization?

  2. Are reprocessed face masks an alternative for new ones?

  3. What effect does sterilization have on the materials?

Methods

A sterilization facility of a Dutch CSSD (ISO 7 validated, Van Straten Medical, De Meern, the Netherlands, operated by CSA services) was set-up for the purpose of reprocessing used (potentially COVID-19 contaminated) FFP2 face masks. New testing methods were developed to test the filter material quality after sterilization [4, 9, 16]. The testing was carried out for any hospital, reseller or manufacturer wanting to check the quality of sterilized or new, imported face masks.

Reprocessing by 121°C steam at CSA services sterilization

Within this new reprocessing approach, decontamination was done solely by steam sterilization. To implement the 121°C sterilization process, a special logistical routing was set-up to collect and process face masks. Upon receipt, the masks were removed from their double wrapping and inspected individually for visual damage. In case of deformations, dirt, lipstick, hairs, black streaks, stains or other deviations, the masks were discarded. The visually approved face masks were marked with a dot and packaged in autoclavable impermeable sterilization laminate bags (type CLFP150X300WI-S20, Halyard, UK) (Fig 1). A mask was disposed after it was marked with a maximum of 5 dots. A maximum of five face masks were packaged per bag to ensure proper sterilization. The autoclaves (GSS6713H-E, Getinge, Sweden) were activated with a 121°C program and re-validated. The autoclave cycle was set for 48 minutes with a 15 min holding time (high vacuum 121 °C; ≥15 min HT, total CT 48 min). Face masks with a higher class (FFP3/N95) were treated as FFP2/KN95. The PFE for the average FFP2/KN95 mask material has to be 94% or higher for a pass and under 94% for a fail [5]. The performance of the mask material was determined by measuring the PFE and breathing resistance. Fig 2 shows the particle counter with a custom-made particle chamber (Lighthouse Solair 3100, San Francisco, www.golighthouse.com). The machine drew air through the mask into the chamber and to the particle counter. The diameter of the chamber was chosen such that it guaranteed sufficient airflow through the filter material for the particle counter [9, 16, 17]. The PFE was determined by measuring the difference in the number of particles before and after filtration by the mask. First, the particle concentration in a standard volume of room air was determined by measuring the number of particles (sizes 0.3, 0.5, 1.0, and 5.0 μm) in a volume of surrounding air. Second, the mask was installed on the chamber to measure the number of particles after filtration.

Fig 1. Autoclave procedure with Halyard laminate bags.

Fig 1

Left, laminated bags entering the autoclave. Middle, masks are wrapped in laminate. Right, the 121 °C steam sterilization program as used for face mask sterilization.

Fig 2. Lighthouse Solair 3100 particle counter connected to a particle chamber.

Fig 2

When very dense filter materials are used with a very high PFE for the smallest particles, it causes a breathing resistance for the user [16, 17]. This resistance causes a pressure drop which was measured using an analog differential pressure sensor, type SDP2000-L (Sensirion AG, Staefa ZH, Switzerland) connected to the particle chamber. The pressure sensor is temperature compensated, calibrated and has a resolution of 11 Pa with a repeatability of 0.3% and accuracy of 1% [17]. The breathability requirements for respiratory protective devices are provided in a European standard [18]. The maximum permitted resistance (mbar) differs for the FFP1, FFP2, and FFP3 masks, ranging from 0.6–1.0 for inhalation at 30 l/min, 2.1–3.0 for 95 l/min and 3.0 for exhalation at 160 l/m. The norm for a FFP2-mask at 30 l/min is 0.7 mbar.

Test setup validation to the European Norm EN 13274–7

The accuracy of the developed particle test setup was evaluated by comparing results from known face masks, tested on (our) particle setup, with the results of the same brand and type masks, tested on a continuous flow system. The continuous flow test system used NaCl particles and was built at the Delft University of Technology according to NEN-EN 13274–7:2019 [19].

Additionally, the standard EN 149:2001+A1:2009 describes experiments to determine if a mask creates a proper fit on the face without leakages. The inward leakage is determined by means of a fit test and strap test [17, 18]. In this study, inspection of the materials were conducted and leakage tests were performed on all reprocessed masks. Although this study focused on the material properties, only masks that showed no change in fit or material properties were included. The types that did deteriorate were registered and disposed after arrival. Although we followed the EN-149 standard as much as possible, we did reference our outcomes with the NaCl test since we used a custom-made test setup as a non-standard EN-149 methodology.

121 °C steam sterilization consistency between CSSDs

The consistency of sterilization results, caused by different processes and equipment was compared between 19 CSSDs. Samples of masks representing the most commonly used brands and types were selected and measured with the PFE setup. Only CSSDs were included that provided a minimum of four masks that were sterilized once. Face masks were not cleaned after visual inspection and prior to sterilization. A student’s t-test (two tailed, unequal variance, SPSS 17.0) was used for comparison, and a probability of p<0.05 was considered to be statistically significant.

Face mask material differences

Differences in mask material were analysed by chemically and thermally comparing the fabric of the two most common types. Therefore, a differential scanning calorimetry (DSC), an X-ray diffraction (XRD) and a transform infrared spectroscopy (FTIR) were conducted (S1 File).

Testing new masks

To determine how many samples of imported face masks were needed, the variance of the PFE and pressure drop was determined on three imported face masks. Ten measurements conducted on each face mask type (S2 File) showed that the largest variance was found in the 0.3 μm particle size category of 0.6%, 1.1% and 0.3% of the mean values respectively. The pressure drop measurements showed a variance of 0.7%, 1.8% and 1.8% of the mean values respectively. This low variance indicated consistent behavior of the filter materials. Combined with the importance of a short processing time it was determined to measure a minimum of two masks of each type that was provided by the clients. The averaged values are listed in S3 File. In case a deviation of more than 10% was found in the 0.3 μm category, two additional masks were tested and the supplier was notified. This data was excluded from the study.

Samples were selected for PFE measurement from batches of imported masks. The PFE results of those new face masks were compared to the PFE results of the sterilized face masks from the 19 CSSDs. New imported face masks that scored above 98% PFE in the particle range were further investigated for breathability by measuring the pressure drop.

Results

Reprocessing by 121°C steam sterilization at CSA services

A total of 74,834 masks were processed by the CSSD of CSA services. Of these masks, 56,668 were disposed after incoming inspection due to visual damage, deformities or dirt. The remaining 18,166 face masks were steam sterilized at 121 °C. Table 1 shows the top five brands that were sterilized and returned to hospitals for use.

Table 1. Top five reprocessed face masks.

Brand (Type) Percentage
3M (1862+) 42%
3M (1872+) 21%
My-T-Gear 8%
IMG Europe (R620) 5%
Kimberly Clarc Corp 5%
Rest 19%

Test setup validation to the European Norm EN 13274–7

Preliminary tests conducted with 84 different masks, tested on the PFE dry particle test setup and a NaCl test setup, following the EN 13274–7:2019 standard, indicated an outcome deviation of 2.3±2% (mean±SD) on average with a max of 7% (S4 File). A measurement test conducted with another ten different masks indicated that an average of 19 s ±21% (mean±SD) (SD 21) is needed to install and inspect the mask on the particle counter and an additional 15 s ±13% (mean±SD) is needed to take the mask from the system after 1 minute of measurement. None of the masks showed visual signs of deformation or damage after being measured.

121 °C steam sterilization consistency between CSSDs

The reprocessing methods by means of steam sterilization were adopted by 19 hospitals (Amsterdam University Medical Center (VUmc and AMC locations), Holendrecht Medical Center, Franciscus Hospital, CombiSter RDGG & Haga, Spaarne Hospital, Erasmuc MC, University Medical Center Groningen, Leiden University Medical Center, Flevo Hospital, Isala Hospital, Diakonessenhuis Utrecht, VieCuri, Rode Kruis Hospital, Noordwest Hospital Group, Amphia Hospital, and Tweesteden Hospital). The PFE results of 444 reprocessed FFP2/KN95 face masks from the CSSDs of 19 different hospitals in the Netherlands are provided in Fig 3. Of the 444 masks, 371 masks were sterilized with steam sterilization and 73 with H2O2 plasma (S5 File).

Fig 3. PFE values after sterilization with 121 °C steam or H2O2 plasma sterilization in chronological order from worst to best.

Fig 3

The red dotted line indicates the FFP2 level at 94% PFE. Each mask number represents a sample of a sterilized batch from one type only.

From these 444 tested face masks, 58 3M 1862+ face masks were provided by seven CSSDs of four university hospitals, one one general hospital and one general practitioner which were only sterilized once at 121 °C using steam sterilization (S6 File). The influence of different installations, protocols or staff on the PFE is shown in Fig 4. The “N” value indicates how many 3M 1862+ face masks were included in the study that were only sterilized once. The statistical tests reveal differences in outcome mainly for the CSSD of University Hospital 2.

Fig 4. PFE values with standard deviation of different 3M 1862+ coming from 7 different CSSDs.

Fig 4

Statistical differences are indicated with P values above the figure.

Face mask material differences

The 444 face masks consisted of 101 different types. Of the 101 different types, 3M 1862 and Kolmi Op-Air were mostly tested. The PFE results of 89 3M 1862 and 26 Kolmi Op-Air are provided in Table 2 for 0.3, 0.5, 1 and 5 μm particles.

Table 2. Particle filter efficiency of two commonly used mask after either 121 °C steam or H2O2 plasma sterilization.
Brand & type Number of masks Sterilization method 0.3 μ 0.5 μ 1 μ 5 μ Mean
% PFE (SD) % PFE (SD) % PFE (SD) % PFE (SD) % PFE
3M 1862 5 H2O2 86,4 93,8 97,4 99,5 94
Sterrad (12,5) (6,2) (2,7) (0,5)
3M 1862 72 121 °C 93,6 97,3 99,0 99,7 97
steam (4,1) (2,1) (0,8) (0,7)
3M 1862 4 2 x H2O2 41,3 66,9 83,9 99,5 73
Sterrad (1,7) (1,6) (1,3) (0,4)
3M 1862 8 2 x 121 °C 91,6 96,2 98,3 100 97
steam (3,2) (1,8) (0,8) (0,1)
Kolmi OP-Air M52010 11 H2O2 89,8 96,4 98,4 99,8 96
Sterrad (1,4) (1,4) (0,5) (0,3)
Kolmi OP-Air M52010 15 121 °C 21,2 56,3 78,4 99,8 64
steam (6,8) (8,5) (8,2) (0,5)

The results indicate that 3M 1862 shows low PFE values after 2x H2O2 plasma processing and that Kolmi Op-Air shows low and inconsistent PFE values after 1x 121 °C processing (S3 File).

Thermal properties of 3M Aura 1862+ and Kolmi Op-Air M52010 face masks using DSC

The three tests, differential scanning calorimetry (DSC), X-ray diffraction (XRD) and transform infrared spectroscopy (FTIR), confirmed that both masks consisted of 5 layers with the profile of Polypropylene (PP) material (S7 File).

Tests of new, imported masks

The PFE results of 471 different types of new, imported FFP2/KN95 face masks from collaborating hospitals and resellers are shown in Fig 5. Of these, 27 face masks scored above 98% PFE for the 0.3 micron particle size category. These masks were tested for breathability by measuring the pressure drop (S8 File). Fig 6 shows the breathing potential of the 27 face masks. The material of 27 face masks with high PFE values showed pressure drops between 251 and 3976 Pa on the measurement setup. When calculated for the total mask areas A and B, five out of 27 masks showed a total pressure drop higher than the EU standard of 0.7 mbar [18]. Finally, four masks showed readings at approximately 3.7 mbar, which is very close to the maximum measurable pressure drop of 4500 Pa. In two occasions the PFE data of the two tested masks of the same type deviated more than 10% (i.e. PFE of 67% vs 84% at 0.3 μm). Closer inspection revealed that the two masks had a different appearance as one had an additional logo in the shape of a heart.

Fig 5. PFE values of new imported FFP2/KN95 face masks in order from worst to best.

Fig 5

The red dotted line indicates the FFP2 level at 94% PFE. Each mask number represents a sample of a new batch from one type only.

Fig 6. Pressure drop of 27 new face mass with PFE>0,7 mbar.

Fig 6

Four masks performed really low (red), 5 performed around the EU norm of 0,7 mbar and 18 performed well according to the EU norm of 0,7 mbar.

Discussion

Regarding the research questions, it can be confirmed that FFP2 masks can be safely reprocessed with 121 °C steam sterilization if appropriate testing facilities are available. The data from Figs 3 and 5 indicate that reprocessed face masks can act as alternatives for new face masks as sterilization of well-known brand often gives better PFE results compared to newly imported masks. Although the base materials are similar, the manufacturing, preparation, and use of coatings have a large effect on the PFE of mainly the smaller particles.

The cross validation with the NaCl continuous flow setup built according to EN 13274–7 standard showed that the most important requirements for determining the filter material properties were met. After nineteen hospitals adopted the steam sterilization process, a nationwide data field experiment was initiated that informed multiple international NGOs, universities, and industry members about the advantages and disadvantages of sterilization of face masks [1721], which led to a Dutch standard for sterilization of face masks. After the first results were shared by request [22], general practitioners, dental practices and pharmacies claimed to successfully adopt the 121 °C sterilization process in their smaller sized autoclaves with sufficient results [16].

Sterilization with the purpose of reusing medical devices is often driven by cost savings [23]. However, some studies also report the reuse of medical devices to realize environmental benefits [24]. In this study, steam sterilization is used to prevent shortages. In 1986, a survey was conducted including Canadian hospitals reusing disposable medical devices [25]. Forty-one percent of the hospitals confirmed that they reused disposable medical devices, with respiratory therapy equipment as the most reused medical device.

Testing by particle counting seems to be essential for both new and sterilized single-use face masks as it indicates the quality of the mask in terms of filtration capacity. This became evident as our data revealed large differences in PFE despite the similar appearance of the mask material. Our results in S6 and S8 Files indicate the presence of coatings that improve the electrostatic behaviour of the mask. As the presence of these kinds of coatings is very difficult to demonstrate, it is advised to test the PFE with a particle counter at all times. To rule out that reprocessed and new face masks do not meet the stated FFP standard, a particle test as a ‘quick and dirty’ test could be applied on every batch, as the test method described in this study can give a quick indication of the quality.

As high quality FFP2 masks react differently to different sterilization methods, it is expected that the electrostatic charge of a mask has a major effect on the PFE especially for smaller lighter particles. Although not part of this study, it would be interesting to investigate how either 121 °C sterilization or H2O2 plasma affects the mask’s electrostatic charges and how this is related to the fiber orientation, pore size and openings between the stacked layers.

Face masks sterilized with the intention of reuse could furthermore undergo a “fit test”. This test may be regarded as a fit validation conforming to a proper fit on the face without leakages around the mask. To assure a decreased risk of spreading other diseases, the bio efficacy of a face mask should also be considered. Tests regarding this aspect were conducted previously and appeared negative for bacteria on steam-sterilized face masks that were tested at the dept. of Microbiology at Franciscus Hospital in the Netherlands [9].

Testing face masks for particles is important in the quality assurance of the sterilization process. Our data shows that despite the implementation of similar 121 °C sterilization protocols, mean PFE outcomes can differ up to 6%. As the types of masks and sterilization methods are similar, the only unknown variable is the wearing/processing influence on the mask during use, transport and inspection. University Hospital 2 in Fig 4 seems to show much lower PFE outcomes. It could be that stretching and bending of the mask can influence the integrity. However, it is also expected that the confidence interval would have been larger as the intensity of the stretching and bending is human dependent. As the confidence interval of the mean PFE outcomes of University Hospital 2 seems similar or even smaller than those of other hospitals, it is advisable to perform validation tests at all hospitals.

In the CSSD at De Meern, a 10% tolerance was accepted for sterilized face masks after testing. Therefore, an 84% filtration capacity on a 0.3 μm particle level was the minimum limit. Although not based on any evidence in the literature, this percentage was considered to be sufficient with respect to the shortages of face masks, taking into consideration that the Coronavirus (SARS-CoV-2) is mainly spread through 0.3 μm or larger droplets. However, a consensus needs to be made to actually define the minimal allowable PFE values in times of crisis.

The DSC, XRD and FTIR test results in S6 File conducted on each of the five layers of the 3M Aura 1862+ and Kolmi op-Air M52010 masks reveal that all layers are made of the same Polypropylene material. The differences in behavior when sterilized cannot be explained by chemical composition. A detailed interpretation of the results can be found in S9 File.

The data of 410 sterilized and 471 newly imported KN/N95 or FFP2 face masks reveal that, despite the differences in PFE between different sterilization processes, still approximately 75% of the face masks of known brands reach the FFP2 standard after sterilization when compared to only 50% of newly imported, less known brands. Our results suggest that the technology needed to manufacture a good mask is not easy. Manufacturing and quality assurance should be monitored and controlled by the government. During the study period, it was observed two times that within a single batch of imported face masks, the quality and layout of the masks were different, despite being wrapped in the same packaging with the same printed PFE standard. This suggests that multiple factories were supplying to one brand. In other cases, some masks (Fig 6) showed almost complete lack of air penetration due to the use of wrong materials or manufacturing processes.

Although our results indicate that sterilized face masks can be used if the filter material can be properly tested, it might be considered that wearing a used mask can have a psychological impact on healthcare workers. To overcome this issue, masks could be marked with the user’s initials so that it can be returned to the same person.

Study limitations

It is of utmost importance that the reprocessing of single-use PPE, as described in this study, is equivalent to existing standards. Each deviation or omission of such standards needs a clear demonstrated equivalence with the applying standards. In our setup, only environmentally dry particles were used in the developed rapid test setup. Although we validated the dry particle setup with an aerosol testing setup (NaCl test, paraffin oil setup) according to the EN 13274–7 standard, it was only possible to compare the PFE for a limited range of particle sizes. Therefore, in-depth knowledge about the PFE related to particle size was not generated. To identify other potential differences between the dry particle and continuous flow setups, a ‘gap’ analysis should be conducted. Other than testing the basic material of the filter layers, we were not able to indicate the presence of surface active coatings. Therefore, it was not possible to investigate the role of surface active coatings on the melting or oxidation of the fibres. Although the study used a validated reprocessing method based on 121°C sterilization to inactivate the virus in the mask, the retention of the inactivated virus has not been studied and should be investigated further in future studies.

Conclusions

Sterilization of disposable face masks by means of standardized steam sterilization at 121°C could be an alternative during face mask shortages due to COVID-19 as long as the fit does not change and the filter materials are not significantly affected by heat. The varying efficiency after reprocessing amongst different brands shows that only quality masks of particular brands such as 3M Aura 1862, 3M Aura 1873 and My-T-Gear 301 are suitable for limited reuse. The data show that the 121°C sterilization process can be safely implemented as long as proper testing of each batch is possible and the process and logistics is well controlled. The new PFE testing method proved to be accurate enough to determine degeneration of the mask material after sterilization and to determine the material quality of imported face masks. FTIR, XRD and DSC measurements indicate that all layers from both masks are made from Polypropylene. Future testing is needed to determine if differences in PFE outcomes after sterilization with 121 °C steam or H2O2 plasma can be explained by the level of crystallinity, or by the orientation and dimensions of the fibres and potential proprietary treatment in the layers of the face mask. PFE comparison between sterilized masks and imported face masks with varying filter qualities indicates that health care professionals in some cases can better reuse a known reprocessed brand rather than an imported face mask from a reseller with an unknown brand.

Supporting information

S1 File. DSC, XRD and FTIR tests on face mask materials.

(PDF)

S2 File. Measurement reports.

(PDF)

S3 File. Validation PFE outcome of particle counter setup against continues flow system of Delft.

(PDF)

S4 File. Particle counter, MISIT.

(PDF)

S5 File. Comparison between CSSD’s.

(PDF)

S6 File. New foreign masks tested on particle counter setup.

(PDF)

S7 File. Material analysis.

(PDF)

S8 File. Testing pressure drop over filter material.

(PDF)

S9 File. Discussion material test results.

(PDF)

Acknowledgments

The authors would like to thank the CSSD personnel and sterilization specialists from all participating hospitals for sending us samples of their sterilized batches for testing. Special thanks to the personnel and management from CSA Services for transforming their CSSD into a high volume face mask processing line. The students and colleagues of TU-Delft who helped test during ‘ProjectMask’ are thanked for their effort. Finally, resellers are thanked for testing their masks at our test facility at De Meern and Delft University of Technology.

Data Availability

All relevant data are within the manuscript and its Supporting information files.

Funding Statement

The author(s) received no specific funding for this work.

References

  • 1.Oberfeld B, Achanta A, Carpenter K, Chen P, Gilette N, Langat P et al. SnapShot: COVID-19. Cell. 2020. doi: 10.1016/j.cell.2020.04.013 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.ROSALES-MENDOZA Sergio, et al. What does plant-based vaccine technology offer to the fight against COVID-19?. Vaccines, 2020;8.2: 183. doi: 10.3390/vaccines8020183 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.ADDI Rachid Ait; BENKSIM Abdelhafid; CHERKAOUI Mohamed. Easybreath Decathlon Mask: An Efficient Personal Protective Equipment (PPE) against COVID-19 in Africa. Journal of Clinical & Experimental Investigations/Klinik ve Deneysel Arastirmalar Dergisi. 2020;11.3. [Google Scholar]
  • 4.Van Den Dobbelsteen J. J., Van Straten B. J., Horeman T. A Comparison of Particle Filter Efficiency Measurements for Protective Masks using Particle Counters with Different Flow Rates. 2020. https://repository.tudelft.nl/islandora/object/uuid%3A604ed9c2-c218-45e8-8c62-7d669865056c (accessed on 28 November 2020).
  • 5.European Committee for Standardization. CEN/TC 79—Respiratory protective devices. 2009;https://standards.cen.eu/dyn/www/f?p=204:110:0::::FSP_PROJECT,FSP_ORG_ID:32928,6062&cs=1FC98AD34A5EE26A0CB5A6155ED4D6E5E
  • 6.Lee SA, Hwang DC, Li HY, Tsai CF, Chen CW, Chen JK. Particle Size-Selective Assessment of Protection of European Standard FFP Respirators and Surgical Masks against Particles-Tested with Human Subjects. J Healthc Eng. 2016;8572493. doi: 10.1155/2016/8572493 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.CDC. The National Institute for Occupational Safety and Health (NIOSH) 2014 https://www.cdc.gov/niosh/docs/96-101/default.html
  • 8.Code of China. Chinese Standard GB 2626–2006. Respiratory protective equipment–China Safety 2006. https://www.codeofchina.com/standard/GB2626-2006.html. 2006.
  • 9.de Man P., van Straten B., van den Dobbelsteen J., van der Eijk A., Horeman T., & Koeleman H. (2020). Sterilization of disposable face masks by means of standardized dry and steam sterilization processes: an alternative in the fight against mask shortages due to COVID-19. Journal of Hospital Infection, 2020;105(2), 356–357. doi: 10.1016/j.jhin.2020.04.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Schöpe H. J., & Klopotek M. Strategies for the re-use of FFP3 respiratory masks during the COVID-19 pandemic 2020. arXiv preprint arXiv:2004.00769.
  • 11.Viscusi Dennis J., Bergman Michael S., Eimer Benjamin C. and Shaffer Ronald E. Evaluation of Five Decontamination Methods for Filtering Facepiece Respirators, Annals of Occupational Hygiene 2009;53,815–827 doi: 10.1093/annhyg/mep070 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Bergman Michael S. et al. Evaluation of Multiple (3-Cycle) Decontamination Processing for Filtering Facepiece Respirators, Journal of Engineered Fibers and Fabrics 20105,33–41 [Google Scholar]
  • 13.Kumar A, Kasloff S. B, Leung A, Cutts T, Strong J.E., Hills K, et al. N95 mask decontamination using standard hospital sterilization technologies. medRxiv. 2020.
  • 14.Henwood T. Coronavirus disinfection in histopathology. Journal of Histotechnology 2020;1–3. doi: 10.1080/01478885.2020.1734718 [DOI] [PubMed] [Google Scholar]
  • 15.Zhang, Qinxin & Zhao, Qingshun. Inactivating porcine coronavirus before nucleic acid isolation at temperatures higher than 56 °C seriously damages its genome integrity. 2020.
  • 16.Harskamp R. E, Van Straten B, Bouman J, Van Maltha-Van Santvoort B, Van den Dobbelsteen J. J., Van der Sijp J.R., et al. Reprocessing filtering facepiece respirators in primary care using medical autoclave: prospective, bench-to-bedside, single-centre study. BMJ open 2020;10(8), e039454. doi: 10.1136/bmjopen-2020-039454 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Teesing G.R., van Straten B, de Man P, & Horeman-Franse T. Is there an adequate alternative to commercially manufactured face masks? A comparison of various materials and forms. Journal of Hospital Infection 2020;106(2), 246–253. doi: 10.1016/j.jhin.2020.07.024 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.European Committee for Standardization. NEN-EN 149+A1 (en): Respiratory protective devices—Filtering half masks to protect against particles—Requirements, testing, marking. 2009.
  • 19.National Institute for Public Health and the Environment, RIVM. Reprocessing FFP-type masks. 2020; data https://www.rivm.nl/coronavirus-covid-19/professionals/binnen-ziekenhuis/mondmaskers
  • 20.3M. Decontamination of 3M Filtering Facepiece Respirators: Global Considerations, technical Bulletin, revision 3. 2020; https://multimedia.3m.com/mws/media/1851918O/decontamination-of-3m-filtering-facepiece-respirators-global-considerations.pdf
  • 21.Markus A, Contingency Reprocessing of Single-Use Personal Protective Equipment (PPE), Issue 2 White paper. 2020. https://www.belimed.com/en-us/success-stories/case-studies.
  • 22.Van Straten B, Van den Dobbelsteen J, Horeman T. Steam sterilization of used disposable face masks with respect to COVID-19 shortages 2020; http://resolver.tudelft.nl/uuid:078a3733-84d6-4d4a-81e6-74210c7fed78 [DOI] [PMC free article] [PubMed]
  • 23.Panta Gopal & Richardson Ann & Shaw Ian. Effectiveness of autoclaving in sterilizing reusable medical devices in healthcare facilities. The Journal of Infection in Developing Countries. 2019; 13. 858–864. doi: 10.3855/jidc.11433 [DOI] [PubMed] [Google Scholar]
  • 24.van Straten B., Dankelman J., van der Eijk A., & Horeman T. A Circular Healthcare Economy; a feasibility study to reduce surgical stainless steel waste. Sustainable Production and Consumption, 27, 169–175. 2020. [Google Scholar]
  • 25.Campbell B.A. & Wells G.A. & Palmer W & Martin D. Reuse of disposable medical devices in Canadian hospitals. American journal of infection control. 1987;15. 196–200. doi: 10.1016/0196-6553(87)90095-2 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Mahendra Singh Dhaka

27 Jul 2021

PONE-D-21-13083

Sterilization of disposable face masks with respect to COVID-19 shortages: a nationwide field study including 19 sterilisation departments

PLOS ONE

Dear Dr. van Straten,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Reviewer's comments are appended below.

Please submit your revised manuscript by Sep 09 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Mahendra Singh Dhaka, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1.Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. In your Methods section, please clarify how imported masks were selected to be included in the analysis; and how many samples were tested for each brand

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: No

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The present manuscript discusses about the efficiency of different methods to reuse the masks, which is the critical need of the hour and can save both, the time and money during this pandemic time. The manuscript, in my opinion is acceptable for publication.

Reviewer #2: The test conducted in this paper tests the used face masks for particle size range 0.3 to 5 micron. The low efficiency of the processed face masks at 0.3 microns is concerning and contradicts the conclusion. Furthermore, the difference in efficiency after reprocessing among the different brands concludes that only masks of a particular brand could be reused.

I am also concerned about the sample size and whether it accurately represents the entire population. There is no sample size test conducted to identify the required sample size to detect the variance to be checked.

Reviewer #3: The manuscript titled “sterilization of disposable face masks with respect to COVID-19 shortage: a nationwide field study including 19 sterilization departments” may be accepted after major revision. The comments are as follows:

1. The language of the manuscript should be improved throughout.

2. The authors need to strengthen the scientific interpretations of the facts thoroughly.

3. Disinfection processes for reuse have not been described for all types of masks.

4. Studies not showing how to remove retention of the virus in the mask.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachment

Submitted filename: PLOS.docx

PLoS One. 2021 Sep 14;16(9):e0257468. doi: 10.1371/journal.pone.0257468.r003

Author response to Decision Letter 0


7 Aug 2021

We would like to thank all of the reviewers for their comments. This round of revision resulted in valuable feedback from the reviewers. We have processed the feedback accordingly and carefully addressed each comment.

We fully agree with the comments raised and improved the manuscript to be more technically sound, with more thorough analysis and improved English grammar. Please find our responses in the document 'Response to Reviewers'.

Attachment

Submitted filename: Response to Reviewers_Rebuttal_ID PONE-D-21-05058_R3.docx

Decision Letter 1

Mahendra Singh Dhaka

2 Sep 2021

Can sterilization of disposable face masks be an alternative for imported face masks? A nationwide field study including 19 sterilization departments and 471 imported brand types during COVID-19 shortages.

PONE-D-21-13083R1

Dear Dr. van Straten,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Mahendra Singh Dhaka, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Mahendra Singh Dhaka

6 Sep 2021

PONE-D-21-13083R1

Can sterilization of disposable face masks be an alternative for imported face masks? A nationwide field study including 19 sterilization departments and 471 imported brand types during COVID-19 shortages.

Dear Dr. van Straten:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Mahendra Singh Dhaka

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File. DSC, XRD and FTIR tests on face mask materials.

    (PDF)

    S2 File. Measurement reports.

    (PDF)

    S3 File. Validation PFE outcome of particle counter setup against continues flow system of Delft.

    (PDF)

    S4 File. Particle counter, MISIT.

    (PDF)

    S5 File. Comparison between CSSD’s.

    (PDF)

    S6 File. New foreign masks tested on particle counter setup.

    (PDF)

    S7 File. Material analysis.

    (PDF)

    S8 File. Testing pressure drop over filter material.

    (PDF)

    S9 File. Discussion material test results.

    (PDF)

    Attachment

    Submitted filename: Reviewer Comments_ID PONE-D-21-05058_R1.docx

    Attachment

    Submitted filename: PLOS.docx

    Attachment

    Submitted filename: Response to Reviewers_Rebuttal_ID PONE-D-21-05058_R3.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES